| Literature DB >> 32429549 |
Neha Sharma1,2, Shuaishuai Cao1,2, Bilal Msallem1,2, Christoph Kunz1, Philipp Brantner2,3, Philipp Honigmann2,4, Florian M Thieringer1,2.
Abstract
Computer-assisted surgery with three-dimensional (3D) printed surgical guides provides more accurate results than free-hand surgery. Steam sterilization could be one of the factors that affect the dimensions of surgical guide resin materials, leading to inaccuracies during surgeries. The purpose of this study was to evaluate the effects of steam sterilization on the dimensional accuracy of indication-specific hollow cube test bodies, manufactured in-house using Class IIa biocompatible resin materials (proprietary and third-party). To evaluate the pre- and post-sterilization dimensional accuracy, root mean square (RMS) values were calculated. The results indicate that, in all the groups, steam sterilization resulted in an overall linear expansion of the photopolymeric resin material, with an increase in outer dimensions and a decrease in inner dimensions. The effects on the dimensional accuracy of test bodies were not statistically significant in all the groups, except PolyJet Glossy (p > 0.05). The overall pre- and post-sterilization RMS values were below 100 and 200 µm, respectively. The highest accuracies were seen in proprietary resin materials, i.e., PolyJet Glossy and SLA-LT, in pre- and post-sterilization measurements, respectively. The dimensional accuracy of third-party resin materials, i.e., SLA-Luxa and SLA-NextDent, were within a comparable range as proprietary materials and can serve as an economical alternative.Entities:
Keywords: 3D printing; additive manufacturing; biocompatible resin; dimensional accuracy; in-house; steam sterilization; surgical guide
Year: 2020 PMID: 32429549 PMCID: PMC7291001 DOI: 10.3390/jcm9051506
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Study groups and printing profile specifications.
| Group | Resin Code Name | Printing Profile Setting |
|---|---|---|
| 1 | PolyJet Matte | Matte profile |
| 2 | PolyJet Glossy | Glossy profile |
| 3 | SLA-LT | Dental LT resin profile |
| 4 | SLA-Luxa | Clear V2 resin-experimental profile |
| 5 | SLA-NextDent | Dental SG resin-experimental profile |
Figure 1Projected and front view of a simplified experimental indication-specific test body design displaying the landmark points: blue points represent the landmark points for inner measurement (IM = 16 mm), while green points represent the landmark points for outer measurement (OM = 20 mm).
Figure 23D printed test bodies after removal of support material: (A) Group 1 (PolyJet Matte) test bodies (T1–T9); and (B) Group 2 (PolyJet Glossy) test bodies (T1–T9).
Figure 33D printed test bodies in a cured state: (A) Group 3 (SLA-LT) test bodies (T1–T9) after removal of support structures; (B) Group 4 (SLA-Luxa) test bodies (T1–T9) before removal of support structures; and (C) Group 5 (SLA-NextDent) test bodies (T1–T9) before removal of support structures.
Figure 4Boxplot demonstrating dimensional accuracy measurements (pre- and post-sterilization) reflecting overall precision RMS values in all groups. Pre, pre-sterilization; Post, post-sterilization; RMS, root mean square. The dotted line represents the CAD reference line.
Figure 5Boxplot demonstrating dimensional accuracy measurements (pre- and post-sterilization) reflecting overall trueness RMS values in all groups. Pre, pre-sterilization; Post, post-sterilization; RMS, root mean square. The dotted line represents the CAD reference line.
Pre- sterilization RMS values (mean ± SD) (µm) of precision and trueness for outer, inner, and overall measurements of test bodies fabricated using different biocompatible resin materials.
| PolyJet Matte | PolyJet Glossy | SLA-LT | SLA-Luxa | SLA-NextDent | ||
|---|---|---|---|---|---|---|
| Measurement | Mean ± SD | Mean ± SD | Mean ± SD | Mean ± SD | Mean ± SD | |
| Precision | OM | 31.0 ± 20.1 a | 9.25 ± 8.4 b | 8.5 ± 4.1 b | 26.7 ± 21.3 a,b | 23.5 ± 17.7 a,b |
| IM | 14.5 ± 11.3 a | 16.88 ± 4.4 a | 20.5 ± 11.2 a | 23.4 ± 10.8 a,b | 38.4 ± 15.2 b | |
| Overall | 24.4 ± 18.7 a,b | 12.3 ± 7.9 a | 13.3 ± 9.6 a | 25.4 ± 17.5 a,c | 29.5 ± 17.9 b,c | |
| Trueness | OM | 132.2 ± 13.9 a | 41.1 ± 16.2 b | 48.9 ± 12.7 b | 116.7 ± 30.0 a,c | 101.1 ± 19.0 c |
| IM | 85.6 ± 8.8 a,b | 54.4 ± 7.3 a,c | 74.4 ± 32.8 b,c | 88.9 ± 41.1 b | 128.9 ± 10.5 d | |
| Overall | 108.9 ± 26.5 a | 47.8 ± 13.9 b | 61.7 ± 27.5 b | 102.8 ± 37.7 a | 115.0 ± 20.6 a |
a–c Within a row, RMS values with a common superscript letter indicate no statistically significant difference (p > 0.05). RMS, root mean square; OM, outer measurement; IM, inner measurement.
Statistical comparison of overall pre- and post-sterilization precision.
| PRE | PolyJet Matte | PolyJet Glossy | SLA-LT | SLA-Luxa | SLA-NextDent |
|---|---|---|---|---|---|
| PolyJet Matte | |||||
| PolyJet Glossy | 0.08 | ||||
| SLA-LT | 0.14 | 0.99 | |||
| SLA-Luxa | 0.99 | 0.06 | 0.09 | ||
| SLA-NextDent | 0.82 | <0.05 * | <0.05 * | 0.91 | |
|
|
|
|
|
|
|
| PolyJet Matte | |||||
| PolyJet Glossy | 0.32 | ||||
| SLA-LT | 0.06 | <0.05 * | |||
| SLA-Luxa | 0.43 | <0.05 * | 0.85 | ||
| SLA-NextDent | 0.99 | 0.15 | 0.15 | 0.67 | |
* p < 0.05 suggests a statistically significant difference between the indicated groups. Pre, pre-sterilization; Post, post-sterilization.
Post-sterilization RMS values (mean ± SD) (µm) of precision and trueness for outer, inner, and overall measurements of test bodies fabricated using different biocompatible resin materials.
| PolyJet Matte | PolyJet Glossy | SLA-LT | SLA-Luxa | SLA-NextDent | ||
|---|---|---|---|---|---|---|
| Measurement | Mean ± SD | Mean ± SD | Mean ± SD | Mean ± SD | Mean ± SD | |
| Precision | OM | 50.8 ± 29.7 a | 38.0 ± 24.7 a,b | 16.8 ± 12.0 b | 35.3 ± 15.8 a,b | 46.0 ± 19.4 a |
| IM | 17.5 ± 11.1 a | 70.5 ± 8.0 | 20.5 ± 17.6 a | 10.3 ± 6.4 a | 17.8 ± 14.5 a | |
| Overall | 37.5 ± 28.9 a,b | 51.0 ± 25.4 b | 18.3 ± 14.2 a | 25.3 ± 17.8 a | 34.7 ± 22.3 a,b | |
| Trueness | OM | 186.7 ± 18.0 a | 138.9 ± 11.7 b | 72.2 ± 19.9 | 127.8 ± 33.1 b | 158.9 ± 27.6 a,b |
| IM | 81.1 ± 19.0 a,c | 214.4 ± 15.1 | 90.0 ± 25.5 b,c | 61.1 ± 22.1 a | 76.7 ± 20.0 a,b | |
| Overall | 133.9 ± 57.2 a | 176.7 ± 41.0 | 81.1 ± 23.9 b | 94.4 ± 43.8 a,b | 117.8 ± 48.3 a,b |
a–c Within a row, RMS values with a common superscript letter indicate no statistically significant difference (p > 0.05). RMS, root mean square; OM, outer measurement; IM, inner measurement.
Statistical comparison of overall pre- and post-sterilization trueness.
| PRE | PolyJet Matte | PolyJet Glossy | SLA-LT | SLA-Luxa | SLA-NextDent |
|---|---|---|---|---|---|
| PolyJet Matte | |||||
| PolyJet Glossy | <0.05 * | ||||
| SLA-LT | <0.05 * | 0.52 | |||
| SLA-Luxa | 0.95 | <0.05 * | <0.05 * | ||
| SLA-NextDent | 0.96 | <0.05 * | <0.05 * | 0.64 | |
|
|
|
|
|
|
|
| PolyJet Matte | |||||
| PolyJet Glossy | <0.05 * | ||||
| SLA-LT | <0.05 * | <0.05 * | |||
| SLA-Luxa | 0.61 | <0.05 * | 0.89 | ||
| SLA-NextDent | 0.8 | <0.05 * | 0.15 | 0.52 | |
* p < 0.05 suggests a statistically significant difference between the indicated groups. Pre, pre-sterilization; Post, post-sterilization.
Figure 6Scatter plot demonstrating the deviations distribution pattern of test bodies regarding their position on 3D printer’s build platform: (A) pre-sterilization outer measurements; (B) pre-sterilization inner measurements; (C) post-sterilization outer measurements; and (D) post-sterilization inner measurements.