Literature DB >> 29626366

Six-year follow-up of the initial Dutch subcutaneous implantable cardioverter-defibrillator cohort: Long-term complications, replacements, and battery longevity.

Anne-Floor B E Quast1, Vincent F van Dijk2, Sing-Chien Yap3, Alexander H Maass4, Lucas V A Boersma2, Dominic A Theuns3, Reinoud E Knops1.   

Abstract

INTRODUCTION: Experience with the subcutaneous implantable cardioverter-defibrillator (S-ICD) is expanding rapidly. However, data on long-term performance or complications related to elective generator replacement are lacking.
METHODS: Follow-up (FU) data of all patients implanted between December 2008 and April 2011 were collected. Complications were defined as those requiring surgical intervention. Kaplan-Meier estimates for complication and shock rates, with corresponding 95% confidence intervals (CI), were calculated.
RESULTS: One hundred and eighteen patients were included. Median FU was 6.1 years (IQR 5.6-6.5 years). Short-term complication rate (0-30 days) was 3% (CI 0-6%). Long-term complication rate at 6 years was 19% (CI 12-26%), corresponding with an annual complication rate of 3%. One patient in this cohort developed a need for a transvenous ICD (TV-ICD) in order to provide pacing for bradycardia (1%). Six patients were implanted with a TV-ICD after experiencing an S-ICD complication for which extraction was necessary. In total, 10 S-ICDs were extracted; none resulted in a complication. Eight patients had a nonsystemic ICD-related infection and no lead failures were observed. The majority, 68 (58%) patients, received an elective generator replacement. Two patients had a complication related to generator replacement (3%). Battery longevity was 5.6 years (IQR 5.2-6.1). Appropriate and inappropriate shock rates of 6-year estimates were 17% (CI 9-25%) and 21% (CI 15-27%), respectively.
CONCLUSIONS: This cohort represents the longest follow-up to date and shows a low annual complication rate without lead failures or systemic infections. Battery longevity of the first S-ICD generation results in relative early generator replacement procedures.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  battery longevity; complications; long-term performance; replacement; subcutaneous ICD; ventricular tachycardia

Mesh:

Year:  2018        PMID: 29626366     DOI: 10.1111/jce.13498

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  8 in total

1.  Feasibility of entirely subcutaneous ICD™ systems in patients with coronary artery disease.

Authors:  Kevin Willy; Markus Bettin; Florian Reinke; Nils Bögeholz; Christian Ellermann; Benjamin Rath; Patrick Leitz; Julia Köbe; Lars Eckardt; Gerrit Frommeyer
Journal:  Clin Res Cardiol       Date:  2019-03-21       Impact factor: 5.460

Review 2.  Non-transvenous ICD therapy: current status and beyond.

Authors:  Vincent F van Dijk; Lucas V A Boersma
Journal:  Herz       Date:  2021-11-09       Impact factor: 1.443

3.  Efficacy, safety, and in-hospital outcomes of subcutaneous versus transvenous implantable defibrillator therapy: A meta-analysis and systematic review.

Authors:  Chao-Feng Chen; Chao-Lun Jin; Mei-Jun Liu; Yi-Zhou Xu
Journal:  Medicine (Baltimore)       Date:  2019-05       Impact factor: 1.817

4.  A novel screening test for inappropriate shocks due to myopotentials from the subcutaneous implantable cardioverter-defibrillator.

Authors:  Yuji Ishida; Shingo Sasaki; Yuichi Toyama; Kimitaka Nishizaki; Yoshihiro Shoji; Takahiko Kinjo; Taihei Itoh; Daisuke Horiuchi; Masaomi Kimura; Michael R Gold; Hirofumi Tomita
Journal:  Heart Rhythm O2       Date:  2020-04-27

Review 5.  An Overview of Clinical Outcomes in Transvenous and Subcutaneous ICD Patients.

Authors:  S W E Baalman; A B E Quast; T F Brouwer; R E Knops
Journal:  Curr Cardiol Rep       Date:  2018-07-10       Impact factor: 2.931

6.  Diagnosis and management of subcutaneous implantable cardioverter-defibrillator infections based on process mapping.

Authors:  Larry M Baddour; Raul Weiss; George E Mark; Mikhael F El-Chami; Mauro Biffi; Vincent Probst; Pier D Lambiase; Marc A Miller; Timothy McClernon; Linda K Hansen; Bradley P Knight
Journal:  Pacing Clin Electrophysiol       Date:  2020-04-23       Impact factor: 1.976

7.  Complications related to elective generator replacement of the subcutaneous implantable defibrillator.

Authors:  Willeke van der Stuijt; Anne-Floor B E Quast; Sarah W E Baalman; Koen C de Wilde; Tom F Brouwer; Arthur A M Wilde; Reinoud E Knops
Journal:  Europace       Date:  2021-03-08       Impact factor: 5.214

8.  The potential impact of acute coronary syndromes on automatic sensing system in Subcutaneous-ICDs.

Authors:  M L Narducci; R Scacciavillani; G Pinnacchio; G Bencardino; F Perna; G Comerci; M Campisi; I Ceccarelli; C Pavone; F Spera; A Bisignani; F Crea; G Pelargonio
Journal:  Int J Cardiol Heart Vasc       Date:  2021-07-20
  8 in total