| Literature DB >> 32886714 |
Job G Godino1,2, David Wing1,2, Massimiliano de Zambotti3, Fiona C Baker3, Kara Bagot4, Sarah Inkelis4, Carina Pautz1, Michael Higgins1,2, Jeanne Nichols1,2, Ty Brumback5, Guillaume Chevance1,2, Ian M Colrain3, Kevin Patrick1,2, Susan F Tapert4.
Abstract
PURPOSE: This study sought to assess the performance of the Fitbit Charge HR, a consumer-level multi-sensor activity tracker, to measure physical activity and sleep in children.Entities:
Mesh:
Year: 2020 PMID: 32886714 PMCID: PMC7473549 DOI: 10.1371/journal.pone.0237719
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Laboratory and field tests conducted for comparison of the Fitbit Charge HR and research-grade measures.
| Category | # | Activity | Time | Description |
|---|---|---|---|---|
| 1 | Quiet | 5 minutes | Sitting quietly | |
| 2 | Music | 5 minutes | Listening to music | |
| 3 | Game | 5 minutes | Playing iPad game | |
| 4 | Moderate (0.8W/kg) | 6 minutes | Cycling @ 55+ rpm | |
| 5 | Vigorous (1.2W/kg) | 6 minutes | Cycling @ 55+ rpm | |
| 6 | 3 mph | 6 minutes | Walking on a treadmill | |
| 7 | 4 mph | 6 minutes | Running/fast walking on a treadmill | |
| 8 | 3 mph with backpack | 6 minutes | Walking on a treadmill with backpack weighing 10% of body weight | |
| 9 | Up | 5 | Going up stairs | |
| 10 | Down | 5 | Going down stairs | |
| 11 | Uphill | 200m | Walking up a marked 200m uphill | |
| 12 | Flat | 400m | Walking a marked 400m flat course | |
| 13 | Downhill | 200m | Walking down a marked 200m downhill | |
| 14 | Ladder Drills | 5 minutes | Agility Ladder Drills | |
| Flag/Cones Drills | Shuttle runs and agility drills |
Fig 1Bland-Altman plots for average steps (A), heart rate (B), energy expenditure (kcal) (C) and METs (D) per minute.
The central dotted line represents the Bland-Altman biases; the thin dotted lines represent the Bland-Altman agreement limits.
Fig 2Bland-Altman plots for Total Sleep Time (TST), Sleep Onset Latency (SOL), and Wake After Sleep Onset (WASO).
The central dotted lines represent the Bland-Altman biases; the thin dotted lines represent the Bland-Altman agreement limits. *PSG is considered the gold standard for sleep assessment, therefore, mean bias was plotted against the PSG measurement alone and not the average of measures from PSG and the Fitbit Charge HR.
Fig 3Bland-Altman plots for nocturnal Heart Rate (HR).
The thick dotted lines represent the Bland-Altman bias; the thin dotted lines represent the Bland-Altman upper and lower agreement limits.