Effect of canal curvature location on the cyclic fatigue resistance of reciprocating files.

Tyler Sobotkiewicz1, Xiangya Huang2, Markus Haapasalo1, Christophe Mobuchon3, Ahmed Hieawy1, Jinghao Hu1, Huimin Zhou4, Zhejun Wang1, Ya Shen5.   

Abstract

OBJECTIVES: To determine the effect of the location of the canal curvature on the fatigue resistance of WaveOne (WO), WaveOne Gold (WOG), Reciproc (Rec), and Reciproc Blue (RecB) files, and to examine the phase transformation behaviors of the reciprocating file systems.
MATERIAL AND METHODS: The instruments were subjected to fatigue testing in five artificial canals with a curvature of 60° angle and a 3-mm radius. The location of the curvature was unique for each canal. Each file was inserted 16 mm into the canal and operated until fracture occurred. The time to fracture was recorded and the length of the fragment was measured. Differential scanning calorimetry was used to characterize the thermal behavior of the files. The number of cycles to failure was analyzed using two-way analysis of variance and the post hoc Tukey test. The Kruskal-Wallis test was used to compare the mean fragment lengths between groups.
RESULTS: The instruments had significantly lower fatigue resistance in canals with curvatures in the middle and coronal canals compared with those with apical curvatures (p < 0.05). At all tested curvature locations, RecB had superior fatigue resistance compared with WO and Rec (p < 0.05). There were no significant differences between WOG and Rec in canals with curvatures in the middle and coronal canals. The DSC thermograms for RecB exhibit a single exothermic peak during cooling but double endothermic peaks during heating indicating that a two-step phase transformation from martensite to R-phase to austenite takes place.
CONCLUSIONS: The reciprocating instruments experience decreased cyclic fatigue resistance when operated in canals with coronal- and middle-third curvatures when compared with curvatures in the apical-third. Instrumenting coronally positioned curvatures with reciprocating files needs to be performed with caution. CLINICAL RELEVANCE: The location of the root canal's curvature influences the fracture resistance of rotary files that are used with reciprocating movements. Therefore, caution needs to be exercised when using reciprocating instruments in canals with coronal or middle curvatures.

Entities:  

Keywords:  Curvature location; Fatigue resistance; Reciproc; Reciproc Blue; WaveOne; WaveOne Gold

Year:  2020        PMID: 32488488     DOI: 10.1007/s00784-020-03348-8

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  32 in total

1.  Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments.

Authors:  Y Haïkel; R Serfaty; G Bateman; B Senger; C Allemann
Journal:  J Endod       Date:  1999-06       Impact factor: 4.171

2.  Cyclic fatigue testing of nickel-titanium endodontic instruments.

Authors:  J P Pruett; D J Clement; D L Carnes
Journal:  J Endod       Date:  1997-02       Impact factor: 4.171

3.  Cyclic fatigue of nickel-titanium rotary instruments in a double (S-shaped) simulated curvature.

Authors:  Dina Al-Sudani; Nicola M Grande; Gianluca Plotino; Giorgio Pompa; Stefano Di Carlo; Luca Testarelli; Gianluca Gambarini
Journal:  J Endod       Date:  2012-05-08       Impact factor: 4.171

4.  The impact of instrument fracture on outcome of endodontic treatment.

Authors:  Peter Spili; Peter Parashos; Harold H Messer
Journal:  J Endod       Date:  2005-12       Impact factor: 4.171

5.  A retrospective clinical study of incidence of root canal instrument separation in an endodontics graduate program: a PennEndo database study.

Authors:  Mian K Iqbal; Meetu R Kohli; Jessica S Kim
Journal:  J Endod       Date:  2006-09-25       Impact factor: 4.171

6.  Defects in nickel-titanium instruments after clinical use. Part 2: Fractographic analysis of fractured surface in a cohort study.

Authors:  Ya Shen; Gary Shun-Pan Cheung; Bin Peng; Markus Haapasalo
Journal:  J Endod       Date:  2009-01       Impact factor: 4.171

7.  Prevalence and management of instrument fracture in the postgraduate endodontic program at the Dental School of Athens: a five-year retrospective clinical study.

Authors:  Giorgos N Tzanetakis; Evangelos G Kontakiotis; Dimitra V Maurikou; Maria P Marzelou
Journal:  J Endod       Date:  2008-04-11       Impact factor: 4.171

8.  Influence of curvature location along an artificial canal on cyclic fatigue of a rotary nickel-titanium endodontic instrument.

Authors:  Hélio Pereira Lopes; Wanderson Miguel Maia Chiesa; Norbi Rodrigues Correia; Núbia Carvalho de Souza Navegante; Carlos Nelson Elias; Edson Jorge Lima Moreira; Bianca Esther Cabral Chiesa
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2011-03-27

9.  Defects in ProTaper S1 instruments after clinical use: fractographic examination.

Authors:  G S P Cheung; B Peng; Z Bian; Y Shen; B W Darvell
Journal:  Int Endod J       Date:  2005-11       Impact factor: 5.264

10.  Influence of the geometry of curved artificial canals on the fracture of rotary nickel-titanium instruments subjected to cyclic fatigue tests.

Authors:  Hélio P Lopes; Márcia V B Vieira; Carlos N Elias; Lucio S Gonçalves; José F Siqueira; Edson J L Moreira; Victor T L Vieira; Letícia C Souza
Journal:  J Endod       Date:  2013-02-20       Impact factor: 4.171

View more
  1 in total

1.  Influence of needle working length and root canal curvature on irrigation: a computational fluid dynamics analysis based on a real tooth.

Authors:  Na Zhou; Zhengqiu Huang; Mingzhou Yu; Shuli Deng; Baiping Fu; Hanhui Jin
Journal:  BMC Oral Health       Date:  2022-05-14       Impact factor: 2.757

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.