| Literature DB >> 33197266 |
Willeke van der Stuijt1, Anne-Floor B E Quast1, Sarah W E Baalman1, Koen C de Wilde1, Tom F Brouwer1, Arthur A M Wilde1, Reinoud E Knops1.
Abstract
AIMS: To guarantee uninterrupted function of the subcutaneous implantable cardioverter-defibrillator (S-ICD), the pulse generator needs to be surgically replaced before the battery is depleted. The risks related to this replacement substantially impact long-term outcome for S-ICD recipients, as the majority will undergo one or several of these procedures in their lifetime. We aim to describe the procedural characteristics of the replacement procedure and to provide an insight in the complications associated with these replacements. METHODS ANDEntities:
Keywords: Battery longevity; Complication rate; PRAETORIAN score; Pulse generator replacement; Shock impedance; Subcutaneous implantable cardioverter-defibrillator
Year: 2021 PMID: 33197266 PMCID: PMC7947576 DOI: 10.1093/europace/euaa263
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Patient characteristics
|
| |
|---|---|
| Age at replacement (years), mean ± SD | 46 ± 15 |
| Follow-up after replacement (years), median (IQR) | 1.9 (0.6–3.25) |
| Male, | 44 (61) |
| LVEF (%), mean ± SD | 49.6 ± 11.6 |
| Implanted using the two-incision technique, | 47 (65) |
| ICD Indication, | |
| Primary indication | 44 (61) |
| Secondary indication | 28 (39) |
| Clinical disease, | |
| Ischaemic cardiomyopathy | 11 (15) |
| Non-ischaemic cardiomyopathy | 7 (9) |
| Genetic arrhythmic disease | 39 (54) |
| Congenital heart disease | 3 (4) |
| Idiopathic VF | 10 (14) |
| Other | 2 (3) |
| NYHA functional class, | |
| Class I | 61 (85) |
| Class II | 9 (13) |
| Class III | 2 (3) |
| PRAETORIAN score at implant, | |
| Low | 64 (89) |
| Intermediate | 4 (6) |
| High | 1 (1) |
LVEF, left ventricular ejection fraction; IQR, interquartile range; NYHA, New York Heart Association; SD, standard deviation; VF, ventricular fibrillation.
Practical aspects of replacement procedure
| Battery longevity (years), mean ± SD | 5.9 ± 0.7 |
| Procedural time (min), mean ± SD | 38 ± 13 |
| Defibrillation testing, | |
| Successful first shock | 53 (91.4) |
| Unsuccessful first shock | 5 (8.6) |
| Unable to induce VF | 5 of 63 (7.9) |
| Shock impedance during implant, mean ± SD | 77 ± 26 Ω |
| Shock impedance during replacement, mean ± SD | 86 ± 28 Ω |
SD, standard deviation; VF, ventricular fibrillation.