Literature DB >> 32670906

Irradiance of Different Curing Modes of Common Light Cure Devices: An In Vitro Study.

Hani M Nassar1, Mahmoud Almutairi2, Albaraa Makhdom2.   

Abstract

AIM: The aim of this study was to test the irradiance values of different curing modes of commonly available light cure devices (LCDs).
MATERIALS AND METHODS: An in vitro investigation was carried out to compare the irradiance output of 10 brands of LCDs available in Saudi Arabia measured using a digital radiometer. Values were recorded for three time points when applicable (0, 10, and 20s). This technique was repeated five times for each LCD. Normal, high-intensity, and soft-start modes were evaluated for all brands with the features available. Irradiance values between brands were analyzed using one-way analysis of variance followed by Bonferroni method. Changes in irradiance between different time points were analyzed using one sample t test for normal and high-intensity modes and using paired t test for soft-start mode. All comparisons were carried out at 0.05 significance level.
RESULTS: The highest values were reported for Ortholux Luminous, Elipar DeepCure-S, Elipar DeepCure, and KaVo mini-LED with values above 1000 mW/cm2. All LCDs showed values above 600 mW/cm2. Three LCDs had high-intensity mode and only one device had soft-start mode. Changes over the different time points were not statistically significant exept for soft-start mode.
CONCLUSION: All tested LCDs had irradiance values sufficient for adequate polymerization of resin composite. Only four of these are capable of curing bulk-fill composites. Copyright:
© 2020 Journal of International Society of Preventive and Community Dentistry.

Entities:  

Keywords:  Curing modes; irradiance; light cure

Year:  2020        PMID: 32670906      PMCID: PMC7339987          DOI: 10.4103/jispcd.JISPCD_496_19

Source DB:  PubMed          Journal:  J Int Soc Prev Community Dent        ISSN: 2231-0762


INTRODUCTION

Resin composites are used widely in the dental field, due to their excellent aesthetic and physical properties.[123] Resin composite formulations have advanced drastically to cope with clinical applications and patients’ demands.[4] The selection of a good light cure device (LCD) is an essential step for curing composite restoration.[35,6] The polymerization of composite particles from monomer to polymer is initiated by photoinitiator activation; light cure quality will affect the degree of polymerization reaction. The number of photons generated by LCDs affect the light intensity and subsequent polymerization reaction.[7] The success and longevity of resin composite restorations can be affected by the quality of light generated by LCDs. A common metric to depict this is the irradiance value (measured in mW/cm2), which is described as the amount of light reaching a particular surface area of composite.[5] In general, irradiance values of <300 mW/cm2 are not recommended to cure dental composites.[8] Irradiance is affected by factors such as radiant power of the LCD and distance from composite surface.[9] The longevity of the resin composite restorations can be affected by LCD performance owing to their irradiance values.[10] This is especially true with the newly introduced bulk-fill composite materials that have different chemistry and photoinitiator formulations requiring greater irradiance values (>1000 mW/cm2) to achieve adequate polymerization.[11] Further, the polymerization reaction of resin composites can be affected by the different modes of light curing as these modes alter the light output of the device. In the traditional mode, a burst of light with high irradiance values will lead to fast polymerization and possibly high polymerization shrinkage stress. To minimize this problem, gradual polymerization mode is found that could potentially help in reducing the formation of marginal gap.[4] High-intensity mode is sometimes used in orthodontic application where a relatively thin amount of resin is used to bond brackets. Still, irradiance values of different light-curing modes are of a great importance to the practicing dentist. Therefore, the aim of this study was to measure the irradiance values of different brands of LCDs and their curing modes.

MATERIALS AND METHODS

STUDY DESIGN

This investigation was an in vitro cross-sectional study involving 10 LCDs available in the Saudi Arabian market. The irradiance values of different curing modes were recorded.

SAMPLING CRITERIA

Light irradiance of 10 light-emitting diode (LED) LCDs available in the Saudi market [Figure 1], Elipar DeepCure (3M, St. Paul, Minnesota), Elipar DeepCure-S (3M), Ortholux Luminous (3M), Bluephase N M (Ivoclar Vivadent, Schaan, Liechtenstein), VALO Cordless (Ultradent Products, South Jordon, Utah), Demi Ultra (KaVo Kerr, Orange, California), Dr’s Light Clever (Good Doctors, Incheon, Republic of Korea), KaVo mini-LED (KaVo Dental GmbH, Bismarckring, Germany), DTE-iLED (Woodpecker, Guilin, Guangxi, China), and FlashMax P3 (Ragnagade 7, Copenhagen, Denmark), was measured using a digital radiometer (Bluephase Meter II, Ivoclar Vivadent).
Figure 1

The different light cure devices included in the study. (A) Ivoclar Vivadent Bluephase N M. (B) 3M Elipar DeepCure-S. (C) Ultradent VALO Cordless. (D) 3M Ortholux Luminous. (E) Dr’s Light Clever. (F) DTE-iLED. (G) FlashMax P3. (H) 3M Elipar DeepCure-L. (I) KaVo mini-LED. (J) Kerr Demi Ultra

The different light cure devices included in the study. (A) Ivoclar Vivadent Bluephase N M. (B) 3M Elipar DeepCure-S. (C) Ultradent VALO Cordless. (D) 3M Ortholux Luminous. (E) Dr’s Light Clever. (F) DTE-iLED. (G) FlashMax P3. (H) 3M Elipar DeepCure-L. (I) KaVo mini-LED. (J) Kerr Demi Ultra

OBSERVATIONAL PARAMETERS

The tip of the LCD was measured using the radiometer’s built in gauge and the value was plugged into the radiometer. The curing tip was laid flat on the radiometer sensor and the device was operated for 20s. Irradiance values were recorded for 0, 10, and 20s. Some devices did not support continuous curing for 20s; only values for 0 and 10s were recorded for those devices. This technique was repeated five times for each LCD and an average value was calculated. A similar approach was used for LCDs with soft-start mode. For high-intensity mode, irradiance value after 3s was reported.

STATISTICAL ANALYSIS

Data were collected, tabulated, and analyzed using Statistical Package for the Social Sciences software program, version 17.0 (IBM, Armonk, New York). Quantitative variables were described using the mean, standard deviation, and 95% confidence interval. For irradiance data among different brands, one-way analysis of variance (ANOVA) was used to test differences in mean values followed by Bonferroni multiple comparison method to determine which devices are different from each other. Changes of irradiance values at different time points within the same brand were analyzed using one-sample t test for normal and high-intensity modes and using paired t test for soft-start mode. All comparisons were carried out at 0.05 significance level.

RESULTS

Overall, all LCDs tested registered irradiance values above the threshold of 600 mW/cm2 [Figure 2]. However, only four of these were associated with irradiance values >1000 mW/cm2 required to adequately polymerize bulk-fill formulation. Statistical testing of mean irradiance values for normal mode showed statistically significant differences among the tested brands [Tables 1 and 2; P < 0.001].
Figure 2

Mean values of light intensities with normal mode of various light cure unit brands. Error bars indicate standard deviation. Devices with the same letter are not statistically different (P > 0.05)

Table 1

One-way analysis of variance: tests between subjects effects for normal curing mode

Statistical testSum of squaresdfMean squareFP value
Between groups2198061.38274757.7160.0624280.000000
Within groups113293.33661716.566
Total2311354.774

df = degrees of freedom

Dependent variable: light intensity mW/cm2

Table 2

Bonferroni test: tests whose devices are different for normal curing mode

Light cure device 1Light cure device 2Mean differenceSEP valueSignificance
Ivoclar Bluephase N MDTE-iLED110.6721.400.000084Yes
Ivoclar Bluephase N M3M Elipar DeepCure-S–225.3321.400.000000Yes
Ivoclar Bluephase N M3M Elipar DeepCure–251.3321.400.000000Yes
Ivoclar Bluephase N MDr’s Light Clever144.6721.400.000000Yes
Ivoclar Bluephase N MUltradent VALO Cordless 102.6721.400.000342Yes
Ivoclar Bluephase N MKaVo mini-LED–255.3315.130.000000Yes
Ivoclar Bluephase N MKerr Demi Ultra56.6715.130.013689Yes
Ivoclar Bluephase N M3M Ortholux Luminous–399.3321.400.000000Yes
DTE-iLED3M Elipar DeepCure-S–336.0026.200.000000Yes
DTE-iLED3M Elipar DeepCure–362.0026.200.000000Yes
DTE-iLEDDr’s Light Clever34.0026.201.000000No
DTE-iLEDUltradent VALO Cordless –8.0026.201.000000No
DTE-iLEDKaVo mini-LED–366.0021.400.000000Yes
DTE-iLEDKerr Demi Ultra–54.0021.400.504681No
DTE-iLED3M Ortholux Luminous–510.0026.200.000000Yes
3M Elipar DeepCure-S3M Elipar DeepCure–26.0026.201.000000No
3M Elipar DeepCure-SDr’s Light Clever370.0026.200.000000Yes
3M Elipar DeepCure-SUltradent VALO Cordless 328.0026.200.000000Yes
3M Elipar DeepCure-SKaVo mini-LED–30.0021.401.000000No
3M Elipar DeepCure-SKerr Demi Ultra282.0021.400.000000Yes
3M Elipar DeepCure-S3M Ortholux Luminous–174.0026.200.000000Yes
3M Elipar DeepCureDr’s Light Clever396.0026.200.000000Yes
3M Elipar DeepCureUltradent VALO Cordless 354.0026.200.000000Yes
3M Elipar DeepCureKaVo mini-LED–4.0021.401.000000No
3M Elipar DeepCureKerr Demi Ultra308.0021.400.000000Yes
3M Elipar DeepCure3M Ortholux Luminous–148.0026.200.000013Yes
Dr’s Light CleverUltradent VALO Cordless –42.0026.201.000000No
Dr’s Light CleverKaVo mini-LED–400.0021.400.000000Yes
Dr’s Light CleverKerr Demi Ultra–88.0021.400.003976Yes
Dr’s Light Clever3M Ortholux Luminous–544.0026.200.000000Yes
Ultradent VALO Cordless KaVo mini-LED–358.0021.400.000000Yes
Ultradent VALO Cordless Kerr Demi Ultra–46.0021.401.000000No
Ultradent VALO Cordless 3M Ortholux Luminous–502.0026.200.000000Yes
KaVo mini-LEDKerr Demi Ultra312.0015.130.000000Yes
KaVo mini-LED3M Ortholux Luminous–144.0021.400.000000Yes
Kerr Demi Ultra3M Ortholux Luminous–456.0021.400.000000Yes

SE = standard error

Dependent variable: light intensity mW/cm2

Mean values of light intensities with normal mode of various light cure unit brands. Error bars indicate standard deviation. Devices with the same letter are not statistically different (P > 0.05) One-way analysis of variance: tests between subjects effects for normal curing mode df = degrees of freedom Dependent variable: light intensity mW/cm2 Bonferroni test: tests whose devices are different for normal curing mode SE = standard error Dependent variable: light intensity mW/cm2 Irradiance values can be categorized into four groups: Group A with irradiance values between 600 and 800 mW/cm2 that includes Dr’s Light Clever, DTE-iLED, VALO Cordless, and Demi Ultra; Group B with values between 800 and 1000 mW/cm2 that includes Bluephase N M; Group C with values between 1000 and 1200 mW/cm2 that includes Elipar DeepCure-S, Elipar DeepCure, and KaVo mini-LED; Group D with values >1200 mW/cm2 that includes Ortholux Luminous. Only three LCDs had a high-intensity mode [Figure 3] with FlashMax P3 showing significantly higher irradiance values as compared to DTE-iLED and VALO Cordless [Tables 3 and 4; P < 0.001]. Regarding changes in irradiance, no statistical significant changes were observed for 0, 10, and 20s in all brands for normal and high-intensity modes (data not shown; P > 0.05).
Figure 3

Mean values of light intensities with high-intensity mode of various light cure unit brands. Error bars indicate standard deviation. Devices with the same letter are not statistically different (P > 0.05)

Table 3

One-way analysis of variance: tests between subjects effects for high-intensity curing mode

Statistical testSum of squaresdfMean squareFP value
Between groups266606932133303473501.842380.000000
Within groups45680123806.667
Total2670637314

df = degrees of freedom

Dependent variable: light intensity mW/cm2

Table 4

Bonferroni test: tests whose devices are different for high-intensity curing mode

Light cure device 1Light cure device 2Mean differenceSEP value
DTE-iLEDUltradent VALO Cordless –176.0039.020.002141
DTE-iLEDFlashMax P3 460–2912.0039.020.000000
Ultradent VALO Cordless FlashMax P3 460–2736.0039.020.000000

SE = standard error

Dependent variable: light intensity mW/cm2

Mean values of light intensities with high-intensity mode of various light cure unit brands. Error bars indicate standard deviation. Devices with the same letter are not statistically different (P > 0.05) One-way analysis of variance: tests between subjects effects for high-intensity curing mode df = degrees of freedom Dependent variable: light intensity mW/cm2 Bonferroni test: tests whose devices are different for high-intensity curing mode SE = standard error Dependent variable: light intensity mW/cm2 For soft-start mode [Figure 4], only one device had this feature, which is KaVo mini-LED. Values for 0s were <300 mW/cm2 and then increased significantly at 10 and 20s (P < 0.05).
Figure 4

Mean light intensity for KaVo mini-light emitting device with soft-start mode. Error bars indicate standard deviation

Mean light intensity for KaVo mini-light emitting device with soft-start mode. Error bars indicate standard deviation

DISCUSSION

The use of composite resin restorations is increasing nowadays due to superior esthetics and good mechanical properties.[212] The irradiance values of light-curing devices are an integral part in achieving predictable composite resin polymerization and subsequent restoration life span.[131415] Due to the introduction of new bulk-fill composite formulations, the need to ensure adequate light irradiance of curing units is even greater. The objective of the current investigation was to test the light irradiance of different light cure units in different modes. The technology of light cure has been improved in the last decade due to foundation of high-intensity LED, quartz–tungsten–halogen (QTH) light, and plasma arc lights. These devices are manufactured to generate less heat and to cure the resin faster. Due to its advantages, LED recently is the most popular LCD as compared to halogen light.[4] Among LEDs, 10 of the most used LCDs were chosen. Some of these have high-intensity mode, which is important in orthodontic brackets placement. We used a digital radiometer because it gives reliable results and takes into consideration the tip dimeter. In addition for testing light output for conventional composite curing, we wanted to investigate which LCDs in the market can do proper polymerization of bulk-fill composite materials. We found three devices (Ortholux Luminous, Elipar DeepCure, and Elipar DeepCure-S), which can do proper polymerization for bulk-fill composite due to their irradiance values being >1000 mW/cm2. In a recent study, only 10% of the tested 166 LCDs produced values capable of polymerizing bulk-fill formulations.[16] The bulk-fill formulation usually contains new photoinitiators that are more sensitive but still require higher irradiance values to produce adequate degree of conversion with 4 mm or more increments.[1718] In general, bulk-fill composites require 10s of curing at >1000 mW/cm2 according to most manufacturers.[19] Three LCUs have the high irradiance (DTE-iLED, VALO Cordless, and FlashMax P3 460), which can be used in placing orthodontic brackets. All tested LCDs can do proper polymerization of conventional composite with irradiance values of 600 mW/cm2 and above. Of course, this setup must be accompanied by using incremental technique to reduce the inherent polymerization shrinkage of resins.[2021] Each layer has to be light cured for at least 20s to achieve proper degree of conversion[10] and to avoid negative repercussions of inadequate polymerization that include discoloration, sensitivity, and pulpal irritation.[35,622,23] In a clinical survey, Bansal et al.[24] found that 54% of the 1000 tested devices gave output values <400 mW/cm2. On the contrary of this and in accordance with our findings, Soares et al.[10] reported that only 2 of 22 tested LCDs produced irradiance values of <400 mW/cm2. Although all the tested devices of this study were new, frequent maintenance and check is required to ensure adequate output of LCDs. As with other in vitro studies, this investigation has some limitations. The choice of LCDs was limited to devices available in the Saudi market. Also, effect of irradiance on actual composite was not tested. Still, dentists practicing in Saudi Arabia and neighboring countries would benefit from the current instigation’s results as it has direct impact on the predictability and mechanical properties of composite restorations they place.[322,23] Further, as depicted from previous investigations, mechanical characteristics of cured composite depend on the irradiance values of LCDs and subsequent degree of conversion.[1415] On the basis of our results, the majority of investigated LCDs produced adequate irradiance values in their normal mode to achieve adequate polymerization of resin composite. Only three devices were able of producing values >1000 mW/cm2 and are capable of curing bulk-fill composite formulations.

CONCLUSION

Using light-curing units with adequate irradiance and exposing each composite resin increment for the recommended curing time are important factors to achieve restorations with predictable properties and long service life.

FINANCIAL SUPPORT AND SPONSORSHIP

Nil.

CONFLICTS OF INTEREST

There are no conflicts of interest.

AUTHOR CONTRIBUTIONS

Study conception (HN), data collection (AM, MA), data acquisition and analysis (AM, MA), data interpretation (HN, AM, MA), manuscript writing (HN, AM, MA).

ETHICAL POLICY AND INSTITUTIONAL REVIEW BOARD STATEMENT

This project is exempted from ethical approval due to lack of human and/or animal samples. All the procedures have been performed as per the ethical guidelines laid down by Declaration of Helsinki.

DATA AVAILABILITY STATEMENT

The data that support the study results are available from the author (Dr. Albaraa Makhdom, e-mail: fz1994@live.com) on request.
  22 in total

1.  Light transmittance characteristics and refractive indices of light-activated pit and fissure sealants.

Authors:  Hiroyuki Arikawa; Naoyuki Shinohara; Hideo Takahashi; Takahito Kanie; Koichi Fujii; Seiji Ban
Journal:  Dent Mater J       Date:  2010-01       Impact factor: 2.102

2.  Influence of physical assessment of different light-curing units on irradiance and composite microhardness top/bottom ratio.

Authors:  Susana Morimoto; Renata Aló Maluza Zanini; Josete Barbosa Cruz Meira; Carlos Martins Agra; Fernanda Calabró Calheiros; Denis Yudi Nagase
Journal:  Odontology       Date:  2016-01-13       Impact factor: 2.634

3.  Emission Characteristics and Effect of Battery Drain in "Budget" Curing Lights.

Authors:  M M AlShaafi; J E Harlow; H L Price; F A Rueggeberg; D Labrie; M Q AlQahtani; R B Price
Journal:  Oper Dent       Date:  2015-12-11       Impact factor: 2.440

Review 4.  Light-Curing Units: A Review of What We Need to Know.

Authors:  R B Price; J L Ferracane; A C Shortall
Journal:  J Dent Res       Date:  2015-07-08       Impact factor: 6.116

5.  Assessing the irradiance delivered from light-curing units in private dental offices in Jordan.

Authors:  Ghada A Maghaireh; Hanan Alzraikat; Nessrin A Taha
Journal:  J Am Dent Assoc       Date:  2013-08       Impact factor: 3.634

6.  Ability of four dental radiometers to measure the light output from nine curing lights.

Authors:  Carlos Alberto Kenji Shimokawa; Jessie Eudora Harlow; Míriam Lacalle Turbino; Richard Bengt Price
Journal:  J Dent       Date:  2016-09-03       Impact factor: 4.379

7.  An Evaluation of the Light Output from 22 Contemporary Light Curing Units.

Authors:  Carlos José Soares; Monise de Paula Rodrigues; Laís Rani Sales Oliveira; Stella Sueli Lourenço Braga; Luciana Mendes Barcelos; Gisele Rodrigues da Silva; Marcelo Giannini; Richard Bengt Price
Journal:  Braz Dent J       Date:  2017 May-Jun

8.  Efficiency of light curing units in a government dental school.

Authors:  Hani M Nassar; Reem Ajaj; Fatin Hasanain
Journal:  J Oral Sci       Date:  2018       Impact factor: 1.556

9.  Light output from six battery operated dental curing lights.

Authors:  Carlos Alberto Kenji Shimokawa; Míriam Lacalle Turbino; Jessie Eudora Harlow; Hannah Louise Price; Richard Bengt Price
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2016-07-21       Impact factor: 7.328

Review 10.  Depth of cure of bulk fill resin composites: A systematic review.

Authors:  Renally Bezerra Wanderley Lima; Cristhian Camilo Madrid Troconis; Marina Barrêto Pereira Moreno; Fabián Murillo-Gómez; Mario Fernando De Goes
Journal:  J Esthet Restor Dent       Date:  2018-10-29       Impact factor: 2.843

View more
  1 in total

1.  Effects of Different Light-curing Modes on the Compressive Strengths of Nanohybrid Resin-based Composites: A Comparative In Vitro Study.

Authors:  Franco Mauricio; Julia Medina; Luzmila Vilchez; Oscar Sotomayor; César Muricio-Vilchez; Frank Mayta-Tovalino
Journal:  J Int Soc Prev Community Dent       Date:  2021-04-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.