Comparison of Cepheid Xpert Xpress and Abbott ID Now to Roche cobas for the Rapid Detection of SARS-CoV-2.

Marie C Smithgall1, Ioana Scherberkova2, Susan Whittier1, Daniel A Green3.   

Abstract

BACKGROUND: The SARS-CoV-2 pandemic has created an urgent and unprecedented need for rapid large-scale diagnostic testing to inform timely patient management. However, robust data are lacking on the relative performance of available rapid molecular tests across a full range of viral concentrations.
OBJECTIVE: This study aimed to compare two recently-authorized rapid tests, Cepheid Xpert Xpress SARS-CoV-2 and Abbott ID Now SARS-CoV-2, to the Roche cobas SARS-CoV-2 assay for samples with low, medium, and high viral concentrations. STUDY
DESIGN: A total of 113 nasopharyngeal swabs from remnant patient samples were tested, including 88 positives spanning the full range of observed Ct values on the cobas assay.
RESULTS: Compared to cobas, the overall positive agreement was 73.9% with ID Now and 98.9% with Xpert. Negative agreement was 100% and 92.0% for ID Now and Xpert, respectively. Both ID Now and Xpert showed 100% positive agreement for medium and high viral concentrations (Ct value <30). However, for Ct values >30, positive agreement was 34.3% for ID Now and 97.1% for Xpert.
CONCLUSIONS: While Xpert showed high agreement with cobas across a wide range of viral concentrations, this study highlights an important limitation of ID Now for specimens collected in viral or universal transport media with low viral concentrations. Further studies are needed to evaluate the performance of ID Now for direct swabs.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  COVID-19; Molecular Diagnostics; PCR; Point of Care; SARS-CoV-2

Mesh:

Substances:

Year:  2020        PMID: 32434706      PMCID: PMC7217789          DOI: 10.1016/j.jcv.2020.104428

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Introduction

Severe acute respiratory virus coronavirus 2 (SARS-CoV-2) emerged in Wuhan, China in December 2019 and has since rapidly spread across the world, causing a global pandemic of coronavirus disease (COVID-19). The majority of cases are mild to moderate, but severe infections have overwhelmed healthcare systems in the United States, particularly in New York City. Real-time polymerase chain reaction (RT-PCR) of viral RNA from nasal or nasopharyngeal swabs has become the standard method used to confirm diagnosis. The first quantitative RT-PCR test for detecting SARS-CoV2 was designed and distributed in January 2020 by the World Health Organization (WHO) (1). In the United States and many other countries, however, the slow rollout of large-scale diagnostic testing and the long turnaround times associated with laboratory tests, particularly those sent to reference laboratories, have significantly hampered public health efforts to contain the outbreak. In contrast, commercially-available rapid diagnostic assays can better inform timely patient management decisions to guide the need for quarantine, isolation, contact tracing, and therapeutic management. Beginning in March 2020, multiple SARS-CoV-2 rapid tests received Emergency Use Authorization (EUA) from the U.S. Food and Drug Administration (FDA). However, manufacturer submissions for EUA only require evaluation of the limit of detection and cross-reactivity of their assays, and do not address other important performance characteristics such as accuracy, precision, and reproducibility. In addition, most manufacturer submissions include studies of contrived positive samples with spiked viral RNA, and do not assess performance on clinical patient specimens. Two recently-authorized rapid tests, Xpert Xpress SARS-CoV-2 (Cepheid, Sunnyvale, CA) and ID Now SARS-CoV-2 (Abbott, Chicago, IL) have been manufactured at wide-scale and distributed to numerous medical centers around the country. While initial studies on these two assays have been recently published, the number of patient samples evaluated to date has been relatively small, and significant questions remain about the accuracy of these tests across the full spectrum of viral loads ([2], [3], [4], [5]). Utilizing the high volume of patient testing performed at our medical center in New York City, we sought to evaluate and compare the performance of these two rapid assays across a wide range of clinical samples.

Materials and Methods

Study Design and Data Analysis

Deidentified remnant patient samples that underwent routine clinical testing with the cobas SARS-CoV-2 assay on the 6800 platform (Roche Diagnostics, Indianapolis, IN) were used to evaluate the Xpert and ID Now assays. Residual nasopharyngeal (NP) swabs in transport media were held at 4 °C prior to testing on the Xpert and ID NOW platforms, with all testing completed within 48 hours of sample collection. Testing was performed according to the manufacturers’ instructions on two separate ID Now instruments and a single GeneXpert Infinity instrument. A total of 113 NP swabs collected in 3 mL of viral transport media (M4RT VTM; ThermoFisher Scientific, Waltham, MA) or universal transport media (UTM; Becton Dickinson and Co., Franklin Lakes, NJ) were included. The specimens were collected from April 8 to April 13, 2020 and included 111 adult and 2 pediatric patients who were all seen in inpatient or emergency room locations. To evaluate assay performance at varying viral concentrations, 88 positive specimens were selected to represent the full range of observed Ct values on the cobas assay, ranging from 14 – 38 cycles. Positive agreement and 95% confidence intervals for the Xpert and ID Now assays were calculated using cobas as the reference test. An additional 25 negative specimens were selected to evaluate negative agreement.

Assay Descriptions

The cobas assay utilizes RT-PCR to amplify and detect two viral targets: ORF1 a/b, a non-structural region that is unique to SARS-CoV-2 and a conserved region in the E-gene, which is a structural protein envelope for pan-Sarbecovirus detection. The Xpert assay is an automated RT-PCR that amplifies and detects two viral targets: N2, a nucleocapsid recombinant protein unique to SARS-CoV-2 and a region in the structural envelope E-gene. The ID Now assay uses proprietary isothermal nucleic acid amplification technology for qualitative detection of SARS-CoV-2 RdRp gene using fluorescent reporter probes. This study was approved by the Columbia University Irving Medical Center Institutional Review Board.

Results

Of the 113 patient specimens, 111 were from adults ranging in age from 23 to 101 years and two were from pediatric patients, aged 1 day and 5 days old. The average age was 65 years for positive samples and 43 years for negative samples. Overall, the majority of positive samples were from males (60.2%) and negative samples from females (68.0%) (Table 1 ).
Table 1

Demographics of included patients.

Cobas Ct CategoryAverage Age (years)Male (%)Female (%)
Total Positive64.953 (60.2)35 (39.8)
Low (>30)60.919 (54.3)16 (45.7)
Medium (20-30)67.124 (63.2)14 (36.8)
High (<20)68.410 (66.6)5 (33.3)
Negative42.68 (32.0)17 (68.0)
Demographics of included patients. Testing results by Abbott ID Now and Cepheid Xpert are shown in Table 2 and Fig. 1 . Compared to cobas, the overall positive agreement with ID Now was 73.9% (95% Confidence Interval (CI): 63.2 – 82.3%) and with Xpert was 98.9% (95% CI 92.9 – 100%). Negative agreement was 100% (95% CI 83.4 – 100%) and 92.0% (95% CI 72.4 – 98.6%) for ID Now and Xpert, respectively. Both ID Now and Xpert showed 100% positive agreement for medium and high viral concentrations, defined as Ct value <30. However, for Ct values >30, positive agreement for ID Now was 34.3% (95% CI 19.7 – 52.2%), whereas it was 97.1% (95% CI 83.4 – 99.8%) for Xpert. Notably, one sample detected by Xpert was a presumptive positive based on detection of the E-gene target but not the N2 target. There were also two samples that tested negative by cobas but positive by the Xpert. These samples had Ct values >40 for the N2 target only without detection of the E-gene target. For the E-gene target, Ct values were generally 1 cycle lower for Xpert than cobas (Supplemental Materials, Table S1 and Fig. S1).
Table 2

Positive and negative agreement of Abbott ID Now SARS-CoV-2 and Cepheid Xpert Xpress SARS-CoV-2 with Roche cobas SARS-CoV-2.

Cobas Ct CategoryID Now (%, 95% CI)Xpert (%, 95% CI)Total
Total Positive65 (73.9, 63.2-82.3)87 (98.9, 92.9-99.9)88
Low (>30)12 (34.3, 19.7-52.2)34* (97.1, 83.4-99.8)35
Medium (20-30)38 (100, 88.6-100)38 (100, 88.6-100)38
High (<20)15 (100, 74.7-100)15 (100, 74.7-100)15
Negative25 (100, 83.4-100)23 (92.0, 72.4-98.6)25

One specimen was presumptively positive by Xpert.

Fig. 1

A: Frequency distribution of cycle threshold (Ct) values for positive patient samples by Roche cobas SARS-CoV-2 (blue), Cepheid Xpert Xpress SARS-CoV-2 (orange) and Abbott ID Now SARS-CoV-2(gray) assays. Roche cobas Target 1 (ORF1) Ct values were rounded to the nearest whole number. B: Frequency distribution of cycle threshold (Ct) values for positive patient samples by Roche cobas SARS-CoV-2 (blue), Cepheid Xpert Xpress SARS-CoV-2 (orange) and Abbott ID Now SARS-CoV-2(gray) assays. Roche cobas Target 1 (ORF1) Ct values were rounded to the nearest whole number.

Positive and negative agreement of Abbott ID Now SARS-CoV-2 and Cepheid Xpert Xpress SARS-CoV-2 with Roche cobas SARS-CoV-2. One specimen was presumptively positive by Xpert. A: Frequency distribution of cycle threshold (Ct) values for positive patient samples by Roche cobas SARS-CoV-2 (blue), Cepheid Xpert Xpress SARS-CoV-2 (orange) and Abbott ID Now SARS-CoV-2(gray) assays. Roche cobas Target 1 (ORF1) Ct values were rounded to the nearest whole number. B: Frequency distribution of cycle threshold (Ct) values for positive patient samples by Roche cobas SARS-CoV-2 (blue), Cepheid Xpert Xpress SARS-CoV-2 (orange) and Abbott ID Now SARS-CoV-2(gray) assays. Roche cobas Target 1 (ORF1) Ct values were rounded to the nearest whole number.

Discussion

To meet the urgent need for wide-scale diagnostic testing during the COVID-19 pandemic, multiple rapid molecular tests have recently been authorized by the US FDA, some of which are available in point-of-care (POC) or near patient settings. However, very few studies have been published to date on the relative performance characteristics of these assays, especially for patient specimens representing a wide range of viral concentrations ([2], [3], [4], [5]). In this comparative analysis, the Xpert assay showed a very high level of agreement with the cobas assay across the entire range of tested Ct values, including low-level positives. These findings confirm those published by Moran et al. (2) and show a high level of agreement between these two assays using an expanded number of positive clinical samples. In contrast, the ID Now assay reliably detected specimens with Ct values ≤30, but did not detect the majority of specimens with Ct values ≥ 30. Whereas Rhoades et al. (3) found an overall high level of agreement between ID Now and the modified CDC assay, Hogan et al. (5) demonstrated lower agreement between ID Now and Panther Fusion (Hologic, Inc., Marlborough, MA). Our findings further highlight an important limitation of the ID Now for low-level positives. While all studies evaluated nasopharyngeal swabs eluted in transport media, it is important to note that the EUA for ID NOW was recently updated to remove the indication for swabs in transport media (6). Our data support that the EUA was appropriately modified, as samples may become too dilute in VTM and low-level positives may falsely test negative. In contrast to batch testing and the higher complexity required for the cobas assay, both Xpert and ID Now offer shorter turnaround times and availability in near-patient settings. However, the two assays differ in throughput capacity and run time. Each ID Now platform can run only a single specimen at a time, with results available in 13 minutes or less. Xpert can be run on larger, random-access platforms that allow for significantly higher throughput, with results available in 45 minutes. Both assays are available for use in POC settings, which introduces both benefits and drawbacks. On the one hand, POC molecular testing delivers the shortest possible interval between sample collection and result, which can facilitate faster clinical decision-making. However, concerns related to assay performance, quality management, and safety in the POC setting still remain. Studies of POC molecular testing for influenza and respiratory syncytial virus have shown promising results, but also highlight some of these concerns ([7], [8], [9], [10]). The risk of contamination and false positives is also much higher when testing is performed outside of a controlled environment and by non-laboratory trained personnel. Nevertheless, the unique circumstances of a rapidly-spreading pandemic may ultimately necessitate the widescale adoption of POC assays in completely new patient settings, such as walk up or drive-thru testing centers. Limitations of this study include the relatively few number of samples from pediatric patients, as only two samples from patients aged 1 day and 5 days were included, and both of these were negative on all three testing methods. We were also only able to evaluate ID Now for specimens in transport media. The performance of this assay with direct nasal swabs requires further evaluation in subsequent studies. Another limitation is the use of the cobas assay as the comparator assay. Two samples that were identified as positive only by Xpert on the basis of N2 nucleocapsid gene detection were negative for both targets on cobas. Whether these samples were truly positive or truly negative could not be determined. Fast, readily available, and reliable test results are critically important during this pandemic, and each of the three assays evaluated in this study holds promise to deliver valuable clinical information. Further head-to-head comparisons of molecular tests will be important in order to establish the usefulness of each method and to help medical providers determine the most appropriate diagnostic tests to best serve their patients and communities.

Declaration of Competing Interest

None.

CRediT authorship contribution statement

Marie C. Smithgall: Data curation, Formal analysis, Investigation, Methodology, Validation, Writing - original draft. Ioana Scherberkova: Investigation, Validation, Writing - review & editing. Susan Whittier: Conceptualization, Methodology, Supervision, Writing - review & editing. Daniel A. Green: Conceptualization, Methodology, Formal analysis, Supervision, Writing - review & editing.
  9 in total

1.  Direct Comparison of Alere i and cobas Liat Influenza A and B Tests for Rapid Detection of Influenza Virus Infection.

Authors:  Frederick S Nolte; Lori Gauld; Susan B Barrett
Journal:  J Clin Microbiol       Date:  2016-08-31       Impact factor: 5.948

2.  Coronavirus and the race to distribute reliable diagnostics.

Authors:  Cormac Sheridan
Journal:  Nat Biotechnol       Date:  2020-04       Impact factor: 54.908

3.  Evaluation of the molecular Xpert Xpress Flu/RSV assay vs. Alere i Influenza A & B assay for rapid detection of influenza viruses.

Authors:  J H Chen; H Y Lam; C C Yip; V C Cheng; J F Chan; T H Leung; S Sridhar; K H Chan; B S Tang; K Y Yuen
Journal:  Diagn Microbiol Infect Dis       Date:  2017-11-24       Impact factor: 2.803

4.  Multicenter Clinical Evaluation of the Alere i Respiratory Syncytial Virus Isothermal Nucleic Acid Amplification Assay.

Authors:  Ferdaus Hassan; Lindsay M Hays; Aleta Bonner; Bradley J Bradford; Ruffin Franklin; Phyllis Hendry; Jed Kaminetsky; Michael Vaughn; Kristin Cieslak; Mary E Moffatt; Rangaraj Selvarangan
Journal:  J Clin Microbiol       Date:  2018-02-22       Impact factor: 5.948

5.  Detection of influenza A and B with the Alere ™ i Influenza A & B: a novel isothermal nucleic acid amplification assay.

Authors:  Briony Hazelton; Timothy Gray; Jennifer Ho; V Mala Ratnamohan; Dominic E Dwyer; Jen Kok
Journal:  Influenza Other Respir Viruses       Date:  2015-02-27       Impact factor: 4.380

6.  Clinical Evaluation of the cobas SARS-CoV-2 Test and a Diagnostic Platform Switch during 48 Hours in the Midst of the COVID-19 Pandemic.

Authors:  Mario Poljak; Miša Korva; Nataša Knap Gašper; Kristina Fujs Komloš; Martin Sagadin; Tina Uršič; Tatjana Avšič Županc; Miroslav Petrovec
Journal:  J Clin Microbiol       Date:  2020-05-26       Impact factor: 5.948

7.  Detection of SARS-CoV-2 by Use of the Cepheid Xpert Xpress SARS-CoV-2 and Roche cobas SARS-CoV-2 Assays.

Authors:  Angelica Moran; Kathleen G Beavis; Scott M Matushek; Carol Ciaglia; Nina Francois; Vera Tesic; Nedra Love
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

8.  Five-minute point-of-care testing for SARS-CoV-2: Not there yet.

Authors:  Catherine A Hogan; Malaya K Sahoo; ChunHong Huang; Natasha Garamani; Bryan Stevens; James Zehnder; Benjamin A Pinsky
Journal:  J Clin Virol       Date:  2020-05-01       Impact factor: 3.168

9.  Comparison of Abbott ID Now, DiaSorin Simplexa, and CDC FDA Emergency Use Authorization Methods for the Detection of SARS-CoV-2 from Nasopharyngeal and Nasal Swabs from Individuals Diagnosed with COVID-19.

Authors:  Daniel D Rhoads; Sree S Cherian; Katharine Roman; Lisa M Stempak; Christine L Schmotzer; Navid Sadri
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

  9 in total
  72 in total

1.  The PRONTO study: Clinical performance of ID NOW in individuals with compatible SARS-CoV-2 symptoms in walk-in centres-accelerated turnaround time for contact tracing.

Authors:  Isabelle Goupil-Sormany; Jean Longtin; Jeannot Dumar; Marieve Jacob-Wagner; Frédéric Bouchard; Liliana Romero; Julie Harvey; Julie Bestman-Smith; Mathieu Provençal; Stéphanie Beauchemin; Valérie Richard; Annie-Claude Labbé
Journal:  Can Commun Dis Rep       Date:  2021-12-09

Review 2.  Current Trends and Future Approaches in Small-Molecule Therapeutics for COVID-19.

Authors:  Mark Laws; Yasmin M Surani; Md Mahbub Hasan; Yiyuan Chen; Peiqin Jin; Taha Al-Adhami; Madiha Chowdhury; Aqeel Imran; Ioannis Psaltis; Shirin Jamshidi; Kazi S Nahar; Khondaker Miraz Rahman
Journal:  Curr Med Chem       Date:  2021       Impact factor: 4.530

Review 3.  Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.

Authors:  Seyed Hamid Safiabadi Tali; Jason J LeBlanc; Zubi Sadiq; Oyejide Damilola Oyewunmi; Carolina Camargo; Bahareh Nikpour; Narges Armanfard; Selena M Sagan; Sana Jahanshahi-Anbuhi
Journal:  Clin Microbiol Rev       Date:  2021-05-12       Impact factor: 26.132

4.  Clinical Performance of the Point-of-Care cobas Liat for Detection of SARS-CoV-2 in 20 Minutes: a Multicenter Study.

Authors:  Glen Hansen; Zi-Xuan Wang; Kathleen G Beavis; Lars F Westblade; Nam K Tran; Jamie Marino; John Rodrigo; Kylie Labog; Run Jin; Nedra Love; Karen Ding; Sachin Garg; Alan Huang; Joanna Sickler
Journal:  J Clin Microbiol       Date:  2021-01-21       Impact factor: 5.948

5.  Real-life head-to-head comparison of performance of two high-throughput automated assays for detection of SARS-CoV-2 RNA in nasopharyngeal swabs: the Alinity m SARS-CoV-2 and cobas 6800 SARS-CoV-2 assays.

Authors:  Rok Kogoj; Petra Kmetič; Anja Oštrbenk Valenčak; Kristina Fujs Komloš; Katja Seme; Martin Sagadin; Miša Korva; Mario Poljak
Journal:  J Mol Diagn       Date:  2021-05-18       Impact factor: 5.568

Review 6.  Review of Current COVID-19 Diagnostics and Opportunities for Further Development.

Authors:  Yan Mardian; Herman Kosasih; Muhammad Karyana; Aaron Neal; Chuen-Yen Lau
Journal:  Front Med (Lausanne)       Date:  2021-05-07

7.  The diagnostic accuracy of seven commercial molecular in vitro SARS-CoV-2 detection tests: a rapid meta-analysis.

Authors:  Zulvikar Syambani Ulhaq; Gita Vita Soraya
Journal:  Expert Rev Mol Diagn       Date:  2021-06-01       Impact factor: 5.225

Review 8.  Laboratory tests for the detection of SARS-CoV-2 infection: basic principles and examples.

Authors:  Khaled R Alkharsah
Journal:  Ger Med Sci       Date:  2021-05-27

9.  Diagnosing SARS-CoV-2 with Antigen Testing, Transcription-Mediated Amplification and Real-Time PCR.

Authors:  Sascha Dierks; Oliver Bader; Julian Schwanbeck; Uwe Groß; Michael S Weig; Kemal Mese; Raimond Lugert; Wolfgang Bohne; Andreas Hahn; Nicolas Feltgen; Setare Torkieh; Fenja R Denker; Peer Lauermann; Marcus W Storch; Hagen Frickmann; Andreas Erich Zautner
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

10.  Detection of SARS-CoV-2 at the point of care.

Authors:  Michael J Loeffelholz; Yi-Wei Tang
Journal:  Bioanalysis       Date:  2021-07-22       Impact factor: 2.681

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.