Yoshikazu Yamaguchi1,2,3, Tetsuya Miyashita1, Yuko Matsuda1, Makoto Sasaki1, Shunsuke Takaki1, Stephani S Kim2, Joseph D Tobias2,3, Takahisa Goto1.
Abstract
BACKGROUND: Precise control of tidal volume is one of the keys in limiting ventilator-induced lung injury and ensuring adequate ventilation in mechanically ventilated neonates. The aim of the study was to compare the tidal volume (mVT) measured from the expiratory limb of the ventilator with the actual tidal volume (aVT) that would be delivered to the patient using a lung model to simulate a neonate.Entities:
Keywords: flow sensor; lung protective ventilation; mechanical ventilation; pediatric anesthesia; tidal volume; volutrauma
Year: 2020 PMID: 32765126 PMCID: PMC7367738 DOI: 10.2147/MDER.S259760
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Lung Simulator Settings
| Group | Resistance (cmH2O/L/sec) | Compliance (mL/cmH2O) |
|---|---|---|
| 1 | 50 | 2 |
| 2 | 100 | 1 |
| 3 | 150 | 0.5 |
Figure 1Diagram of the set-up for this study. The ASL 5000 was connected to the computer which recorded the data. Ventilators were connected to ASL 5000 using a standard pediatric circuit. Set tidal volume (sVT) was 30 mL. Actually delivered tidal volume (aVT) was measured by the syringe in ASL5000. Ventilator displayed tidal volume (mVT) was measured by the flow sensor in expiratory limb of the ventilator. aVT = delivered tidal volume measured by the ASL 5000 lung model; mVT = ventilator displayed or measured tidal volume; sVT = ventilator set tidal volume of 30 mL.
Ventilator Displayed or Measured Tidal Volume and Actually Delivered Tidal Volume
| Group 1 | Group 2 | Group 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|
| aVT | mVT | P value | aVT | mVT | P value | aVT | mVT | P value | |
| Fabius GS VCV | 26.6 ± 0.56 | 23.0 ± 0.58 | <0.001 | 26.7 ± 0.18 | 15.6 ± 0.9 | <0.001 | 25.7 ± 0.49 | 17.6 ± 3.9 | <0.001 |
| Servo-I, VCV | 31.4 ± 0.12 | 29.4 ± 0.53 | <0.001 | 30.1 ± 0.13 | 27.7 ± 0.49 | <0.001 | 29.2 ± 0.26 | 26 ± 0.0 | <0.001 |
| Servo-I, DCV | 30.1 ± 0.17 | 26.6 ± 0.53 | <0.001 | 28.9 ± 0.24 | 24.9 ± 0.38 | <0.001 | 27.4 ± 0.14 | 23.5 ± 0.55 | <0.001 |
| Infinity V500, VCV | 30.2 ± 0.13 | 29.9 ± 0.9 | 0.28 | 30.1 ± 0.12 | 30.0 ± 0.0 | 0.04 | 28.9 ± 0.08 | 30.0 ± 0 | <0.001 |
| Infinity V500, DCV | 30.6 ± 0.13 | 30.0 ± 0.0 | <0.001 | 30.9 ± 0.18 | 30.0 ± 0.0 | <0.001 | 30.6 ± 0.07 | 30.0± 0.0 | <0.001 |
Notes: For all studies, the set tidal volume was 30 mL. aVT = actual delivered tidal volume measured by the ASL 5000 lung model; mVT = ventilator displayed or measured tidal volume.
Abbreviations: DCV, dual-controlled ventilation; VCV, volume-controlled ventilation.
Difference Between Actual Delivered Tidal Volume (aVT) and Set Tidal Volume in Each Setting (sVT)
| Ventilator | Group | Mean ± SD (mL) | Fabius GS | Servo-I | Servo-I | Infinity | Infinity |
|---|---|---|---|---|---|---|---|
| Fabius GS, VCV | Group1 | −3.4 ± 0.56 | |||||
| Servo-i,VCV | Group1 | 1.4 ± 0.12 | <0.001 | ||||
| Servo-i,DCV | Group1 | 0.14 ± 0.17 | <0.001 | <0.001 | |||
| Infinity | Group1 | 0.24 ± 0.13 | <0.001 | <0.001 | 0.96 | ||
| Infinity | Group1 | 0.59 ± 0.13 | <0.001 | <0.001 | 0.0445 | 0.1776 |
Notes: aVT = delivered tidal volume measured by the ASL 5000 lung model; sVT = ventilator set tidal volume of 30 mL.
Abbreviations: DCV, dual-controlled ventilation; VCV, volume-controlled ventilation.
Figure 2Actual delivered tidal volume measured by the ASL 5000 (aVT) and measured or ventilator displayed tidal volume (mVT) in each setting. Mean value and standard deviation (error bar) are shown. P values are shown above the bar graph. (A) Fabius GS, VCV; (B) Servo-i, VCV; (C) Servo-i, DCV; (D) Infinity V 500, VCV; and (E) Infinity V 500, DCV. aVT = delivered tidal volume measured by the ASL 5000 lung model; mVT = ventilator displayed or measured tidal volume.
Abbreviations: DCV, dual-controlled ventilation; VCV, volume-controlled ventilation.
Difference Between Actual Delivered Tidal Volume (aVT) and Measured Tidal Volume (mVT)
| Ventilator | Group | Mean ± SD (mL) | Fabius GS VCV | Servo-I | Servo-I | Infinity | Infinity |
|---|---|---|---|---|---|---|---|
| Fabius GS | Group1 | 3.6 ± 0.89 | |||||
| Servo-i | Group1 | 2.0 ± 0.55 | <0.001 | ||||
| Servo-i | Group1 | 3.6 ± 0.61 | 1.00 | <0.001 | |||
| Infinity | Group1 | 0.38 ± 0.86 | <0.001 | 0.001 | <0.001 | ||
| Infinity | Group1 | 0.59 ± 0.14 | <0.001 | 0.0043 | <0.001 | 0.98 |
Notes: For all studies, the set tidal volume was 30 mL. aVT = delivered tidal volume measured by the ASL 5000 lung model; mVT = ventilator displayed or measured tidal volume.
Abbreviations: DCV, dual-controlled ventilation; VCV, volume-controlled ventilation.