Literature DB >> 32371853

Assessment of Eye Lens Dose Reduction When Using Lateral Lead Shields on the Patient's Head during Neurointerventional Fluoroscopic Procedures and Cone-beam Computed Tomography (CBCT) Scans.

Zhenyu Xiong, Stephen Rudin1, Daniel R Bednarek1.   

Abstract

The purpose of this study was to evaluate the effect of placing small lead shields on the temple region of the skull to reduce radiation dose to the lens of the eye during interventional fluoroscopically-guided procedures and cone-beam computed tomography (CBCT) scans of the head. EGSnrc Monte-Carlo code was used to determine the eye lens dose reduction when using lateral lead shields for single x-ray projections, CBCT scans with different protocols, and interventional neuroradiology procedures with the Zubal computational head phantom. A clinical C-Arm system was used to take radiographic projections and CBCT scans of anthropomorphic head phantoms without and with lead patches, and the images were compared to assess the effect of the shields. For single lateral projections, a 0.1 (0.3)-mm-thick lead patch reduced the dose to the left-eye lens by 40% to 60% (55% to 80%) from 45° to 90° RAO and to the right-eye lens by around 30% (55%) from 70° to 90° RAO. For different CBCT protocols, the reduction of lens dose with a 0.3-mm-thick lead patch ranged from 20% to 53% at 110 kVp. For CBCT scans of the anthropomorphic phantom, the lead patch introduced streak artifacts that were mainly in the orbital regions but were insignificant in the brain region where most neurointerventional activity occurs. The dose to the patient's eye lens can be reduced considerably by placing small lead shields over the temple region of the head without substantially compromising image quality in neuro-imaging procedures.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32371853      PMCID: PMC7398852          DOI: 10.1097/HP.0000000000001232

Source DB:  PubMed          Journal:  Health Phys        ISSN: 0017-9078            Impact factor:   2.922


  14 in total

1.  Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose.

Authors:  Eike Hein; Patrik Rogalla; Randolph Klingebiel; Bernd Hamm
Journal:  Eur Radiol       Date:  2002-02-02       Impact factor: 5.315

2.  Lens of the eye dose calculation for neuro-interventional procedures and CBCT scans of the head.

Authors:  Zhenyu Xiong; Sarath Vijayan; Vijay Rana; Amit Jain; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-22

3.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

4.  Radioprotection to the eye during CT scanning.

Authors:  K D Hopper; J D Neuman; S H King; A R Kunselman
Journal:  AJNR Am J Neuroradiol       Date:  2001 Jun-Jul       Impact factor: 3.825

5.  Computerized three-dimensional segmented human anatomy.

Authors:  I G Zubal; C R Harrell; E O Smith; Z Rattner; G Gindi; P B Hoffer
Journal:  Med Phys       Date:  1994-02       Impact factor: 4.071

6.  Bismuth shielding, organ-based tube current modulation, and global reduction of tube current for dose reduction to the eye at head CT.

Authors:  Jia Wang; Xinhui Duan; Jodie A Christner; Shuai Leng; Katharine L Grant; Cynthia H McCollough
Journal:  Radiology       Date:  2012-01       Impact factor: 11.105

7.  Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction.

Authors:  Maria Raissaki; Kostas Perisinakis; John Damilakis; Nicholas Gourtsoyiannis
Journal:  Pediatr Radiol       Date:  2010-06-16

8.  Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study.

Authors:  Kostas Perisinakis; Maria Raissaki; Nicholas Theocharopoulos; John Damilakis; Nicholas Gourtsoyiannis
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

9.  [Protection of eye lens in computed tomography--dose evaluation on an anthropomorphic phantom using thermo-luminescent dosimeters and Monte-Carlo simulations].

Authors:  B Keil; J Wulff; R Schmitt; D Auvanis; D Danova; J T Heverhagen; M Fiebich; B Madsack; R Leppek; K J Klose; K Zink
Journal:  Rofo       Date:  2008-11-28

10.  Local skin and eye lens equivalent doses in interventional neuroradiology.

Authors:  Michael Sandborg; Sandro Rossitti; Håkan Pettersson
Journal:  Eur Radiol       Date:  2009-09-02       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.