MRI pulse sequence design with first-order gradient moment nulling in arbitrary directions by solving a polynomial program.

Kurt Majewski1, Oliver Heid, Thomas Kluge.   

Abstract

We suggest a polynomial program for the calculation of optimized gradient waveforms for magnetic resonance tomography pulse sequences. Such non-linear mathematical programs can describe gradient system capabilities, meet k-space trajectory specifications, and capture sequence timing conditions. Moreover they allow the incorporation of gradient moment nulling constraints in one or several arbitrary spatial directions, which can reduce flow motion artifacts in the images. We report first experiences in solving such automatic pulse sequence design programs with the interior point solver Ipopt.

Mesh:

Year:  2010        PMID: 20304730     DOI: 10.1109/TMI.2010.2042723

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048