Evaluation of five portable blood glucose meters for use in dogs.

G Wess1, C Reusch.   

Abstract

OBJECTIVE: To evaluate clinical and analytical accuracy of 5 portable blood glucose meters (PBGM) used to measure blood glucose concentrations in dogs and to determine potential sources of error.
DESIGN: Prospective study. ANIMALS: 221 dogs. PROCEDURE: Venous blood samples were obtained, and results of the 5 PBGM were compared with results of a hexokinase reference method. Agreement among methods was determined by use of error grid analysis and statistical methods.
RESULTS: Accuracy of the PBGM varied with glucose concentration of the sample. The largest differences between results of the PBGM and results of the reference method were obtained with samples with high glucose concentrations; 4 PBGM tended to underestimate and 1 PBGM tended to overestimate the true glucose concentration. Absolute differences between results of the PBGM and results of the reference method were small for samples with low glucose concentrations and samples with concentrations in the reference range. None of the PBGM yielded measurements that would result in clinically unacceptable errors. Within-run and between-day precision was good for all PBGM, and results were not affected by use of EDTA or heparin to anticoagulate blood. Readings of the PBGM were significantly higher for blood samples with low Hct than for samples with normal Hct. For 3 PBGM, samples < 3 microliters resulted in inaccurate measurements. CONCLUSIONS AND CLINICAL RELEVANCE: Results suggest that currently available PBGM are sufficiently accurate for use in clinical practice to determine blood glucose concentrations in dogs.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10649754     DOI: 10.2460/javma.2000.216.203

Source DB:  PubMed          Journal:  J Am Vet Med Assoc        ISSN: 0003-1488            Impact factor:   1.936


  15 in total

1.  Retrospective study of owners' perception on home monitoring of blood glucose in diabetic dogs and cats.

Authors:  I Van de Maele; N Rogier; S Daminet
Journal:  Can Vet J       Date:  2005-08       Impact factor: 1.008

2.  Accuracy of 5 Point-of-Care Glucometers in C57BL/6J Mice.

Authors:  Linnea A Morley; Thomas H Gomez; Julia L Goldman; Rene Flores; Mary A Robinson
Journal:  J Am Assoc Lab Anim Sci       Date:  2018-01-01       Impact factor: 1.232

3.  Comparison of a human portable glucometer and an automated chemistry analyzer for measurement of blood glucose concentration in pet ferrets (Mustela putorius furo).

Authors:  Noémie M Summa; David Eshar; Bridget Lee-Chow; Sylvain Larrat; Dorothy C Brown
Journal:  Can Vet J       Date:  2014-09       Impact factor: 1.008

Review 4.  Hypoglycemia in dogs: Causes, management, and diagnosis.

Authors:  Olutunbi Idowu; Kathryn Heading
Journal:  Can Vet J       Date:  2018-06       Impact factor: 1.008

Review 5.  Monitoring methods for dogs and cats with diabetes mellitus.

Authors:  Audrey K Cook
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

6.  Blood sugar measurement in zebrafish reveals dynamics of glucose homeostasis.

Authors:  Stefani C Eames; Louis H Philipson; Victoria E Prince; Mary D Kinkel
Journal:  Zebrafish       Date:  2010-06       Impact factor: 1.985

7.  Comparison of 2 glucose analytical methodologies in immature Kemp's ridley sea turtles: dry chemistry of plasma versus point-of-care glucometer analysis of whole blood.

Authors:  Justin R Perrault; Michael D Arendt; Jeffrey A Schwenter; Julia L Byrd; Kathryn A Tuxbury; Nicole I Stacy
Journal:  J Vet Diagn Invest       Date:  2021-03-19       Impact factor: 1.279

8.  Obtaining accurate glucose measurements from wild animals under field conditions: comparing a hand held glucometer with a standard laboratory technique in grey seals.

Authors:  Kimberley A Bennett; Lucy M Turner; Sebastian Millward; Simon E W Moss; Ailsa J Hall
Journal:  Conserv Physiol       Date:  2017-02-27       Impact factor: 3.079

9.  Comparison of glucose concentrations in canine whole blood, plasma, and serum measured with a veterinary point-of-care glucometer.

Authors:  Natalie D Suchowersky; Elizabeth A Carlson; Hollie P Lee; Ellen N Behrend
Journal:  J Vet Diagn Invest       Date:  2021-06-02       Impact factor: 1.569

10.  ISO-based assessment of accuracy and precision of glucose meters in dogs.

Authors:  Y Brito-Casillas; P Figueirinhas; J C Wiebe; L López-Ríos; D Pérez-Barreto; C Melián; A M Wägner
Journal:  J Vet Intern Med       Date:  2014-07-02       Impact factor: 3.333

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.