Literature DB >> 32612897

Gap-filling method for suppressing grating lobes in ultrasound imaging: Experimental study with deep-learning approach.

Viksit Kumar1, Po-Yang Lee1,2, Bae-Hyung Kim1, Mostafa Fatemi2, Azra Alizad1,3.   

Abstract

Sparse arrays reduce the number of active channels that effectively increases the inter-element spacing. Large inter-element spacing results in grating lobe artifacts degrading the ultrasound image quality and reducing the contrast-to-noise ratio. A deep learning-based custom algorithm is proposed to estimate inactive channel data in periodic sparse arrays. The algorithm uses data from multiple active channels to estimate inactive channels. The estimated inactive channel data effectively reduces the inter-element spacing for beamforming, thus suppressing the grating lobes. Estimated inactive element channel data was combined with active element channel data resulting in a pseudo fully sampled array. The channel data was beamformed using a simple delay-and-sum method and compared with the sparse array and fully sampled array. The performance of the algorithm was validated using a wire target in a water tank, multi-purpose tissue-mimicking phantom, and in-vivo carotid data. Grating lobes suppression up to 15.25 dB was observed with an increase in contrast-to-noise (CNR) for the pseudo fully sampled array. Hypoechoic regions showed more improvement in CNR than hyperechoic regions. Root-mean-square error for unwrapped phase between fully sampled array and the pseudo fully sampled array was low, making the estimated data suitable for Doppler and elastography applications. Speckle pattern was also preserved; thus, the estimated data can also be used for quantitative ultrasound applications. The algorithm can improve the quality of sparse array images and has applications in small scale ultrasound devices and 2D arrays.

Entities:  

Keywords:  convolutional neural networks; deep learning; gap-filling; sparse array; ultrasound imaging

Year:  2020        PMID: 32612897      PMCID: PMC7328874          DOI: 10.1109/access.2020.2989337

Source DB:  PubMed          Journal:  IEEE Access        ISSN: 2169-3536            Impact factor:   3.367


  15 in total

1.  Thinning and weighting of large planar arrays by simulated annealing.

Authors:  A Trucco
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  1999       Impact factor: 2.725

2.  Minimally redundant 2-D array designs for 3-D medical ultrasound imaging.

Authors:  Mustafa Karaman; Ira O Wygant; Omer Oralkan; Butrus T Khuri-Yakub
Journal:  IEEE Trans Med Imaging       Date:  2009-01-06       Impact factor: 10.048

3.  Design of optimal 2-D nongrid sparse arrays for medical ultrasound.

Authors:  Bakary Diarra; Marc Robini; Piero Tortoli; Christian Cachard; Hervé Liebgott
Journal:  IEEE Trans Biomed Eng       Date:  2013-06-11       Impact factor: 4.538

4.  Development of an ultrasonic method to detect cervical remodeling in vivo in full-term pregnant women.

Authors:  Barbara L McFarlin; Jennifer Balash; Viksit Kumar; Timothy A Bigelow; Xavier Pombar; Jacques S Abramowicz; William D O'Brien
Journal:  Ultrasound Med Biol       Date:  2015-05-23       Impact factor: 2.998

5.  Cross-imaging system comparison of backscatter coefficient estimates from a tissue-mimicking material.

Authors:  Kibo Nam; Ivan M Rosado-Mendez; Lauren A Wirtzfeld; Viksit Kumar; Ernest L Madsen; Goutam Ghoshal; Alexander D Pawlicki; Michael L Oelze; Roberto J Lavarello; Timothy A Bigelow; James A Zagzebski; William D O'Brien; Timothy J Hall
Journal:  J Acoust Soc Am       Date:  2012-09       Impact factor: 1.840

6.  Optimizing circular ring arrays for forward-looking IVUS imaging.

Authors:  Coskun Tekes; Mustafa Karaman; F Levent Degertekin
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2011-12       Impact factor: 2.725

7.  Unambiguous Identification and Visualization of an Acoustically Active Catheter by Ultrasound Imaging in Real Time: Theory, Algorithm, and Phantom Experiments.

Authors:  Viksit Kumar; Richard Liu; Randall R Kinnick; Adriana Gregory; Azra Alizad; Marek Belohlavek; Mostafa Fatemi
Journal:  IEEE Trans Biomed Eng       Date:  2017-09-25       Impact factor: 4.538

8.  Ultrasonic attenuation and backscatter coefficient estimates of rodent-tumor-mimicking structures: comparison of results among clinical scanners.

Authors:  Kibo Nam; Ivan M Rosado-Mendez; Lauren A Wirtzfeld; Alexander D Pawlicki; Viksit Kumar; Ernest L Madsen; Goutam Ghoshal; Roberto J Lavarello; Michael L Oelze; Timothy A Bigelow; James A Zagzebski; William D O'Brien; Timothy J Hall
Journal:  Ultrason Imaging       Date:  2011-10       Impact factor: 1.578

9.  Diagnostic features of quantitative comb-push shear elastography for breast lesion differentiation.

Authors:  Mahdi Bayat; Max Denis; Adriana Gregory; Mohammad Mehrmohammadi; Viksit Kumar; Duane Meixner; Robert T Fazzio; Mostafa Fatemi; Azra Alizad
Journal:  PLoS One       Date:  2017-03-03       Impact factor: 3.240

10.  Correlation of ultrasound bladder vibrometry assessment of bladder compliance with urodynamic study results.

Authors:  Mahdi Bayat; Viksit Kumar; Max Denis; Jeremy Webb; Adriana Gregory; Mohammad Mehrmohammadi; Mathew Cheong; Douglas Husmann; Lance Mynderse; Azra Alizad; Mostafa Fatemi
Journal:  PLoS One       Date:  2017-06-16       Impact factor: 3.240

View more