Literature DB >> 32895860

Rb-82 generator yield.

Adrian Nunn1, Peter Oehlberg2.   

Abstract

Entities:  

Year:  2020        PMID: 32895860      PMCID: PMC7599146          DOI: 10.1007/s12350-020-02287-x

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


× No keyword cloud information.
We read with interest the article by Ahmadi et al.1 reporting the results of an investigation led by the inventors and developers of RUBY-FILL®, one of the two marketed 82Sr/ 82Rb generators being compared. We are concerned that the authors incorrectly state that the CardioGen-82® infusion system does not report the integrated total 82Rb activity as part of daily quality control (QC) procedures, but rather the dose calibrator maximum value. They also provide no support for their statement that neither the Patient Dose nor the End-of-Infusion activity values appear to be accurate in the CardioGen-82®daily QC breakthrough elution reports. Moreover, based on the use of a correction factor of 1.314 applied to the dose calibrator maximum values, they suggest that the RUBY-FILL® generator yield is ~ 7% higher than that of CardioGen-82®. The CardioGen-82® system does provide an integrated dose for each infusion, the Patient Dose, as the authors ultimately acknowledge. For CardioGen-82®, the average difference between the dose calibrator maximum value and the Patient Dose for daily QC is actually a factor of 1.426, not the 1.314 the authors derived—a difference of ~ 8%.2 Thus, if the true correction factor with the dose calibrator maximum values or the actual integrated calibration doses are used, CardioGen-82® yields are similar to RUBY-FILL ® yields. However, the concept of a yield or elution efficiency is arbitrary when the product has such a short half-life as 82Rb, because the 'yield' is dependent on the elution conditions. As the authors note, these are quite different both during daily QC and during normal use for the two systems. What matters is whether the user can get the desired doses over the lifetime of the generator. In this regard, the maximum dose available is constrained for both generators by the half-life of the 82Sr.
  1 in total

1.  Rubidium-82 generator yield and efficiency for PET perfusion imaging: Comparison of two clinical systems.

Authors:  Ali Ahmadi; Ran Klein; Howard C Lewin; Rob S B Beanlands; Robert A deKemp
Journal:  J Nucl Cardiol       Date:  2020-05-20       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.