| Literature DB >> 32937785 |
Luca Lamazza1, Marco Lollobrigida1, Iole Vozza1, Luigi Palmieri2, Claudio Stacchi3, Teresa Lombardi4, Alberto De Biase1.
Abstract
Piezoelectric devices are widely used in oral surgical procedures, including implant site preparation. However, little is known about the influence of working movement on temperature elevation in bone. The aim of this study was to assess the effects of two different working cycles on temperature elevation during piezoelectric implant site preparation. Sixty osteotomies at a depth of 10 mm were performed on bone blocks of bovine ribs using a piezoelectric tip with external irrigation (IM1s, Mectron Medical Technology, Carasco, Italy). A mechanical positioning device was used to guarantee reproducible working and measuring conditions. Two different working cycles, of 4 and 6 s, respectively, were tested, including both longitudinal and rotational movements. Temperature was recorded in real time with a fiber optic thermometer and applied pressure was maintained under 150 g. For each test, the highest recorded temperature (Tmax) and the mean temperature recorded from 30 s before to 30 s after the highest recorded temperature (T±30) were extrapolated. Tests duration was also recorded. Both Tmax and T±30 were significantly higher in the '6 s cycles' group than the '4 s cycles' group (42.44 ± 7.3 °C vs. 37.24 ± 4.6 °C, p = 0.002; 37.24 ± 4.6 °C vs. 33.30 ± 3.3 °C, p = 0.003). Test duration was also significantly higher using 6 s cycles compared to 4 s cycles (143.17 ± 29.4 s vs. 119.80 ± 36.4 s, p = 0.002). The results of this study indicate that working cycles of 4 s effectively reduce heat generation and working time during piezoelectric implant site preparation.Entities:
Keywords: heat generation; implant site preparation; piezoelectric surgery; temperature rise; working movements
Year: 2020 PMID: 32937785 PMCID: PMC7560358 DOI: 10.3390/ma13184072
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1Mechanical positioning device used for the study with graphical representation of the transmission tool responsible for the handpiece movements. (a) Load cell, (b) bone sample, (c) piezoelectric handpiece, (d) transmission tool, (e) drill used to create the hole for the thermometer probe, and (f) micrometer screws to obtain three-dimensional movements of both the bone sample and the drill used to create the holes for the probe.
Figure 2Schematic section of the bone sample (a) with the ultrasonic tip (b) and the thermometer probe (c) at completion of osteotomy.
Means of the study parameters.
| 4 s Cycles Group | 6 s Cycles Group | ||||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
|
| 119.80 | 36.4 | 143.17 | 29.4 | 0.002 |
|
| 29.95 | 9.1 | 23.80 | 4.9 | 0.005 |
|
| 37.24 | 4.6 | 42.44 | 7.3 | 0.002 |
|
| 33.30 | 3.3 | 37.01 | 5.2 | 0.003 |
SD, standard deviation. # Mann–Whitney U test.
Medians of the study parameters.
| 4 s Cycles Group | 6 s Cycles Group | ||
|---|---|---|---|
| Median | Median | ||
|
| 108.67 | 142.50 | <0.0001 |
|
| 27.17 | 23.75 | 0.020 |
|
| 37.82 | 40.71 | 0.020 |
|
| 32.94 | 36.32 | 0.020 |
SD, standard deviation. # Median test.
Comparison of variances in the two experimental groups.
| F * | ||
|---|---|---|
|
| 0.13 | 0.72 |
|
| 4.35 | 0.04 |
|
| 2.37 | 0.13 |
|
| 3.48 | 0.07 |
* Equal variances assumed.
Figure 3Box plots with median values for each study parameter. (a) Duration (s): time necessary to perform each osteotomy to a 10 mm depth; (b) Cycles: number of cycles necessary to perform each osteotomy to a 10 mm depth; (c) Tmax (°C): Highest recorded temperature; (d) T±30 (°C): Arithmetic mean of all temperature data points recorded from 30 s before to 30 s after the highest recorded temperature.