Assessing the irradiance delivered from light-curing units in private dental offices in Jordan.

Ghada A Maghaireh1, Hanan Alzraikat, Nessrin A Taha.   

Abstract

BACKGROUND: The authors conducted a study to examine the irradiance from light-curing units (LCUs) used in dental offices in Jordan.
METHODS: Two of the authors visited 295 private dental offices (15 percent) in Jordan and collected the following information about the LCUs: age, type (quartz-tungsten-halogen or light-emitting diode), date of last maintenance, type of maintenance, last date of use, number of times used during the day, availability of a radiometer, exposure time for each resin-based composite increment, size of light-curing tips and presence of resin-based composite on the tips. The authors used a radiometer to measure the irradiance from the LCUs. They used linear regression with stepwise correlation for the statistical analysis. The authors set the minimum acceptable irradiance at 300 milliwatts/square centimeter.
RESULTS: The mean irradiance of the 295 LCUs examined was 361 mW/cm(2), and 136 LCUs (46.1 percent) delivered an irradiance of less than 300 mW/cm(2). The unit's age, type and presence of resin-based composite on the light-curing tips had a significant effect on the irradiance (P ≤ .001).
CONCLUSIONS: Only 37 of the 141 quartz-tungsten-halogen units (26.2 percent) and 122 of the 154 light-emitting diode units (79.2 percent) delivered at least 300 mW/cm(2). Resin contamination on the light-curing tips had a significant effect on the irradiance delivered. The irradiance from the LCUs decreased with use. Practical Implications. The irradiance from many of the units in this study was less than 300 mW/cm(2), which may affect the quality of resin-based composite restorations. Dentists should monitor the performance of the LCUs in their offices weekly.

Entities:  

Keywords:  Resin-based composites; dental offices; irradiance; light output; light-curing units; survey

Mesh:

Substances:

Year:  2013        PMID: 23904579     DOI: 10.14219/jada.archive.2013.0210

Source DB:  PubMed          Journal:  J Am Dent Assoc        ISSN: 0002-8177            Impact factor:   3.634


  13 in total

1.  Influence of physical assessment of different light-curing units on irradiance and composite microhardness top/bottom ratio.

Authors:  Susana Morimoto; Renata Aló Maluza Zanini; Josete Barbosa Cruz Meira; Carlos Martins Agra; Fernanda Calabró Calheiros; Denis Yudi Nagase
Journal:  Odontology       Date:  2016-01-13       Impact factor: 2.634

2.  Guidelines for the selection, use, and maintenance of LED light-curing units - Part 1.

Authors:  A C Shortall; R B Price; L MacKenzie; F J T Burke
Journal:  Br Dent J       Date:  2016-10-21       Impact factor: 1.626

3.  The Effect of Different Light Curing Units and Tip Distances on Translucency Parameters of Bulk fill Materials.

Authors:  Dana Al-Senan; Hend Al-Nahedh
Journal:  Saudi Dent J       Date:  2022-04-12

4.  Irradiance of Different Curing Modes of Common Light Cure Devices: An In Vitro Study.

Authors:  Hani M Nassar; Mahmoud Almutairi; Albaraa Makhdom
Journal:  J Int Soc Prev Community Dent       Date:  2020-03-10

Review 5.  Factors affecting polymerization of resin-based composites: A literature review.

Authors:  Maan M AlShaafi
Journal:  Saudi Dent J       Date:  2017-03-07

6.  Light intensity output of visible light communication units and clinicians' knowledge and attitude among Riyadh private clinics.

Authors:  Lama Mana Alqabbaa; Mounira Sami Alsenani; Norah Saad Alsaif; Ruba Abdulrahman Alsaif; Sultan Rukayb Binalrimal
Journal:  J Conserv Dent       Date:  2018 Nov-Dec

7.  Types of polymerisation units and their intensity output in private dental clinics of twin cities in eastern province, KSA; a pilot study.

Authors:  Theeb Alquria; Mohammed Al Gady; Abdul Khabeer; Saqib Ali
Journal:  J Taibah Univ Med Sci       Date:  2018-12-14

Review 8.  Utilizing Light Cure Units: A Concise Narrative Review.

Authors:  Fatin A Hasanain; Hani M Nassar
Journal:  Polymers (Basel)       Date:  2021-05-15       Impact factor: 4.329

9.  Knowledge and Attitude of Dental Clinicians towards Light-Curing Units: A Cross-Sectional Study.

Authors:  Dana Al-Senan; Fatin Ageel; Amani Aldosari; Haifa Maktabi
Journal:  Int J Dent       Date:  2021-06-14

Review 10.  High-Power LED Units Currently Available for Dental Resin-Based Materials-A Review.

Authors:  Rita Almeida; Patricia Manarte-Monteiro; Joana Domingues; Carlos Falcão; Mariano Herrero-Climent; Blanca Ríos-Carrasco; Bernardo Ferreira Lemos
Journal:  Polymers (Basel)       Date:  2021-06-30       Impact factor: 4.329

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.