Optimization methods for magnetic resonance imaging gradient waveform design.

Matthew J Middione1,2, Michael Loecher1,2, Kévin Moulin1,2, Daniel B Ennis1,2.   

Abstract

The development and implementation of novel MRI pulse sequences remains challenging and laborious. Gradient waveforms are typically designed using a combination of analytical and ad hoc methods to construct each gradient waveform axis independently. This strategy makes coding the pulse sequence complicated, in addition to being time inefficient. As a consequence, nearly all commercial MRI pulse sequences fail to maximize use of the available gradient hardware or efficiently mitigate physiological effects. This results in expensive MRI systems that underperform relative to their inherent hardware capabilities. To address this problem, a software solution is proposed that incorporates numerical optimization methods into MRI pulse sequence programming. Examples are shown for rotational variant vs. invariant waveform designs, acceleration nulled velocity encoding gradients, and mitigation of peripheral nerve stimulation for diffusion encoding. The application of optimization methods to MRI pulse sequence design incorporates gradient hardware limits and the prescribed MRI protocol parameters (e.g. field-of-view, resolution, gradient moments, and/or b-value) to simultaneously construct time-optimal gradient waveforms. In many cases, the resulting constrained gradient waveform design problem is convex and can be solved on-the-fly on the MRI scanner. The result is a set of multi-axis time-optimal gradient waveforms that satisfy the design constraints, thereby increasing SNR-efficiency. These optimization methods can also be used to mitigate imaging artifacts (e.g. eddy currents) or account for peripheral nerve stimulation. The result of the optimization method is to enable easier pulse sequence gradient waveform design and permit on-the-fly implementation for a range of MRI pulse sequences.
© 2020 John Wiley & Sons, Ltd.

Entities:  

Keywords:  acquisition methods; applications; diffusion MR sequences; diffusion methods; flow imaging methods; flow quantitation; methods and engineering; other applications

Year:  2020        PMID: 32342560      PMCID: PMC7606655          DOI: 10.1002/nbm.4308

Source DB:  PubMed          Journal:  NMR Biomed        ISSN: 0952-3480            Impact factor:   4.044


  33 in total

1.  MRI pulse sequence design with first-order gradient moment nulling in arbitrary directions by solving a polynomial program.

Authors:  Kurt Majewski; Oliver Heid; Thomas Kluge
Journal:  IEEE Trans Med Imaging       Date:  2010-03-18       Impact factor: 10.048

2.  Image formation by induced local interactions. Examples employing nuclear magnetic resonance. 1973.

Authors:  P C Lauterbur
Journal:  Clin Orthop Relat Res       Date:  1989-07       Impact factor: 4.176

3.  Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging].

Authors:  J Pauly; P Le Roux; D Nishimura; A Macovski
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

4.  Elimination of eddy current artifacts in diffusion-weighted echo-planar images: the use of bipolar gradients.

Authors:  A L Alexander; J S Tsuruda; D L Parker
Journal:  Magn Reson Med       Date:  1997-12       Impact factor: 4.668

5.  Application of linear optimization techniques to MRI phase contrast blood flow measurements.

Authors:  V L Morgan; R R Price; C H Lorenz
Journal:  Magn Reson Imaging       Date:  1996       Impact factor: 2.546

Review 6.  Multishot rosette trajectories for spectrally selective MR imaging.

Authors:  D C Noll
Journal:  IEEE Trans Med Imaging       Date:  1997-08       Impact factor: 10.048

7.  Motion-robust and blood-suppressed M1-optimized diffusion MR imaging of the liver.

Authors:  Yuxin Zhang; Óscar Peña-Nogales; James H Holmes; Diego Hernando
Journal:  Magn Reson Med       Date:  2019-03-12       Impact factor: 4.668

8.  A low power radiofrequency pulse for simultaneous multislice excitation and refocusing.

Authors:  Cornelius Eichner; Lawrence L Wald; Kawin Setsompop
Journal:  Magn Reson Med       Date:  2014-08-07       Impact factor: 4.668

9.  Optimization of diffusion-weighted single-refocused spin-echo EPI by reducing eddy-current artifacts and shortening the echo time.

Authors:  Manoj Shrestha; Pavel Hok; Ulrike Nöth; Bianca Lienerth; Ralf Deichmann
Journal:  MAGMA       Date:  2018-03-30       Impact factor: 2.310

10.  Maxwell-compensated design of asymmetric gradient waveforms for tensor-valued diffusion encoding.

Authors:  Filip Szczepankiewicz; Carl-Fredrik Westin; Markus Nilsson
Journal:  Magn Reson Med       Date:  2019-05-31       Impact factor: 4.668

View more
  3 in total

1.  Quantitative diffusion MRI of the abdomen and pelvis.

Authors:  Diego Hernando; Yuxin Zhang; Ali Pirasteh
Journal:  Med Phys       Date:  2021-10-08       Impact factor: 4.506

2.  Flow-compensated diffusion encoding in MRI for improved liver metastasis detection.

Authors:  Frederik B Laun; Tobit Führes; Hannes Seuss; Astrid Müller; Sebastian Bickelhaupt; Alto Stemmer; Thomas Benkert; Michael Uder; Marc Saake
Journal:  PLoS One       Date:  2022-05-26       Impact factor: 3.752

3.  Gradient nonlinearity correction in liver DWI using motion-compensated diffusion encoding waveforms.

Authors:  Sean McTavish; Anh T Van; Johannes M Peeters; Kilian Weiss; Marcus R Makowski; Rickmer F Braren; Dimitrios C Karampinos
Journal:  MAGMA       Date:  2021-12-11       Impact factor: 2.533

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.