Tangent-based volumetric modulated arc therapy for advanced left breast cancer.

Pei-Chieh Yu1,2, Ching-Jung Wu1,3,4, Hsin-Hua Nien1,5, Louis Tak Lui1, Suzun Shaw6, Yu-Lun Tsai7.   

Abstract

PURPOSE: To introduce the benefits of tangent-based volumetric modulated arc therapy (TVMAT), an innovative radiotherapy planning technique, compared with traditional volumetric modulated arc therapy (VMAT) for advanced left breast cancer needing nodal irradiation.
MATERIALS AND METHODS: Twenty-three patients with advanced left breast cancer who had received modified radical mastectomy (MRM) and needed adjuvant radiotherapy including nodal irradiation were assessed. Among 23 radiotherapy treatment plans, 17 plans were designed by using TVMAT technique and 6 plans were designed by using traditional VMAT. The main difference of TVMAT from VMAT was that the area of avoidance sector within specific degrees of angle that had no monitor unit (MU) delivery was used in the arc planning, including a total of 5 sectors in 5 partial arcs. The dosimetries of planning target volume (PTV), right breast, bilateral lungs, and heart between TVMAT and VMAT were compared.
RESULTS: The conformity index (CI) and homogeneity index (HI) of PTV between two groups were statistically equivalent (CI: 0.98 ± 0.02 and 0.98 ± 0.03, P = 0.431; HI: 0.12 ± 0.03 and 0.11 ± 0.05, P = 0.177), which indicated that the treatment efficacy of the plans regarding TVMAT was compatible with VMAT. However, all neighboring organs at risk (OAR) showed a great percentage of reduction in mean doses (right breast: 53.1%, right lung: 37.7%, left lung: 8.8%, heart: 21.2%) and low dose parameters (V10: right breast: 72.3%, right lung: 86.1%, left lung: 12.5%, heart: 25.1%; V5: right breast: 56.5%, right lung: 28.3%, left lung: 12.7%, heart: 18.2%) by using TVMAT.
CONCLUSION: TVMAT greatly decreases the radiation doses delivered to the OAR with maintained therapeutic efficacy. It is highly recommended for treating breast cancer, especially for difficult cases with left side disease needing nodal irradiation.

Entities:  

Keywords:  Breast cancer; Radiation dose; Tangent-based volumetric modulated arc therapy

Mesh:

Year:  2018        PMID: 30486829      PMCID: PMC6260755          DOI: 10.1186/s13014-018-1167-y

Source DB:  PubMed          Journal:  Radiat Oncol        ISSN: 1748-717X            Impact factor:   3.481


Introduction

Breast cancer is one of the most commonly diagnosed malignancies globally. The estimated percentage of breast cancers diagnosed with lymph node involvement is around 30% [1]. Nowadays, the general principle of treating breast cancers with nodal disease is to irradiate adjacent lymph node regions along with the primary site after surgery. Including nodal irradiation in the treatment plans can improve survival and reduce mortality from breast cancer [2-4]. When intending to treat lymph node regions, treatment planning is a challenge because the geometry of treatment volumes is much more complicated than breast irradiation alone. Planning is harder for left side diseases, considering the risk of ischemic heart disease after radiotherapy for breast cancer increases linearly with the mean dose to the heart by 7.4% per gray (Gy) with no apparent threshold [5]. Among planning techniques, VMAT allows for a reduction of the maximum doses to OAR, especially to the heart, while retaining target homogeneity and coverage [6]. Otherwise, it reduces the number of MU by 30% and the treatment time by 55% compared with conventional intensity-modulated radiation therapy (IMRT) [7]. Although VMAT planning has the best intermediate-high dose OAR sparing, it compromises the mean and low doses received by the OAR, and it comes also at the expense of more doses to the contralateral lung and breast [8]. This trade-off even presents at the application of deep inspiration breath-hold (DIBH), which shows that a significant increase in the low dose to the heart as well as low and mean doses to the contralateral lung and breast still exists in VMAT [9]. This drives the motivation to explore a better planning technique to improve the disadvantages of VMAT. The study focuses on this issue and introduces TVMAT, an innovative planning technique based on the modifications of VMAT, with the planning objectives, design details, and dosimetry comparisons with VMAT, for treating advanced left breast cancer needing nodal irradiation.

Materials and methods

Patient and target volume delineation

A total of 23 patients with advanced left breast cancer who had undergone MRM and needed adjuvant radiotherapy including nodal irradiation were assessed. Among them, 17 patients received radiotherapy designed by using TVMAT, and the other 6 plans were designed by using traditional VMAT. During computed tomography (CT) simulation, all patients were lying in the supine position with their arms over their heads and immobilized with an extended wing board with T-bar handgrip immobilization devices (CIVCO Medical Solutions, IA, USA). CT images were acquired using the Discovery CT590 RT 16-slice scanner (GE Healthcare, WI, USA) with a slice thickness of 2.5 mm. The scanning range stretched from the second vertebral body of the cervical spine to the forth vertebral body of the lumbar spine. Regarding target volume delineation, Fig. 1 illustrates the locations and contours with respect to different target volumes. The clinical target volume (CTV) included the left chest wall (CW), axillary lymph node region (AXLN), internal mammary lymph node region (IMLN), supraclavicular fossa (SCF), and infraclavicular fossa (ICF). The PTV was generated by expanding 0.5 cm from the CTV in all directions, but cropped out the area with distance less than 0.3 cm from the surface.
Fig. 1

Target volume delineation of advanced left breast cancer. The CTV included a chest wall (CW, blue line), axillary lymph node region (AXLN, brown line), and internal mammary lymph node region (IMLN, dark blue line) and b supraclavicular fossa (SCF, cyan line) and infraclavicular fossa (ICF, brown line). The PTV (magenta line) was generated by expanding 0.5 cm from the CTV in all directions. c overview of target volume delineation

Target volume delineation of advanced left breast cancer. The CTV included a chest wall (CW, blue line), axillary lymph node region (AXLN, brown line), and internal mammary lymph node region (IMLN, dark blue line) and b supraclavicular fossa (SCF, cyan line) and infraclavicular fossa (ICF, brown line). The PTV (magenta line) was generated by expanding 0.5 cm from the CTV in all directions. c overview of target volume delineation

TVMAT and VMAT planning

All treatment plans were designed with 5 partial arcs with 2 centers using RapidArc® (Varian Medical Systems, Palo Alto, CA, USA). In TVMAT planning, an additional avoidance sector within specific degrees of angle that had no MU delivery was used in each partial arc. Two partial arcs with half-beam fields (ArcN1 and ArcN2) were designed to irradiate SCF and ICF. ArcN1 rotated from 179 to 335 degrees with an avoidance sector from 120 to 60 degrees, and ArcN2 rotated from 35 to 181 degrees with an avoidance sector from 300 to 240 degrees (Fig. 2). Figure 3 shows the delivery of MU per gantry degree of ArcN1, which was a tangent distribution without MU delivered from the left sides of the patients in TVMAT but a semi-circle distribution in VMAT. Another 2 partial arcs (ArcC1 and ArcC2) were designed for CW and AXLN. ArcC1 rotated from 147 (range 130–155, depending on the curve of CW) to 295 (range 282–306) degrees, and ArcC2 rotated from 156 (range 139–179) to 295 (range 282–306) degrees. Both arcs had an avoidance sector from 90 to 0 degrees to reduce lung doses (Fig. 4). Figure 5 demonstrates the delivery of MU per gantry degree of ArcC1, which had no MU delivery from the left-anterior of the patients in TVMAT. ArcC3 was designed to treat IMLN. The field size was just enough to cover the volume of IMLN. It rotated from 270 (range 270–300) to 179 degrees with an avoidance sector from 60 to 120 degrees. The same degrees of partial arcs without avoidance sectors were used in VMAT planning.
Fig. 2

Two partial arcs (ArcN1 and ArcN2) in TVMAT were designed for SCF and ICF. Both were half-beam fields (orange rectangle in (a)). ArcN1 rotated from 179 to 335 degrees with an avoidance sector from 120 to 60 degrees. ArcN2 rotated from 35 to 181 degrees with an avoidance sector from 300 to 240 degrees. a Beam’s eye view of ArcN1. b Angles with radiation in ArcN1 and ArcN2 (red dash lines)

Fig. 3

The delivery of monitor units per gantry degree (MU/deg) of ArcN1. a TVMAT with avoidance sector. b VMAT without avoidance sector

Fig. 4

Two partial arcs (ArcC1 and ArcC2) in TVMAT were designed for CW and AXLN. ArcC1 rotated from 147 (range 130–155) to 295 (range 282–306) degrees. ArcC2 rotated from 156 (range 139–179) to 295 (range 282–306) degrees. Both had an avoidance sector from 90 to 0 degrees to reduce lung doses. a Beam’s eye view of ArcC1. b Angles with radiation in ArcC1 and ArcC2 (red dash lines)

Fig. 5

The delivery of monitor units per gantry degree (MU/deg) of ArcC1. a TVMAT with avoidance sector. b VMAT without avoidance sector

Two partial arcs (ArcN1 and ArcN2) in TVMAT were designed for SCF and ICF. Both were half-beam fields (orange rectangle in (a)). ArcN1 rotated from 179 to 335 degrees with an avoidance sector from 120 to 60 degrees. ArcN2 rotated from 35 to 181 degrees with an avoidance sector from 300 to 240 degrees. a Beam’s eye view of ArcN1. b Angles with radiation in ArcN1 and ArcN2 (red dash lines) The delivery of monitor units per gantry degree (MU/deg) of ArcN1. a TVMAT with avoidance sector. b VMAT without avoidance sector Two partial arcs (ArcC1 and ArcC2) in TVMAT were designed for CW and AXLN. ArcC1 rotated from 147 (range 130–155) to 295 (range 282–306) degrees. ArcC2 rotated from 156 (range 139–179) to 295 (range 282–306) degrees. Both had an avoidance sector from 90 to 0 degrees to reduce lung doses. a Beam’s eye view of ArcC1. b Angles with radiation in ArcC1 and ArcC2 (red dash lines) The delivery of monitor units per gantry degree (MU/deg) of ArcC1. a TVMAT with avoidance sector. b VMAT without avoidance sector The treating linear accelerator was Truebeam STx (Varian Medical Systems, Palo Alto, CA, USA), equipped with a high-definition 120-leaf multileaf collimator with 2.5 mm leaves in the middle and 5 mm leaves on the sides. The treatment plans were designed using Eclipse™ treatment planning system version 11.5 (Varian Medical Systems, Palo Alto, CA, USA). The grid size for dose calculation was 2.5 mm. The maximum dose rate was 600 MU per minute. The radiation beams were 6 MV photon beams. The Anisotropic Analytical Algorithm version 11.0.31 was used for volume dose calculation. The Progressive Resolution Optimizer version 11.0.31 was used for planning optimization.

Planning optimization

The prescribed dose was 50.0 Gy in 2.0 Gy per fraction. Figure 6 illustrates the delineation for planning optimization. It included a 2.0-cm wide ring structure to control the dose falloff, 1 part of the whole right lung, and 2 parts of the left lung, heart, and right breast. Bilateral shoulders were also delineated for optimization to prevent dose attenuation at the shoulders. Table 1 summarizes the objectives for planning optimization regarding each structure. The main scheme was to achieve the D98% of PTV more than 47.5 Gy, which was 95% of the prescribed dose, while maintaining the V20 of the left lung at less than 30% and keeping the mean dose to the right lung less than 5 Gy. An experienced medical physicist made all plans with minimal individual adjustments for different patients.
Fig. 6

Delineation for planning optimization. All structures were at least 0.6 cm away from PTV (magenta line). A 2.0-cm wide ring structure (between two yellow lines) was designed for optimization to control the dose falloff. It was 0.6–2.6 cm away from PTV towards left lung, 1.6–3.6 cm away from PTV towards right breast, and 2.6–4.6 cm away from PTV towards back. Left lung and heart were both designed into two parts depending on whether they were within the ring structure (LungL1 and Heart2) or not (LungL2 and Heart1). The breast tissue within the 5 cm from the medial side was defined as BreastR2, and the rest was BreastR1

Table 1

Dose parameters and priorities for planning optimization

StructureVolume [%]Dose [cGy]PriorityStructureVolume [%]Dose [cGy]PriorityStructureVolume [%]Dose [cGy]Priority
PTV05163950LungL1410550BreastR117668450
1551627103027755011723450
1005160800235805508774450
9851617501412285503838450
817425500870450
42281550
Ring421697100LungL2280500BreastR214891400
31221410020415009906400
192883100122225005975400
103342100741550021050400
34080100171150001089400
LungR27314300Heart1431135300ShoulderL6843300
1840230035124830031009300
1062530028148430001142300
4862300191801300
01139300122209300
62710300
Heart2231042350ShoulderR7552300
1711713503617300
1313113500756300
81465350
51598350
Delineation for planning optimization. All structures were at least 0.6 cm away from PTV (magenta line). A 2.0-cm wide ring structure (between two yellow lines) was designed for optimization to control the dose falloff. It was 0.6–2.6 cm away from PTV towards left lung, 1.6–3.6 cm away from PTV towards right breast, and 2.6–4.6 cm away from PTV towards back. Left lung and heart were both designed into two parts depending on whether they were within the ring structure (LungL1 and Heart2) or not (LungL2 and Heart1). The breast tissue within the 5 cm from the medial side was defined as BreastR2, and the rest was BreastR1 Dose parameters and priorities for planning optimization

Plan evaluation

The CI and HI with respect to the whole PTV between TVMAT and VMAT groups were compared [10, 11]:where V95%PTV is the volume receiving at least 95% of prescribed dose in PTV, VPTV is the whole PTV volume, and V95% is the whole volume receiving at least 95% of the prescribed dose.where D2%, D98%, and D50% are the minimum doses that the most exposed 2, 98, and 50% of PTV receive, respectively. The D98% and D2% of PTV regarding left chest wall and lymph node regions were also assessed. Several dose parameters concerning the dose received by each OAR were evaluated, including right breast (mean dose, D2%, V10, and V5), right lung (mean dose, D2%, V10, and V5), left lung (mean dose, D2%, V20, V10, and V5), and heart (mean dose, D2%, V30, V10, and V5). The statistical analysis of the dosimetry data was performed using independent-sample t test. A P value less than 0.05 was considered significant.

Results

PTV dose evaluation

The average PTV volume was 857 ± 252 cm3 for TVMAT and 881 ± 223 cm3 for VMAT, respectively. Table 2 shows the comparison of PTV doses between TVMAT and VMAT. The CI and HI of the whole PTV showed high conformity (CI: 0.98) and homogeneity (HI: 0.12) in TVMAT plans, which were compatible with VMAT plans (P = 0.431 and 0.177). The D98% and D2% of PTV regarding left chest wall (P = 0.449 and 0.747) and lymph node regions (P = 0.319 and 0.117) were all statistically equivalent as well. The overall results indicated that the treatment efficacies of TVMAT and VMAT were compatible.
Table 2

PTV dose evaluation

TVMATVMATP value
meanSDmeanSD
PTVCI0.980.020.980.030.431
HI0.120.030.110.050.177
PTV-CWD98%[Gy]49.20.948.82.30.449
D2%[Gy]55.00.654.60.80.747
PTV-LND98%[Gy]48.91.349.70.70.319
(SCF, ICF, AXLN, IMLN)D2%[Gy]55.00.754.40.90.117

Abbreviations: TVMAT Tangent-based volumetric modulated arc therapy, VMAT Volumetric modulated arc therapy, SD Standard deviation, PTV Planning target volume, CW Chest wall, LN Lymph node region, SCF Supraclavicular fossa, ICF Infraclavicular fossa, AXLN Axillary lymph node region, IMLN Internal mammary lymph node region, CI Conformity index, HI Homogeneity index

PTV dose evaluation Abbreviations: TVMAT Tangent-based volumetric modulated arc therapy, VMAT Volumetric modulated arc therapy, SD Standard deviation, PTV Planning target volume, CW Chest wall, LN Lymph node region, SCF Supraclavicular fossa, ICF Infraclavicular fossa, AXLN Axillary lymph node region, IMLN Internal mammary lymph node region, CI Conformity index, HI Homogeneity index

OAR dose evaluation

Table 3 demonstrates the dose reductions of OAR in TVMAT compared with VMAT, which were especially large for the contralateral organs (right breast and right lung) and under low dose parameters (V10 and V5). The right lung dose at V10 in TVMAT accounted for the greatest reduction by a percentage regarding the assessed parameters of all organs and was only 13.7% of that in VMAT. The right breast also showed significantly lower doses of all assessed parameters in TVMAT. Figure 7 illustrates the area of low dose reduction of the right breast. Moreover, the mean heart doses of every patient in TVMAT were lower than 15 Gy and at least 3.2 Gy less than that in VMAT. Otherwise, the dose at V5 was significantly reduced in TVMAT. Figure 8 illustrates the area where the heart had lower doses. The percentages of dose reduction in the left lung were not so tremendous, but still had statistical significance at V20, V10, and V5.
Table 3

OAR dose evaluation

TVMATVMATP value
meanSDmeanSD
Right breastmean dose [Gy]3.7 (46.8%)1.47.93.00.001*
D2% [Gy]12.9 (63.5%)4.820.37.40.029*
V10 [%]7.4 (27.6%)7.426.821.40.011*
V5 [%]25.9 (43.6%)13.559.417.1< 0.001*
Right lungmean dose [Gy]3.6 (63.2%)1.15.71.3< 0.001*
D2% [Gy]10.4 (67.1%)2.215.54.90.004*
V10 [%]2.9 (13.7%)2.321.116.50.003*
V5 [%]23.9 (71.6%)12.233.422.40.431
Left lungmean dose [Gy]15.5 (91.2%)1.417.01.10.052
D2% [Gy]50.2 (101.0%)1.149.71.90.354
V20 [%]27.3 (89.5%)2.430.52.20.013*
V10 [%]45.7 (87.5%)5.152.25.50.030*
V5 [%]74.6 (87.4%)11.585.45.00.016*
Heartmean dose [Gy]11.7 (78.5%)3.014.94.50.062
D2% [Gy]40.4 (99.5%)8.340.69.11.000
V30 [%]7.2 (69.2%)4.010.49.70.708
V10 [%]41.7 (74.9%)16.355.717.80.062
V5 [%]75.7 (81.7%)15.492.610.10.030*

*Statistical significance

Abbreviations: TVMAT Tangent-based volumetric modulated arc therapy, VMAT Volumetric modulated arc therapy, SD Standard deviation

Fig. 7

Dose distribution at the level of SCF and ICF. The red circles indicate the reduction in low dose of right breast by using a TVMAT compared with b VMAT

Fig. 8

Dose distribution at the level of CW. The red circles indicate the reduction in low dose of heart by using a TVMAT compared with b VMAT

OAR dose evaluation *Statistical significance Abbreviations: TVMAT Tangent-based volumetric modulated arc therapy, VMAT Volumetric modulated arc therapy, SD Standard deviation Dose distribution at the level of SCF and ICF. The red circles indicate the reduction in low dose of right breast by using a TVMAT compared with b VMAT Dose distribution at the level of CW. The red circles indicate the reduction in low dose of heart by using a TVMAT compared with b VMAT

Plan objectives

The average treatment time of TVMAT was 2.3 ± 0.4 min with an average of 973 ± 189.6 MU being delivered, which was less than VMAT with 1305.5 ± 137.6 MU being delivered in 3.8 ± 0.9 min.

Discussion

VMAT is generally associated with a larger volume of low dose spreading to healthy tissues, which leads to concerns about increasing the risk of secondary malignancies. Although the actual doses received by the contralateral breast and lung are low, the probabilities of secondary cancers are notable. The value of excess relative risk per Gy for secondary cancer in the contralateral breast following radiotherapy of breast cancer is 0.86 Gy− 1, which is reduced by a factor of 2 and set to 0.43 Gy− 1 in fractionated radiotherapy [12-14]. For patients diagnosed with a second primary lung cancer 5 or more years after breast cancer treatment, the rate of lung cancer increases linearly with 8.5% per Gy. This rate is even enhanced for ever-smokers with an excess rate of 17.3% per Gy [15]. However, it is difficult to abandon VMAT because it generally has a better performance at intermediate-high dose sparing and mean heart dose, which is related to ischemic heart disease [6-9]. In the present study, TVMAT comprehensively improved the disadvantages of VMAT. The analysis demonstrated a statistically significant reduction in almost all low dose parameters and contralateral organ parameters. The only contralateral organ parameter that had a reduced but insignificant result was V5 of the right lung. It is probably due to the relatively large standard deviations, which are related to fewer doses delivered to the right lung, and hence large variations between patients. However, in fact, the average V5 of the right lung was reduced from 33.4% in VMAT to 23.9% in TVMAT. Using TVMAT may relieve the worries about secondary cancers while preserving the advantages of VMAT. Some earlier studies used tangential arcs or avoidance sectors in order to improve the dose distribution of VMAT for treating left breast cancer. Fogliata et al. applied an avoidance sector from 0 to 105 degrees to do whole breast irradiation, which presented a reduction of the mean doses for all critical structures with trade-offs of high-dose spillage in the healthy tissue and higher skin dose [16]. Pasler et al. conducted a technique with 2 small tangential arc segments less than 60 degrees and compared it to large-angle VMAT more than 230 degrees for patients needing left breast, supraclavicular lymph nodes, and infraclavicular lymph nodes irradiation. They concluded that the technique provided considerably reduced low dose, especially in the heart and contralateral structures, with the cost of a slightly inferior target coverage and homogeneity [17]. Kuo et al. introduced 5 partial arcs with 3 from the anterior and 2 from the side for the patients with expander or implant reconstructions. The modified VMAT technique avoiding the ipsilateral arm can produce acceptable clinical plans regarding post-mastectomy radiotherapy [18]. The application of 2 centers in TVMAT is unique and noteworthy. Because the anatomy geometries at the upper level over SCF and ICF and the lower level over CW and AXLN are different, it is warranted to separate these 2 parts in treatment planning. The beam angles are essential for taking care of both target coverage and OAR sparing. To make a rainbow-shaped isodose curve over CW and AXLN, the tangent-based beams from the right-anterior and left-posterior of the patients are essential for prevention of the beam penetrating through the body, which is quite different to the beams arranged for SCF and ICF in the anterior-posterior direction to avoid dose to the contralateral breast. Therefore, 2 centers were used in TVMAT other than the feature of avoidance sectors in specific degrees. Furthermore, for a better dose control at the junction of 2 fields, ArcN1 and ArcN2 were set up in half-beam fields refraining beam diversion. IMLN irradiation is strongly considered for nodal positive patients [2-4]. Although sampling of internal mammary sentinel nodes is not performed routinely in contemporary practice, it has to be mentioned that the first filter stations for the lymphatic drainage of breasts are AXLN and IMLN [19]. The incidence of IMLN metastasis is highly correlated with tumor location and AXLN status, which is only 4.4% with no AXLN metastasis, but increases to 18.8% with 1–3 AXLN metastases, 28.1% with 4–6 AXLN metastases, and 41.5% with ≧7 AXLN metastases [20]. In order to treat IMLN, which may generally increase ipsilateral lung dose, DIBH with wide tangents or 3-dimensional conformal radiotherapy (3DCRT) was most employed [21-23]. Osman et al. further compared VMAT with 3DCRT in patients with or without DIBH. They discovered the dose benefits in a better target coverage and a reduction of ipsilateral lung dose and cumulative heart dose with VMAT in DIBH, but at the expense of increasing the dose to the contralateral breast [24]. We used ArcC3 to solve the problem caused by IMLN irradiation. It rendered a field just wide enough to cover the volume of IMLN with specific rotation degrees designed for IMLN. Concerning IMLN locating at the anterior and leaning to the left of the body, the beam contributed more from the anterior than posterior, not directly from the anterior-posterior direction but slightly from the right-anterior to the left-posterior. As the beam angles are essential for taking care of both target coverage and OAR sparing, we took them into account for every individual target towards a more specific approach. This study has some limitations. First, the CT simulation was performed under free-breathing conditions but not in a DIBH setting. The concern was the feasibility of DIBH with VMAT-based planning in clinical implementation. To perform a successful treatment with DIBH, patients have to keep their breath at a consistent and stable level, not only during CT simulation but throughout the whole treatment process. The average treatment time of TVMAT and VMAT was 2.3 and 3.8 min in the study, respectively, which is much longer than people can hold their breath. Therefore, a 4-dimensional CT-based planning and respiratory gating treatment system is mandatory to assist in DIBH treatment with VMAT-based planning. In an institution with cooperative patients and adequate equipment, DIBH with TVMAT planning may further reduce OAR doses compared with VMAT planning. Second, although 2 centers were essential in TVMAT for different beam angles regarding anatomy geometries and half-beam fields were used to avoid beam diversion, setup errors, particularly in the longitudinal axis, may have resulted in dose deviations. In our assessment, hot or cold spots at the junction with about 30% dose deviations presented in recalculation by shifting a center by 0.5 cm in the longitudinal axis after planning. In clinical application, this warrants combining 2 centers to make a single plan, recording the distance between the 2 centers, and moving a patient exactly according to the plan using an automated moving couch with quality assurance. Finally, it is necessary to pay attention to the bolus used in the study. A thick and wide enough bolus, 1 cm in thickness and covering the whole left chest wall, was applied to each patient. One benefit of the bolus was to achieve our prescribed skin dose, which was 50.0 Gy in the study, because skin involvement is a risk in advanced breast cancer. The other benefit was to solve the problem of breathing-induced shifts of the chest wall. The shifts were considered in the target delineation because CTV can directly expand into the bolus on the image to obtain PTV. Because it is not easy to open up multileaf collimators in VMAT-based planning and because TVMAT forces the radiation beams to be mainly tangential, such respiratory organ motions have to be taken into account beforehand in target delineation. For the patients without a bolus, we recommend expanding the contour into the air and designing TVMAT with a replaced virtual bolus in planning. The above considerations are the necessary precautions when employing TVMAT in clinical practice. TVMAT can be applied properly with attention to the details.

Conclusion

TVMAT is an innovative radiotherapy planning technique designed for advanced left breast cancer. It comprehensively improves the disadvantages of traditional VMAT with respect to low dose spreading and detriments of contralateral organs on treatment plan dosimetry while maintaining target coverage and homogeneity. It creates a paradigm for management of difficult cases needing nodal irradiation in clinical applications.
  22 in total

1.  Volumetric modulated arc therapy and breath-hold in image-guided locoregional left-sided breast irradiation.

Authors:  Sarah O S Osman; Sandra Hol; Philip M Poortmans; Marion Essers
Journal:  Radiother Oncol       Date:  2014-05-10       Impact factor: 6.280

2.  Dosimetric trade-offs in breast treatment with VMAT technique.

Authors:  Antonella Fogliata; Jan Seppälä; Giacomo Reggiori; Francesca Lobefalo; Valentina Palumbo; Fiorenza De Rose; Davide Franceschini; Marta Scorsetti; Luca Cozzi
Journal:  Br J Radiol       Date:  2017-01-03       Impact factor: 3.039

3.  Internal Mammary and Medial Supraclavicular Irradiation in Breast Cancer.

Authors:  Philip M Poortmans; Sandra Collette; Carine Kirkove; Erik Van Limbergen; Volker Budach; Henk Struikmans; Laurence Collette; Alain Fourquet; Philippe Maingon; Mariacarla Valli; Karin De Winter; Simone Marnitz; Isabelle Barillot; Luciano Scandolaro; Ernest Vonk; Carla Rodenhuis; Hugo Marsiglia; Nicola Weidner; Geertjan van Tienhoven; Christoph Glanzmann; Abraham Kuten; Rodrigo Arriagada; Harry Bartelink; Walter Van den Bogaert
Journal:  N Engl J Med       Date:  2015-07-23       Impact factor: 91.245

4.  Radiation during deep inspiration allows loco-regional treatment of left breast and axillary-, supraclavicular- and internal mammary lymph nodes without compromising target coverage or dose restrictions to organs at risk.

Authors:  Mari H B Hjelstuen; Ingvil Mjaaland; Johan Vikström; Kjell Ivar Dybvik
Journal:  Acta Oncol       Date:  2011-12-16       Impact factor: 4.089

5.  Estimated risk for secondary cancer in the contra-lateral breast following radiation therapy of breast cancer.

Authors:  Safora Johansen; Turi Danielsen; Dag Rune Olsen
Journal:  Acta Oncol       Date:  2008       Impact factor: 4.089

6.  Left-sided breast cancer loco-regional radiotherapy with deep inspiration breath-hold: Does volumetric-modulated arc radiotherapy reduce heart dose further compared with tangential intensity-modulated radiotherapy?

Authors:  Trang T Pham; Rachel Ward; Drew Latty; Catherine Owen; Val Gebski; Jacek Chojnowski; Christopher Kelly; Verity Ahern; Kenneth Tiver; Kirsty Stuart; Wei Wang
Journal:  J Med Imaging Radiat Oncol       Date:  2016-04-20       Impact factor: 1.735

7.  Internal mammary lymph drainage and sentinel node biopsy in breast cancer - A study on 1008 patients.

Authors:  E M Heuts; F W C van der Ent; M F von Meyenfeldt; A C Voogd
Journal:  Eur J Surg Oncol       Date:  2008-08-05       Impact factor: 4.424

8.  Significant reductions in heart and lung doses using deep inspiration breath hold with active breathing control and intensity-modulated radiation therapy for patients treated with locoregional breast irradiation.

Authors:  Vincent M Remouchamps; Frank A Vicini; Michael B Sharpe; Larry L Kestin; Alvaro A Martinez; John W Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-02-01       Impact factor: 7.038

9.  Wide tangential fields including the internal mammary lymph nodes in patients with left-sided breast cancer. Influence of respiratory-controlled radiotherapy (4D-CT) on cardiac exposure.

Authors:  Heidi Stranzl; Brigitte Zurl; Tanja Langsenlehner; Karin S Kapp
Journal:  Strahlenther Onkol       Date:  2009-03-28       Impact factor: 3.621

10.  A planning comparison of dose patterns in organs at risk and predicted risk for radiation induced malignancy in the contralateral breast following radiation therapy of primary breast using conventional, IMRT and volumetric modulated arc treatment techniques.

Authors:  Safora Johansen; Luca Cozzi; Dag Rune Olsen
Journal:  Acta Oncol       Date:  2009       Impact factor: 4.089

View more
  7 in total

1.  Three-dimensional conformal radiotherapy (3D-CRT) vs. volumetric modulated arc therapy (VMAT) in deep inspiration breath-hold (DIBH) technique in left-sided breast cancer patients-comparative analysis of dose distribution and estimation of projected secondary cancer risk.

Authors:  Iga Racka; Karolina Majewska; Janusz Winiecki
Journal:  Strahlenther Onkol       Date:  2022-08-09       Impact factor: 4.033

2.  In search of a one plan solution for VMAT post-mastectomy chest wall irradiation.

Authors:  T T Monajemi; P A K Oliver; A Day; M Yewondwossen
Journal:  J Appl Clin Med Phys       Date:  2020-06-27       Impact factor: 2.102

3.  Non-coplanar volumetric modulated arc therapy for locoregional radiotherapy of left-sided breast cancer including internal mammary nodes.

Authors:  Yuan Xu; Pan Ma; Zhihui Hu; Yuan Tian; Kuo Men; Shulian Wang; Yingjie Xu; Jianrong Dai
Journal:  Radiol Oncol       Date:  2021-11-19       Impact factor: 2.991

4.  Half-beam volumetric-modulated arc therapy in adjuvant radiotherapy for gynecological cancers.

Authors:  Pei-Chieh Yu; Ching-Jung Wu; Hsin-Hua Nien; Louis Tak Lui; Suzun Shaw; Yu-Lun Tsai
Journal:  J Appl Clin Med Phys       Date:  2021-11-16       Impact factor: 2.102

5.  Feasibility study: spot-scanning proton arc therapy (SPArc) for left-sided whole breast radiotherapy.

Authors:  Sheng Chang; Gang Liu; Lewei Zhao; Joshua T Dilworth; Weili Zheng; Saada Jawad; Di Yan; Peter Chen; Craig Stevens; Peyman Kabolizadeh; Xiaoqiang Li; Xuanfeng Ding
Journal:  Radiat Oncol       Date:  2020-10-07       Impact factor: 3.481

6.  Comprehensive nodal breast VMAT: solving the low-dose wash dilemma using an iterative knowledge-based radiotherapy planning solution.

Authors:  Cameron Stanton; Linda J Bell; Andrew Le; Brooke Griffiths; Kenny Wu; Jessica Adams; Leigh Ambrose; Denise Andree-Evarts; Brian Porter; Regina Bromley; Kirsten van Gysen; Marita Morgia; Gillian Lamoury; Thomas Eade; Jeremy T Booth; Susan Carroll
Journal:  J Med Radiat Sci       Date:  2021-08-12

7.  Auxiliary Structures-Assisted Radiotherapy Improvement for Advanced Left Breast Cancer.

Authors:  Runhong Lei; Xile Zhang; Jinna Li; Haitao Sun; Ruijie Yang
Journal:  Front Oncol       Date:  2021-07-08       Impact factor: 6.244

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.