Tolga Karapinar1, Kenan Cagri Tumer1, Sébastien Buczinski2.
Abstract
BACKGROUND: Data on the performance of a glucometer in calves with different diseases are currently lacking.Entities:
Keywords: accuracy; bovine; glucometer; precision; sensitivity; specificity
Mesh:
Substances:
Year: 2020 PMID: 32420677 PMCID: PMC7379022 DOI: 10.1111/jvim.15794
Source DB: PubMed Journal: J Vet Intern Med ISSN: 0891-6640 Impact factor: 3.333
FIGURE 1The Passing‐Bablok regression analysis (top) of the blood glucose concentrations determined using the POC glucometer versus plasma glucose concentrations measured using the autoanalyzer (n = 196). The diagonal dashed line is the identity line, the solid line represents the best fit, and the dashed lines represent the 95% confidence interval (CI). The Bland‐Altman plot analysis (bottom) of differences between blood glucose concentrations determined using the POC glucometer and plasma glucose concentrations determined using the autoanalyzer against their averages. The solid horizontal, straight line is the mean bias, the thick dashed line is the best fit line, the short horizontal lines represent the 95% CI of the bias, and the horizontal dashed lines represent the 95% CI of agreement
Receiver operating characteristic analysis for the Freestyle Optium Neo H in sick calves (samples, n = 196)
| Threshold glucose concentration | Optimized Cutoff | Sensitivity (95% CI) | Specificity (95% CI) | AUC |
| J |
|---|---|---|---|---|---|---|
| 40 mg/dL (n = 25) | ≤43 | 100 (83.9‐100) | 96 (91.9‐98.4) | 0.997 (0.975‐1) | <.0001 | 0.96 |
| 58 mg/dL (n = 45) | ≤54 | 92.7 (80.1‐98.5) | 97.4 (93.5‐99.3) | 0.991 (0.965‐0.999) | <.0001 | 0.90 |
| 72 mg/dL (n = 68) | ≤72 | 92.2 (82.7‐97.4) | 86.4 (79.3‐91.7) | 0.966 (0.930‐0.987) | <.0001 | 0.79 |
Plasma glucose concentration determined by the reference method.
Optimized cutoff value for the Freestyle Optium Neo H.
Area under the ROC curve.
Youden index.
FIGURE 2Percentage difference (%) between the point of care and reference analyzer ((GluPOC‐GluRef)/GluRef) and hematocrit value (%) in 196 calves. The line represents the significant linear association (Table 2) between hematocrit and glycemia percent difference and its associated SE shaded in gray. GluPOC, glycemia as evaluated by the point of care analyzer; GluRef, glycemia as evaluated by the reference analyzer
Results of multivariable regression analysis characterizing the effect of the Hct on the percent difference between the point of care (POC) Freestyle Optium Neo H and the reference Cobas analyzer (DGlu (%) = (GluPOC‐GluRef)/GluRef)). (n = 196)
| Estimated value | SE |
| |
|---|---|---|---|
| Intercept | 7.117 | 4.049 | .08 |
| Hematocrit (%) | −0.270 | 0.117 | .02 |