An Evaluation of the Light Output from 22 Contemporary Light Curing Units.

Carlos José Soares1, Monise de Paula Rodrigues1, Laís Rani Sales Oliveira1, Stella Sueli Lourenço Braga1, Luciana Mendes Barcelos1, Gisele Rodrigues da Silva1, Marcelo Giannini2, Richard Bengt Price3.   

Abstract

This study measured the radiant power (mW), irradiance (mW/cm2) and emission spectra (mW/cm2/nm) of 22 new, or almost new, light curing units (LCUs): - Alt Lux II, BioLux Standard, Bluephase G2, Curing Light XL 3000, Demetron LC, DX Turbo LED 1200, EC450, EC500, Emitter C, Emitter D, KON-LUX, LED 3M ESPE, Led Lux II, Optilight Color, Optilight Max, Optilux 501, Poly Wireless, Radii cal, Radii plus, TL-01, VALO Cordless. These LCUs were either monowave or multiple peak light emitting diode (LED) units or quartz-tungsten-halogen LCUs used in anterior and posterior teeth. The radiant power emitted by the LCUs was measured by a laboratory grade laser power meter. The tip area (cm²) of the LCUs was measured and used to calculate the irradiance from the measured radiant power source. The MARC-Patient Simulator (MARC-PS) with a laboratory grade spectrometer (USB4000, Ocean Optics) was used to measure the irradiance and emission spectrum from each LCU three times at the sensor located on the facial of the maxillary central incisors and then separately at the occlusal of a maxillary second molar. The minimum acceptable irradiance level was set as 500 mW/cm2. Irradiance data was analyzed using two-way ANOVA and the radiant power data was analyzed by one-way ANOVA followed by Tukey test (a=0.05). In general, the irradiance was reduced at the molar tooth for most LCUs. Only the Valo, Bluephase G2 and Radii Plus delivered an irradiance similar to the anterior and posterior sensors greater than 500 mW/cm2. KON-LUX, Altlux II, Biolux Standard, TL-01, Optilux 501, DX Turbo LED 1200 LCUs delivered lower irradiance values than the recommended one used in molar region, KON-LUX and Altlux II LCUs used at the maxillary incisors. Bluephase G2 and Optilight Max delivered the highest radiant power and KON-LUX, Altlux II and Biolux Standard delivered the lowest power. The emission spectrum from the various monowave LED LCUs varied greatly. The multi-peak LCUs delivered similar emission spectra to both sensors.

Entities:  

Mesh:

Year:  2017        PMID: 29297558     DOI: 10.1590/0103-6440201601466

Source DB:  PubMed          Journal:  Braz Dent J        ISSN: 0103-6440


  8 in total

1.  Influence of irradiation distance on the mechanical performances of resin composites polymerized with high-irradiance light curing units.

Authors:  Soram Oh; Hyun Ju Kim; Hyun-Jung Kim; Sibel A Antonson; Sun-Young Kim
Journal:  Biomater Res       Date:  2022-05-20

2.  The light-curing unit: An essential piece of dental equipment.

Authors:  Richard B Price; Jack L Ferracane; Reinhard Hickel; Braden Sullivan
Journal:  Int Dent J       Date:  2020-07-21       Impact factor: 2.607

3.  Irradiance of Different Curing Modes of Common Light Cure Devices: An In Vitro Study.

Authors:  Hani M Nassar; Mahmoud Almutairi; Albaraa Makhdom
Journal:  J Int Soc Prev Community Dent       Date:  2020-03-10

4.  Assessment of the radiant emittance of damaged/contaminated dental light-curing tips by spectrophotometric methods.

Authors:  Abdulrahman A Balhaddad; Isadora Garcia; Fabrício Collares; Cristopher M Felix; Nisha Ganesh; Qoot Alkabashi; Ward Massei; Howard Strassler; Mary Anne Melo
Journal:  Restor Dent Endod       Date:  2020-11-03

5.  Physical-chemical characterization and bond strength to zirconia of dental adhesives with different monomer mixtures and photoinitiator systems light-activated with poly and monowave devices.

Authors:  Constantino Fernandes Neto; Mayara Hana Narimatsu; Pedro Henrique Magão; Reginaldo Mendonça da Costa; Carmem Silvia Pfeifer; Adilson Yoshio Furuse
Journal:  Biomater Investig Dent       Date:  2022-04-28

6.  Influence of Tip Diameter and Light Spectrum of Curing Units on the Properties of Bulk-Fill Resin Composites.

Authors:  Igor Oliveiros Cardoso; Alexandre Coelho Machado; Luísa de Oliveira Fernandes; Paulo Vinícius Soares; Luís Henrique Araújo Raposo
Journal:  Eur J Dent       Date:  2021-12-14

Review 7.  Assessing Dental Light-curing Units' Output Using Radiometers: A Narrative Review.

Authors:  Cendrella Assaf; Jean-Claude Fahd; Joseph Sabbagh
Journal:  J Int Soc Prev Community Dent       Date:  2020-01-24

8.  Effect of infection control barriers on the light output from a multi-peak light curing unit.

Authors:  Carlos José Soares; Stella Sueli Lourenço Braga; Maria Tereza Hordones Ribeiro; Richard Bengt Price
Journal:  J Dent       Date:  2020-10-19       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.