PURPOSE: Our purpose was to assess our proposed new synthetic MRI (synMRI) technique, combined with T2-based water suppression (T2wsup), to reduce cerebral spinal fluid (CSF)-partial volume effects (PVEs). These PVEs are problematic in the T2-weighted fluid-attenuation inversion recovery (FLAIR) images obtained by conventional synMRI techniques. METHODS: Our T2wsup was achieved by subtracting additionally acquired long TE spin echo (SE) images of water signals dominant from the originally acquired images after T2 decay correction and a masking on the long TE image using the water volume (Vw) map to preserve tissue SNR, followed by quantitative mapping and then calculation of the synthetic images. A simulation study based on a two-compartment model including tissue and water in a voxel and a volunteer MR study were performed to assess our proposed method. Parameters of long TE and a threshold value in the masking were assessed and optimized experimentally. Quantitative parameter maps of standard and with T2wsup were generated, then wsup-synthetic FLAIR and SE images were calculated using those suitable combinations and compared. RESULTS: Our simulation clarified that the CSF-PVE artifacts in the standard synthetic FLAIR increase T2 as the water volume increases in a voxel, and the volunteer MR brain study demonstrated that the hyperintense artifacts on synthetic images were reduced to <10% of Vw in those with the standard synMRI while keeping the tissue SNR by selecting optimal masking parameters on additional long TE images of TE = 300 ms. In addition, the wsup-synthetic SE provided better gray-white matter contrasts compared with the wsup-synthetic FLAIR while keeping CSF suppression. CONCLUSION: Our proposed T2wsup-synMRI technique makes it easy to reduce the CSF-PVE artifacts problematic in the synthetic FLAIR images using the current synMRI technique by adding long TE images and simple processing. Although further optimizations in data acquisition and processing techniques are required before actual clinical use, we expect our technique to become clinically useful.
PURPOSE: Our purpose was to assess our proposed new synthetic MRI (synMRI) technique, combined with T2-based water suppression (T2wsup), to reduce cerebral spinal fluid (CSF)-partial volume effects (PVEs). These PVEs are problematic in the T2-weighted fluid-attenuation inversion recovery (FLAIR) images obtained by conventional synMRI techniques. METHODS: Our T2wsup was achieved by subtracting additionally acquired long TE spin echo (SE) images of water signals dominant from the originally acquired images after T2 decay correction and a masking on the long TE image using the water volume (Vw) map to preserve tissue SNR, followed by quantitative mapping and then calculation of the synthetic images. A simulation study based on a two-compartment model including tissue and water in a voxel and a volunteer MR study were performed to assess our proposed method. Parameters of long TE and a threshold value in the masking were assessed and optimized experimentally. Quantitative parameter maps of standard and with T2wsup were generated, then wsup-synthetic FLAIR and SE images were calculated using those suitable combinations and compared. RESULTS: Our simulation clarified that the CSF-PVE artifacts in the standard synthetic FLAIR increase T2 as the water volume increases in a voxel, and the volunteer MR brain study demonstrated that the hyperintense artifacts on synthetic images were reduced to <10% of Vw in those with the standard synMRI while keeping the tissue SNR by selecting optimal masking parameters on additional long TE images of TE = 300 ms. In addition, the wsup-synthetic SE provided better gray-white matter contrasts compared with the wsup-synthetic FLAIR while keeping CSF suppression. CONCLUSION: Our proposed T2wsup-synMRI technique makes it easy to reduce the CSF-PVE artifacts problematic in the synthetic FLAIR images using the current synMRI technique by adding long TE images and simple processing. Although further optimizations in data acquisition and processing techniques are required before actual clinical use, we expect our technique to become clinically useful.
Synthetic MRI (synMRI) generates several kinds of contrast-weighted images by synthesizing quantitative parameter maps of proton density (PD), longitudinal relaxation time (T1), and transverse relaxation time (T2) obtained using acquired fast-spin echo (FSE) data. The synMRI is beginning to be used clinically because of the advantages of saving total examination time and the freedom to select MRI acquisition parameters of TR, TI, and TE. Also, using synMRI, quantitative parameter maps can be obtained.[1-7] However, synthetic T2-weighted images (T2WIs) of fluid-attenuation inversion recovery (FLAIR) or a double-inversion recovery (DIR) introduces hyperintense artifacts at the border zone of tissue and ventricle or the surface of the brain cortex. This is thought to be caused by partial volume effects (PVE) introduced by including brain tissue and cerebral spinal fluid (CSF) in a single voxel.[3-6] Nevertheless, there is a quantitative relationship between the CSF–PVE and the hyperintense artifacts between the standard synthetic FLAIR (std-synFLAIR) image, and therefore fundamental solutions have not been clarified yet. FLAIR, including DIR imaging, is very important for clinical MR brain imaging to separate lesions from CSF; however, it requires a relatively long acquisition time. Therefore, the first priority is to solve the problem of CSF–PVE artifacts in synthetic FLAIR images with a minimum time load, in order to further encourage this technique in clinical use.The longitudinal magnetization (Mz) of water in the acquired FLAIR is ideally zero; however, the quantitative maps of T1, T2, and PD obtained by the std-synMRI are calculated from the mixed signals of tissue and water. Therefore, the T2 decay behavior differs depending on the rate of water in a voxel between the acquired FLAIR and std-synFLAIR images. It is thought that those T2 differences introduce the CSF–PVE hyperintense artifacts in the std-synFLAIR image, and this problem is not easily solved unless tissue and water components in each voxel can be separated.This study was done to quantitatively clarify the cause of CSF–PVE-dependent hyperintense artifacts and then to assess a new synMRI technique to reduce those artifacts, especially in FLAIR, without loss of the tissue SNR. The new technique combines synMRI and a T2-based water suppression technique (T2wsup). The assessment was done using a simulation and MR volunteer studies.
Materials and Methods
Model
As shown in Fig. 1, when a unit voxel of volume V consists of two-components of water and tissue, the MR signal in the voxel is based on a two-compartment model of water volume Vw, tissue volume Vt, and the voxel volume is
and the voxel mean proton density is
Likewise, the suffix w and t denote, respectively, the parameters for water and for tissue.
Fig. 1
A two-compartment voxel model for containing water and tissue components in a single voxel. PD, proton density; T1, longitudinal relaxation time; T2, transverse relaxation time.
A two-compartment voxel model for containing water and tissue components in a single voxel. PD, proton density; T1, longitudinal relaxation time; T2, transverse relaxation time.Signal intensities (SIs) of spin echo (SE) and inversion recovery (IR)-SE for a two-compartment model are, respectively,
and
where
When analyzing in the std-synMRI based on the single-compartment model shown in Eq. (5) or Eq. (6) without suffix w or t, the averaged voxel values are used for calculating those quantitative parameters, resulting in parameters that become close to the water values with increasing Vw. In contrast, if water suppression (wsup) is perfectly achieved, Eqs. (3) and (4) become respectively as
and
These mean that the quantitative parameters of tissue can be obtained even analyzing based on a single-compartment model when using wsup data.
Data acquisition and processing flow
Our proposed water suppression (wsup) technique is based on the T2 difference between tissue and water, similar to the high-intensity REduction (HIRE) technique,[8] where the additionally acquired long TE image of the dominant water signal is subtracted from the acquired standard TE image. In our method, the water suppressed quantitative maps are calculated using these wsup images, and finally synthetic images with arbitrary combinations of acquisition parameter maps are obtained. A general processing flow for our proposed technique of wsup-synMRI and the example of using minimum number of data is shown in Fig. 2. The basic idea is that the wsup for the acquired images is performed before calculating quantitative parameters, then followed by calculating quantitative parameter maps and synthetic images. In this, a key flow is wsup for acquired images, as shown in Fig. 3.
Fig. 2
General whole process flow for our proposed technique of T2wsup-synMRI. On the right are example images of each stage using minimum number (L = 1, M = 3) of acquired images when T1wsup is not required. T2wsup-synMRI, T2-based water-suppressed synthetic MRI.
Fig. 3
Detailed process flow for water suppression for acquired images in Fig. 2, where the representative flow is shown. The details are given in the main text.
General whole process flow for our proposed technique of T2wsup-synMRI. On the right are example images of each stage using minimum number (L = 1, M = 3) of acquired images when T1wsup is not required. T2wsup-synMRI, T2-based water-suppressed synthetic MRI.Detailed process flow for water suppression for acquired images in Fig. 2, where the representative flow is shown. The details are given in the main text.Acquired image data at the generalized condition are Sacq(TEm, r), TE1 < TE2 ---< TEM, and water image is Sacq(TElong, r), TElong = TEM, where r denotes the spatial position of each image in all imaging space, R. Below, our proposed technique is explained in 3 steps.First, a water volume ratio map, is obtained, normalizing by dividing the pure water (CSF) SI of the long TE image, The easiest way to normalize is to use a maximum value of the long TE image as
where is an operator to obtain maximum value of [ ] for all r defined in R.Optionally, it is possible to make robust by enlarging the pure CSF portion as
where the Th Vwmax is a threshold to include pure CSF portions with a margin by setting Th Vwmax < 1 If Th Vwmax = 1 in Eq. (11′), then Eq. (11′) becomes Eq. (11).Second, a scaling for T2 decay correction, α(TE), is obtained by using T2 of pure water, T2w, as
Or, when the T2w is unknown, the ratio of the averaged SI within the same pure water regions among different TEs is
where
is an operator to obtain averaged value of [ ] for all r satisfying r ∈ Vw(r) = 1Third, a spatial mask, to keep tissue SNR after subtraction is obtained as
where Th Vwmin is a threshold value (0–1) when pure (100%) water is 1, decided experimentally dependent on TElong but it can be commonly used once it is decided. Then, a spatial smoothing filter is applied to the to smooth the boundary of 0 and 1 in the mask.Fourth, the water images after T2 decay correction Sacq.mask (TE, r), and masking, are obtained from the corresponding standard acquired images with the same condition asFinally, water-suppressed acquired images, Sacp.wsup(TE, r), are obtained by subtracting the corrected water images from the originally acquired images of the same TE as
Quantitative parameter mapping
After obtaining those wsup images for acquired images, quantitative maps are calculated using those wsup images in addition to the standard images. When multiple images (data points) along TE or TI are acquired, unknown quantitative parameters for wsup: Parawsup = (T2wsup, T1wsup, PDwsup) {∼ (T2t, T1t, PDt)}, and for standard: Parastd = (T2std, T1std, PDstd), can be respectively calculated using acquired images of wsup (Sacq.wsup) and standard (Sacq) by least squares method based on the single-compartment model equations in Eq. (5) or Eq. (6).The quantitative parameters can also be obtained using the minimum number (4 or 5) of the acquired images of wsup and standard, as shown in Table 1, based on the following simple equations as
where the MR parameters of TR, TI, and TE are selected as TR1 > T1w > T1t, TI1 ∼ T1t, and TE2 > TE1 to be close to the correct values. Here, if the TE1 is selected to be shortest, the PDwsup can be obtained using T1std because of . When the T1wsup image is required, another long TE image (#5 in Table 1) must be acquired, but it is not necessary for calculating the wsup-synthetic images of FLAIR and SE.
Table 1
Simulation conditions of quantitative parameters and imaging parameters
Parameters
#
Water (CSF)
Tissue
Quantitativea
PD
0.97
0.8
T1 [ms]
4000
1000
T2 [ms]
1910
100
Imagingb
Acquired data
1
SSE (TR1,TE1)
2
SSE (TR1,TE2)
3
SSE (TR1,TE3)
4
SIR (TR1,TI1,TE1)
5
SIR (TR1,TI1,TE3)
TR1 [ms]
10000
TE1,TE2, TE3 [ms]
20, 100, 500
TI1 [ms]
1000
Synthetic
TR [ms]
10000
TE [ms]
20
TI [ms] (for FLAIR)
2350
Quantitative parameters were referred to Warntjes et al.[1]
Simulation conditions of quantitative parameters and imaging parametersQuantitative parameters were referred to Warntjes et al.[1]TE3 correspond to TElong. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; TE, echo time.
Synthetic contrast-weighted imaging
Synthetic images, both of the standard and wsup, are calculated based on the signal models in Eq. (5) for SE and Eq. (6) for FLAIR, respectively, by the following 3 parameter combinations: std-synFLAIR/SE: (PDstd, T1std, T2std), wsup-synFLAIR: (PDstd, T1std, T2wsup), and wsup-synSE: (PDwsup, T1std, T2wsup). Here, the parameter combination for the wsup-synFLAIR is selected to be equivalent to the acquired FLAIR contrast, and that for the wsup-synSE is selected to provide a new wsup contrast, enabling tissue SNR without inversion.
Simulation
First, to compare the degree of PVE quantitatively for the standard and proposed wsup methods, the quantitative parameters of PD, T1, and T2 based on the single-compartment model were calculated as a parameter of a water volume fraction, Vw. The parameters in this simulation are summarized in Table 1. Voxel mean values of quantitative parameters were calculated from those averaged SIs as a parameter of Vw and with a parameter of subtracting weight α(TE) for ideal and those scaled versions with 0.8 and 0.9.Second, the synFLAIR and synSE SIs for the standard and wsup were calculated as a function of TE, each with a parameter of Vw.
MRI experiments
In the MR experiments, the first 4 data points in Fig. 2 were acquired for our proposed synMRI. A healthy volunteer study was performed on an MRI machine (Galan 3T [ZGO]; Canon Medical Systems, Tochigi, Japan) with a 32-channel head coil after obtaining informed consent. An FSE sequence was used, and the acquisition parameters were the following: parallel imaging (SPEEDER; Canon Medical Systems) of speed-up factor 2; acquisition matrix of 256 × 256; display matrix of 512 × 512 after sinc interpolation; FOV = 23 cm; slice thickness = 5 mm; number of slices was commonly selected at maximum for TElong = 500 ms; number of average, 1; TR1 = 10,000 ms, TE1 = 20 ms, TE2 = 100 ms, TElong = 300 or 500 ms, TI1 = 1000 ms, where each different contrast data were acquired with the separated sequence. An adiabatic inversion pulse was used for IR to reduce RF field (B1) inhomogeneity. The scaling factor of α(TE) obtained with the SI ratio of ThVwmax = 1, which corresponds to fitted water T2 of T2w = 990 ms, was used for T2 decay correction of the water SI of each TE. The acquired FLAIR with a coverage interleaved multi-slice was used for comparison. This is introduced to improve both the inversion efficiency and CSF in-flow artifacts by setting the inversion thickness as nearly twice the standard; however, the acquisition time requires 2 times the standard.For assessments, masking effects as a parameter of ThVwmin and the difference of TE for long TE images were compared, then the optimal parameters were used for the following steps. Quantitative maps and synthetic images of FLAIR and SE each with the standard and wsup were compared as a parameter of TE, and then the acquired FLAIR and wsup-synthetic images of FLAIR and SE were also compared.
Results
The quantitative parameters for the water-suppressed (wsup) data, compared with the standard (w/o wsup) data, became close to those for the tissue components when the water signals were ideally suppressed. The wsup effects were weaker when decreasing the subtracting weight, α(TE), from the ideal (Fig. 4).
Fig. 4
Simulation results of signal intensities (left) and quantitative parameters (center and right) versus Vw for our T2-based wsup technique. The standard (std) data became close to water value with increasing Vw but the wsup data showed the values for tissue components by eliminating the water component. The effects of water suppression were weaker as α was smaller, where the T2 and T1 for ideal suppression (α * 1) cannot be obtained at Vw = 1 due to zero signal. wsup, water suppression.
Simulation results of signal intensities (left) and quantitative parameters (center and right) versus Vw for our T2-based wsup technique. The standard (std) data became close to water value with increasing Vw but the wsup data showed the values for tissue components by eliminating the water component. The effects of water suppression were weaker as α was smaller, where the T2 and T1 for ideal suppression (α * 1) cannot be obtained at Vw = 1 due to zero signal. wsup, water suppression.For the SIs as a function of TE (Fig. 5), the hyperintense signals in std-synFLAIR were because the averaged T2 values in a voxel increased as the water volume, Vw, increased, and those effects were stronger with increasing TE. However, those effects were reduced in our proposed wsup-synFLAIR. Furthermore, our proposed wsup-synSE provided SNR improvements by 10–20% in this condition in addition to CSF suppression, because the tissue signals were not IR signals but SE signals. Although this simulation was performed without noise, when considering a noise propagation in actual synMRI, the noise for SE is smaller than the IR-SE due to the smaller contribution of T1 in SE compared to that in IR-SE.
Fig. 5
Simulation results of SIs versus TE with FLAIR (top) and SE (bottom) as parameters of Vw for 3 cases. Note that the SI in standard synthetic-FLAIR increased with increasing TE due to slower T2 decay with increasing Vw (arrow). These effects were corrected by our proposed wsup-syn technique. Furthermore, the wsup-synSE provided better SNR than the wsup-synFLAIR in addition to CSF suppression, that is, the SIs for wsup-synSE, compared with the wsup-synFLAIR, improved 23%, 17%, and 12%, respectively at Vw = 0%, 25%, 50%. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; SE, spin echo; SIs, signal intensities; TE, echo time; Vw, water volume; wsup-syn, water suppression synthetic.
Simulation results of SIs versus TE with FLAIR (top) and SE (bottom) as parameters of Vw for 3 cases. Note that the SI in standard synthetic-FLAIR increased with increasing TE due to slower T2 decay with increasing Vw (arrow). These effects were corrected by our proposed wsup-syn technique. Furthermore, the wsup-synSE provided better SNR than the wsup-synFLAIR in addition to CSF suppression, that is, the SIs for wsup-synSE, compared with the wsup-synFLAIR, improved 23%, 17%, and 12%, respectively at Vw = 0%, 25%, 50%. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; SE, spin echo; SIs, signal intensities; TE, echo time; Vw, water volume; wsup-syn, water suppression synthetic.
MR experiments
The effects of water masking against ThVwmin are shown in Fig. 6. The CSF signals were comparably suppressed without loss of tissue SNR by using the wsup-images with masking compared with those without masking. Regarding the relationship with ThVwmin, the tissue SNRs increased as the ThVwmin increased until a certain ThVwmin, while the hyperintensity CSF–PVE artifacts were increased over those ThVwmin. Thus, the optimal ThVwmin was decided considering the trade-off between the tissue SNR and the CSF–PVE artifacts. In our experiments, the optimal ThVwmin values were ThVwmin = 0.10 at TElong = 300 ms and ThVwmin = 0.05 at TElong = 500 ms, and the tissue SNRs were increased by ∼80% and ∼50% in those conditions, respectively. Comparing the long TE images of TElong = 300 ms with those of TElong = 500 ms, the residual tissue signals of GM and WM were greater for TElong = 300 ms; however, the tissue SNRs of subtracted T2WI even for TElong = 300 ms were used without subtraction by using the optimal ThVwmin. In this volunteer study, all the wsup images of acquired, quantitative maps, and synthetic images were obtained using the optimal ThVwmin = 0.05 and the long TE image of TElong = 500 ms.
Fig. 6
Results of masking effects as a parameter of ThVwmin for TElong of 300 ms (a) and 500 ms (b), each shown by 2 images of long TE and T2WI of TE = 100 ms, signal intensities of long TE images, and SNR of corrected T2WIs. Although the cerebral spinal fluid (CSF)–partial volume effects (PVE) hyperintensity artifacts remained at ThVwmin ≥ 0.15 (arrows), those were reduced comparably at 0.05 ≤ ThVwmin ≤ 0.1 while the SNRs of GM and WM were improved, compared to those w/o masking (ThVwmin = 0). The SNR improvements due to masking were greater for shorter TElong. CSF, cerebral spinal fluid; GM, gray matter; PVE, partial volume effects; T2WIs, T2-weighted images; TE, echo time; WM, white matter.
Results of masking effects as a parameter of ThVwmin for TElong of 300 ms (a) and 500 ms (b), each shown by 2 images of long TE and T2WI of TE = 100 ms, signal intensities of long TE images, and SNR of corrected T2WIs. Although the cerebral spinal fluid (CSF)–partial volume effects (PVE) hyperintensity artifacts remained at ThVwmin ≥ 0.15 (arrows), those were reduced comparably at 0.05 ≤ ThVwmin ≤ 0.1 while the SNRs of GM and WM were improved, compared to those w/o masking (ThVwmin = 0). The SNR improvements due to masking were greater for shorter TElong. CSF, cerebral spinal fluid; GM, gray matter; PVE, partial volume effects; T2WIs, T2-weighted images; TE, echo time; WM, white matter.Acquired images and quantitative maps for standard and wsup are shown in Fig. 7. The CSF signals in our proposed wsup-images with optimal masking were suppressed compared with those in the standard.
Fig. 7
Acquired images and quantitative maps for standard (left) and wsup (right). The CSF signals in our proposed wsup-images with optimal masking (with ThVwmin = 0.05 using TElong = 500 ms) were suppressed, while the tissue signals were almost comparable with those in the standard. Those numerical results are shown in Table 2. CSF, cerebral spinal fluid; TE, echo time; wsup, water suppression.
Acquired images and quantitative maps for standard (left) and wsup (right). The CSF signals in our proposed wsup-images with optimal masking (with ThVwmin = 0.05 using TElong = 500 ms) were suppressed, while the tissue signals were almost comparable with those in the standard. Those numerical results are shown in Table 2. CSF, cerebral spinal fluid; TE, echo time; wsup, water suppression.
Table 2
Numerical volunteer study results of acquired images each with 3 methods of std, wsup w/o, and with masking, where 2 kinds of TE long were used for wsup
Subject
Method
TElong [ms]
SNR
GM–WM CRa
GM–WM CNRa
noiseSD
CSF
Temporal (tissue + CSF)
GM
WM
Acquired SNR, CR, CNR [a.u.]
SE (10000, 20)
std
-
311
221
230
175
0.137
55.6
6.5
wsup_w/o mask
300
-
132
148
117
0.116
30.8
9.2
500
-
140
159
122
0.131
36.8
9.2
wsup_mask
300
-
199
230
175
0.137
55.6
6.5
500
-
203
230
175
0.137
55.6
6.5
SE (10000, 100)
std
-
249
87
93
58
0.234
35.3
6.5
wsup_w/o mask
300
-
42
53
35
0.203
18.1
9.2
500
-
48
62
39
0.226
23.0
9.2
wsup_mask
300
-
69
93
58
0.234
35.3
6.5
500
-
72
93
58
0.234
35.3
6.5
FLAIR (10000, 2700, 100)
std
-
44
58
67
52
0.128
15.1
6.5
Note that the all data in the wsup w/o masking were restored by the wsup with masking comparably as in the standard data, and the data for TElong of 300 ms were greatly improved by masking subtraction. Analysis parameters were: ThVwmax = 1; ThVwmin = 0.1 (TElong = 300 ms), = 0.05 (TElong = 500 ms). GM–WM CR = (SGM − SWM)/(SGM + SWM), GM–WM CNR = (SGM − SWM)/noiseSD, where SGM and SWM are signal intensities of GM and WM. CNR, contrast to noise ratio; CR, contrast ratio; CSF, cerebral spinal flow; GM, gray matter; FLAIR, fluid-attenuation inversion recovery; noiseSD, noise standard deviation; std, standard; TE, echo time; WM, white matter; wsup, water suppression.
Synthetic images of FLAIR and SE each for standard and wsup as a parameter of TE are shown in Fig. 8. The hyperintense artifacts shown in the std-synFLAIR images, especially at the CSF–PVE portions such as the temporal lobe or cerebellum, were almost reduced in the wsup-synFLAIR images with increasing TE. In addition, the CSF signals in the std-synSE images were perfectly reduced in the wsup-synSE images.
Fig. 8
Synthetic images of FLAIR of (TR, TI) = (10,000 ms, 2700 ms) (a) and SE of (TR = 10,000 ms) (b) as a parameter of TE for the standard and wsup. The CSF–PVE artifacts in the std-synFLAIR images were reduced dominantly with increasing TE in the wsup-synFLAIR images. In addition, the CSF signals in the std-synSE images were perfectly reduced in the wsup-synSE images. Here, all images are displayed with the same window level and width. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; PVE, partial volume effect; SE, spin echo; TE, echo time; wsup-syn, water suppression synthetic.
Synthetic images of FLAIR of (TR, TI) = (10,000 ms, 2700 ms) (a) and SE of (TR = 10,000 ms) (b) as a parameter of TE for the standard and wsup. The CSF–PVE artifacts in the std-synFLAIR images were reduced dominantly with increasing TE in the wsup-synFLAIR images. In addition, the CSF signals in the std-synSE images were perfectly reduced in the wsup-synSE images. Here, all images are displayed with the same window level and width. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; PVE, partial volume effect; SE, spin echo; TE, echo time; wsup-syn, water suppression synthetic.Comparison of the acquired FLAIR images, wsup-synFLAIR, and wsup-synSE is shown in Fig. 9. When comparing the wsup-synFLAIR with the acquired FLAIR images, the tissue contrasts were almost equivalent, while the CSF inflow or motion artifacts shown in the acquired FLAIR image were not introduced in the wsup-synFLAIR images.
Fig. 9
Comparison of 3 kinds of wsup images of 4 different slices each with acquired FLAIR (top), wsup-syn FLAIR (middle), and wsup-synSE (bottom). Note that the wsup-synFLAIR provided almost equivalent CSF suppression effects and GM–WM contrasts but no artifacts due to CSF inflow (closed arrow) and those motion (open arrows) from the Monro foramen as shown in the acquired FLAIR. In addition, the wsup-synSE provided better GM–WM contrasts than those in the wsup-synFLAIR. Here, TE = 100 ms and the other parameters of FLAIR and SE were same as in Fig. 8, and all images were displayed with the same window level and width. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; GM, gray matter; SE, spin echo; syn, synthetic; WM, white matter; wsup, water suppression.
Comparison of 3 kinds of wsup images of 4 different slices each with acquired FLAIR (top), wsup-syn FLAIR (middle), and wsup-synSE (bottom). Note that the wsup-synFLAIR provided almost equivalent CSF suppression effects and GM–WM contrasts but no artifacts due to CSF inflow (closed arrow) and those motion (open arrows) from the Monro foramen as shown in the acquired FLAIR. In addition, the wsup-synSE provided better GM–WM contrasts than those in the wsup-synFLAIR. Here, TE = 100 ms and the other parameters of FLAIR and SE were same as in Fig. 8, and all images were displayed with the same window level and width. CSF, cerebral spinal fluid; FLAIR, fluid-attenuation inversion recovery; GM, gray matter; SE, spin echo; syn, synthetic; WM, white matter; wsup, water suppression.Numerical results of acquired are summarized in Table 2 and those quantitative parameters and synthetic images are summarized in Table 3, each with 3 methods of the standard, wsup w/o and with masking, where 2 kinds of TElong were used for wsup. Note that all the tissue SNRs reduced in the wsup w/o masking were perfectly restored by the wsup with masking, comparable to the standard data, and the wsup with masking data using TElong = 300 ms was greatly improved compared with that using TElong = 500 ms.
Table 3
Numerical volunteer study results of quantitative parameter, and synthetic images, each with 3 methods of std, wsup w/o, and with masking, using the acquired data shown in Table 2
Subject
Method
TElong [ms]
CSF
Temporal (tissue + CSF)
GM
WM
GM–WM CRa
Quatitative parameter
PD [a.u.]
std
-
2238
1829
1878
1497
0.101
wsup_w/o mask
300
-
1621
1746
1444
0.086
500
-
1682
1837
1480
0.097
wsup_mask
300
-
1688
1878
1497
0.101
500
-
1709
1878
1497
0.101
T2 [ms]
std
-
987
88
88
72
0.090
wsup_w/o mask
300
-
70
79
67
0.075
500
-
76
86
71
0.087
wsup_mask
300
-
76
88
72
0.090
500
-
78
88
72
0.090
T1 [ms]
std
-
3201
1248
1314
819
0.188
Synthetic S.I, CR [a.u.]
FLAIR (10000, 2700, 100)
std
-
314
424
451
348
0.114
wsup_w/o mask
300
-
334
392
311
0.104
500
-
370
434
337
0.112
wsup_mask
300
-
374
451
348
0.114
500
-
385
451
348
0.114
SE (10000, 100)
std
-
1696
568
606
376
0.189
wsup_w/o mask
300
-
386
490
324
0.169
500
-
443
571
360
0.185
wsup_mask
300
-
451
606
376
0.189
500
-
469
606
376
0.189
GM–WM CR = (SGM − SWM)/(SGM + SWM), where SGM and SWM are signal intensities of GM and WM. CR, contrast ratio; CSF, cerebral spinal flow; GM, gray matter; std, standard; TE, echo time; WM, white matter; wsup, water suppression.
Numerical volunteer study results of acquired images each with 3 methods of std, wsup w/o, and with masking, where 2 kinds of TE long were used for wsupNote that the all data in the wsup w/o masking were restored by the wsup with masking comparably as in the standard data, and the data for TElong of 300 ms were greatly improved by masking subtraction. Analysis parameters were: ThVwmax = 1; ThVwmin = 0.1 (TElong = 300 ms), = 0.05 (TElong = 500 ms). GM–WM CR = (SGM − SWM)/(SGM + SWM), GM–WM CNR = (SGM − SWM)/noiseSD, where SGM and SWM are signal intensities of GM and WM. CNR, contrast to noise ratio; CR, contrast ratio; CSF, cerebral spinal flow; GM, gray matter; FLAIR, fluid-attenuation inversion recovery; noiseSD, noise standard deviation; std, standard; TE, echo time; WM, white matter; wsup, water suppression.Numerical volunteer study results of quantitative parameter, and synthetic images, each with 3 methods of std, wsup w/o, and with masking, using the acquired data shown in Table 2GM–WM CR = (SGM − SWM)/(SGM + SWM), where SGM and SWM are signal intensities of GM and WM. CR, contrast ratio; CSF, cerebral spinal flow; GM, gray matter; std, standard; TE, echo time; WM, white matter; wsup, water suppression.
Discussion
We proposed and assessed a new synthetic MRI technique called T2wsup-synMRI. This technique was based on a simple T2-based wsup using additional long TE data to reduce the CSF–PVE artifacts in synthetic FLAIR images obtained by the conventional synthetic MRI technique. We clarified the cause of those artifacts quantitatively with a simulation and then demonstrated that the CSF–PVE artifacts in quantitative maps and synthetic images of FLAIR, as well as SE obtained by our proposed method, were reduced in the MR volunteer study, especially in longer TE for synthetic images.A limitation of our study was that there were only a few healthy volunteers and a single 3T MRI scanner for the MR study, and the results were processed with the minimum number of acquired data. Below, we discuss our experimental results and the optimization for those parameters, the modification of our proposed technique, alternative techniques, and the future.
Parameter optimization
There are several parameters to be determined in our proposed method, such as TE of water dominant image, TElong, thresholding values for making Vw, ThVwmax, and water mask, ThVwmin.Regarding TElong, it must be selected to be water signal dominant. The tissue signals were almost perfectly reduced below image noises at TElong = 500 ms, but remained at TElong = 300 ms in our healthy volunteer study, as shown in Fig. 6. However, when applied to the long T2 legions for patients even when using TElong = 500 ms, there is a risk of underestimation of tissue signals after subtraction if the tissue signals remain in long TE images. In such a case, the TElong is set to be even longer, or the greater ThVwmin is used at the price of remaining water signals, as discussed below.As for the threshold value, ThVwmax, to normalize Vw map, here we used ThVwmax = 1, but it becomes more robust to noises when set to 0.8–0.9 so as to use averaged SI of pure CSF portions. The method for obtaining scaling, α(TEm) using T2w requires the same TR among all T2WI of different TEs, and the same value can be used so far as the acquisition condition is fixed. In contrast, the method of using SI ratios of the same water portions with ThVwmax can be applied even among different TR data. A long TR is better for long TE images to measure the PDw and T1w correctly or to increase the multiple slices in a manner similar to the standard T2WI, and therefore the latter method is suitable in that case, and the fixed values can also be applied if the acquisition condition is fixed.In terms of the threshold value, ThVwmin was used to decide water Vw rate for subtraction. With increasing ThVwmin, the area including water signals becomes wider and the longer T2 tissue signals become easily to be remained, but the risk of SNR reduction becomes smaller. In our experiments with healthy volunteers (Fig. 6), the additional image noises due to subtraction in the wsup image without masking were almost perfectly avoided in the wsup image, with the optimal ThVwmin value dependent on the TElong. In addition, if the optimum ThVwmin is set to be greater when the TElong is shorter, the signal of the longer T2 tissue is reserved without subtraction. The degree of wsup effects and the ratio of remaining tissue signals is a trade-off. Therefore, the ThVwmin needs to be decided experimentally considering the SNR, imaging conditions including TElong, and clinical requirements for the extent of wsup or T2 values in target tissue or lesion. Once it is decided, the same value can be used in different studies once the TElong, tissue of maximum T2t to be kept, and minimum Vw to be suppressed are decided.
Modification of the proposed method
In our proposed method, the water images of the pure water portions such as ventricles can also be obtained in addition to tissue images. The pure water portions in our wsup quantitative maps become zero or noise. If the correct values are required on pure water portions, the standard quantitative parameter maps should be used. When another parameter to control the ratio of water components, α
(0 ≤ α
≤ 1), is applied and it is varied in real time, it will become clinically useful to evaluate the portion of pure or close-to-pure CSF portions. In addition, the long TE image or that quantitative Vw map is also useful for clinical use such as an assessment for glymphatic systems,[9] which have recently gained attention. The Vw map after masking is better than before masking as a quantitative map because it is free of tissue signals or Rician noise.In addition to the Vw map for the masking to separate tissue and water portions, the standard T2 map was also studied. However, the Vw map was better than the T2 map for separating tissue and water portions into small Vw portions. That may be why the sensitivity for separating water from the voxels is higher for the Vw map than for the T2 map when Vw is small, as shown in our simulation of the standard T2 versus Vw in Fig. 4. When using the long TE image for the Vw map, high SI noises, the selection of pure CSF portions or the signal inhomogeneity due to B1 must be considered. A combination of T2 map with long TE image will be useful to make Vw map furthermore robust against those artifacts.
Comparison with alternative approaches
The problem with the original HIRE method[8] was the SNR reduction in longer T2 lesions due to a simple subtraction of the long TE image of the relatively shorter TE of 240 ms as water-dominant TE. To solve this, a modified method was proposed,[10] where a non-linear conversion of the long TE image was performed before subtraction to suppress the lower SI. However, that method is not suitable for quantitative mapping because the degree of wsup varies depending on the TE or spatial position. In contrast, in our proposed method, as shown in Table 2, the tissue T2 is not varied after our wsup with masking because the same scaling of 0 or 1 by spatial masking then-suitable TE-dependent scalings to correct SIs of long TE images is performed before subtraction.Other alternative approaches to reduce the CSF–PVE artifacts in synMRI are a 3D acquisition[11,12] or a combination with deep learning (DL) technique to use acquired FLAIR image as a teacher.[13,14] However, the PVE artifacts cannot be principally avoided even when using small voxels, and DL is a complicated procedure and the robustness of suppression is unclear. In contrast, our proposed technique is very simple and easily combined with the conventional synthetic MRI technique.More generally, unknown parameters for such a bi-exponential model can be principally obtained by applying a least-square algorithm to the different TE data of multiple points more than the number of unknown parameters; however, this is usually ill-posed problem and difficult to obtain a stable solution under noise contamination. Bayesian approaches[15,16] have succeeded in improving the robustness of stabilities to noises for bi-exponential decay data based on a two-compartment model. In this case, two parameters of (Vw, T2t) can be calculated with fixing T2w in Eq. (3) or Eq. (4) if the number of data points is over 3. Here we used the minimum number of different TE images (M = 3 including long TE) to obtain wsup maps of tissue. However, those approaches require many data points and longer processing time due to the use of a non-linear solution. Better separation is not expected unless longer TElong or some thresholding technique is used to improve the accuracy of separating components of longer T2 tissue and water even when using multiple data points.More recently, a novel method called “MR fingerprinting (MRF),” which enables quantitative maps of T1, T2, and PD to be obtained very fast, was proposed,[17,18] and it is beginning to be applied widely. The MRF uses data obtained by steady-state free precession (FISP) sequence with varying TR, flip-angle (FA), and k-space under-sampling, then matches the data with the dictionary-described relationship between the data and parameters based on the Bloch equation. In addition, a machine-learning approach using DL combined with the MRF technique was proposed[19] and solved the shortcomings of dictionary size and computing load in current MRF techniques. It is generally difficult for the FISP to obtain pure water images compared with the FSE we used. However, in the future, those approaches might solve the multi-parameter model problem,[20] which is the difficulty of current non-linear fitting by combination with the acquisition technique to reflect a wider range of quantitative parameters, including tissue to water components, while keeping quantization resolution.
Conclusion
Our proposed T2wsup-synMRI technique makes it easy to reduce the CSF–PVE artifacts problematic in the synthetic FLAIR images using the current synMRI technique by adding long TE images and simple processing. Although further optimizations in data acquisition and processing techniques are required before actual clinical use, we expect our technique to become clinically useful.
Authors: L N Tanenbaum; A J Tsiouris; A N Johnson; T P Naidich; M C DeLano; E R Melhem; P Quarterman; S X Parameswaran; A Shankaranarayanan; M Goyen; A S Field Journal: AJNR Am J Neuroradiol Date: 2017-04-27 Impact factor: 3.825
Authors: A Hagiwara; Y Otsuka; M Hori; Y Tachibana; K Yokoyama; S Fujita; C Andica; K Kamagata; R Irie; S Koshino; T Maekawa; L Chougar; A Wada; M Y Takemura; N Hattori; S Aoki Journal: AJNR Am J Neuroradiol Date: 2019-01-10 Impact factor: 3.825
Authors: A Hagiwara; M Hori; K Yokoyama; M Y Takemura; C Andica; K K Kumamaru; M Nakazawa; N Takano; H Kawasaki; S Sato; N Hamasaki; A Kunimatsu; S Aoki Journal: AJNR Am J Neuroradiol Date: 2016-10-27 Impact factor: 3.825
Authors: T Granberg; M Uppman; F Hashim; C Cananau; L E Nordin; S Shams; J Berglund; Y Forslin; P Aspelin; S Fredrikson; M Kristoffersen-Wiberg Journal: AJNR Am J Neuroradiol Date: 2016-01-21 Impact factor: 3.825
Authors: Antoine Louveau; Igor Smirnov; Timothy J Keyes; Jacob D Eccles; Sherin J Rouhani; J David Peske; Noel C Derecki; David Castle; James W Mandell; Kevin S Lee; Tajie H Harris; Jonathan Kipnis Journal: Nature Date: 2015-06-01 Impact factor: 49.962
Authors: Dan Ma; Vikas Gulani; Nicole Seiberlich; Kecheng Liu; Jeffrey L Sunshine; Jeffrey L Duerk; Mark A Griswold Journal: Nature Date: 2013-03-14 Impact factor: 49.962