T T Monajemi1,2, P A K Oliver1,2, A Day1, M Yewondwossen1,2.
Abstract
PURPOSE: This study was designed to evaluate skin dose in both VMAT and tangent treatment deliveries for the purpose of identifying suitable bolus use protocols that should produce similar superficial doses.Entities:
Keywords: VMAT; bolus; brass mesh; chest wall radiation therapy; skin dose; tangents
Mesh:
Year: 2020 PMID: 32592451 PMCID: PMC7484836 DOI: 10.1002/acm2.12948
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Picture of the three‐dimensional (3D) printed phantom used in this study. Location of the detectors is shown (a). The phantom consists of two segments for ease of handling (b). VMAT plan with 1 cm bolus, 95% dose wash is shown. The arrows point to the location of OSLDs and strip of film. CTV is shown (c). Tangent plan with 1 cm bolus, 95% dose wash is shown. The arrows point to the location of OSLDs and strip of film. CTV is shown (d).
Planning objectives used in this study.
| PTV_Eval |
V95% > 95% D10cc < 4280 cGy D0.03cc < 4400 cGy |
| Lung_Ipsi |
V500 cGy < 60% V1000 cGy < 45% V2000 cGy < 25% Mean < 1000 cGy |
| Lungs_Total |
V2000 cGy < 10% V750 cGy < 25% V500 cGy < 50% Mean < 600 cGy |
| Lung_Contra |
V500 cGy < 10% Mean < 300 cGy |
| Heart | Mean < 300 cGy |
| Breast_Contra |
V1000 cGy < 15% Mean < 400 cGy |
| Esophagus | D0.03cc < 4000 cGy |
Fig. 2Three‐dimensional field in field superficial doses. Skin dose measured with EBT3 where negative distance is in the lateral direction. For comparison skin doses measured by bare OSLDs are shown for the case of no bolus (a). Average skin doses in the three OLSD locations as measured by jacketed, bare OSLD, EBT3, and calculated by Eclipse (b). All doses are expressed as a fraction of prescription dose. The bars indicated the range of measurements.
Fig. 3VMAT superficial doses. Skin dose measured with EBT3 where negative distance is in the lateral direction. For comparison skin doses measured by bare OSLDs are shown for the case of no bolus (a). Average skin doses in the three OLSD locations as measured by jacketed, bare OSLD, EBT3, and calculated by Eclipse (b). All doses are expressed as a fraction of prescription dose. The bars indicated the range of measurements.
Fig. 4Superficial doses measured by EBT3 with no bolus in the presence of hospital gown and blanket used in our institute for tangents (a) and VMAT (b). All doses are expressed as a fraction of prescription dose.