| Literature DB >> 32326885 |
Marcus Wieczorek1, Kiarash Sassani2, Reinhard Hoeltgen2.
Abstract
BACKGROUND: Pulmonary vein isolation (PVI) using phased radiofrequency (RF) energy has been shown to be effective in the treatment of atrial fibrillation (AF).Entities:
Keywords: Atrial fibrillation; Cryoballoon; Phased radiofrequency; Pulmonary vein isolation; Pulmonary vein reconnection
Mesh:
Year: 2020 PMID: 32326885 PMCID: PMC7181531 DOI: 10.1186/s12872-020-01459-4
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Clinical characteristics of the enrolled patients (N = 151)
| Group 1 (PVAC | Group 2 (Cryo | ||
|---|---|---|---|
| Age (years) | 58 ± 17 [31–72] | 63 ± 12 [29–83] | ns |
| Male | 56 (66) | 39 (59) | ns |
| CHA2DS2-VASc-score | 2.2 [0–5] | 2.4 [0–5] | ns |
| Paroxysmal AF | 58 (68) | 40 (60) | ns |
| Left ventricular ejection fraction (%) | 57 ± 8 | 54 ± 7 | ns |
| Left atrial diameter (mm) | 43 ± 4 [37–50] | 42 ± 6 [36–53] | ns |
| No. of antiarrhythmic drugs | 1.4 ± 1.3 [0–3] | 1.1 ± 1.2 [0–3] | ns |
| Amiodarone therapy | 17 (20) | 12 (18) | ns |
| No structural heart disease | 44 (52) | 28 (42) | ns |
| Hypertension | 42 (49) | 38 (58) | ns |
| Coronary artery disease | 22 (26) | 15 (23) | ns |
| Chronic obstructive pulmonary disease | 9 (11) | 7 (11) | ns |
| Implantable cardiac devicesa | 6 (7) | 7 (11) | ns |
Data are numbers with mean ± SD, range [] and % in (); ns: no significant difference, a pacemaker, implantable cardioverter/defibrillator, event recorder
Electrophysiological characteristics at redo procedure
| Group 1 (PVAC | Group 2 (CB | ||
|---|---|---|---|
| Angiographically delineated PVs / CTs | 308/17 | 244/11 | – |
| Targeted PVs / CTs at initial procedure | 302/17 | 242/11 | 1.0000 |
| Isolated PVs / CTs at initial procedure | 298/17 | 240/10 | 1.0000 |
| Pts. with PV reconnections | 79 (93%) | 61 (92%) | 1.0000 |
| Reconnected PVs and CTs at redo procedure | 237/315 (75%) | 130/250 (52%) | < 0.0001 |
| Reconnected left-sided PVs/CTs | 125/152 (82%) | 75/119 (63%) | 0.0005 |
| Reconnected CTs | 16/17 (94%) | 6/10 (60%) | 0.0473 |
| Reconnected LSPV | 57/67 (85%) | 36/54 (67%) | 0.0170 |
| Reconnected LIPV | 52/68 (76%) | 33/55 (60%) | 0.0533 |
| Reconnected right-sided PVs | 112/163 (69%) | 55/131 (42%) | < 0.0001 |
| Reconnected RSPV | 63/79 (80%) | 29/65 (45%) | < 0.0001 |
| Reconnected RIPV | 49/84 (58%) | 26/66 (39%) | 0.0321 |
| Reconnected superior PVs | 120/146 (82%) | 65/119 (55%) | < 0.0001 |
| Reconnected inferior PVs | 101/152 (66%) | 59/121 (49%) | 0.0044 |
| PV reconnections per patient #: | 2.79 ± 1.2 | 1.97 ± 0.8 | < 0.0001 |
| Pts. with 4 reconnected PVs | 28 (33%) | 0 (0%) | < 0.0001 |
| Pts. with 3 reconnected PVs | 28 (33%) | 17 (26%) | 0.2899 |
| Pts. with 2 reconnected PVs | 18 (21%) | 35 (53%) | 0.0001 |
| Pts. with 1 reconnected PV | 5 (6%) | 9 (14%) | 0.2723 |
| Pts. without reconnected PVs | 6 (7%) | 5 (8%) | 1.0000 |
PTs Patients, CB 2nd Generation cryoballoon, PVs Pulmonary veins, CTs Common trunks, LSPV Left superior pulmonary vein, LIPV Left inferior pulmonary vein, RSPV Right superior pulmonary vein and RIPV Right inferior pulmonary vein. #: in patients with a reconnected common trunk, CT was considered and counted as one PV
Fig. 1Schematic left anterior oblique view to LA with percentage of reconnection observed for each pulmonary vein / left common trunk after PVAC (left) and Cryoballoon PVI (right)
Percentage of patients with reconnection at redo procedure calculated for each vein
| Group 1 (PVAC | Group 2 (CB | ||
|---|---|---|---|
| Reconnected CTs | 16 (19%) | 6 (9%) | 0.1077 |
| Reconnected LSPV | 57 (67%) | 36 (55%) | 0.1313 |
| Reconnected LIPV | 52 (61%) | 33 (50%) | 0.1885 |
| Reconnected RSPV | 63 (74%) | 29 (44%) | 0.0002 |
| Reconnected RIPV | 49 (58%) | 26 (39%) | 0.0330 |
No. of reconnections and percentage () related to each group, CTs Common trunks, LSPV Left superior pulmonary vein, LIPV Left inferior pulmonary vein, RSPV Right superior pulmonary vein and RIPV Right inferior pulmonary vein, CB 2nd Generation cryoballoon
Comparison of PVAC and CB reconnection patterns
| a) | Reconnected left-sided vs right-sided PVs/CTs | Reconnected superior vs inferior PVs/CTs |
|---|---|---|
| PVAC | 125/152 (82%) vs 112/163 (69%) | 136/163 (83%) vs 101/152 (66%) |
| p value | ||
| CB | 75/119 (63%) vs 55/131 (42%) | 71/129 (55%) vs 59/121 (49%) |
| p value | 0.3754 | |
| b) | ||
| PVAC | 109/135 (81%) vs 112/163 (69%) | 120/146 (82%) vs 101/152 (66%) |
| p value | ||
| CB | 69/109 (63%) vs 55/131 (42%) | 65/119 (55%) vs 59/121 (49%) |
| p value | 0.3696 |