Skin dose effects of postmastectomy chest wall radiation therapy using brass mesh as an alternative to tissue equivalent bolus.

Erin Healy1, Shawnee Anderson2, Jing Cui3, Laurel Beckett4, Allen M Chen3, Julian Perks3, Robin Stern3, Jyoti Mayadev5.   

Abstract

PURPOSE: The use of brass mesh as a bolus is relatively uncommon in postmastectomy chest wall radiation therapy (PMRT). This study aimed to characterize the skin dose effects of using 2-mm fine brass mesh as an alternative to the traditional tissue-equivalent bolus during chest wall PMRT. METHODS AND MATERIALS: Data were collected from patients who received PMRT using brass mesh at the University of California Davis Department of Radiation Oncology between January 2008 and June 2011. Several patient characteristics including age, body habitus, and ethnicity were analyzed along with several disease and treatment characteristics to determine whether or not they had an impact on the skin reaction observed during radiation treatment. Additionally, in vivo surface dose measurements were obtained for 16 of the 48 patients (33%).
RESULTS: Forty-eight female patients aged 28-83 received PMRT using brass mesh. As expected, the severity of skin toxicity increased with subsequent doses of radiation with all patients beginning treatment with no skin reaction (National Cancer Institute scores [NCIS] = 0) and the majority of patients completing treatment with either faint to moderate erythema (n = 19, 40%, NCIS = 1) or moderate to brisk erythema (n = 23, 48%, NCIS = 2). In vivo dosimetry analysis revealed surface doses between 81% and 122% of the prescribed dose, with an average of 99% of the prescribed radiation dose and standard deviation of 10% being delivered.
CONCLUSIONS: For postmastectomy chest wall radiation therapy, brass mesh is an effective alternative to tissue-equivalent bolus. The brass mesh achieved moderate erythema in the majority of patients at the end of treatment and the surface dose was validated using in vivo dosimetry. Published by Elsevier Inc.

Entities:  

Year:  2012        PMID: 24674320     DOI: 10.1016/j.prro.2012.05.009

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  9 in total

1.  A single plan solution to chest wall radiotherapy with bolus?

Authors:  C Ordonez-Sanz; S Bowles; A Hirst; N D MacDougall
Journal:  Br J Radiol       Date:  2014-02-26       Impact factor: 3.039

2.  The accuracy of treatment planning system dose modelling in the presence of brass mesh bolus.

Authors:  Neil Richmond
Journal:  Rep Pract Oncol Radiother       Date:  2017-07-22

3.  Dosimetric assessment of brass mesh bolus for postmastectomy photon radiotherapy.

Authors:  Ryan Manger; Adam Paxton; Laura Cerviño
Journal:  J Appl Clin Med Phys       Date:  2016-07-25       Impact factor: 2.102

4.  Treatment planning for metals using an extended CT number scale.

Authors:  John P Mullins; Michael P Grams; Michael G Herman; Debra H Brinkmann; John A Antolak
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

5.  An Innovative Risk-Reducing Approach to Postmastectomy Radiation Delivery after Autologous Breast Reconstruction.

Authors:  Merisa L Piper; Maristella Evangelista; Dominic Amara; David Daar; Robert D Foster; Barbara Fowble; Hani Sbitany
Journal:  Plast Reconstr Surg Glob Open       Date:  2017-04-25

6.  Characterization of a novel scale maille contralateral breast shield: SMART Armor.

Authors:  Macinley Butson; Susan Carroll; Martin Butson; Robin Hill
Journal:  J Appl Clin Med Phys       Date:  2017-08-11       Impact factor: 2.102

7.  In search of a one plan solution for VMAT post-mastectomy chest wall irradiation.

Authors:  T T Monajemi; P A K Oliver; A Day; M Yewondwossen
Journal:  J Appl Clin Med Phys       Date:  2020-06-27       Impact factor: 2.102

8.  A prospective feasibility study of a 1-mm bolus for postmastectomy radiotherapy.

Authors:  Terufumi Kawamoto; Naoto Shikama; Chie Kurokawa; Naoya Hara; Masaki Oshima; Keisuke Sasai
Journal:  BMC Cancer       Date:  2021-02-02       Impact factor: 4.430

9.  Surface Dose Measurements in Chest Wall Postmastectomy Radiotherapy to Achieve Optimal Dose Delivery with 6 MV Photon Beam.

Authors:  Dilson Lobo; Sourjya Banerjee; Challapalli Srinivas; M S Athiyamaan; Shreyas Reddy; Johan Sunny; Ramamoorthy Ravichandran; Himani Kotian; P U Prakash Saxena
Journal:  J Med Phys       Date:  2021-11-20
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.