| Literature DB >> 32632827 |
Antonio C Traino1, Patrizio Barca2,3, Rocco Lamastra2,3, Raffaele M Tucciariello3, Chiara Sottocornola4, Carolina Marini5, Giacomo Aringhieri6, Davide Caramella6, Maria E Fantacci3.
Abstract
BACKGROUND: To propose a practical and simple method to individually evaluate the average absorbed dose for digital breast tomosynthesis.Entities:
Keywords: Digital breast tomosynthesis; Mammography; Phantoms (imaging); X-rays
Mesh:
Year: 2020 PMID: 32632827 PMCID: PMC7338293 DOI: 10.1186/s41747-020-00165-2
Source DB: PubMed Journal: Eur Radiol Exp ISSN: 2509-9280
Comparison between measured and calculated (Eq. (2)) incident air kerma for the reference Hologic Selenia Dimensions (device A)
| Phantom thickness (cm) | Tube load (mAs) | Tube voltage (kVp) | Measured air kerma (mGy) | Calculated air kerma (mGy) | Relative difference (%) |
|---|---|---|---|---|---|
| 2.0 | 37.5 | 26 | 2.11 ± 0.08 | 2.06 ± 0.15 | -2.6 |
| 2.5 | 40 | 27 | 2.61 ± 0.10 | 2.54 ± 0.19 | -2.8 |
| 3.0 | 37.5 | 28 | 2.74 ± 0.11 | 2.71 ± 0.20 | -1.3 |
| 3.5 | 40 | 29 | 3.33 ± 0.13 | 3.24 ± 0.24 | -2.7 |
| 4.0 | 50 | 29 | 4.14 ± 0.17 | 4.12 ± 0.31 | -0.4 |
| 4.5 | 50 | 30 | 4.66 ± 0.19 | 4.59 ± 0.34 | -1.6 |
| 5.0 | 55 | 31 | 5.80 ± 0.23 | 5.60 ± 0.42 | -3.6 |
| 5.5 | 60 | 32 | 7.08 ± 0.28 | 6.73 ± 0.50 | -5.3 |
| 6.0 | 65 | 33 | 8.18 ± 0.33 | 7.99 ± 0.59 | -2.4 |
| 6.5 | 70 | 34 | 9.78 ± 0.39 | 9.40 ± 0.70 | -4.1 |
| 7.0 | 80 | 35 | 12.03 ± 0.48 | 11.69 ± 0.87 | -2.9 |
| 7.5 | 80 | 36 | 13.02 ± 0.52 | 12.68 ± 0.94 | -2.7 |
| 8.5 | 80 | 38 | 15.19 ± 0.61 | 14.78 ± 1.10 | -2.8 |
| 8.5 | 90 | 42 | 21.52 ± 0.86 | 20.55 ± 1.53 | -4.7 |
Data for measured air kerma and calculated air kerma are presented as mean ± standard deviation. Note that the exposure parameters were chosen as close as possible to the corresponding automatic exposure control settings for a given phantom thickness
Values of the coefficients of Eq. (2), (3), and (5) evaluated for the reference Hologic Selenia Dimensions (device A)
| Parameter | Y0 at 32 kVp (mGy/mAs) | α (mGy/(kVp2 ·mAs)) | β (mGy/(kVp·mAs)) | γ (mGy/mAs) | a (kVp/cm) | b (a.u.) |
|---|---|---|---|---|---|---|
| Value | 0.094 ± 0.005 | (5.70 ± 0.86)·10−5 | (3.77 ± 0.56)·10−3 | (− 8.44 ± 0.89)·10−2 | 20.32 ± 1.97 | − 1.04 ± 0.03 |
Data are presented as mean ± standard deviation. These values were employed to calculate the incident air kerma and 2ABD. Notice that these values depend on the anode-filter combination (W-Al in our case)
Fig. 1Beam intensity as a function of the phantom depth for different tube voltage values (26–48 kVp). An exponential relationship (Eq. (4)) was employed to fit the data and obtain the m values for different tube voltages
Estimation of the m values for different tube voltage values
| kVp | m (cm-1) |
|---|---|
| 26 | 0.65 ± 0.03 |
| 28 | 0.60 ± 0.02 |
| 30 | 0.55 ± 0.02 |
| 32 | 0.52 ± 0.02 |
| 34 | 0.48 ± 0.02 |
| 36 | 0.46 ± 0.01 |
| 38 | 0.44 ± 0.01 |
| 40 | 0.42 ± 0.01 |
| 44 | 0.39 ± 0.01 |
| 48 | 0.36 ± 0.01 |
Data for m values are presented as mean ± standard deviation. The m values were obtained by fitting experimental data through Eq. (4), as shown in Fig. 1. The fitting parameters of Eq. (4) are presented in Table 2
Fig. 2Coefficient m estimated from measurements in the phantom, expressed as a function of tube voltage. A power function (Eq. (5)) was applied to model the m as a function of the tube voltage values
Comparison between 2ABD computed from Eq. (1) and the average glandular dose (AGD) calculated by the Dance approach for 100% glandularity
| Mammographic device | Phantom thickness (cm) | Tube load (mAs) | Tube voltage (kVp) | 2ABD (mGy) | AGD 100% glandularity (mGy) | Relative difference (%) |
|---|---|---|---|---|---|---|
| Hologic-A | 3.0 | 37.5 | 28 | 0.94 ± 0.19 | 1.00 ± 0.20 | -6.0 |
| Hologic-A | 3.5 | 40 | 29 | 1.05 ± 0.20 | 1.07 ± 0.21 | -1.9 |
| Hologic-A | 4.0 | 50 | 29 | 1.21 ± 0.23 | 1.21 ± 0.24 | 0.0 |
| Hologic-A | 4.5 | 50 | 30 | 1.26 ± 0.23 | 1.27 ± 0.25 | -0.8 |
| Hologic-A | 5.0 | 55 | 31 | 1.45 ± 0.26 | 1.46 ± 0.29 | -0.7 |
| Hologic-A | 5.5 | 60 | 32 | 1.66 ± 0.30 | 1.67 ± 0.33 | -0.6 |
| Hologic-A | 6.0 | 65 | 33 | 1.87 ± 0.33 | 1.89 ± 0.38 | -1.1 |
| Hologic-A | 6.5 | 70 | 34 | 2.12 ± 0.37 | 2.14 ± 0.43 | -0.9 |
| Hologic-A | 7.0 | 80 | 35 | 2.51 ± 0.44 | 2.57 ± 0.51 | -2.3 |
| Hologic-B | 3.5 | 30 | 29 | 1.07 ± 0.20 | 1.14 ± 0.23 | -6.1 |
| Hologic-B | 4.5 | 35 | 30 | 1.20 ± 0.22 | 1.22 ± 0.24 | -1.6 |
| Hologic-B | 5.5 | 42.5 | 32 | 1.61 ± 0.29 | 1.64 ± 0.33 | -1.8 |
| Hologic-B | 6.5 | 50 | 34 | 2.06 ± 0.37 | 2.09 ± 0.42 | -1.4 |
| Fuji-C | 3.5 | 28 | 29 | 1.19 ± 0.24 | 1.21 ± 0.24 | -1.7 |
| Fuji-C | 4.5 | 32 | 32 | 1.63 ± 0.30 | 1.56 ± 0.31 | 4.5 |
| Fuji-C | 5.5 | 42 | 33 | 2.06 ± 0.37 | 2.04 ± 0.41 | 1.0 |
| Fuji-C | 6.5 | 50 | 35 | 2.66 ± 0.47 | 2.65 ± 0.53 | 0.4 |
| Fuji-C | 4.5* | 45 | 32 | 2.30 ± 0.43 | 2.15 ± 0.43 | 7.0 |
| Fuji-C | 6.5* | 63 | 35 | 3.35 ± 0.59 | 3.27 ± 0.65 | 2.4 |
| Fuji-D | 3.5 | 28 | 30 | 1.30 ± 0.26 | 1.38 ± 0.28 | -5.8 |
| Fuji-D | 4.5 | 32 | 32 | 1.59 ± 0.30 | 1.60 ± 0.32 | -0.6 |
| Fuji-D | 5.5 | 50 | 33 | 2.40 ± 0.43 | 2.43 ± 0.49 | -1.2 |
| Fuji-D | 6.5 | 56 | 35 | 2.91 ± 0.50 | 2.87 ± 0.57 | 1.4 |
| Fuji-D | 3.5* | 32 | 30 | 1.49 ± 0.28 | 1.55 ± 0.31 | -3.9 |
| Fuji-D | 4.5* | 45 | 32 | 2.24 ± 0.43 | 2.20 ± 0.44 | 1.8 |
| Fuji-D | 5.5* | 63 | 34 | 3.35 ± 0.61 | 3.22 ± 0.64 | 4.0 |
| Fuji-D | 6.5* | 71 | 35 | 3.69 ± 0.65 | 3.56 ± 0.71 | 3.7 |
Data for 2ABD and AGD 100% glandularity are presented as mean ± standard deviation. The comparison was carried out by considering all mammographic devices. The exposure parameters were chosen as close as possible to the corresponding automatic exposure control settings for a given phantom thickness. For the AGD calculation by the Dance method, the measured air kerma was employed. For the Amulet Innovality devices both the standard acquisition mode and high-resolution acquisition mode (*) were employed
Fig. 3Comparison between the 2ABD computed for digital mammography (DM) and digital breast tomosynthesis (DBT) for the same phantom thickness. The exposure parameters and the complete set of data are presented in Table 5. DM values are referred to a single view procedure
Comparison between 2ABD in digital mammography and digital breast tomosynthesis for different phantom thicknesses on the reference Hologic Selenia Dimensions (device A)
| Phantom thickness (cm) | Digital mammography (single view) (W/Rh and W/Ag*) | Digital breast tomosynthesis (W/Al) | ||||
|---|---|---|---|---|---|---|
| Tube load (mAs) | Tube voltage (kVp) | 2ABD (mGy) | Tube load (mAs) | Tube voltage (kVp) | 2ABD (mGy) | |
| 2.5 | 50 | 26 | 0.49 ± 0.11 | 39 | 27 | 0.95 ± 0.21 |
| 3.0 | 67 | 26 | 0.61 ± 0.15 | 38 | 28 | 0.95 ± 0.19 |
| 3.5 | 72 | 27 | 0.67 ± 0.15 | 40 | 29 | 1.05 ± 0.20 |
| 4.0 | 82 | 28 | 0.77 ± 0.16 | 49 | 29 | 1.18 ± 0.23 |
| 4.5 | 98 | 28 | 0.85 ± 0.18 | 50 | 30 | 1.26 ± 0.23 |
| 5.0 | 134 | 29 | 1.20 ± 0.24 | 55 | 31 | 1.45 ± 0.26 |
| 5.5 | 138 | 30 | 1.26 ± 0.23 | 60 | 32 | 1.66 ± 0.30 |
| 6.0 | 155 | 31 | 1.43 ± 0.26 | 65 | 33 | 1.86 ± 0.33 |
| 6.5 | 161 | 32 | 1.51 ± 0.28 | 71 | 34 | 2.13 ± 0.38 |
| 7.0* | 159 | 30 | 1.73 ± 0.36 | 78 | 35 | 2.43 ± 0.43 |
| 7.5* | 174 | 31 | 1.96 ± 0.40 | 82 | 36 | 2.70 ± 0.47 |
Data for 2ABD are presented as mean ± standard deviation. In DM modality, the W-Rh anode-filter combination was selected by the AEC system for a range of 2.5–6.5 cm of phantom thickness, while for a thickness of 7 cm and 7.5 cm the W-Ag (*) was automatically set. On the other hand, only the W-Al option was available for DBT
Comparison between AGD computed from 2ABD and the AGD calculated by the Dance approach for different glandularities
| Glandularity (%) | Phantom thickness (cm) | Tube load (mAs) | Tube voltage (kVp) | AGD from 2ABD (mGy) | AGD by Dance (mGy) | Relative difference (%) |
|---|---|---|---|---|---|---|
| 100 | 3 | 37.5 | 28 | 0.94 ± 0.19 | 1.00 ± 0.20 | -6.0 |
| 100 | 4 | 50 | 29 | 1.21 ± 0.23 | 1.21 ± 0.24 | 0.0 |
| 100 | 5 | 55 | 31 | 1.45 ± 0.26 | 1.46 ± 0.29 | -0.7 |
| 100 | 6 | 65 | 33 | 1.87 ± 0.33 | 1.89 ± 0.38 | -1.1 |
| 100 | 7 | 80 | 35 | 2.57 ± 0.44 | 2.57 ± 0.31 | -2.3 |
| 80 | 3 | 37.5 | 28 | 0.96 ± 0.19 | 1.06 ± 0.21 | -9.4 |
| 80 | 4 | 50 | 29 | 1.24 ± 0.24 | 1.30 ± 0.26 | -4.6 |
| 80 | 5 | 55 | 31 | 1.48 ± 0.27 | 1.58 ± 0.32 | -6.3 |
| 80 | 6 | 65 | 33 | 1.91 ± 0.34 | 2.04 ± 0.41 | -6.4 |
| 80 | 7 | 80 | 35 | 2.56 ± 0.45 | 2.78 ± 0.56 | -7.9 |
| 60 | 3 | 37.5 | 28 | 0.98 ± 0.20 | 1.12 ± 0.22 | -12.5 |
| 60 | 4 | 50 | 29 | 1.26 ± 0.24 | 1.40 ± 0.28 | -10.0 |
| 60 | 5 | 55 | 31 | 1.51 ± 0.27 | 1.70 ± 0.34 | -11.2 |
| 60 | 6 | 65 | 33 | 1.95 ± 0.34 | 2.21 ± 0.44 | -11.8 |
| 60 | 7 | 80 | 35 | 2.62 ± 0.46 | 3.01 ± 0.60 | -13.0 |
| 40 | 3 | 37.5 | 28 | 1.00 ± 0.20 | 1.19 ± 0.24 | -16.0 |
| 40 | 4 | 50 | 29 | 1.29 ± 0.25 | 1.50 ± 0.30 | -14.0 |
| 40 | 5 | 55 | 31 | 1.55 ± 0.28 | 1.84 ± 0.37 | -15.8 |
| 40 | 6 | 65 | 33 | 2.00 ± 0.35 | 2.41 ± 0.48 | -17.0 |
| 40 | 7 | 80 | 35 | 2.68 ± 0.47 | 3.29 ± 0.66 | -18.5 |
| 20 | 3 | 37.5 | 28 | 1.03 ± 0.21 | 1.26 ± 0.25 | -18.3 |
| 20 | 4 | 50 | 29 | 1.32 ± 0.25 | 1.61 ± 0.32 | -18.0 |
| 20 | 5 | 55 | 31 | 1.59 ± 0.28 | 2.00 ± 0.40 | -20.5 |
| 20 | 6 | 65 | 33 | 2.04 ± 0.36 | 2.62 ± 0.52 | -22.1 |
| 20 | 7 | 80 | 35 | 2.74 ± 0.48 | 3.58 ± 0.72 | -23.5 |
Data for AGD from 2ABD and for AGD by Dance are presented as mean ± standard deviation. The comparison was carried out by considering the reference mammographic devices (Hologic, device A). The exposure parameters were chosen as close as possible to the corresponding AEC settings for a given phantom thickness. For the AGD by Dance, the measured air kerma was employed