Literature DB >> 32869656

Comparison of Insulin Pump Bolus Calculators Reveals Wide Variation in Dose Recommendations.

Jeanne Buchanan1, Jennifer A Zabinsky1, Christine Ferrara-Cook1, Saleh Adi1, Jenise C Wong1.   

Abstract

BACKGROUND: The introduction of insulin pumps with bolus calculators (BCs) has improved glycemic outcomes and quality of life for those with type 1 diabetes. Despite the increased reliance on BCs, the formulas used to derive recommended boluses are not standardized. Our objective was to examine whether recommendations from different pump BCs vary significantly for identical clinical scenarios.
METHODS: Three commercially available insulin pump BCs were programmed with identical settings and then presented with combinations of blood glucose (BG) and carbohydrates (CHOs) to generate a 4-unit bolus. At one- and two-hour time points, while there was insulin-on-board (IOB) present, we simulated various BG and CHO scenarios in order to compare BC-recommended doses.
RESULTS: Differences in suggested doses were noted between BCs, as well as within the same brand. The greatest variation was apparent when BG was below target. Doses suggested by one BC varied depending on whether the IOB resulted from a previous dose given for BG or CHO, while the other two BCs adjusted for total IOB regardless of the source.
CONCLUSIONS: In this simulation study, there were large differences in recommended doses between BCs due to the unique way each manufacturer incorporates IOB into their formulas as well as the pharmacokinetics used to derive the IOB amount. Providers should be aware that identical pump settings will result in a different dose recommendation for each pump brand and advise patients accordingly.

Entities:  

Keywords:  bolus calculator; insulin bolus; insulin pump; insulin pump comparison; type 1 diabetes

Mesh:

Substances:

Year:  2020        PMID: 32869656      PMCID: PMC8655273          DOI: 10.1177/1932296820951855

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  12 in total

1.  Bolus calculator improves long-term metabolic control and reduces glucose variability in pump-treated patients with Type 1 diabetes.

Authors:  G Lepore; A R Dodesini; I Nosari; C Scaranna; A Corsi; R Trevisan
Journal:  Nutr Metab Cardiovasc Dis       Date:  2012-06-04       Impact factor: 4.222

2.  Clinical performance of three bolus calculators in subjects with type 1 diabetes mellitus: a head-to-head-to-head comparison.

Authors:  Howard Zisser; Robin Wagner; Stefan Pleus; Cornelia Haug; Nina Jendrike; Chris Parkin; Matthias Schweitzer; Guido Freckmann
Journal:  Diabetes Technol Ther       Date:  2010-12       Impact factor: 6.118

3.  Relevance of Bolus Calculators in Current Hybrid Closed Loop Systems.

Authors:  Gregory P Forlenza
Journal:  Diabetes Technol Ther       Date:  2017-07-07       Impact factor: 6.118

4.  Important determinants of diabetes control in insulin pump therapy in patients with type 1 diabetes mellitus.

Authors:  Asma Deeb; Samar Abu-Awad; Salima Abood; Mohamed El-Abiary; Jamal Al-Jubeh; Hana Yousef; Laila AbdelRahman; Ahlam Al Hajeri; Huda Mustafa
Journal:  Diabetes Technol Ther       Date:  2014-12-16       Impact factor: 6.118

Review 5.  The effect of using the insulin pump bolus calculator compared to standard insulin dosage calculations in patients with type 1 diabetes mellitus - systematic review.

Authors:  A Ramotowska; D Golicki; K Dżygało; A Szypowska
Journal:  Exp Clin Endocrinol Diabetes       Date:  2013-01-17       Impact factor: 2.949

Review 6.  Bolus calculator: a review of four "smart" insulin pumps.

Authors:  Howard Zisser; Lauren Robinson; Wendy Bevier; Eyal Dassau; Christian Ellingsen; Francis J Doyle; Lois Jovanovic
Journal:  Diabetes Technol Ther       Date:  2008-12       Impact factor: 6.118

7.  Assessment of use of specific features of subcutaneous insulin infusion systems and their relationship to metabolic control in patients with type 1 diabetes.

Authors:  Carmen Quirós; Ioana Patrascioiu; Marga Giménez; Irene Vinagre; Mercè Vidal; Margarita Jansà; Ignacio Conget
Journal:  Endocrinol Nutr       Date:  2014-02-16

Review 8.  7. Diabetes Technology: Standards of Medical Care in Diabetes-2020.

Authors: 
Journal:  Diabetes Care       Date:  2020-01       Impact factor: 19.112

9.  Continuous glucose monitoring and glycemic control among youth with type 1 diabetes: International comparison from the T1D Exchange and DPV Initiative.

Authors:  Daniel J DeSalvo; Kellee M Miller; Julia M Hermann; David M Maahs; Sabine E Hofer; Mark A Clements; Eggert Lilienthal; Jennifer L Sherr; Martin Tauschmann; Reinhard W Holl
Journal:  Pediatr Diabetes       Date:  2018-07-01       Impact factor: 4.866

Review 10.  Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range.

Authors:  Tadej Battelino; Thomas Danne; Richard M Bergenstal; Stephanie A Amiel; Roy Beck; Torben Biester; Emanuele Bosi; Bruce A Buckingham; William T Cefalu; Kelly L Close; Claudio Cobelli; Eyal Dassau; J Hans DeVries; Kim C Donaghue; Klemen Dovc; Francis J Doyle; Satish Garg; George Grunberger; Simon Heller; Lutz Heinemann; Irl B Hirsch; Roman Hovorka; Weiping Jia; Olga Kordonouri; Boris Kovatchev; Aaron Kowalski; Lori Laffel; Brian Levine; Alexander Mayorov; Chantal Mathieu; Helen R Murphy; Revital Nimri; Kirsten Nørgaard; Christopher G Parkin; Eric Renard; David Rodbard; Banshi Saboo; Desmond Schatz; Keaton Stoner; Tatsuiko Urakami; Stuart A Weinzimer; Moshe Phillip
Journal:  Diabetes Care       Date:  2019-06-08       Impact factor: 19.112

View more
  3 in total

1.  An In Silico Head-to-Head Comparison of the Do-It-Yourself Artificial Pancreas Loop and Bio-Inspired Artificial Pancreas Control Algorithms.

Authors:  Ryan Armiger; Monika Reddy; Nick S Oliver; Pantelis Georgiou; Pau Herrero
Journal:  J Diabetes Sci Technol       Date:  2021-12-03

2.  Automated Bolus Calculators and Connected Insulin Pens: A Smart Combination for Multiple Daily Injection Insulin Therapy.

Authors:  David T Ahn
Journal:  J Diabetes Sci Technol       Date:  2021-12-22

3.  Continuous Glucose Monitors and Automated Insulin Dosing Systems in the Hospital Consensus Guideline.

Authors:  Rodolfo J Galindo; Guillermo E Umpierrez; Robert J Rushakoff; Ananda Basu; Suzanne Lohnes; James H Nichols; Elias K Spanakis; Juan Espinoza; Nadine E Palermo; Dessa Garnett Awadjie; Leigh Bak; Bruce Buckingham; Curtiss B Cook; Guido Freckmann; Lutz Heinemann; Roman Hovorka; Nestoras Mathioudakis; Tonya Newman; David N O'Neal; Michaela Rickert; David B Sacks; Jane Jeffrie Seley; Amisha Wallia; Trisha Shang; Jennifer Y Zhang; Julia Han; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2020-09-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.