ICP-MS trace element analysis in serum and whole blood.

Nico Laur1,2,3, Ralf Kinscherf2, Karolina Pomytkin3, Lars Kaiser1,4, Otto Knes3, Hans-Peter Deigner1,5,6.   

Abstract

Trace elements and minerals are compounds that are essential for the support of a variety of biological functions and play an important role in the formation of and the defense against oxidative stress. Here we describe a technique, allowing sequential detection of the trace elements (K, Zn, Se, Cu, Mn, Fe, Mg) in serum and whole blood by an ICP-MS method using single work-up, which is a simple, quick and robust method for the sequential measurement and quantification of the trace elements Sodium (Na), Potassium (K), Calcium (Ca), Zinc (Zn), Selenium (Se), Copper (Cu), Iron (Fe), Manganese (Mn) and Magnesium (Mg) in whole blood as well as Copper (Cu), Selenium (Se), Zinc (Zn), Iron (Fe), Magnesium (Mg), Manganese (Mn), Chromium (Cr), Nickel (Ni), Gold (Au) and Lithium (Li) in human serum. For analysis, only 100 μl of serum or whole blood is sufficient, which make this method suitable for detecting trace element deficiency or excess in newborns and infants. All samples were processed and analyzed by ICP-MS (Agilent Technologies). The accuracy, precision, linearity and the limit of quantification (LOQ), Limit of Blank (LOB) and the limit of detection (LOD) of the method were assessed. Recovery rates were between 80-130% for most of the analyzed elements; repeatabilities (Cv %) calculated were below 15% for most of the measured elements. The validity of the proposed methodology was assessed by analyzing a certified human serum and whole blood material with known concentrations for all elements; the method described is ready for routine use in biomonitoring studies.

Entities:  

Year:  2020        PMID: 32433650      PMCID: PMC7239469          DOI: 10.1371/journal.pone.0233357

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

During the last years, interest in the role of trace elements in biological systems has increased. Almost every metabolic process operates by involving trace elements, ultra-trace elements, or minerals, making them essential for human metabolism; biochemical- and metabolic interactions, very relevant to human medicine as well as in diagnostics and therapy, have already been described in the context of the role of these elements. Especially for trace elements such as Se, Fe, Zn and Cu, but also for minerals such as Na, K, Ca and Mg and other, specific, binding- and transport-proteins could be identified in different membranes of cells, intracellularly and in body fluids [1, 2]. Blood is a body fluid that can be obtained from patients and analyzed to assess the current physiological state of the body. Consequently, serum samples have long been used for identification of disease-related biomarkers and for clinical diagnosis. Whole blood and serum samples have become established for trace element analysis. Elements to be measured were chosen based on their relevance to biological processes. Sodium (Na), for example, plays a key role in a variety of nerve and muscle functions. Furthermore, potassium (K) and calcium (Ca) are essential for cellular signaling, present in a variety of biological processes, as both are involved in ion channels and pumps [3]. An overload of potassium, called hyperkalemia, for example, can lead to dangerous and possibly deadly changes in heart rhythm, which can also be caused by an overload of calcium. Lithium (Li), in turn, has recently been associated with oxidative stress and high concentrations of lithium obviously exert a positive effect on the extent of it. [4] Magnesium (Mg) is found in a variety of enzymes and cells, such as muscle cells. [5] Also, high levels of manganese (Mn) can lead to toxicity and a neurological disease, similar to Parkinson, called Manganism, making evaluation of actual blood levels beneficial for diagnosis and treatment [6]. Furthermore, haemodialysis patients can experience accumulation of chromium (Cr) and nickel (Ni), but may become deficient in values of manganese, making assessment of blood concentrations essential for supplementation or clearance of corresponding elements. Ni, in turn, facilitates iron (Fe) absorption and may play an important role in the production of red blood cells [7]. The toxicity of Ni excess is, however, associated with cardiovascular, renal and respiratory problems [8]. In addition, chromium (Cr) is important in the metabolism of fats and carbohydrates, as it stimulates fatty acid and cholesterol biosynthesis [9]. Studies like Garbuz et al. [10] demonstrated, that the concentration of Cr can also be used as a monitoring parameter during clinical trials of metal-on-metal hip replacements, as the hip replacements contain Cr and therefore the level of metal ion release may be due to the hip replacements. Zinc (Zn), in turn, is also found in enzymes, catalyzing many synthetic biological reactions [11, 12]. A Zn deficiency is characterized by acral dermatitis, alopecia and diarrhea [13]. Also, iron (Fe) is a trace element with an essential role in oxygen transportation e.g. in hemoglobin of erythrocytes, an Fe deficiency ends in anemia [14]. On the other hand, an Fe overload causes severe consequences in the human body, including liver damage, fibrosis, heart failure, or even cancer [13]. Fe, as well as, Mn are further involved in numerous redox reactions [14,15]. Signs of Mn deficiency include impaired growth, impaired reproductive function, skeletal abnormalities, impaired glucose tolerance and altered carbohydrate and lipid metabolism. Nevertheless, Mn deficiency is not common. It is more of concern for toxicity related to Mn overexposure, which can cause an inflammatory response in the lung, resulting in impaired lung function [16]. In addition, two of the most important trace elements, playing a major role in the development of oxidative stress, are selenium (Se) and copper (Cu). They are mandatory to numerous oxidase and superoxide dismutase (SOD) reactions in the cytosol, especially Cu is a critical functional component of a number of enzymes related to oxidative stress e.g. SOD, cytochrome C oxidase (CCO) in the mitochondria and tyrosinase. [13, 17] As well as Fe and Mn, an excess amount of Cu, is toxic for the human body. An unregulated Cu homeostasis may lead to diseases like Menkes disease, characterized by a deficiency of Cu, or to a disease called Wilson, characterized by excessive overload of Cu [13]. Se, in turn, is an essential micronutrient crucial for many biological functions including the regulation of cell death signaling, in endogenous metabolism and redox reactions [2]. However, an excessive Se overload may lead to selenosis, which is characterized in nausea, chronic fatigue, myasthenia and diarrhea [18]. Indeed, Gold (Au) so far has hardly appeared as an environmental toxin. It is known, however, that some patients have contact allergy due to wearing gold spruce. Furthermore, in human medicine Au salts have been used for the treatment of rheumatoid arthritis since about 60 years [19]. Au has recently also been added to many cosmetic products. An overload with this element causes rashes, gastrointestinal bleeding, nausea, vomiting or even bone marrow damages. Unintended imbalance in the concentrations of these trace elements can lead to deficiency or overload disorders as described above, which may cause some life-threatening diseases such as inflammation, arteriosclerosis, myocardial infarction, or even cancer [20]. In addition to partially obsolete methods, such as atomic absorption spectrophotometry (AAS), or flame emission spectrophotometry, modern methods, for instance inductively coupled plasma coupled with a mass spectrometry (ICP-MS) for the detection of these classes of substances, are now available since about the late 1990s. Of course, methods like AAS are still robust, high sensitive and low-cost methods for the measurement of single elements and can be used as reference methods. But the key advantages of the ICP-MS method with its speed, accuracy and simplicity preponderance the AAS. All these methods, however, are not necessarily suitable for routine diagnostics. ICP-MS is a type of mass spectrometry that is highly sensitive enabling determination of a range of elements including trace elements. Nowadays, turnaround time plays an increasingly important role in diagnostics. The faster and more precise a deficiency of these substances is detected and corrected, the better the prognosis for the patients affected. Recently various protocols for processing biological samples have been described, e.g. a work-up of serum samples under acidic conditions using HNO3 [21]. For processing whole blood, however, a method in an alkaline range with ammonia or other alkaline reagents is required [22]. Here we describe a technique, allowing sequential detection of the trace elements (K, Zn, Se, Cu, Mn, Fe, Mg) in serum and whole blood by an ICP-MS method using single work-up. Because of the small amount of serum or whole blood, this method is also suitable for detecting a lack or an overload in newborns or infants. This paper describes the development and the validation of a method for extracting a range of trace elements using 100 μl serum or whole blood only, as well as the subsequent quantification by ICP-MS.

Materials and method

All chemicals and acids were obtained from Sigma Aldrich (Switzerland). Control samples and Calibrators were ordered from Recipe® (Munich, Germany). Consumable materials like tubes, gloves and vessels were purchased from VWR (Switzerland) laboratories.

Samples

We used commercially available control sample material, produced for quality assurance purposes (ClinCheck–Control) manufactured by Recipe® Chemicals and Instruments GmbH in Munich, Germany. Therefore, no ethic approval is necessary.

Standard and calibration

For comparative analysis of the trace element concentrations of Na, K, Zn, Se, Cu, Mn, Ca, Fe, Mg in whole blood and Cu, Se, Zn, Fe, Mg, Mn, Cr, Ni, Au and Li in serum, a serial dilution of the Recipe® control samples and Recipe® calibration in decreasing amounts as calibration standards had been prepared in water. Dilution and sample preparation was performed under a clean hood to prevent contamination by atmospheric particulates. Calibration standards were employed to prepare daily diluted calibration solutions. As internal standard, a mixture of Ir and Rh in a specific concentration were added to all calibration points and samples.

Instrumentation

An Agilent 7900 ICP-MS (Agilent Technologies, Tokyo, Japan) equipped with standard nickel cones, connected to a SPS 4 Autosampler (Agilent Technologies, Tokyo, Japan) was used for all measurements. The optimization of ICP-MS was carried out using a tuning solution consisting of Cs (cesium, 55), Co (cobalt, 27), Li (lithium, 3), Mg (magnesium, 12), Tl (thallium, 81), and Y (yttrium, 39) (Agilent Technologies, Palo Alto, CA, USA). All measurements were performed in triplicates from each vial. The instrument parameters are described in Table 1. Full data were recorded with the Agilent MassHunter Data software (version 4.2).
Table 1

Setting parameters for the ICP-MS method.

Operating ConditionsValues
Extract 2-250 [V]
Omega Bias-90 [V]
Omega Lens9,8 [V]
Deflect-2,2 [V]
He flow3,6 ml / min
H2 flow4,8 ml / min
OctP RF170 [V]
Forward Power1550 W
Auxiliary gas flow0,9 L / min
Nebulizer gas flow1 L / min
Nebulizer typeMicroMist
Sample introductionPeriPump
Replicates3
Table 2 shows the acquisition parameters for the trace elements. Elements were measured in three different modes; performance check of the ICP-MS-System was performed daily.
Table 2

Acquisition parameters of the ICP-MS system.

Integration m / z [sec]
MassElement NameNo Gas–ModeH2—ModeHe—Mode
7Li0,800N/AN/A
24MgN/AN/A0,500
39KN/AN/A0,500
40CaN/A1,000N/A
52CrN/AN/A1,000
55MnN/AN/A0,500
56FeN/A1,000N/A
60Ni1,000N/AN/A
63CuN/AN/A1,000
66ZnN/AN/A0,500
78SeN/A1,000N/A
103Rh1,0001,0001,000
193Ir1,0001,0001,000
197Au1,000N/AN/A

N/A = not available

N/A = not available

Preparation and extraction protocol

The processing protocol of the samples were optimized and modified according to the publication of Meyer et al. [23]. 100 μl of serum or whole blood ClinCheck control were added to a 15 ml polypropylene tube with screw cap (VWR laboratories). 200 μl of approximately 65% nitric acid (HNO3) and 100 μl of hydrogen peroxide (H2O2) were added and the mixture vortexed briefly. For dissolving, tubes then were set for 90 min at 60°C in a heating system. After incubation, samples were cooled down by adding 2100 μl ultrapure water (RT). Approximately 18 MΩ cm-1 ultrapure water was used for sample preparation and analysis. After vortexing and centrifugation (3 min/ 2500 rpm), samples were ready to measure.

Method validation

The method presented was validated according to the ICH-Guidelines [24] and implemented on three different days. Validation of the method was carried out to determine performance, comprising precision in series (Cv in %), accuracy, limits of blank (LOB), limits of detection (LOD) and limits of quantification (LOQ). LOB, LOD and LOQ were calculated by quantifying blank values in three independent replicates of 3% HNO3 samples subsequent to sample preparation as described above. For this purpose, the corresponding amount of internal standard was added to all samples. Precision was assessed using the coefficient of variation and accuracy by analyte recovery determination.

Particiption in an external quality assessment

External quality assessments are used to determine the performance of individual laboratories for specific measurements, and to monitor the continuing performance of laboratories. In addition, participation in an interlaboratory (external) quality program is an essential component of quality assurance and also provides evidence of laboratory competence for clinics, researchers and regulatory agencies [25]. After validating our method, we participated in an external quality assessment conducted by INSTAND e.V. (Düsseldorf Germany) for testing our validated method. The freeze-dried samples (plasma and whole blood) were transported to our laboratory at room temperature. After reconstituting the samples according to the guidelines of INSTAND e.V. the samples were ready for preparation. Descriptive statistics were calculated after standardisation of all laboratory results to percentages with reference to the true value. By subtracting 100% of these percentages, the percentage bias of the true concentration (inaccuracy) was calculated. 20% limits around the true values were considered to be appropriate threshold values for satisfactory measurements. All participating laboratories were informed about their performance within 2 months after transmitting their results.

Results and discussion

Validation range and linearity of the method are reported in Table 3. The R2 value for the response function was exceeding 0,922 for all trace elements. Results of the quality control samples perfectly agree with the calculated mean values. The element composition of the serum and whole blood samples and the inter-day analysis of the samples are shown in S2 File of S1,S2 Tables. The values shown in S2 File of S1,S2 Tables are the average of triplicate preparations within a day analysis (n = 3). The final average concentration is an average concentration of the average values from each day.
Table 3

Linearity and validation range of serum and whole blood.

Serum
R2Validation Range
Copper0,9988246,67–1360,00 μg/L
Selenium0,998918,63–112,00 μg/L
Zinc0,9967350,00–1510,00 μg/L
Iron0,9872303,33–1480,00 μg/L
Magnesium0,997530,00–21500,00 μg/L
Manganese0,98070,80–6,04 μg/L
Chromium0,9830,53–5,66 μg/L
Nickel0,96980,81–5,96 μg/L
Gold0,99732,47–483,00 μg/L
Lithium0,99771213,33–3640,00 μg/L
Whole Blood
Sodium0,98731095000,00–2210000,00 μg/L
Magnesium0,998713250,00–43300,00 μg/L
Potassium0,9673670000,00–2180000,00 μg/L
Calcium0,977523800,00–47600,00 μg/L
Manganese0,9227,70–22,10 μg/L
Iron0,9811187500,00–375000,00 μg/L
Copper0,9981339,00–1680,00 μg/L
Zinc0,98652290,00–7980,00 μg/L
Selenium0,988837,65–168,00 μg/L
Unintended imbalance in the concentrations of these trace elements can lead to deficiency or overload disorders as described above, which may cause some diseases. Routinely, blood samples from patients are collected by puncturing a vein using a stainless steel needle, raising concerns regarding possible contaminations by such. The major components of stainless steel needles are iron, nickel, chromium and manganese. However, there is a huge variability between stainless steel needles from different suppliers [26]. Studies like Pineau et al. [27] demonstrated that these trace metals on the stainless steel needles can be reduced by rinsing them with water. Since 1995 the international Union of Pure and Applied Chemistry have been recommended using intravenous plastic needle rather than stainless steel needle when taking blood samples for analyzing trace elements [28]. Nevertheless, studies have not shown explicit verifications of significant contaminations with trace elements using stainless steel needles or plastic needles. Bro et al. [29] performed a study including 20 healthy volunteers. They conclude no differences in the content of chromium and nickel between samples taking using stainless steel needles or plastic needles. Penny and Overgaard [30] found only marginally significant differences in the contamination of chromium between samples taken using stainless steel needles and plastic needles. However, in order to exclude possible contamination, the recommendations of the international Union of Pure and Applied Chemistry should be taken into account. Another important aspect is the impact of sample haemolysis of red blood cells in relation of some trace elements, which may have an influence on result reliability due to biological and analytical interference [31]. Some trace elements like manganese, iron or potassium show higher intracellulary concentration than extracellulary. Haemolysis of red blood cells induces the release of intracellulary trace elements. Therefore, haemolytic samples should not be analysed, or should be analysed with caution, because this can lead to falsely high values. The processing protocol taken from Meyer et al. [23] was optimized and modified towards a more practical and more time saving method by reducing sample volume and shortening incubation time. This makes this method more suitable for routine use in biomonitoring. Compared to other studies e.g. Luna D et al. [32] we showed better results in accuracy (%) and LOD for most of the elements. In our study we only used 100 μl of serum or whole blood, which make this method also suitable for detecting trace element deficiency or excess in newborns and infants, using capillary blood sampling techniques (collection volume typically > 500 μl). [33] Basically, it should always be questioned which laboratory requirements are actually necessary, because the blood volume of a newborn is only 80–85 ml/ Kg. Several trace elements are indispensable and essential for life, others and their compounds are simply inert and others exhibit a high toxicity even in low concentrations. Mn, Fe, Cu, Zn and Se play important roles in fetal cell growth. On the other hand, overloading of these trace elements may cause toxicity to fetus and newborns [34, 35]. In this study the LOD, LOB and LOQ-values for the measurements of 7Li, 24Mg, 39K, 52Cr, 55Mn, 56Fe, 60 Ni, 63Cu 66Zn, 78Se, 197Au are shown in Table 4; LOD, LOB and LOQ were calculated according to Armbruster and Pry [36]. The accuracy and repeatability were calculated by analyzing replicated dilutions of the Recipe® control material.
Table 4

LOD, LOQ and LOB of the ICP-MS method of serum and whole blood.

Limit of blank (μg/l) (n = 9)Limit of detection (μg/l) (n = 9)Limit of quantification (μg/l) (n = 9)
Serum
Copper0,090,110,23
Selenium0,150,190,33
Zinc1,962,524,78
Iron0,931,403,29
Magnesium2,993,967,93
Manganese0,110,150,30
Chromium0,090,130,27
Nickel0,140,170,32
Gold0,010,010,03
Lithium0,150,180,33
Whole blood
Sodium1206,501341,301891,40
Magnesium477,70759,701909,80
Potassium303,00473,001166,30
Calcium83,50127,80308,20
Manganese0,500,701,50
Iron23,5036,8091,20
Copper1,402,205,50
Zinc6,507,109,50
Selenium9,0012,0024,20
Linearity of the calibration curves was evaluated within the range of the Recipe® calibration, control samples affording determination coefficients with R2 ≥0,922. Regression analysis indicated significant linear correlation between concentration and signal intensity (p < 0,05). The method shows very good accuracy and repeatability (Tables 5 & 6) for the trace elements quantified in both, serum and whole blood.
Table 5

Accuracy and repeatability of the measured trace elements in serum.

AnalyteTheoretical Concentration (μg/L)Measured Concentration (μg/L) Mean ± SDAccuracy [%]Repeatability Cv [%]
Copper246,67255,11 ± 4,75103,421,86
370,00374,02 ± 1,99101,090,53
453,33475,34 ± 9,14104,851,92
680,00717,52 ± 22,78105,523,17
740,00766,16 ± 22,29103,532,91
1360,001362,18 ± 58,52100,164,30
Selenium18,6318,25 ± 1,1297,966,14
27,9526,20 ± 0,7293,732,75
37,3335,43 ± 1,2494,903,50
55,9051,61 ± 5,0992,339,86
56,0054,57 ± 2,7797,445,08
112,00108,49 ± 5,7296,875,27
Zinc350,00304,56 ± 15,6987,025,15
503,33466,76 ± 18,1992,733,90
525,00484,95 ± 31,3592,376,46
755,00735,73 ± 14,8697,452,02
1050,001044,59 ± 27,5499,492,64
1510,001428,00 ± 65,5294,574,59
Iron303,33331,81 ± 12,64109,393,81
455,00487,69 ± 17,17107,183,52
493,33588,55 ± 47,94119,308,15
740,00824,84 ± 14,63111,461,77
910,00867,97 ± 65,1895,387,51
1480,001479,20 ± 62,5099,954,23
Magnesium5300,005186,97 ± 89,1097,871,72
7166,677106,15 ± 127,8899,101,80
7950,007878,62 ± 321,3099,164,08
10750,0010780,78 ± 343,13100,293,18
15900,0016107,24 ± 425,26101,302,64
21500,0020534,65 ± 753,9795,513,67
Manganese0,800,83 ± 0,06103,907,23
1,211,20 ± 0,1499,9811,67
2,012,49 ± 0,55123,6922,09
2,412,10 ± 0,3587,0116,67
3,023,31 ± 0,40109,7212,08
6,045,85 ± 0,3896,946,50
Chromium0,591,28 ± 0,17214,9813,28
0,891,38 ± 0,16155,1511,59
1,782,10 ± 0,26117,9912,38
1,893,75 ± 1,88118,4550,13
2,834,53 ± 1,12124,7624,72
5,665,65 ± 0,3099,765,31
AnalyteTheoretical concentrationsMeasured Concentration (μg/L) Mean ± SDAccuracy [%]Repeatability Cv [%]
Nickel0,811,05 ± 0,26129,4924.76
1,221,90 ± 0,18156,539.47
1,992,14 ± 0,27107,8712.62
2,433,33 ± 0,16136,944.80
2,983,34 ± 0,34111,9310.18
5,966,00 ± 0,64100,6410.67
Gold32,4738,97 ± 5,17120,0213.27
48,7064,29 ± 8,29132,0212.89
97,40101,57 ± 2,24104,282.21
161,00171,99 ± 13,59106,837.90
241,50267,24 ± 21,33110,667.98
483,00512,87 ± 54,24106,1910.58
Lithium1213,331161,43 ± 26,5995,722.29
1820,001736,70 ± 34,9195,422.01
2406,672194,99 ± 17,1491,200.78
3610,003437,43 ± 89,6795,222.61
3640,003678,20 ± 120,63101,053.28
7220,006889,48 ± 265,2595,423.85
Table 6

Accuracy and repeatability of the measured trace elements in whole blood.

AnalyteTheoretical Concentration (μg/L)Measured Concentration (μg/L) Mean ± SDAccuracy [%]Repeatability Cv [%]
Sodium1095000938069,75 ± 22445,1885,672,39
1095000963072,28 ± 31214,2987,953,24
11050001073121,30 ± 2964,0797,120,28
21900001889717,80 ± 47427,5086,292,51
21900002005786,19 ± 29856,5391,591,49
22100002034499,42 ± 64855,4492,063,19
Magnesium1325012897,04 ± 628,0297,344,87
1745016734,75 ± 789,7595,904,72
2165021105,16 ± 382,0597,481,81
2650024478,03 ± 543,4492,372,22
3490032592,08 ± 463,5693,391,42
4330039513,01 ± 518,3991,251,31
Potassium670000634366,81 ± 21801,6994,683,44
880000833121,09 ± 23972,6594,672,88
1090000986309,60 ± 19680,1390,492,00
13400001172996,44 ± 52933,8287,544,51
17600001511923,57 ± 75412,8685,904,99
21800002223448,50 ± 106555,26101,994,79
Calcium2380019600,41 ± 599,3682,353,06
2380021608,76 ± 1304,4690,796,04
2385021477,18 ± 729,1590,053,39
4760044225,27 ± 2226,5592,915,03
4760043826,45 ± 609,0692,071,39
4770049460,14 ± 393,61103,690,80
Manganese4,445,01 ± 1,24112,9224,75
7,707,46 ± 1,1196,8614,88
8,878,38 ± 1,8194,4821,60
11,0510,20 ± 1,6992,3516,57
15,4012,55 ± 0,2681,502,07
22,1026,54 ± 1,56120,105,88
Iron187500164985,76 ± 11946,2687,997,24
187500171830,79 ± 9332,2291,645,43
187500183436,39 ± 4125,2697,832,25
375000297550,28 ± 5688,7779,351,91
375000310154,58 ± 4684,5082,711,51
375000324810,47 ± 17702,8086,625,45
Copper339,00311,05 ± 77,1691,7624,81
555,00501,58 ± 93,3090,3718,60
678,00650,26 ± 27,6495,914,25
840,00748,40 ± 32,0289,094,28
1110,001025,60 ± 67,8592,406,62
1680,001505,50 ± 61,7689,614,10
Zinc2290,002138,61 ± 64,6993,393,02
3135,003013,66 ± 219,7896,137,29
3990,003630,23 ± 145,6890,984,01
4580,004328,68 ± 368,8494,518,52
6270,006119,93 ± 132,8797,612,17
7980,006995,66 ± 50,8987,660,73
Selenium37,6550,48 ± 4,73134,089,37
75,3084,07 ± 12,03111,6414,31
84,0093,43 ± 2,97111,233,18
84,0082,86 ± 4,9798,646,00
168,00163,17 ± 2,6297,131,61
168,00174,01 ± 12,58103,587,23
N/A = not available The method presented was assessed by analyzing aliquots and dilutions of human blood material with known concentrations of several elements (ClinCheck-Control, serum and whole blood, RECIPE Chemicals + Instruments GmbH, Munich, Germany); all samples were analyzed in triplicates. The selected validation areas for each element were based on the physiological areas published in literature (Table 7). With our described method and our chosen validation points, we cover the physiological areas described in the literature for most of the elements.
Table 7

Comparison of the reference ranges established in this work with values in the literature.

Analyte in serumvalidation range in this studyReference ranges in literature
Copper246,67–1360,00 μg/L600,00–1400,00 μg/L [37]
Selenium18,63–112,00 μg/L30,00–105,00 μg/L [37]
Zinc350,00–1510,00 μg/L600,00–1200,00 μg/L [37]
Iron303,33–1480,00 μg/L1100,00–1300,00 μg/L [37]
Magnesium530,00–21500,00 μg/L16044,60–26011,70 μg/L [38]
Manganese0,80–6,04 μg/L0,10–2,90 μg/L [37]
Chromium0,53–5,66 μg/L< 0,40 μg/L [38]
Nickel0,81–5,96 μg/L< 2,00 μg/L [38]
Gold32,47–483,00 μg/LN/A
Lithium1213,33–3640,00 μg/L4164,00–5205,00 μg/L [39]
Analyte in whole bloodvalidation range in this studyReference ranges in literature
Sodium1095000,00–2210000,00 μg/LN/A
Magnesium13250,00–43300,00 μg/L30000,00–40000,00 μg/L [38]
Potassium670000,00–2180000,00 μg/LN/A
Calcium23800,00–47600,00 μg/LN/A
Manganese7,70–22,10 μg/L8,30–15,00 μg/L [38]
Iron187500,00–375000,00 μg/L425000,00–500000,00 μg/L [40]
Copper339,00–1680,00 μg/L700,00–1390,00 μg/L [38]
Zinc2290,00–7980,00 μg/L4500,00–7500.00 μg/L [38]
Selenium37,65–168,00 μg/L90,00–230,00 μg/L [38]

N/A = not available

Effective reduction of sample matrix extraction and optimization procedures made it possible to avoid most of the interference problems and matrix-induced ionization effects in analysis. Accuracy was found to be between 80–130%, for all elements analysed (except for Selenium in whole blood, Chromium, Nickel and Gold in serum at lower concentrations) in both, serum and whole blood. In fact, the range is in agreement with data presented in the study of Burrows et al. [41] while trace element determination in serum shows slightly better accuracies for copper, zinc, selenium and magnesium compared to whole blood samples which may be due to less matrix effects in serum samples. Whole blood contains high concentrations of inorganic ions. It also contains a numerous of diverse organic compounds such as lipids, carbohydrates and proteins of a wide range of molecular weights. These components can cause analytical interferences when the matrix is not completely decomposed. Repeatabilities (Cv) of all elements were below 15% except for manganese, nickel and copper at a low concentration in whole blood and serum samples as being close to the LOQ; calculated concentrations, however, were in agreement with certified values for all elements. Validation ranges of elements (Table 3) were linear across concentrations measured for each element with R2 values exceeding 0,922. Ir and Rh were used as internal standards to eliminate or reduce variations such as power fluctuations, temperature changes, differences in solution parameters (e.g. density, viscosity, or surface tension) and changes in the sample introduction system [42,43]. After validating the method, we participated in an external quality assessment for most of the validated elements, conducted by INSTAND e.V.. The results are shown in Table 8.
Table 8

Results of the external quality assessment.

Analyte in serumUnitSample Nr.Measured valueTarget valuedeviation
Copperμg/L211550,621652,30-6,20%
22724,47775,31-5,80%
Seleniumμg/L21153,00137,0011,80%
22161,00156,003,30%
Zincμg/L2124602560-3,90%
2219702070-4,80%
Ironμg/L21955,04977,38-2,30%
221541,461731,35-11,00%
Manganeseμg/L214,205,63-25,40%
229,6010,70-10,30%
Analyte in whole bloodUnitSample Nr.Measured valueTarget valuedeviation
Magnesiumμg/L6136200,0037200,00-2,70%
6240200,0041900,00-4,10%
Manganeseμg/L6114,1013,603,30%
6217,4016,108,30%
Seleniumμg/L61127,00123,003,00%
62150,00146,002,50%
Zincμg/L614910,004970,00-1,20%
6254600,0046500,0017,40%
If the trace element analysis are used to check a lack or an overload in the human body, or are used to determine the compliance with the regulations, any error in the result can have extremely serious consequences. For an independent laboratory, producing unreliable test results taints its reputation and the trust of its customers. It may lose them for ever. After measuring and transmitting the values, INSTAND e.V. issued us with a certificate of good standard for each trace element we recorded. Our data and the subsequent participation in an external quality assessment demonstrate, that the method is valid for quantification of the respective elements in both, human serum and whole blood, requiring only one-step preparation with 100 μl of sample; it is cost-effective and fast without noteworthy matrix effects for all trace elements measured and is also suitable for the measurement of newborns and infants because of its less sample volume.

Conclusions

Serum and whole blood are the two most widely used matrices in clinical laboratories. During the last years, interest in the role of trace elements in biological systems has increased. Unintended imbalance in the concentrations of these trace elements can lead to deficiency or overload disorders as described above. The described method allows the sequential detection of the trace elements (K, Zn, Se, Cu, Mn, Fe, Mg) in serum and whole blood by an ICP-MS method using a single work-up out of 100 μl only. Results show that trace element concentrations measured with the validated ICP-MS method can be valuable to research and routine diagnostics with the method offering excellent, rapid, precise, inexpensive and timely quantitation in low-concentration multielement analysis.

Graphical abstract.

(TIF) Click here for additional data file. (DOCX) Click here for additional data file.
  33 in total

1.  Comparison of capillary blood sampling using an automated incision device with and without warming the heel.

Authors:  Marianne Janes; Janet Pinelli; Shannon Landry; Susan Downey; Bosco Paes
Journal:  J Perinatol       Date:  2002-03       Impact factor: 2.521

Review 2.  Adverse health effects of selenium in humans.

Authors:  M Vinceti; E T Wei; C Malagoli; M Bergomi; G Vivoli
Journal:  Rev Environ Health       Date:  2001 Jul-Sep       Impact factor: 3.458

3.  Limit of blank, limit of detection and limit of quantitation.

Authors:  David A Armbruster; Terry Pry
Journal:  Clin Biochem Rev       Date:  2008-08

4.  The influence of storage age on iron status, oxidative stress and antioxidant protection in paediatric packed cell units.

Authors:  Keith Collard; Desley White; Adrian Copplestone
Journal:  Blood Transfus       Date:  2013-11-29       Impact factor: 3.443

Review 5.  Absorption Mechanisms of Iron, Copper, and Zinc: An Overview.

Authors:  Yukina Nishito; Taiho Kambe
Journal:  J Nutr Sci Vitaminol (Tokyo)       Date:  2018       Impact factor: 2.000

Review 6.  Copper chaperones. The concept of conformational control in the metabolism of copper.

Authors:  Peep Palumaa
Journal:  FEBS Lett       Date:  2013-05-16       Impact factor: 4.124

7.  Correlation of oxidative stress with serum trace element levels and antioxidant enzyme status in Beta thalassemia major patients: a review of the literature.

Authors:  Q Shazia; Z H Mohammad; Taibur Rahman; Hossain Uddin Shekhar
Journal:  Anemia       Date:  2012-05-09

8.  Simultaneous quantification of 49 elements associated to e-waste in human blood by ICP-MS for routine analysis.

Authors:  Ana González-Antuña; María Camacho; Luis Alberto Henríquez-Hernández; Luis D Boada; Maira Almeida-González; Manuel Zumbado; Octavio P Luzardo
Journal:  MethodsX       Date:  2017-10-15

Review 9.  Trace Element Provision in Parenteral Nutrition in Children: One Size Does Not Fit All.

Authors:  Boutaina Zemrani; Zoe McCallum; Julie E Bines
Journal:  Nutrients       Date:  2018-11-21       Impact factor: 5.717

10.  Validation of a simple sample preparation method for multielement analysis of bovine serum.

Authors:  Diego Luna; Marta Miranda; Antonio Humberto Hamad Minervino; Verónica Piñeiro; Carlos Herrero-Latorre; Marta López-Alonso
Journal:  PLoS One       Date:  2019-02-05       Impact factor: 3.240

View more
  9 in total

1.  A caged imidazopyrazinone for selective bioluminescence detection of labile extracellular copper(ii).

Authors:  Justin J O'Sullivan; Valentina Medici; Marie C Heffern
Journal:  Chem Sci       Date:  2022-03-23       Impact factor: 9.825

2.  Diet and dog characteristics affect major and trace elements in hair and blood of healthy dogs.

Authors:  Sarah Rosendahl; Johanna Anturaniemi; Kristiina A Vuori; Robin Moore; Manal Hemida; Anna Hielm-Björkman
Journal:  Vet Res Commun       Date:  2021-11-06       Impact factor: 2.816

3.  Evaluation of Serum Trace Element Levels and Biochemical Parameters of COVID-19 Patients According to Disease Severity.

Authors:  Omer Faruk Kocak; Fatma Betul Ozgeris; Emine Parlak; Yucel Kadıoglu; Neslihan Yuce; Mehmet Emrah Yaman; Ebubekir Bakan
Journal:  Biol Trace Elem Res       Date:  2021-10-04       Impact factor: 4.081

4.  Inductively-Coupled Plasma Mass Spectrometry-Novel Insights From an Old Technology Into Stressed Red Blood Cell Physiology.

Authors:  Daniel Stephenson; Travis Nemkov; Syed M Qadri; William P Sheffield; Angelo D'Alessandro
Journal:  Front Physiol       Date:  2022-02-07       Impact factor: 4.566

5.  Facile transmetallation of [SbIII(DOTA)]- renders it unsuitable for medical applications.

Authors:  Catherine Chen; Charlotte Sommer; Helge Thisgaard; Vickie McKee; Christine J McKenzie
Journal:  RSC Adv       Date:  2022-02-16       Impact factor: 3.361

Review 6.  Facets of ICP-MS and their potential in the medical sciences-Part 1: fundamentals, stand-alone and hyphenated techniques.

Authors:  David Clases; Raquel Gonzalez de Vega
Journal:  Anal Bioanal Chem       Date:  2022-08-27       Impact factor: 4.478

7.  Factors Influencing Trace Element Levels in the Blood of Tin Smelting Workers.

Authors:  Ludi Zhang; Boshen Wang; Huanxi Shen; Hengdong Zhang; Xin Liu; Lixin Zhong; Deye Liu; Dong Jiang; Yong Zhu; Baoli Zhu; Lei Han
Journal:  J Occup Environ Med       Date:  2022-06-08       Impact factor: 2.306

8.  Determination of lithium in human serum by isotope dilution atomic absorption spectrometry.

Authors:  Alexander Winckelmann; Dalia Morcillo; Silke Richter; Sebastian Recknagel; Jens Riedel; Jochen Vogl; Ulrich Panne; Carlos Abad
Journal:  Anal Bioanal Chem       Date:  2021-09-10       Impact factor: 4.142

9.  ICP-MS Multi-Elemental Analysis of the Human Meninges Collected from Sudden Death Victims in South-Eastern Poland.

Authors:  Jacek Baj; Grzegorz Teresiński; Beata Kowalska; Tomasz Krajka; Grzegorz Buszewicz; Alicja Forma; Wojciech Flieger; Kaja Hanna Karakuła; Paweł Kędzierawski; Tomasz Cywka; Jolanta Flieger
Journal:  Molecules       Date:  2022-03-15       Impact factor: 4.411

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.