Mary A Rodgers1, Ana Olivo2, Barbara J Harris2, Chris Lark2, Xinxin Luo2, Michael G Berg2, Todd V Meyer2, Aurash Mohaimani2, Gregory S Orf2, Yitz Goldstein3, Amy S Fox3, Julie Hirschhorn4, William B Glen4, Frederick Nolte4, Alan Landay5, Cheryl Jennings5, James Moy5, Venice Servellita6, Charles Chiu6, Rahul Batra7, Luke B Snell7, Gaia Nebbia7, Sam Douthwaite7, Amilcar Tanuri8, Lavanya Singh9, Tulio de Oliveira9, Ambroise Ahouidi10, Souleymane Mboup10, Gavin A Cloherty2.
Abstract
BACKGROUND: Viral diversity presents an ongoing challenge for diagnostic tests, which need to accurately detect all circulating variants. The Abbott Global Surveillance program monitors severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) variants and their impact on diagnostic test performance.Entities:
Keywords: Antibody assays; Molecular diagnostics; Rapid antigen tests; SARS-CoV-2; Variant of concern
Mesh:
Substances:
Year: 2022 PMID: 35086043 PMCID: PMC8770247 DOI: 10.1016/j.jcv.2022.105080
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 3.168
m2000 and Alinity m SARS-CoV-2 assay results with B.1.1.7, B.1.351, and P.1 virus culture dilutions.
| B.1.1.7 variant | BEI NR-54,011 Dilution 1 | 28 | 6.25 | Detected | Detected | Detected |
| BEI NR-54,011 Dilution 2 | 2.8 | 5.19 | Detected | Detected | Detected | |
| BEI NR-54,011 Dilution 3 | 0.28 | 4.00 | Detected | Detected | Detected | |
| BEI NR-54,011 Dilution 4 | 0.028 | 2.85 | Detected | Detected | Detected | |
| B.1.351 variant | BEI NR-54,008 Dilution 1 | 28 | 4.63 | Detected | Detected | Detected |
| BEI NR-54,008 Dilution 2 | 2.8 | 3.83 | Detected | Detected | Detected | |
| BEI NR-54,008 Dilution 3 | 0.28 | 2.51 | Detected | Detected | Detected | |
| BEI NR-54,008 Dilution 4 | 0.028 | 1.27 | Detected | Detected | Detected | |
| BEI NR-54,009 Dilution 1 | 28 | 4.35 | Detected | Detected | Detected | |
| BEI NR-54,009 Dilution 2 | 2.8 | 3.27 | Detected | Detected | Detected | |
| BEI NR-54,009 Dilution 3 | 0.28 | 2.11 | Detected | Detected | Detected | |
| BEI NR-54,009 Dilution 4 | 0.028 | 0.73 | Detected | Detected | Detected | |
| P.1 variant | BEI NR-54,982 Dilution 1 | 28 | 4.56 | Detected | Detected | Detected |
| BEI NR-54,982 Dilution 2 | 2.8 | 3.45 | Detected | Detected | Detected | |
| BEI NR-54,982 Dilution 3 | 0.28 | 2.39 | Detected | Detected | Detected | |
| BEI NR-54,982 Dilution 4 | 0.028 | 1.32 | Detected | Detected | Detected |
Log genome equivalents (GE)/mL were calculated from a standard curve plot of TCID50 vs GE with an R2 value of 0.99. Four out of four replicates were detected for all dilutions tested.
Comparison of GE*/TCID50 Ratios.
| 1 | 28 | 6.25 | 65,189 | 4.63 | 1540 | 4.35 | 809 | 4.56 | 1296 | 42 | 81 | 50 |
| 2 | 2.8 | 5.19 | 55,963 | 3.83 | 2454 | 3.27 | 672 | 3.45 | 1020 | 23 | 83 | 55 |
| 3 | 0.28 | 4.00 | 36,169 | 2.51 | 1193 | 2.11 | 471 | 2.39 | 913 | 30 | 77 | 40 |
| 4 | 0.028 | 2.85 | 25,294 | 1.27 | 674 | 0.73 | 248 | 1.32 | 755 | 38 | 102 | 33 |
| Average | NA | NA | 45,654 | NA | 1465 | NA | 550 | NA | 996 | 31 | 83 | 46 |
*Genome equivalents (GE)/mL were calculated from a standard curve plot of TCID50 vs GE with an R2 value of 0.99.
Fig. 1Characteristics of clinical specimen panel. The percentage of N = 228 total remnant respiratory samples collected from each of the indicated countries (country two letter codes, BR – Brazil, SA – South Africa, SN – Senegal, UK – United Kingdom, US – United States) and US states (CA – California, FL – Florida, IL – Illinois, NY – New York, SC – South Carolina) are shown in panel A. The total number of specimens (N) for each lineage in the panel of 228 remnant respiratory samples are shown in panel B.
Fig. 2Clinical variant specimen panel. Remnant clinical specimens from patients with sequence-confirmed VOI/VOC infections were tested on at least one diagnostic assay –m2000 RealTime SARS-CoV-2 (A), Abbott ID NOW COVID-19 (B), Alinity m SARS-CoV-2 (C), the BinaxNOW COVID-19 Ag Card (D) and Panbio COVID-19 Ag Rapid Test Device (E-F). Genome equivalents (GE).
SARS-CoV-2 Serology Assay Results with B.1.1.7 Clinical Samples.
| Patient 1 | 18.36 | 18.68 | 6.76 | 6.97 | 13,359.5 | 11,595.4 | Positive | 26 |
| Patient 2 | 1.22 | 1.26 | 0.22 | 0.22 | 178.5 | 189.5 | Negative | 20 |
| Patient 3 | 11.68 | 12.05 | 5.62 | 6.01 | 5288.1 | 6570.1 | Positive | 19 |
| Patient 4 | 11.71 | 11.04 | 0.8 | 0.87 | 149.4 | 153.7 | Negative | 25 |
| Patient 5 | 6.47 | 6.58 | 5.29 | 5.44 | 5472.9 | 5468.1 | Positive | 20 |
| Patient 6 | 1.31 | 1.32 | 6.6 | 6.58 | > 40,000.0 | > 40,000.0 | Positive | 16 |
| Patient 7 | 14.16 | 13.51 | 6.8 | 6.98 | 12,034.7 | 11,191.5 | Positive | 22 |
| Patient 8 | 7.41 | 7.2 | 6.39 | 6.36 | 14,489.9 | 12,718.3 | Positive | 21 |
| Patient 9 | 15.15 | 14.86 | 4.73 | 4.82 | 2745.6 | 2603.7 | Positive | 20 |
| Patient 10 | 0.03 | 0.04 | 0.02 | 0.02 | 0.9 | 0.2 | Negative | 15 |
| Patient 11 | 1.27 | 1.22 | 0.22 | 0.23 | 186.4 | 186.8 | Negative | 20 |
| Patient 12 | 26.79 | 26.67 | 6.62 | 6.6 | 22,392.8 | 19,367.7 | Positive | 23 |
| Patient 13 | 5.81 | 6.06 | 4.31 | 4.33 | 1255.4 | 1183.3 | Positive | 19 |
| Patient 14 | 88.25 | 88.59 | 6.67 | 6.42 | 6811.4 | 6250 | Positive | 16 |
| Patient 15 | 88.27 | 88.14 | 6.66 | 6.58 | 6815.2 | 6587.5 | Positive | 21 |
| Patient 16 | 88.87 | 89.11 | 6.39 | 6.5 | 6924.4 | 6401.6 | Positive | 18 |
| Patient 17 | 2.09 | 2.2 | 6.23 | 6.19 | 1626.4 | 1630.5 | Positive | 17 |
| Patient 18 | 91.31 | 87.56 | 6.62 | 6.55 | 7459.3 | 6330 | Positive | 19 |
| Patient 19 | 4.56 | 4.58 | 7.36 | 7.31 | 2345.6 | 2274.9 | Positive | 19 |
| Patient 20 | 4.55 | 4.61 | 7.18 | 7.53 | 2491.1 | 2236.1 | Positive | 20 |
An index value of ≥1.4 is positive for IgG assays, ≥1.0 is positive for IgM assays, a result of ≥50 AU/ml per mL is positive and <22 AU/mL is not reported in a clinical setting for IgG II assays.