Claudio De Lazzari1, Beatrice De Lazzari2, Attilio Iacovoni3, Silvia Marconi4, Silvia Papa5, Massimo Capoccia6, Roberto Badagliacca7, Carmine Dario Vizza8. 1. National Research Council, Institute of Clinical Physiology (IFC-CNR), Via Palestro, 32 (00185) Rome, Italy; National Institute for Cardiovascular Research (I.N.R.C.), Bologna, Via Irnerio, 48 (40126) Bologna, Italy. Electronic address: claudio.delazzari@ifc.cnr.it. 2. Department of Engineering, Roma Tre University, Italy. Electronic address: beatrice.delazzari@gmail.com. 3. Papa Giovanni XIII Hospital, Bergamo, Italy. Electronic address: aiacovoni@asst-pg23.it. 4. National Research Council, Institute of Clinical Physiology (IFC-CNR), Via Palestro, 32 (00185) Rome, Italy. Electronic address: silvia.marconi@ifc.cnr.it. 5. Department of Cardiovascular Respiratory Nephrologic and Geriatric Sciences, Sapienza University of Rome, Italy. Electronic address: silvia.papa@uniroma1.it. 6. Royal Brompton Hospital, Royal Brompton & Harefield NHS Foundation Trust, UK; Department of Biomedical Engineering, University of Strathclyde, Glasgow, UK. Electronic address: capoccia@doctors.org.uk. 7. Department of Cardiovascular Respiratory Nephrologic and Geriatric Sciences, Sapienza University of Rome, Italy. Electronic address: roberto.badagliacca@uniroma1.it. 8. Department of Cardiovascular Respiratory Nephrologic and Geriatric Sciences, Sapienza University of Rome, Italy. Electronic address: dario.vizza@uniroma1.it.
Aortic blood pressure [mmHg]End diastolic blood pressure [mmHg]Peak diastolic blood pressure [mmHg]Left ventricular pressure [mmHg]Body surface area [m2]Cardiac index [L/min/m2]Cardiac output [L/min]End-diastolic volume [ml]End-systolic volume [ml]Stroke volume [ml]Arterial elastance [mmHg/ml]Slope of ESPVR [mmHg/ml]End-diastolic (end-systolic) pressure volume relationshipEjection fractionMean coronary blood flow [ml/min]Heart rate [bpm]Right atrial pressure [mmHg]Pulmonary arterial pressure [mmHg]Pulmonary capillary wedge pressure [mmHg]End-diastolic (end-systolic) ventricular pressure [mmHg]Pulmonary vascular resistance [mmHg · s/ml]Left ventricular stroke work [g · m−1]Right ventricular stroke work [g · m−1]Left ventricular stroke work index [g · m · m−2]Right ventricular stroke work index [g · m · m−2]Intra-aortic balloon pumpMechanical ventilator assistance
Introduction
The intra-aortic balloon pump (IABP) is a widely available in-series cardiac assist device. It consists of a double-lumen catheter with a polyurethane balloon attached at its distal end and a mobile pump console, which shuttles helium through the main lumen of the catheter. The tip of the catheter has a pressure sensor to monitor aortic blood pressure. The IABP is now inserted percutaneously through the femoral artery and positioned just below the origin of the left subclavian artery either under fluoroscopic guidance in the cath lab or under trans-oesophageal guidance in theatre.The IABP is based on the principle of counterpulsation, which aims to optimize the balance between myocardial oxygen supply and demand in terms of endocardial viability ratio (EVR) [1],[2].The functional relationship between stroke volume (SV), aortic mean diastolic pressure (MDP), tension time index (TTI), aortic end diastolic pressure (EDP), balloon inflation/deflation timing and heart rate (HR) is a key element for optimal pump control [1],[2].The physiological advantage of intra-aortic balloon counterpulsation is increased aortic diastolic blood pressure. Rapid inflation of the balloon at the beginning of diastole generates proximal and distal blood displacement, which is proportional to the volume of the balloon. Diastolic blood pressure augmentation increases the intrinsic windkessel effect leading to storage of extra potential energy in the aorta and conversion to kinetic energy following the elastic recoil of the vessel [3]. This event has the potential to increase coronary blood flow. Rapid deflation of the balloon in early systole leads to afterload reduction or, to be more precise, to reduction of impedance to ventricular ejection and cardiac work [4],[5]. This is in accordance with a previous analytical model [5].The ability to model the interactions between IABP and the cardiovascular system and how alterations of specific parameters such as timing can affect their coupling remains a key element for clinical application [6],[7].Simulations of combined VA-ECMO and IABP support show an increase in pulsatility and LV stroke volume between 5% and 10% due to afterload reduction although PCWP and left ventricular EDV are only marginally affected. Significant LV unloading is achieved during combined VA-ECMO and Impella support although aortic valve opening and improved diastolic coronary perfusion pressure are not observed in comparison with IABP [8]. Nevertheless, the pulse contour is higher and more similar to the physiological pattern during partial ECMO support where some degree of LV ejection is allowed [9].The concomitant use of IABP and VA-ECMO shows reduced in-hospital mortality in patients with cardiogenic shock secondary to post-cardiotomy failure, ischaemic heart disease and myocarditis [10], [11], which is in contrast with the questionable outcome of the SHOCK II trial [12], [13], [14]. The study was designed as a multicentre, randomised, open-label trial. Between 2009 and 2012, 600 patients with cardiogenic shock following acute myocardial infarction and requiring early revascularisation were randomised to IABP versus control. Long-term follow-up (6.2 years) showed no difference in mortality, recurrent myocardial infarction, stroke, repeat revascularisation or hospital readmission for cardiac reasons between the two groups. Nevertheless, the use of IABP in cardiogenic shock remains the subject of significant debate and controversy.
Electrical analogue of the cardiovascular network. The key compartments and their relationship are highlighted: pulmonary (systemic) venous section to left (right) atrium via mitral (tricuspid) valve to left (right) ventricle; aortic (pulmonary) valve to systemic (pulmonary) arterial section. The coronary network is modelled using RC elements as already described in [22], [23], [24], [25], [26]. Pt is the mean intrathoracic pressure.
Figure 2 shows the electrical analogue of the systemic circulation and the IABP. The systemic bed is divided in aortic, thoracic and two abdominal tracts. When the IABP is “OFF” (SW1=ON and SW2=OFF), the aortic (thoracic) tract is modelled using resistance RAT (RTT), compliance CAT (CTT) and inertance LAT (LTT). The behavior of the first abdominal tract is reproduced by RLC elements (RABT1, LABT1 and CABT1) when the IABP is disabled. The second abdominal tract is modelled by RABT2, LABT2, CABT2 and by variable systemic arterial resistance (Ras). The switches SW1 and SW2 are set to OFF (ON) and ON (OFF) respectively, when the device is (not) working. The effect induced by the balloon is reproduced in each tract of the aorta by the presence of compliances (CIABP1, CIABP2 andCIABP3) connected to PIABP generator and resistances (RIABP1, RIABP2 and RIABP3). PIABP generator reproduces the balloon pressure with the option to change the IABP timing. The resistances (compliances) RIABP1 (CIABP1), RIABP2 (CIABP2) and RIABP3 (CIABP3) are connected in series (parallel) to RAT (CAT), RTT (CTT) and RABT1 (CABT1) in the aortic, thoracic and first abdominal tract. Figure 3
shows the waveform of the pressure produced by the generator that is defined as follows:
Fig. 2
Electrical analogue of systemic arterial section and IABP. The systemic compartment consists of aortic, thoracic and abdominal tract. The abdominal bed is divided in two parts modelled with RABD1, LABD1, CABD1 and RABD2, LABD2, CABD2 and Ras elements respectively. The aortic, thoracic and first abdominal tract is directly influenced by IABP activation. Qi (i=AT, TT, ABT1 and ABT2) represents the flow inside each compartment. RAT, LAT, and CAT (RTT, LTT, and CTT) reproduce the aortic (thoracic) bed. The compliances CIABP1, CIABP2 and CIABP3 and the resistances RIABP1, RIABP2 and RIABP3 with the generator PIABP allow simulating the effects of the intra-aortic counterpulsation.
Electrical analogue of systemic arterial section and IABP. The systemic compartment consists of aortic, thoracic and abdominal tract. The abdominal bed is divided in two parts modelled with RABD1, LABD1, CABD1 and RABD2, LABD2, CABD2 and Ras elements respectively. The aortic, thoracic and first abdominal tract is directly influenced by IABP activation. Qi (i=AT, TT, ABT1 and ABT2) represents the flow inside each compartment. RAT, LAT, and CAT (RTT, LTT, and CTT) reproduce the aortic (thoracic) bed. The compliances CIABP1, CIABP2 and CIABP3 and the resistances RIABP1, RIABP2 and RIABP3 with the generator PIABP allow simulating the effects of the intra-aortic counterpulsation.Schematic representation of the balloon pressure. T0 starting balloon inflation, T1 ending balloon inflation, T1-T2 inflation overshoot interval (PDRIVE), T3-T4 maximal inflation interval, T4 starting balloon deflation, T5 ending balloon deflation, T5-T6 deflation overshoot interval, T6 starting filling pressure baseline, T7 ending filling pressure baseline (PVACUUM).When the IABP is OFF, the network showed in Fig. 2 is solved by the equations:If the IABP is ON, the equations are:where Plv is the left ventricular pressure. The resistance Rlo and the diode D1 model the mitral valve (Fig. 2). The software simulator allows the IABP to be synchronized with either the ECG or the aortic pressure waveform. The frequency of balloon-assisted beats can be set from the maintenance 1:1 ratio to a weaning 1:2 ratio (every other systole is assisted). Depending on the clinician's judgment, weaning modes of 1:4 or even 1:8 may be initiated if a more gradual approach is needed. In addition, IABP driving and vacuum pressures can be changed.The hemodynamic effects induced by the IABP may vary with assisting frequency and depend on balloon inflation/deflation timing. A range of settings T1-T7 is available in the software simulator. The IABP can be triggered to deflate during systole once the peak of the R wave is sensed. IABP inflation may be triggered to occur in the middle of the T wave, which corresponds to diastole. The simulator allows the setting of different delays. Changing IABP compliance and resistance allows the balloon volume to be modified.
Cardiogenic Shock Patients
Patients with acute myocardial infarction (AMI) and cardiogenic shock (CS) may require intra-aortic balloon counterpulsation as an adjunct to medical treatment [27]. For the purposes of our study, literature data [28], [29], [30] were used to reproduce the baseline conditions of CS patients and those following IABP assistance. The hemodynamic data used in this study have been listed in Table 1
.
Table 1
Literature data for pathological and assisted conditions.
Baseline conditions
IABP on [1:1]
PDP [mmHg]
67±7
75±7
EDP [mmHg]
55±6
42±9
HR [beat/min]
99±27
100±22
LV EDV [cc]
293±35
285±36
LV ESV [cc]
263±32
259±36
SV [cc]
29±4
26±4
PAP [mmHg]
42±11
32±6
PCWP [mmHg]
27±8
23±6
RA [mmHg]
10±5
10±5
CI [l·min−1/m2]
1.56±0.29
1.57±0.3
SVR [Wood]
21.67±5.2
16.42±11.8
PVR [Wood]
4.95±2.7
1.46±1.1
BSA [m2]
1.8±0.13
1.8±0.13
CO [l · min−1]
2.8±1.25
2.91±1.25
Literature data for pathological and assisted conditions.
Simulation Protocol
The duration of the whole cardiac cycle was set at 1000 ms for all the simulations. Starting from the reproduced baseline conditions, the IABP was activated with a driving (vacuum) pressure of PDRIVE = 240 mmHg (PVACUUM= -10 mmHg). The plateau IABP pressure was set to PPLATEAU = 150 mmHg. During the simulations, the IABP was synchronized with the ECG and its ratio was set to 1:1, 1:2, 1:4 and 1:8. During baseline conditions and when the IABP ratio was set to 1:1, the mean values for pressure, flow, EDV and ESV (for both ventricles) were calculated for one cardiac cycle. When the IABP ratio was set to 1:2 (1:4 or 1:8), the mean values for pressure, flow, EDV and ESV (for both ventricles) were calculated for two (four or eight) cardiac cycles.
We have developed an innovative IABP numerical model with potential for clinical application. The focus has been on IABP timing and the evaluation of the most appropriate weaning strategy in terms of device assist ratio. Its value as a training tool in a clinical setting has been proposed. Although based on literature data, the outcome of the simulations is encouraging. Additional work is ongoing with a view to further validate its features.
Authors: Claudio De Lazzari; D Neglia; G Ferrari; F Bernini; M Micalizzi; A L'Abbate; M G Trivella Journal: Methods Inf Med Date: 2009-02-27 Impact factor: 2.176
Authors: Stéphanie Schampaert; Marcel C M Rutten; Marcel van T Veer; Lokien X van Nunen; Pim A L Tonino; Nico H J Pijls; Frans N van de Vosse Journal: ASAIO J Date: 2013 Jan-Feb Impact factor: 2.872