| Literature DB >> 33284392 |
Sonia D'Arrigo1, Claudio Sandroni2,3, Sofia Cacciola2, Antonio Maria Dell'Anna2, Mauro Pittiruti4, Maria Giuseppina Annetta2, Cesare Colosimo5, Massimo Antonelli2,3.
Abstract
BACKGROUND: Single-lumen 4Fr or double-lumen 5Fr power injectable peripherally inserted central catheters (PICCs) are not accurate for trans-pulmonary thermodilution (TPTD), since they overestimate cardiac index and other TPTD-derived parameters when compared with centrally inserted central catheters (CICCs) because of the smaller size of their lumen. We hypothesize that PICCs with larger lumen size may be reliable for the cardiac index assessment using the TPTD.Entities:
Keywords: Cardiac output; Centrally inserted central catheter; Hemodynamic monitoring; Intensive care unit; Peripherally inserted central catheter; Trans-pulmonary thermodilution
Year: 2020 PMID: 33284392 PMCID: PMC7720030 DOI: 10.1186/s13613-020-00785-2
Source DB: PubMed Journal: Ann Intensive Care ISSN: 2110-5820 Impact factor: 6.925
Fig. 1Flowchart of study population
Patient population: demographic and clinical characteristics
| Characteristics | All patients ( |
|---|---|
| Age, years, mean (SD) | 62.5 (14.4) |
| Male, | 8 (53.3) |
| Body Mass Index, kg/m2, mean (SD) | 26.3 (5.5) |
| SAPS II, mean (SD) | 59.8 (17.1) |
| Diagnosis on admission in ICU, | |
| Septic shock | 10 (66.6) |
| Acute respiratory failure | 3 (20) |
| Cardiogenic shock | 2 (13.4) |
| CICC characteristics | |
| Triple-lumen 7Fr, | 15 (100) |
| Catheter length, cm | 20 |
| Extension length, cm | 14 |
| Approach, | |
| Internal jugular vein | 15 (100) |
| PICC characteristics | |
| Single-lumen 5Fr, | 8 (53.3) |
| Trimmed length 5Fr, cm, mean (SD) | 40.6 (3) |
| Extension length 5Fr, cm | 11.5 |
| Triple-lumen 6Fr, | 7 (46.6) |
| Trimmed length 6Fr, cm, mean (SD) | 38.8 (2.9) |
| Extension length 6Fr, cm | 11.5 |
| Approach, | |
| Brachial vein | 3 (20) |
| Basilic vein | 8 (53.3) |
| Axillary vein | 4 (26.7) |
CICC: centrally inserted central catheter; ICU: intensive care unit; PICC: peripherally inserted central catheter; SAPS: simplified acute physiology score
Manual phase: Comparison between single-lumen 5Fr PICC vs. CICC
| Variables | 5 Fr single-lumen PICC | CICC | Bias [LoA] | Bias, % | Error, % | |
|---|---|---|---|---|---|---|
| Measurements, | 40 | 40 | – | – | – | – |
| CI, L/min/m2 | 3.2 (1.04) | 3.2 (1.06) | 0.824 | 0.005 [− 0.42; 0.52] | 2.2 | 14.7 |
| GEDVI, mL/m2 | 716 (169) | 694 (156) | 0.539 | 22.5 [− 122; 167] | 3.0 | 20.5 |
| EVLWI, mL/kg | 10.6 (2.5) | 10.4 (2.9) | 0.805 | 0.1 [− 2.0; 2.3] | 2.5 | 21 |
| SVI, mL/m2 | 36.1 (9.9) | 35.0 (9.5) | 0.582 | 1.2 [− 5.2; 7.6] | 3.3 | 18 |
| CVP, mmHg | 10.1 (5.3) | 10.1 (5.8) | 0.823 | – | – | – |
| ΔT, °C | 0.32 (0.006) | 0.32 (0.06) | 0.611 | − 0.007 [− 0.027; 0.041] | – | 10.8 |
Data are reported as mean (standard deviation)
CI: cardiac index; CICC: centrally inserted central catheter; CVP: central venous pressure; EVLWI: extra-vascular lung water index; GEDVI: global end-diastolic volume index; LoA: limits of agreement; PICC: peripherally inserted central catheter; SVI: stroke volume index; ΔT: delta temperature
Fig. 2Bland-Altman plot comparing the difference of cardiac index between CICC and PICC during the manual phase. The solid horizontal line represents the mean bias (%) between the two devices. The two dashed horizontal lines correspond to the limits of agreement (mean bias ± 1.96 times the standard deviation of the differences)
Manual phase: Comparison between triple-lumen 6Fr PICC vs. CICC
| Variables | 6 Fr triple-lumen PICC | CICC | Bias [LoA] | Bias, % | Error, % | |
|---|---|---|---|---|---|---|
| Measurements, | 40 | 40 | – | – | – | – |
| CI, L/min/m2 | 3.3 (0.8) | 3.0 (0.7) | 0.107 | 0.28 [− 0.32; 0.88] | 8.5 | 19 |
| GEDVI, mL/m2 | 685 (133) | 632 (102) | 0.05 | 52.8 [− 75.7; 181] | 7.4 | 19.5 |
| EVLWI, mL/kg | 14.0 (5.1) | 12.2 (4.9) | 0.178 | 1.5 [− 2.2; 5.2] | 12.8 | 28.2 |
| SVI, mL/m2 | 44.4 (10.4) | 40.6 (8.5) | 0.077 | 3.8 [− 4.2; 1.8] | 8.5 | 18.8 |
| CVP, mmHg | 11.3 (4.8) | 11.7 (5.5) | 0.764 | – | – | – |
| Δ | 0.33 (0.07) | 0.34 (0.08) | 0.514 | − 0.01 [− 0.06; 0.03] | – | 12.9 |
Data are reported as mean (standard deviation)
CI: cardiac index; CICC: centrally inserted central catheter; CVP: central venous pressure; EVLWI: extra-vascular lung water index; GEDVI: global end-diastolic volume index; LoA: limits of agreement; PICC: peripherally inserted central catheter; SVI: stroke volume index; ΔT: delta temperature
Automated injection phase: Comparison between single-lumen 5Fr PICC vs. CICC
| Variables | 5 Fr single-lumen PICC | CICC | Bias [LoA] | Bias, % | Error, % | |
|---|---|---|---|---|---|---|
| Measurements, | 40 | 40 | – | – | – | – |
| CI, L/min/m2 | 3.2 (0.9) | 3.06 (0.9) | 0.44 | 0.16 [− 0.29; 0.61] | 5.4 | 14.4 |
| GEDVI, mL/m2 | 721 (171) | 704 (188) | 0.676 | 16.9 [− 83.2; 116.9] | 3.1 | 14 |
| EVLWI, mL/kg | 11.4 (3.1) | 10.9 (3.1) | 0.487 | 0.49 [− 1.14; 2.12] | 5.3 | 14.6 |
| SVI, mL/m2 | 35.7 (8.5) | 33.7 (8.6) | 0.299 | 2.0 [− 3.2; 7.2] | 6.1 | 15 |
| CVP, mmHg | 9.8 (5.6) | 9.7 (4.5) | 0.965 | – | – | – |
| Δ | 0.30 (0.06) | 0.32 (0.06) | 0.224 | − 0.017 [− 0.045; 0.010] | – | 8.7 |
| Psi | 112 | 109 | 0.07 | – | – | – |
Data are reported as mean (standard deviation)
CI: cardiac index; CICC: centrally inserted central catheter; CVP: central venous pressure; EVLWI: extra-vascular lung water index; GEDVI: global end-diastolic volume index; LoA: limits of agreement; PICC: peripherally inserted central catheter; Psi: pound per square inch; SVI: stroke volume index; ΔT: delta temperature
Automated injection phase: Comparison between triple-lumen 6Fr PICC vs. CICC
| Variables | 6 Fr triple-lumen PICC | CICC | Bias [LoA] | Bias, % | Error, % | |
|---|---|---|---|---|---|---|
| Measurements, | 40 | 40 | – | – | – | – |
| CI, L/min/m2 | 3.4 (0.7) | 3.0 (0.7) | 0.012 | 0.39 [− 0.05; 0.83] | 12.5 | 13.8 |
| GEDVI, mL/m2 | 689 (117) | 622 (89) | 0.006 | 66.5 [− 40; 173] | 9.7 | 16.2 |
| EVLWI, mL/kg | 14.3 (5.0) | 12.7 (4.5) | 0.134 | 1.6 [− 0.4; 3.7] | 12 | 15.6 |
| SVI, mL/m2 | 44.6 (8.4) | 39.4 (7.5) | 0.004 | 5.2 [− 1.3; 11.7] | 12.5 | 15.5 |
| CVP, mmHg | 12 (5.3) | 11.1 (4.9) | 0.458 | – | – | – |
| Δ | 0.30 (0.07) | 0.34 (0.08) | 0.008 | − 0.05 [− 0.09; 0.00] | – | 15.6 |
| Psi | 120 | 115 | < 0.0001 | – | – | – |
Data are reported as mean (standard deviation)
CI: cardiac index; CICC: centrally inserted central catheter; CVP: central venous pressure; EVLWI: extra-vascular lung water index; GEDVI: global end-diastolic volume index; LoA: limits of agreement; PICC: peripherally inserted central catheter; Psi: pound per square inch; SVI: stroke volume index; ΔT: delta temperature
Injection pressure, ΔT and CI during automated rapid injection through PICCs of different lumen size: comparison with CICC
| Triple-lumen 7Fr CICC | Single-lumen 5Fr PICC | Triple-lumen 6Fr PICC | Double-lumen 5Fr PICCa | |
|---|---|---|---|---|
| Internal diameter, mma | 1.016 | 1.02 | 0.94 | 0.81 |
| Mean Psi | 110 | 112 | 120 | 130 |
| ΔT bias, °Cb | Reference | − 0.017 | − 0.05 | − 0.12 |
| CI mean difference, L/min/m2 | Reference | 0.16 | 0.39 | 1.55 |
| CI bias, % | Reference | 5.4 | 12.5 | 39.3 |
aData from D’Arrigo et al. [7]
bDifference between mean ΔTPICC and mean ΔTCICC
CI: cardiac index; CICC: centrally inserted central catheter; ΔT: delta temperature; PICC: peripherally inserted central catheter; Psi: pound per square inch
Fig. 3Cross section of catheters. A Single-lumen 5fr; B Triple-lumen 6Fr; C Double-lumen 5Fr