Literature DB >> 12441771

Accuracy of displayed values of tidal volume in the pediatric intensive care unit.

Rosemary A Castle1, Catherine J Dunne, Quen Mok, Angie M Wade, Janet Stocks.   

Abstract

OBJECTIVES: To assess the accuracy of the expired tidal volumes (VT(E)) displayed by one of the most frequently used ventilators that measures exhaled volume at the expiratory valve.
DESIGN: Prospective study.
SETTING: The intensive care units of a pediatric tertiary referral center in London, UK. PATIENTS: A total of 56 intubated children aged between 3 wks and 16.6 yrs who were clinically stable and ventilated with a Servo 300 ventilator.
INTERVENTIONS: The CO2SMO Plus respiratory monitor, which measures flow at the airway opening, was validated using calibrated syringes and appropriate tracheal tubes and connections. Simultaneous in vivo recordings of VT(E) from the Servo 300 and CO2SMO Plus were compared before (displayed Servo VT(E)) and after (effective Servo VT(E)) compensating for ventilator circuit compliance.
MEASUREMENTS AND MAIN RESULTS: The in vitro accuracy of the CO2SMO Plus was within +/-5% over a wide range of volumes and measurement conditions. The displayed Servo 300 VT(E) overestimated the true VT(E) by between 2% and 91%. The magnitude of error varied within and between children, according to pressure change (peak inspiratory pressure minus positive end-expiratory pressure), VT(E), and circuit size. Mean (sd) error was 32% (20%) in 40 children with displayed Servo VT(E) of <160 mL and 18% (6%) in 16 subjects with displayed Servo VT(E) of >/=160 mL. After correcting for gas compression, effective VT(E) from the Servo 300 underestimated the true VT(E) by up to 64% in the smallest infants but continued to overestimate by as much as 29% in older children.
CONCLUSIONS: The accuracy of tidal volume values is crucially dependent on the site of measurement. Unless measured at the airway opening, displayed values are an inconsistent and misleading indicator of the true volumes delivered.

Entities:  

Mesh:

Year:  2002        PMID: 12441771     DOI: 10.1097/00003246-200211000-00027

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  14 in total

1.  Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation.

Authors:  Aissam Lyazidi; Arnaud W Thille; Guillaume Carteaux; Fabrice Galia; Laurent Brochard; Jean-Christophe M Richard
Journal:  Intensive Care Med       Date:  2010-09-23       Impact factor: 17.440

2.  Reliability of measured tidal volume in mechanically ventilated young pigs with normal lungs.

Authors:  Mark J Heulitt; Shirley J Holt; Tracy L Thurman; Renée A Hall; Chan-Hee Jo; Pippa Simpson
Journal:  Intensive Care Med       Date:  2005-07-19       Impact factor: 17.440

3.  Comparison of volumetric capnography and mixed expired gas methods to calculate physiological dead space in mechanically ventilated ICU patients.

Authors:  Pratik Sinha; Neil Soni
Journal:  Intensive Care Med       Date:  2012-08-15       Impact factor: 17.440

4.  Is volume and leak monitoring feasible during nasopharyngeal continuous positive airway pressure in neonates?

Authors:  Hendrik S Fischer; Charles C Roehr; Hans Proquitté; Hannes Hammer; Roland R Wauer; Gerd Schmalisch
Journal:  Intensive Care Med       Date:  2009-11       Impact factor: 17.440

5.  The influence of physiotherapy and suction on respiratory deadspace in ventilated children.

Authors:  Eleanor Main; Janet Stocks
Journal:  Intensive Care Med       Date:  2004-05-08       Impact factor: 17.440

6.  Reproducibility of the respiratory dead space measurements in mechanically ventilated children using the CO2SMO monitor.

Authors:  Y Riou; F Leclerc; V Neve; L Dupuy; O Noizet; S Leteurtre; A Sadik
Journal:  Intensive Care Med       Date:  2004-05-11       Impact factor: 17.440

7.  Ventilator assessment of respiratory mechanics in paediatric intensive care.

Authors:  Gopinathannair Harikumar; Anne Greenough; Gerrard F Rafferty
Journal:  Eur J Pediatr       Date:  2007-03-30       Impact factor: 3.183

Review 8.  An Official American Thoracic Society/European Respiratory Society Workshop Report: Evaluation of Respiratory Mechanics and Function in the Pediatric and Neonatal Intensive Care Units.

Authors:  Stacey Peterson-Carmichael; Paul C Seddon; Ira M Cheifetz; Inéz Frerichs; Graham L Hall; Jürg Hammer; Zoltán Hantos; Anton H van Kaam; Cindy T McEvoy; Christopher J L Newth; J Jane Pillow; Gerrard F Rafferty; Margaret Rosenfeld; Janet Stocks; Sarath C Ranganathan
Journal:  Ann Am Thorac Soc       Date:  2016-02

9.  Effect of lung compliance and endotracheal tube leakage on measurement of tidal volume.

Authors:  Sami I Al-Majed; John E Thompson; Kenneth F Watson; Adrienne G Randolph
Journal:  Crit Care       Date:  2004-10-06       Impact factor: 9.097

10.  Accuracy of currently available neonatal respiratory function monitors for neonatal resuscitation.

Authors:  Charlotte Verbeek; Henriëtte A van Zanten; Jeroen J van Vonderen; Marcus J Kitchen; Stuart B Hooper; Arjan B Te Pas
Journal:  Eur J Pediatr       Date:  2016-06-08       Impact factor: 3.183

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.