| Literature DB >> 32181134 |
Jian-Ming Li1,2, Yanhui Li3, Venkatakrishna Tholakanahalli1,2, David G Benditt1,2.
Abstract
Entities:
Keywords: Defibrillation; Fibrous encapsulation; ICD lead; Impedance; S-ICD; Sudden cardiac death
Year: 2019 PMID: 32181134 PMCID: PMC7064802 DOI: 10.1016/j.hrcr.2019.11.009
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Gross and microscopic view of fibrous sheath around defibrillator shock electrode and lead position. A: The defibrillation shock electrode was completely covered with a translucent fiber sheath. B: Microscope features: A dense fibrous tissue with collagen fibers in parallel (pink) and fibroblast (blue), hematoxylin–eosin (H-E) stain, 400×. C,D: Full cross-section of fibrous sheath in microscope, H-E stain, 40×. E,F: Chest radiograph of lead and device positions before (E) and after (F) lead revision.
Figure 2Peeling off the fibrous encapsulation and repositioning the electrode into the same location through a string preattached to the tip of the electrode. Top left: Failure to sense ventricular fibrillation by the device, and external defibrillator rescue. Top right: The defibrillation shock electrode was completely covered with a fiber sheath. Bottom: Chest radiograph to verify the same lead position before and after revision.
Clinical characteristics and high-voltage lead impedance measured by high-voltage shock at initial implant and before, during, and after subcutaneous implantable cardioverter-defibrillator system revision
| Case 1 | Case 2 | |
|---|---|---|
| Age | 61 | 58 |
| Sex | Male | Male |
| Indication | Primary prevention | Primary prevention |
| ICM | NICM | |
| BMI | 34 | 31 |
| LVEF | 30% | 30% |
| LV thickness (septum/posterior wall) | 1.2 cm/1.2 cm | 1.1 cm/1.1 cm |
| Renal function (GFR, mL/min/1.73 m2) | 34 | >60 |
| DT at initial implant | Not done | 98 ohms @ 65 J |
| Failed DT at 2 years after implant | 153 ohms @65J | 185 ohms @ 10 J |
| 138 ohms@70J | 170 ohms @ 65 J | |
| Revision with successful DT | 85 ohms @65J | 77 ohms @ 65 J |
| Pocket revision | No | No |
| Follow-up after revision | Died of unknown cause at 6 months | Successful DT at 1 year |
| 87 ohms @ 65 J | ||
| 79 ohms @ 65 J |
BMI = body mass index; DT = defibrillation test; GFR = glomerular filtration rate; ICM = ischemic cardiomyopathy; LV = left ventricle; LVEF = left ventricular ejection fraction; NICM = nonischemic cardiomyopathy.
Figure 3Trends of high-voltage impedance measured by low-voltage pacing pulse during 25 months of follow-up. Arrow indicates the high-voltage impedance with 65 J shock at the time of failed defibrillation test.