Adult ICU ventilators to provide neonatal ventilation: a lung simulator study.

Andrew D Marchese1, Daniel Chipman, Pedro de la Oliva, Robert M Kacmarek.   

Abstract

BACKGROUND: Traditionally, specific ventilators have been manufactured to only provide neonatal mechanical ventilation. However, many of the current generation of ICU ventilators also include a neonatal mode.
METHODS: Using the IngMar ASL5000 lung simulator the Puritan Bennett 840, the Maquet Servo i, the Viasys AVEA, the GE Engström, the Drager Evita XL and Babylog 8000 Plus were evaluated during assisted ventilation in the pressure assist/control mode. Three lung mechanics were set: resistance 50 cmH(2)O/L/s, compliance 2 mL/cmH(2)O; resistance 100 cmH(2)O/L/s, compliance 1 mL/cmH(2)O; and resistance 150 cmH(2)O/L/s, compliance 0.5 mL/cmH(2)O. A maximum negative pressure drop of 4 and 7 cmH(2)O was achieved during simulated inspirations. Each ventilator was evaluated with PEEP 5 cmH(2)O, peak pressure 20 cmH(2)O and inspiratory time 0.3 s and with PEEP 10 cmH(2)O, peak pressure 30 cmH(2)O and inspiratory time 0.4 s. Each ventilator setting was then repeated with a leak of 0.3 L/min at a constant pressure of 5 cmH(2)O.
RESULTS: Overall each of the 5 ICU ventilators responded faster or greater than the Babylog with respect to: pressure to trigger (except the Servo i), time to trigger (except the Evita XL), time between trigger and return of pressure to baseline, time from start of breath to 90% of peak pressure (except the Avea) and pressure time product of breath activation. Expiratory tidal volume was also greater with all ICU ventilators except the Avea. Variation in mechanics, leak, PEEP and muscular effort had little effect on these differences.
CONCLUSION: All ICU ventilators tested were able to at least equal the performance of the Babylog 8000 Plus on all variables evaluated.

Mesh:

Year:  2008        PMID: 18931831     DOI: 10.1007/s00134-008-1332-0

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  8 in total

1.  Performance of neonatal ventilators in volume targeted ventilation mode.

Authors:  Atul Sharma; Anthony D Milner; Anne Greenough
Journal:  Acta Paediatr       Date:  2007-02       Impact factor: 2.299

2.  A new pediatric volume ventilator.

Authors:  R R Kirby; E J Robison; J Schulz; R DeLemos
Journal:  Anesth Analg       Date:  1971 Jul-Aug       Impact factor: 5.108

3.  Patient-triggered ventilation in neonates: comparison of a flow-and an impedance-triggered system.

Authors:  H D Hummler; T Gerhardt; A Gonzalez; J Bolivar; N Claure; R Everett; E Bancalari
Journal:  Am J Respir Crit Care Med       Date:  1996-10       Impact factor: 21.405

4.  Performance of neonatal ventilators: the effects of changes in resistance and compliance.

Authors:  G Simbruner; G A Gregory
Journal:  Crit Care Med       Date:  1981-07       Impact factor: 7.598

5.  The response of flow-triggered infant ventilators.

Authors:  M Nishimura; D Hess; R M Kacmarek
Journal:  Am J Respir Crit Care Med       Date:  1995-12       Impact factor: 21.405

Review 6.  Lung function tests in neonates and infants with chronic lung disease: lung and chest-wall mechanics.

Authors:  Monika Gappa; J Jane Pillow; Julian Allen; Oscar Mayer; Janet Stocks
Journal:  Pediatr Pulmonol       Date:  2006-04

7.  Airway leak size in neonates and autocycling of three flow-triggered ventilators.

Authors:  G Bernstein; E Knodel; G P Heldt
Journal:  Crit Care Med       Date:  1995-10       Impact factor: 7.598

8.  Effort and work of breathing in neonates during assisted patient-triggered ventilation.

Authors:  M Kapasi; Y Fujino; M Kirmse; E A Catlin; R M Kacmarek
Journal:  Pediatr Crit Care Med       Date:  2001-01       Impact factor: 3.624

  8 in total
  7 in total

1.  Comment on "Adult ICU ventilators to provide neonatal ventilation: a lung simulator study".

Authors:  Matteo Fontana; Antoine Payot; Sylvain Morneau; Philippe Jouvet
Journal:  Intensive Care Med       Date:  2009-04-02       Impact factor: 17.440

2.  ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers.

Authors:  Jean Christophe Marie Richard; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2009-07-29       Impact factor: 17.440

3.  [Monitoring tidal volumes when using the Ventrain® emergency ventilator].

Authors:  A R Schmidt; K Ruetzler; T Haas; A Schmitz; M Weiss
Journal:  Anaesthesist       Date:  2016-05-31       Impact factor: 1.041

4.  The Difference Between Set and Delivered Tidal Volume: A Lung Simulation Study.

Authors:  Yoshikazu Yamaguchi; Tetsuya Miyashita; Yuko Matsuda; Makoto Sasaki; Shunsuke Takaki; Stephani S Kim; Joseph D Tobias; Takahisa Goto
Journal:  Med Devices (Auckl)       Date:  2020-07-12

5.  Healthcare Technology Management (HTM) of mechanical ventilators by clinical engineers.

Authors:  Jun Yoshioka; Masaki Nakane; Kaneyuki Kawamae
Journal:  J Intensive Care       Date:  2014-04-15

6.  Simulations for mechanical ventilation in children: review and future prospects.

Authors:  Olivier Flechelles; Annie Ho; Patrice Hernert; Guillaume Emeriaud; Nesrine Zaglam; Farida Cheriet; Philippe A Jouvet
Journal:  Crit Care Res Pract       Date:  2013-03-07

7.  Additional work of breathing from trigger errors in mechanically ventilated children.

Authors:  Robert G T Blokpoel; Alette A Koopman; Jefta van Dijk; Martin C J Kneyber
Journal:  Respir Res       Date:  2020-11-10
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.