| Literature DB >> 32825468 |
Josune J Ezenarro1,2, Noemí Párraga-Niño3, Miquel Sabrià3, Fancisco Javier Del Campo4,5, Francesc-Xavier Muñoz-Pascual4, Jordi Mas1, Naroa Uria4.
Abstract
Legionella is a pathogenic bacterium, ubiquitous in freshwater environments and able to colonise man-made water systems from which it can be transmitted to humans during outbreaks. The prevention of such outbreaks requires a fast, low cost, automated and often portable detection system. In this work, we present a combination of sample concentration, immunoassay detection, and measurement by chronoamperometry. A nitrocellulose microfiltration membrane is used as support for both the water sample concentration and the Legionella immunodetection. The horseradish peroxidase enzymatic label of the antibodies permits using the redox substrate 3,3',5,5'-Tetramethylbenzidine to generate current changes proportional to the bacterial concentration present in drinking water. Carbon screen-printed electrodes are employed in the chronoamperometric measurements. Our system reduces the detection time: from the 10 days required by the conventional culture-based methods, to 2-3 h, which could be crucial to avoid outbreaks. Additionally, the system shows a linear response (R2 value of 0.99), being able to detect a range of Legionella concentrations between 101 and 104 cfu·mL-1 with a detection limit (LoD) of 4 cfu·mL-1.Entities:
Keywords: Legionella pneumophilla; amperometry; immunodetection; preconcentration
Mesh:
Substances:
Year: 2020 PMID: 32825468 PMCID: PMC7558583 DOI: 10.3390/bios10090102
Source DB: PubMed Journal: Biosensors (Basel) ISSN: 2079-6374
Figure 1General scheme. The membrane retains the Legionella cells for the subsequent immunoassay and the chronoamperometric transduction of the signal.
Figure 2(A) Cyclic voltammetry of the substrate 3,3′,5,5′-Tetramethylbenzidine (TMB) in a screen-printed electrode (SPE). (B) Cyclic voltammetry of the substrate 3,3′,5,5′-Tetramethylbenzidine (TMB) and horseradish peroxidase (HRP) reaction. Peak potentials are highlighted in red (oxidation) and blue (reduction).
Figure 3(A) Current change response in time, measured for different Legionella concentrations. (B) Current changes measured for different Legionella concentrations at different times (from 0 to 240 s). Standard error bars correspond to the measurements made in three different culture replicates of each concentration measured in two different assays (n = 6).
Figure 4Sensor calibration curve, where the current obtained at a constant potential of 50 mV after 50 s of the chronoamperometry is expressed in absolute values as a function of the logarithm of the increasing concentrations of Legionella from 0 (blank) to 105 cfu·mL−1. The regression line is indicated in blue and error bars represent the standard error (n = 6).