Hani M Nassar1, Reem Ajaj1, Fatin Hasanain1.
Abstract
The light intensity of a light-curing unit is a crucial factor that affects the clinical longevity of resin composites. This study aimed to investigate the efficiency of light-curing units in use at a local governmental dental school for curing conventional and bulk-fill resin materials. A total of 166 light-curing units at three locations were examined, and the brand, type, clinic location, diameter of curing tip, tip cleanliness (using a visual score), and the output (in mW/cm2 using a digital radiometer) were recorded. Only 23.5% of the units examined had clean tips, with the graduate student clinical area containing the highest percentage of clean tips. Further, tips with poor cleanliness score values were associated with significantly lower output intensities. A small percentage (9.4%) of units was capable of producing intensities higher than 1,200 mW/cm2 and lower than 600 mW/cm2 (7.6%). The majority of the low intensity units were located in the undergraduate student area, which also contained the highest number of units with intensities between 900 and 1,200 mW/cm2. The output of all the units in service was satisfactory for curing conventional resin composites, and most units were capable of curing bulk-fill resin materials.Entities:
Keywords: efficiency; light curing; resin-based composites
Mesh:
Year: 2018 PMID: 29576574 DOI: 10.2334/josnusd.17-0071
Source DB: PubMed Journal: J Oral Sci ISSN: 1343-4934 Impact factor: 1.556