Robin de Nijs1, Nienke D Sijtsema1,2, Matthijs F Kruis3, Claus Verner Jensen4, Martin Iversen5, Michael Perch5, Jann Mortensen1.
Abstract
INTRODUCTION: Ventilation and perfusion single-photon emission computed tomography combined with computed tomography (SPECT/CT) is a powerful tool to assess the state of the lungs in chronic obstructive pulmonary disease (COPD). 81mKrypton is a gaseous ventilation tracer and distributes similarly to air, but is not widely available and relatively expensive. 99mTc-Technegas is cheaper and has wider availability, but is an aerosol, which may deposit in hot spots as the severity of COPD increases. In this study, 81mKrypton and 99mTc-Technegas were compared quantitatively in patients with severe COPD.Entities:
Mesh:
Substances:
Year: 2021 PMID: 33105398 PMCID: PMC7808361 DOI: 10.1097/MNM.0000000000001314
Source DB: PubMed Journal: Nucl Med Commun ISSN: 0143-3636 Impact factor: 1.698
Summary of characteristics from the lung function test of the included patients at the first follow-up
| Total number | Age (years) | FEV1 | FVC | FEV1/FVC (%) | TLC | RV | DLCOc |
|---|---|---|---|---|---|---|---|
| 11 (3F) | 60 ± 9 | 27 ± 11 | 62 ± 21 | 35 ± 11 | 130 ± 13 | 254 ± 53 | 33 ± 13 |
All values are shown as the mean ± the sample SD. Unless stated otherwise % of predicted is stated.
DLCOc, diffusing capacity of lungs for carbon monoxide (corrected for hemoglobin); FEV1, forced expired volume in the first second; FVC, forced vital volume; RV, residual volume; TLC, total lung capacity.
Some relevant available collimator specifications according to the vendor for the Philips Precedence single-photon emission computed tomography combined with computed tomography
| Collimator name | Planar resolution, mm at 10 cm | Tc-99m sensitivity (cps/MBq) | Septal thickness (mm) | Septal penetration (%) | Photon energy (keV) |
|---|---|---|---|---|---|
| MEGP | 11.3 | 98 | 1.143 | 3.3 | 300 |
| LEGP | 8.8 | 119 | 0.180 | 1.9 | 140 |
| LEHR | 7.4 | 66 | 0.152 | 1.3 | 140 |
Parameters are defined according to NEMA NU 1-2012.
LEGP, low energy general purpose; LEHR, low energy high resolution; MEGP, medium energy general purpose.
Fig. 1Example of the definition of periphery and center for an axial slice of the left lung for calculation of the penetration ratio. (a) Shows the mask as fits onto the SPECT data. (b) Shows the distribution of fractional radiuses over the mask indicated by the grayscale. The center with a fractional radius of zero is displayed with black inside the lung. (c) Shows the resulting division of the mask in central zone, shown in white, and the peripheral zone, shown in gray.
Summary of the penetration ratio values for 81mKrypton and Technegas for all reconstructions
| Penetration ratio (±SEM) | MEGP | LEGP | ||||
|---|---|---|---|---|---|---|
| 81mKrypton | Technegas | 81mKrypton | Technegas | |||
| ACSC | 1.03 ± 0.10 | 0.84 ± 0.09 | 3·× 10−4 | 0.89 ± 0.03 | 0.83 ± 0.06 | 0.18 |
| AC | 1.06 ± 0.09 | 0.88 ± 0.10 | 7·× 10−4 | 0.94 ± 0.03 | 0.88 ± 0.06 | 0.16 |
| NAC | 0.83 ± 0.06 | 0.66 ± 0.05 | 2·× 10−5 | 0.73 ± 0.02 | 0.68 ± 0.04 | 0.09 |
AC, attenuation correction; ACSC, attenuation and scatter correction; LEGP, low energy general purpose; MEGP, medium energy general purpose; (N)AC, (non)attenuation corrected; SC, scatter corrected.
Indicate the significant differences between 81mKrypton and Technegas.
Summary of the overall heterogeneity index and the filtered heterogeneity index values for 81mKrypton and Technegas for all reconstructions
| MEGP | LEGP | |||||
|---|---|---|---|---|---|---|
| Heterogeneity Index (±SEM) | 81mKrypton | Technegas | 81mKrypton | Technegas | ||
| ACSC | ||||||
| Overall | 1.03 ± 0.04 | 1.18 ± 0.07 | 0.006 | 0.71 ± 0.03 | 1.24 ± 0.07 | 3·× 10−11 |
| 2–4 cm | 0.38 ± 0.02 | 0.55 ± 0.03 | 3·× 10−7 | 0.19 ± 0.01 | 0.57 ± 0.04 | 1·× 10−15 |
| AC | ||||||
| Overall | 0.98 ± 0.04 | 1.11 ± 0.06 | 0.009 | 0.69 ± 0.03 | 1.17 ± 0.06 | 5·× 10−11 |
| 2–4 cm | 0.34 ± 0.02 | 0.51 ± 0.03 | 2·× 10−7 | 0.18 ± 0.01 | 0.53 ± 0.03 | 2·× 10−15 |
| NAC | ||||||
| Overall | 1.02 ± 0.04 | 1.08 ± 0.06 | 0.22 | 0.68 ± 0.02 | 1.12 ± 0.06 | 4·× 10−10 |
| 2–4 cm | 0.37 ± 0.02 | 0.50 ± 0.03 | 1·× 10−5 | 0.19 ± 0.01 | 0.50 ± 0.03 | 2·× 10−15 |
AC, attenuation correction; ACSC, attenuation and scatter correction; LEGP, low energy general purpose; MEGP, medium energy general purpose; (N)AC, (non)attenuation corrected; SC, scatter corrected.
Indicate the significant differences between 81mKrypton and Technegas.
Summary of the paired analysis for HI and PR comparing 81mKrypton with medium energy general purpose collimators and Technegas with low energy general purpose collimators
| (±SEM) | 81mKrypton | Technegas | |
|---|---|---|---|
| MEGP | LEGP | ||
| ACSC | |||
| HI Overall | 1.03 ± 0.04 | 1.20 ± 0.07 | 0.004 |
| HI 2–4 cm | 0.38 ± 0.02 | 0.56 ± 0.04 | 1·× 10−6 |
| PR | 1.03 ± 0.1 | 0.78 ± 0.06 | 3·× 10−5 |
| AC | |||
| HI Overall | 0.98 ± 0.04 | 1.13 ± 0.07 | 0.007 |
| HI 2–4 cm | 0.34 ± 0.02 | 0.51 ± 0.04 | 9·× 10−7 |
| PR | 1.06 ± 0.1 | 0.83 ± 0.07 | 3·× 10−5 |
| NAC | |||
| HI Overall | 1.02 ± 0.04 | 1.08 ± 0.06 | 0.26 |
| HI 2–4 cm | 0.37 ± 0.02 | 0.49 ± 0.03 | 4·× 10−5 |
| PR | 0.83 ± 0.06 | 0.64 ± 0.04 | 5·× 10−6 |
Indicate the significant differences between 81mKrypton and Technegas.
AC, attenuation correction; ACSC, attenuation and scatter correction; HI, heterogeneity index; (N)AC, (non)attenuation corrected; PR, penetration ratio; SC, scatter corrected; LEGP, low energy general purpose; MEGP, medium energy general purpose.
Fig. 2Bland–Altman plot of the PR for 81mKrypton versus Technegas, for MEGP and LEGP collimators. Attenuation and scatter correction were applied. The dashed lines indicate 1.96 SD. LEGP, low energy general purpose; MEGP, medium energy general purpose; PR, penetration ratio.
Fig. 3Bland–Altman plots of 81mKrypton versus Technegas for the overall HI (a) and filtered HI (b) after application of ACSC. ACSC, attenuation and scatter correction; HI, heterogeneity index.
Fig. 4An example of a patient with a severe hot spot in the left lung for Technegas in an attenuation and scatter corrected scan using MEGP collimators. (a) The scatter plot corresponding to the left lung showing 33 outliers. The red line indicates the linear fit, while the orange line indicates the outlier limit; all points above the outlier limit are considered to be statistically significant outliers. (b) An axial slice of the corresponding Technegas ventilation image and (c) the same slice in the corresponding 81mKrypton ventilation image. Note the presence of a severe hot spot in the center of the left lung in the Technegas scan outlined in the ROI marked in green (b). The ROI is an autocontour based on 40% of the maximum Technegas intensity. Points in this hot spot ROI are indicated in green in the scatterplot (a). The other hot spot, indicated in the scatter plot in yellow, is not shown in the axial slices. In (b) and (c), the right lung is outlined in orange and the left lung is outlined in blue. MEGP, medium energy general purpose.