Erlangga Yusuf1, Mireille van Westreenen2, Wil Goessens2, Peter Croughs2.
Abstract
Colistin is considered as one of the last-resort antibiotics and reliable antimicrobial susceptibility testing is therefore crucial. The reference standard for AST according to EUCAST and CLSI is broth microdilution (BMD). However, BMD is labor intensive to perform. Commercial antimicrobial susceptibility tests derived from BMD method are available. We investigated the performance of four different commercial tests: Sensititre™, SensiTest™ Colistin, Micronaut MIC Strip Colistin and UMIC Colistin using 70 clinical isolates (half of them was deemed by VITEK2 as resistant), including isolates from cystic fibrosis patients and mcr-1 bearing isolates. We used two reference standards: BMD and composite MIC as determined by all four tests. Sensititre™ had essential agreement (EA, defined as minimum inhibitory concentration within ± 1 dilution) of 87% and 89% compared to BMD and composite reference standard, respectively. For SensiTest™, the EA's were 93% and 90%. For UMIC, 87% and 90%, and for Micronaut, 83% and 84%. All four tests demonstrated categorical agreement (CA) above 90%. CA for SensiTest™ and Micronaut was both 96%, UMIC 94%, and Sensititre™ 93%. All tests were reproducible as tested in two quality control isolates. In conclusion, in clinical isolates from a large referral center, the four commercial tests for determination of colistin minimum inhibitory concentrations showed acceptable performance.Entities:
Keywords: Broth microdilution; Colistin; Cystic fibrosis; Minimum inhibitory concentration; mcr-1
Year: 2020 PMID: 32928253 PMCID: PMC7488655 DOI: 10.1186/s12941-020-00383-x
Source DB: PubMed Journal: Ann Clin Microbiol Antimicrob ISSN: 1476-0711 Impact factor: 3.944
Fig. 1The MIC distribution of the included isolates as determined by the four commercial tests and broth microdilution reference standard
Essential Agreements of commercial MIC tests compared to reference standard Broth Microdilution, n (%)
| Isolates | n | Sensititre™ | SensiTest™ colistin | UMIC colistine | Micronaut MIC strip–colistin | ||||
|---|---|---|---|---|---|---|---|---|---|
| vs. BMD | vs. composite reference | vs. BMD | vs. composite reference | vs. BMD | vs. composite reference | vs. BMD | vs. composite reference | ||
| Total | 70 | 61 (87.1) | 62 (88.6) | 65 (92.8) | 63 (90.0) | 61 (87.1)a | 62 (88.6)a | 58 (82.9) | 59 (84.3) |
| 44 | 39 (88.6) | 38 (86.4) | 42 (95.5) | 40 (90.9) | 41 (93.2) | 38 (86.4) | 40 (90.9) | 38 (86.4) | |
| | 19 | 19 (100) | 19 (100) | 19 (100) | 18 (94.7) | 19 (100) | 19 (100) | 19 (100) | 19 (100) |
| | 14 | 11 (78.6) | 12 (85.7) | 13 (92.9) | 14 (100) | 13 (92.9) | 13 (92.9) | 12 (85.7) | 12 (85.7) |
| | 3 | 1 (33.3) | 3 (100) | 3 (100) | 3 (100) | 3 (100) | 2 (66.7) | 2 (66.7) | 2 (66.7) |
| | 3 | 3 (100) | 1 (33.3) | 3 (100) | 1 (33.3) | 3 (100) | 1 (33.3) | 2 (66.7) | 1 (33.3) |
| | 1 | 1 (100) | 1 (100) | 0 (0) | 1 (100) | 1 (100) | 0 (0) | 1 (100) | 0 (0) |
| | 1 | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 0 (0) | 1 (100) | 1 (100) | 1 (100) |
| | 1 | 1 (100) | 0 (0) | 1 (100) | 1 (100) | 0 (0) | 0 (0) | 1 (100) | 1 (100) |
| 1 | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | |
| 1 | 1 (100) | 0 (0) | 1 (100) | 0 (0) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | |
| 26 | 22 (84.6) | 24 (92.3) | 23 (88.5) | 23 (88.5) | 20 (76.9) | 24 (92.3) | 18 (69.2) | 21 (80.8) | |
| 23 | 20 (86.9) | 22 (95.7) | 20 (87.0) | 20 (87.0) | 18 (78.6) | 22 (95.7) | 17 (73.9) | 20 (87.0) | |
| 2 | 2 (100) | 1 (50) | 2 (100) | 2 (100) | 1 (50)a | 1 (50)a | 1 (50) | 1 (50) | |
| 1 | 0 (0) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 1 (100) | 0 (0) | ||
a No growth counted as disagreement
Categorical Agreements of commercial MIC tests compared to reference standard Broth Microdilution, n (%)
| Categorical agreement, n (%) | Very major error, n (%) | Major error, n (%) | |
|---|---|---|---|
| Sensititre™ | 65 (92.9) | 5 (14.7) | 0 (0) |
| SensiTest™ | 67 (95.7) | 3 (8.8) | 0 (0) |
| UMIC | 66 (94.3) | 2 (5.9) a | 1 (0) a |
| Micronaut | 67 (95.7) | 3 (8.8) | 0 (0) |
| VITEK2 (n = 56) | 47 (83.9) | 3 (8.8) | 6 (10.7) |
a No growth in one isolate
Isolates with VME or ME (by any commercial BMD test) and their corresponding MICs (µg/ml)
| Isolates | BMD reference standarda | Sensititre™ (VME n = 5, ME n = 0) | SensiTest™ (VME n = 3, ME = 0) | UMIC (VME n = 2, ME = 1) | Micronaut (VME n = 3, ME = 0) |
|---|---|---|---|---|---|
| 0.5 | 0.5 | ≤ 0.25 | 4 | 2 | |
| 8 | 1 | 1 | 1 | 1 | |
| 64 | 4 | > 16 | > 64 | 0,5 | |
| 64 | 16 | 1 | 4 | 4 | |
| 4 | 2 | 4 | 4 | 4 | |
| 4 | 2 | 4 | 8 | 4 | |
| 4 | 2 | 4 | 4 | 4 | |
| 16 | 0.25 | 0.5 | 0.25 | 1 |
VME very major error, ME major error
aBMD: broth microdilution