Karin A Skalina1, D Y Goldstein2, Jaffar Sulail3, Eunkyu Hahm3, Momka Narlieva3, Wendy Szymczak3, Amy S Fox3.
Abstract
With the global outbreak of COVID-19, the demand for testing rapidly increased and quickly exceeded the testing capacities of many laboratories. Clinical tests which receive CE (Conformité Européenne) and Food and Drug Administration (FDA) authorisations cannot always be tested thoroughly in a real-world environment. Here we demonstrate the long-term stability of nasopharyngeal swab specimens for SARS-CoV-2 molecular testing across three assays recently approved by the US FDA under Emergency Use Authorization. This study demonstrates that nasopharyngeal swab specimens can be stored under refrigeration or even ambient conditions for 21 days without clinically impacting the results of the real-time reverse transcriptase-PCR testing. © Author(s) (or their employer(s)) 2022. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.Entities:
Keywords: hospital; microbiology; pathology department; polymerase chain reaction; virology
Mesh:
Year: 2020 PMID: 33144357 PMCID: PMC8685649 DOI: 10.1136/jclinpath-2020-206738
Source DB: PubMed Journal: J Clin Pathol ISSN: 0021-9746 Impact factor: 3.411
Figure 1Cycle threshold of SARS-CoV-2 in viral transport media stored at ambient temperature over time on three automated real-time reverse transcriptase-PCR machines: Luminex ARIES (A), Abbott m2000 (B) and Hologic Panther Fusion (C). Target 1 refers to the N target for the ARIES and both targets for the m2000 and Fusion. Target 2 refers to the ORF target for the ARIES. IC, internal amplification control; ORF, open-reading frame.
Figure 2Cycle threshold of SARS-CoV-2 in viral transport media stored under refrigeration over time on three automated real-time reverse transcriptase-PCR machines: Luminex ARIES (A), Abbott m2000 (B) and Hologic Panther Fusion (C). Target 1 refers to the N target for the ARIES and both targets for the m2000 and Fusion. Target 2 refers to the ORF target for the ARIES. IC, internal amplification control; ORF, open-reading frame.
Comparison of the impact of storage temperature on the average cycle threshold
| Mean | SD | ||
|
| |||
| 20°C | 20.34 | 0.2062 | n=20 |
| 4°C | 20.12 | 0.1196 | n=20 |
| Difference | −0.22 | 0.1795 | P<0.0001 |
|
| |||
| 20°C | 10.96 | 1.793 | n=11 |
| 4°C | 9.45 | 1.658 | n=11 |
| Difference | −1.506 | 2.077 | P=0.037 |
|
| |||
| 20°C | 20.41 | 0.522 | n=20 |
| 4°C | 20.26 | 0.7687 | n=20 |
| Difference | −0.15 | 0.5835 | P=0.2646 |
|
| |||
| 20°C | 22.58 | 0.6348 | n=20 |
| 4°C | 22.53 | 0.794 | n=20 |
| Difference | −0.05 | 0.6295 | P=0.7264 |
Significance was calculated using a paired t-test.
ORF, open-reading frame.
Statistical analysis of the cycle thresholds
| CV | R2 | ||
|
| |||
| IC | 20°C | 0.921 | 0.0117 |
| 4°C | 1.423 | 0.00834 | |
| Target | 20°C | 1.014 | 0.220 |
| 4°C | 0.599 | 0.0535 | |
|
| |||
| IC | 20°C | 2.924 | 0.022 |
| 4°C | 5.521 | 0.144 | |
| Target | 20°C | 16.362 | 0.192 |
| 4°C | 17.55 | 0.139 | |
|
| |||
| IC | 20°C | 2.482 | 0.360 |
| 4°C | 2.323 | 0.0551 | |
| Target N | 20°C | 2.812 | 0.406 |
| 4°C | 3.525 | 0.343 | |
| Target ORF | 20°C | 2.558 | 0.488 |
| 4°C | 3.794 | 0.390 |
CV, coefficient of variation; IC, internal amplification control; ORF, open-reading frame.
Cycle thresholds (Ct) of true patient samples
| Patient | Ct day 0 | Ct day 35 |
| 1 | 21.28* | 30.6 |
| 2 | 24.6/21.5† | 20 |
| 3 | 28.4/24.5† | 22.9 |
| 4 | 16.8/14.8† | 13.2 |
| 5 | 32.2/28.1† | 29.3 |
| 6 | 16.4* | 27.7 |
| 7 | 25.3/23† | 22.8 |
All samples on day 35 were tested on Hologic Panther.
*Samples tested on the m2000.
†Samples tested on ARIES (N/ORF).