Minimizing rubidium-82 tracer activity for relative PET myocardial perfusion imaging.

Eline D Huizing1, Joris D van Dijk, Jorn A van Dalen, Jorik R Timmer, Hester Arkies, Cees H Slump, Pieter L Jager.   

Abstract

OBJECTIVES: Recommended rubidium-82 activities for relative myocardial perfusion imaging (MPI) using present-generation PET scanners may be unnecessarily high. Our aim was to derive the minimum activity for a reliable relative PET MPI assessment.
MATERIALS AND METHODS: We analyzed 140 scans from 28 consecutive patients who underwent rest-stress MPI-PET (Ingenuity TF). Scans of 852, 682, 511, and 341 MBq were simulated from list-mode data and compared with a reference scan using 1023 MBq. Differences in the summed rest score, total perfusion deficit, and image quality were obtained between the reference and each of the simulated rest scans. Combined stress-rest scans obtained at a selected activity of 682 MBq were diagnostically interpreted by experts and outcome was compared with the reference scan interpretation.
RESULTS: Differences in summed rest score more than or equal to 3 were found using 682, 511, and 341 MBq in two (7%), four (14%), and five (18%) patients, respectively. Differences in total perfusion deficit more than 7% were only found at 341 MBq in one patient. Image quality deteriorated significantly only for the 341 MBq scans (P<0.001). Interpretation of stress-rest scans did not differ between 682 and 1023 MBq scans.
CONCLUSION: A significant reduction in administered Rb-82 activity is feasible in relative MPI. An activity of 682 MBq resulted in reliable diagnostic outcomes and image quality, and can therefore be considered for clinical adoption.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28703721     DOI: 10.1097/MNM.0000000000000701

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  2 in total

1.  Comparison of maximal Rubidium-82 activities for myocardial blood flow quantification between digital and conventional PET systems.

Authors:  Joris D van Dijk; Pieter L Jager; Jochen A C van Osch; Maryam Khodaverdi; Jorn A van Dalen
Journal:  J Nucl Cardiol       Date:  2018-01-16       Impact factor: 5.952

2.  Rubidium-82 generator yield and efficiency for PET perfusion imaging: Comparison of two clinical systems.

Authors:  Ali Ahmadi; Ran Klein; Howard C Lewin; Rob S B Beanlands; Robert A deKemp
Journal:  J Nucl Cardiol       Date:  2020-05-20       Impact factor: 5.952

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.