The Incidence, Electrophysiological Characteristics and Ablation Outcome of Left Atrial Tachycardias after Pulmonary Vein Isolation Using Three Different Ablation Technologies.

Patrick Leitz1, Kristina Wasmer1, Christian Andresen1, Fatih Güner1, Julia Köbe1, Benjamin Rath1, Florian Reinke1, Julian Wolfes1, Philipp S Lange1, Christian Ellermann1, Gerrit Frommeyer1, Lars Eckardt1.   

Abstract

BACKGROUND: Left atrial tachycardias (LAT) are a well-known outcome of pulmonary vein isolation (PVI). Few data are available on whether the catheter used to perform PVI influences the incidence, as well as the characteristics of post PVI LAT. We present data on LAT following PVI by the following three ablation technologies: (1) phased multi-electrode radiofrequency catheter (PVAC), (2) irrigated single-tip catheter (iRF), and (3) cryoballoon ablation.
METHODS: Using a prospectively designed single-center database, we analyzed 650 patients (300 iRF, 150 PVAC, and 200 cryoballoon) with paroxysmal (n = 401) and persistent atrial fibrillation (AF), who underwent their first PVI at our center.
RESULTS: The three populations were comparable in their baseline characteristics; however, the cryoballoon group comprised a higher percentage of patients with persistent AF (p = 0.05). The LAT rates were 3.7% in the iRF group (mean follow-up 22 ± 14 months), 0.7% in the PVAC group (mean follow-up 21 ± 14 months), and 4% in the cryoballoon group (mean follow-up 15 ± 8 months). The predominant mechanism of LAT was macro-reentrant tachycardia. Reconnection of at least one pulmonary vein was observed in 87% of the patients who underwent 3D mapping. No predictors for LAT occurrence were identified.
CONCLUSION: The occurrence of LAT post PVI is rare; the predominant mechanism was macro-reentrant tachycardia. Reconnection of at least one pulmonary vein was observed in nearly all the LAT patients. In our retrospective analysis, the lowest rate of LAT was observed with the PVAC. No predictors for LAT occurrence were identified.

Entities:  

Keywords:  atrial fibrillation; left atrial flutter; pulmonary vein isolation

Year:  2022        PMID: 35200703      PMCID: PMC8879099          DOI: 10.3390/jcdd9020050

Source DB:  PubMed          Journal:  J Cardiovasc Dev Dis        ISSN: 2308-3425


1. Introduction

Atrial fibrillation (AF) is the most widespread sustained arrhythmia [1] and a major cause of stroke [2]. In symptomatic patients, pulmonary vein isolation (PVI) has steadily gained importance in comparison to antiarrhythmic drugs [3,4]. Developments such as the cryoballoon and the multipolar phased ablation catheter have allowed for shorter procedural times with comparable efficacy and safety as the established single-tip point-by-point ablation [5,6]. A well-known outcome of PVI is the occurrence of left atrial tachycardias (LAT), which are often more symptomatic than the initial AF. It remains unclear if the technology used for PVI affects the occurrence rates and electrophysiological mechanisms of post PVI LAT. We sought to determine the occurrence rates, as well as mechanisms, of post PVI LAT after ablation, performed with the following three technical approaches: (1) single-tip irrigated RF catheter (iRF), (2) phased multipolar RF ablation catheter (PVAC), and (3) a cryoballoon system.

2. Methods

2.1. Study Design and Patient Collective

We included a total of 650 patients who were referred to our center for an index PVI. Data were collected from a prospectively designed database, as well as our outpatient program. Patients were seen three months after PVI for an in-person visit. After 6 and 12 months, telephone interviews were performed. Further telephone interviews were scheduled in 6–12-month intervals. Additionally, patients were instructed to present to our department in case of symptomatic arrhythmias. Data collection was performed from 2011 to 2020. Exclusion criteria were defined as follows: severe valvular disease, prior surgical- or catheter-based PVI, prior linear ablations, inability to give informed consent, pregnancy, <18 years old.

2.2. Diagnosis of LAT

During the in-person visits and telephone interviews, patients were interrogated on arrhythmia occurrence. In the case of recurrence, ECG documentation was obtained. LATs within 3 months following PVI were discarded from the analysis.

2.3. Index PVI

The catheter used for PVI was chosen at the operator’s discretion. Point-by-point RF ablation (St. Jude Medical, Saint Paul, MN, USA), ablation with the PVAC (Medtronic Inc., Minneapolis, MN, USA) system, as well as the cryoballoon (Medtronic Inc., Minneapolis, MN, USA) performed at our center have been described in detail in previous publications [7,8,9,10]. In brief, the groups are outlined below: iRF group: iRF ablation was guided by the NAVX system (Ensite NavXVelocity; St. Jude Medical, Inc., Saint Paul, MN, USA). A 3D model was created with a circular mapping catheter (Inquiry Optima; St. Jude Medical Inc.), and the acquired geometry was fused with the 3D reconstruction of the LA from computed tomography (CT). Ablation was performed with a 4 mm open-tip irrigated catheter (IBI TherapyCoolpath Duo or CoolFlex; St. Jude Medical, Inc). Antral point-by-point circumferential ablation around ipsilateral PVs, with a distance from 0.5 to 1.0 cm from the ostia, was performed. The power was set to 30 W and the temperature was limited to 43 °C. In case of persistent PV conduction, especially at the anterior ridge border of the lateral PVs, we increased the power to 35 W, to close conduction gaps. Complete electrical isolation was monitored and confirmed by the circular mapping catheter during sinus rhythm and differential pacing maneuvers. Ablation lesions were limited to circumferential PVI. PVAC group: Ablation with the PVAC system was performed using the second-generation catheter, the PVAC GOLD (Medtronic Inc., Minneapolis, MN, USA). The PVAC was connected to the GENius (Medtronic Inc.) RF generator. The target temperature was 60 °C, with a maximum power of 8 W. Duty-cycled energy delivery was exclusively delivered in a 4:1 ratio, and the impulse duration was 60 s. Electrode pairs with insufficient tissue contact were selectively deactivated. After RF applications, the PVAC was rotated around the PV ostium, looking for the earliest pulmonary vein potential (PVP) to completely isolate the vein. Finally, the catheter electrodes were advanced in the PV demonstrating the entrance block; the exit block was not routinely checked for. No other catheters were used for touch-up lesions. No further ablations were performed. Cryoballoon group: All cryoballoon patients were treated with the second-generation 28 mm cryoballoon (Medtronic Arctic Front Advance, Medtronic Inc.). Occlusion of the PV was tested by a contrasting agent. In this cohort, 2 freezes were applied for 180 s per vein, irrespective of the achieved time to isolation (TTI). Additional freezes were applied if PVI was not achieved. Energy delivery was prematurely stopped in the case of inadequate temperature development, or if PVI could not be achieved within the first 90 s of the freeze. Ablation of the right PVs was performed under continuous stimulation of the phrenic nerve with a cycle length of 800 ms. Freezes were immediately stopped if temperatures fell below −60 °C, or if phrenic nerve capture was lost. Complete electrical isolation of the PVs was confirmed using the Achieve mapping catheter in sinus rhythm and during differential pacing maneuvers. At the end of the procedure, PVI was verified in sinus rhythm. No further linear ablations were performed and no supplementary ablation catheters were used.

2.4. Ablation Procedures of Post PVI Left Atrial Tachycardia

Mapping of LAT was performed using either NAVX (St. Jude Medical, Saint Paul, MN, USA) or CARTO (Biosense Webster, Diamond Bar, CA, USA) systems. In the more recent cases, multielectrode high-density catheters were used (Pentaray for CARTO, HD Grid Catheter for NAVX). The ablation catheter was either a 3.5 mm irrigated-tip ablation catheter (NAVISTAR Thermo-Cool, Biosense-Webster) or a 4 mm irrigated-tip ablation catheter (IBI Therapy Coolpath Duo 7F, St. Jude Medical, Inc.). In more recent cases, contact force sensing irrigated-tip catheters were used (ThermoCool SmartTouch, Biosense-Webster/TactiCath, Sensor Enabled, St. Jude Medical). A 3-dimensional activation and voltage map was acquired in all patients. The LATs were characterized upon their activation pattern, the presence of an “early meets late” region, and the differential pacing maneuvers (see video of exemplary activation map of peri-mitral and roof-dependent flutters in Supplementary material). RF ablation was performed at a power of 30 W (maximum 25 W in the coronary sinus and maximum 40 W in the LA, in the case of persistent conduction gaps). Peri-mitral flutter was usually treated by an anterior mitral isthmus line. Upon termination of LAT, conduction block of linear lesions was checked through differential pacing and activation re-mapping. Programmed atrial stimulation was performed in order to reinduce the atrial arrhythmia. Finally, in the case of PV reconnection, re-isolation was performed using the same RF catheter used for LAT ablation.

2.5. Statistical Analysis

Statistical analysis was performed using SPSS (version 25, IBM SPSS Statistics, Armonk, NY, USA). Continuous data were expressed as mean ± SD and compared between groups using a Mann–Whitney U test. Categorical data were summarized by their observed frequencies and percentages, and compared using cross tabulation and a chi-Square test. For all statistical tests, a value of p < 0.05 was considered significant.

3. Results

3.1. Population Characteristics and Antiarrhythmic Drug Treatment

A total of 650 patients (mean age 61 years old, 36% persistent AF) were included in the study. The three groups were comparable in baseline characteristics (Table 1).
Table 1

Patient characteristics (NICM: non-ischemic cardiomyopathy).

iRF (n = 300)Cryo (n = 200)PVAC (n = 150)p-Value
Age at time of PVI (years)58.9 ± 1160.9 ± 1158.4 ± 120.2
CHA2DS2-Vasc1.79 ± 1.51.83 ± 1.51.89 ± 1.30.5
LVEF (%)52.6 ± 1161 ± 954.5 ± 110.5
Persistent AF (%)3245.533.10.05
Diabetes mellitus (%)2717140.1
Arterial hypertension (%)6354.5660.7
Male gender (%)62.769.563.70.25
NICM (%)382.70.45
All the patients were advised to stop all antiarrhythmic drugs 3 months after the index PVI. At the time of LAT ablation, the PVAC patient presented with flecainide. In the iRF group, one patient was on amiodarone, three were on dronedarone, and four were on flecainide/propafenone. In the cryoballoon group, three patients were on amiodarone at the time of LAT ablation.

3.2. Occurrence and Characteristics of LATs

During a mean follow-up of 19 ± 8 months, 20 patients developed LAT. Three patients declined an electrophysiological study and opted for an electrical cardioversion. The diagnosis of LAT in these three patients was made based on the ECG, which was reviewed by at least two experienced electrophysiologists. In two patients, LAT terminated during or prior to 3D mapping; therefore, the LAT diagnosis was made through differential pacing. Of the 150 PVAC patients, 1 patient (0.7%; paroxysmal AF; 48 years old; LAVI 40 mL/m2; 2 PVs reconnected) developed peri-mitral macro-reentry during a mean follow-up of 21 ± 14 months. In the iRF group, 11 of the 300 patients included (3.7%; 4 with persistent AF; mean age 63 years; mean LAVI 38.4 mL/m2; at least one PV reconnected in 81%) developed LAT during a mean follow-up of 22 ± 14 months. In the cryoballoon group, eight LAT patients developed LAT over a mean follow-up of 15 ± 8 months (4%; six males, five with persistent AF; mean age 62 ± 12 years; at least one PV reconnection in 63%). The LAT cycle lengths (CL) ranged between 220 and 370 ms, with a mean of 252 ± 36 ms. Forty-five percent of the LAT patients had persistent AF at the time of the index PVI, and the mean left atrial volume (LAVI) was 40 ± 11 mL/m2 (see Table 2 and Figure 1 for the detailed characteristics and comparative occurrence rates of different LAT mechanisms). Figure 2 shows the percentages of PV reconnections in the LAT patients.
Table 2

Patient characteristics of the LAT patients and mechanisms of LAT (EF: ejection fraction; AF: atrial fibrillation; LA: left atrial; CL: cycle length; LSPV: left superior pulmonary vein; LIPV: left inferior pulmonary vein; RIPV: right inferior pulmonary vein; RSPV: right superior pulmonary vein; RPV: right pulmonary veins; EP study: electrophysiological study).

PatientPVIEFGenderAF TypeLA SizeFlutter CLPVMechanism of LAT
1Cryo65%maleparoxysmal53 mL/m2280 msRSPV reconnectedRoof dependent
2Cryo63%femalepersistent57 mL/m2270 msN/A (No EP study)N/A (No EP study)
3Cryo67%femaleparoxysmal50 mL/m2210 msRIPV reconnectedMicroreentry RSPV
4Cryo52%malepersistent51 mL/m2240 msN/A (No EP study)N/A (No EP study)
5Cryo65%malepersistent19 mL/m2220 msRSPV & RIPV reconnectedMitralisthmus dependent
6Cryo56%malepersistent30 mL/m2220 msN/A (No EP study)N/A (No EP study)
7Cryo55%femaleparoxysmal40 mL/m2370 msNo reconnectionMitralisthmus dependent
8Cryo44%malepersistent36 mL/m2240 msNo reconnectionMacroreentry involving RPVs
9PVAC59%Maleparoxysmal40 mL/m2250 msRSPV & RIPV reconnectedMitralisthmus dependent
10RF60%maleparoxysmalN/A245 msLSPV & LIPV reconnectedMitralisthmus dependent
11RF50%femaleparoxysmal25 mL/m2270 msLSPV & RIPV reconnectedN/A (spontaneous LAT Termination)
12RF65%femaleparoxysmal40 mL/m2215 msLIPV & RSPV reconnectedRoof dependent
13RF59%maleparoxysmal29 mL/m2280 msAll 4 Veins reconnectedN/A (spontaneous LAT Termination)
14RF56%malepersistent36 mL/m2250 msAll 4 Veins reconnectedMicroreentry posterior LSPV
15RF49%femalepersistent34 mL/m2240 msLSPV & LIPV reconnectedRoof dependent
16RF64%femaleparoxysmal23 mL/m2220 msNo reconnectionRoof dependent
17RF58%malepersistent44 mL/m2230 msLSPV& RSPV reconnectedMitralisthmus dependent
18RF64%maleparoxysmal43 mL/m2270 msRSPV reconnectedMultiple atrial tachycardia
19RF61%maleparoxysmal52 mL/m2240 msNo reconnectionMitralisthmus dependent
20RF60%malepersistent58 mL/m2280 msRSPV & RIPV reconnectedMicroreentry posterior RSPV
Figure 1

Mechanisms of LAT following PVI (RPV: right pulmonary veins).

Figure 2

Schematic left anterior oblique view of the LA with the percentage of PV reconnections in the LAT patients.

3.3. Possible Predictors of LAT

We performed a logistic univariate analysis to screen for predictors of LAT after PVI (see Table 3). Considering the entire study population, neither the age at the time of the index procedure (p = 0.76), CHA2DS2-Vasc (p = 0.1), male gender (p = 0.90), the presence of non-ischemic cardiomyopathy (NICM) (p = 0.39), the presence of ischemic cardiomyopathy (ICM) (p = 0.43), nor the left ventricular ejection fraction (LVEF) (p = 0.44) showed a significant correlation with the occurrence of LAT. Further, none of these characteristics reached statistical significance for predicting LAT occurrence in the different subgroups (iRF, cryoballoon and PVAC). For the cryoballoon group, the total freeze duration (p = 0.83), total number of freezes (p = 0.3), time to isolation (TTI) recorded in LSPV (p = 0.73), TTI recorded in RIPV (p= 0.09), and TTI recorded in RSPV (p = 0.63) showed no significant correlation with LAT occurrence. In the iRF and PVAC groups, the total RF durations also failed to reach statistical significance (p = 0.5 resp. p = 0.34).
Table 3

Univariate analysis of possible LAT predictors for the entire population, as well as the cryoballoon, iRF and PVAC populations (NICM: non-ischemic cardiomyopathy; ICM: ischemic cardiomyopathy; RF: radiofrequency; LSPV: left superior pulmonary vein; LIPV: left inferior pulmonary vein; RIPV: right inferior pulmonary vein; RSPV: right superior pulmonary vein; TTI: time to isolation).

Entire PopulationEntire PopulationEntire PopulationCryoballoonCryoballoonCryoballooniRFiRFiRFPVACPVACPVAC
LAT (n = 20)No LAT (n = 630)p-ValueLAT (n = 8)No LAT (n = 192)p-ValueLAT (n = 11)No LAT (n = 289)p-ValueLAT (n = 1)No LAT (n = 149)p-Value
Age (years)60.1 ± 12.859.4 ± 110.7662 ± 1261 ± 100.6359.95 ± 13.5958.84 ± 10.750.494858.54 ± 11.720.3
Male Gender (n)164080.9061330.7391820.531930.44
NICM (n)2270.391150.6180.44040.87
ICM (n)5870.431200.84410.250260.65
CHA2DS2-Vasc2.2 ± 1.51.8 ± 1.40.11.63 ± 0.911.75 ± 1.30.92.44 ± 0.431.75 ± 0.10.1031.89 ± 0.110.29
Ejection Fraction (%)58.5 ± 1155.4 ± 110.4456.5 ± 1156.11 ± 110.355.73 ± 9.6551.6 ± 10.070.15652.9 ± 1.780.92
Total RF Duration (min)----- - 56.91 ± 29.6851.69 ± 24.310.51724.2 ± 10.20.34
Total Freeze Duration (s)---1552 ± 1671399 ± 4000.83-- - -- -
Total Number of Freezes---8.88 ± 1.888.08 ± 1.720.3-- - -- -
Mean of TTI LSPV (s)---50 ± 1449 ± 220.73-- - -- -
Mean of TTI LIPV (s)---N.A.62 ± 45--- - -- -
Mean of TTI RIPV (s)---12053 ± 250.09-- - -- -
Mean of TTI RSPV (s)---60 ± 3743 ± 180.63-- - -- -

4. Discussion

A direct comparison of the three different catheter technologies used for PVI ablation, in terms of their LAT rates, has, to the best of our knowledge, not yet been published. In the entire cohort, we were able to identify an occurrence of LAT of 3%. The lowest rate of LAT was observed with phased RF technology, followed by comparable LAT rates for point-by-point RF ablation and cryoballoon technology. No independent predictors of LAT occurrence were identified. Nearly all the patients with LAT showed mild-to-severe LA dilation, a well-known marker for advanced atrial electrical and structural remodeling [11]. The majority presented with paroxysmal AF at the time of the index PVI. There was no apparent difference in the mechanisms observed, although, in the PVAC group, only one patient developed LAT during follow up, which allows no valid comparison. Macro-reentry was the predominant mechanism of LAT. In most cases, the reentry was LA roof or mitralisthmus dependent. Of note, in the patients with macro-reentry LAT, reconnection of at least one pulmonary vein was observed in most cases. In the patients with micro-reentrant LAT, the circuit was neighboring a reconnected PV in all but one. Thus, one may hypothesize that focal PV activity could be the trigger for reentry in these cases. The fibrosis due to advanced atrial remodeling, as well as supplementary ablation lesions and, especially, gaps in the lesion sets, serves as the arrhythmogenic substrate for re-entry. It is well established that the rate and mechanism of LAT after PVI varies with the lesion set chosen for AF ablation, as well as the lesion quality [12,13,14]. Achieving a contiguous lesion set is the evident goal for durable PVI, and helps reduce the incidence of LAT, especially in normal LA. It is well known that handling these single shot devices is easier and requires less-experienced operators in comparison. Particularly for less-experienced operators, the ease of use of a single shot device may, therefore, help reduce the occurrence of LAT. Our rate of LAT in the iRF group is in line with previous publications, citing rates of 2.9% after isolation of the pulmonary veins alone [15,16](see Table 4 for an overview of the current literature). For the cryoballoon, published data show rates of LAT of up to 11.3% [17]. In our population, iRF and cryoballoon revealed comparable rates of LAT, but the cryoballoon group comprised a higher percentage of patients with persistent AF. Further, in comparison, the cryoballoon patients with LAT showed lower rates of PV reconnection than the iRF group, leading to the assumption that the post cryoballoon LATs in the cryoballoon group were more likely to be substrate dependent.
Table 4

Exert of current literature on left atrial tachycardia after PVI [13,16,18,19,20].

AuthorsIndex PVIFU DurationLAT Rate
Ankerström et al.iRF (n = 415)Cryoballoon (n = 215)38 ± 21 months4.8%2.8%
Wasmer et al.iRF (n = 839)31 ± 17 months4%
Deisenhofer et al.iRF (n = 67)NA31%
Gerstenfeld et al.iRF (n = 341)NA2.9%
Mikhaylov et al.Cryoballoon (n = 181)16 ± 93 months8%
Data on the lesion morphology created with the three studied ablation and energy technologies are scarce. The cryoballoon is known to lead to wider lesions, reaching further in the LA as compared to phased ablation. Magnetic resonance imaging studies comparing cryoballoon and RF ablation lesion contiguity are contradictory [21,22]. Cryoballoon lesions in a human autopsy case showed clearer demarcation compared to RF lesions [23,24]. Animal models showed similar findings [24]. As with all systems, the quality of the lesions is dictated by the tissue contact. The balloon-based system seems to struggle with the inferior parts of the veins. Ablation with multielectrode catheters applying phased RF energy has been shown to create contiguous lesions in different animal models [25]. The circular multipolar design of the catheter seems to create easy, fast and effective PVI. However, the one year follow-up after PVI with this system revealed high rates of reconnection of multiple PVs in those with AF recurrence [6]. Similar findings have been reported in a comparison study of reconnection patterns with the cryoballoon [26]. The authors describe, for example, the reconnection of all four veins in 33% of the PVAC patients compared to 0% in the cryoballoon group, stipulating that our three groups showed comparable degrees of atrial substrate. The lowest rate of LAT after PVI with the PVAC system seems, therefore, surprising, but may be explained by the sample size. Whether RF ablation using high-power short-duration (HPSD) or novel pulse field ablation technology reduces the rates of LAT remains to be determined. A comparative study in a pig model between conventional iRF and HPSD, with 90 W/4 s for PVI, showed HPSD to be superior to conventional iRF, in terms of lesion contiguity, transmurality, as well as uniformity [27]. The first data on pulsed field ablation show promising results in terms of acute and chronic lesion continuity and transmurality. Due to the non-thermal lesion formation, catheter–tissue contact is less important, which seems of particular interest in the quest for lower LAT rates and durable PVI [28,29].

5. Limitations

We present a non-randomized study; the choice of ablation catheter was made at the operator’s discretion. Further, we performed a non-continuous follow-up and may have missed some asymptomatic LATs, although, in our experience, LATs are, in general, symptomatic. Further, the use of newer developments in point-by-point RF ablation, such as high-power, short-duration and contact force sensing catheters, may have affected the LAT occurrence rate in the iRF group.

6. Conclusions

The occurrence of LAT after ablation of AF when limited to PVI only remains low, independent of the catheter technology or the energy source. The predominant mechanism of LAT was macro-reentrant tachycardia. In most patients with LAT, at least one PV was reconnected, which may play a role in LAT development. We did not find any predictors for LAT development.
  28 in total

1.  Pulmonary vein isolation using a second-generation cryoballoon catheter: a randomized comparison of ablation duration and method of deflation.

Authors:  Jason G Andrade; Marc Dubuc; Peter G Guerra; Evelyn Landry; Nicolas Coulombe; Hugues Leduc; Léna Rivard; Laurent Macle; Bernard Thibault; Mario Talajic; Denis Roy; Paul Khairy
Journal:  J Cardiovasc Electrophysiol       Date:  2013-03-13

2.  Lower rate of left atrial tachycardia after pulmonary vein isolation with PVAC versus irrigated-tip circumferential antral ablation.

Authors:  Kristina Wasmer; David Krüsemann; Patrick Leitz; Fatih Güner; Christian Pott; Stephan Zellerhoff; Dirk Dechering; Julia Köbe; Philipp S Lange; Lars Eckardt; Gerold Mönnig
Journal:  Heart Rhythm       Date:  2016-02-24       Impact factor: 6.343

3.  High-Power and Short-Duration Ablation for Pulmonary Vein Isolation: Biophysical Characterization.

Authors:  Eran Leshem; Israel Zilberman; Cory M Tschabrunn; Michael Barkagan; Fernando M Contreras-Valdes; Assaf Govari; Elad Anter
Journal:  JACC Clin Electrophysiol       Date:  2018-02-02

Review 4.  Atrial Fibrillation and Mechanisms of Stroke: Time for a New Model.

Authors:  Hooman Kamel; Peter M Okin; Mitchell S V Elkind; Costantino Iadecola
Journal:  Stroke       Date:  2016-01-19       Impact factor: 7.914

5.  Incidence, characteristics, and outcome of left atrial tachycardias after circumferential antral ablation of atrial fibrillation.

Authors:  Kristina Wasmer; Gerold Mönnig; Alex Bittner; Dirk Dechering; Stephan Zellerhoff; Peter Milberg; Julia Köbe; Lars Eckardt
Journal:  Heart Rhythm       Date:  2012-06-06       Impact factor: 6.343

6.  Randomized study comparing duty-cycled bipolar and unipolar radiofrequency with point-by-point ablation in pulmonary vein isolation.

Authors:  Alex Bittner; Gerold Mönnig; Stephan Zellerhoff; Christian Pott; Julia Köbe; Dirk Dechering; Peter Milberg; Kristina Wasmer; Lars Eckardt
Journal:  Heart Rhythm       Date:  2011-03-30       Impact factor: 6.343

7.  Mechanism, underlying substrate and predictors of atrial tachycardia following atrial fibrillation ablation using the second-generation cryoballoon.

Authors:  Evgeny Lyan; Kivanc Yalin; Amr Abdin; Noureddin Sawan; Spyridon Liosis; Stefan A Lange; Ingo Eitel; Christian-H Heeger; Roza Meyer-Saraei; Charlotte Eitel; Roland Richard Tilz
Journal:  J Cardiol       Date:  2019-03-13       Impact factor: 3.159

8.  Comparison of regular atrial tachycardia incidence after circumferential radiofrequency versus cryoballoon pulmonary vein isolation in real-life practice.

Authors:  Finn Akerström; Hamid Bastani; Per Insulander; Jonas Schwieler; Miguel A Arias; Mats Jensen-Urstad
Journal:  J Cardiovasc Electrophysiol       Date:  2014-05-02

9.  Cryoballoon or Radiofrequency Ablation for Paroxysmal Atrial Fibrillation.

Authors:  Karl-Heinz Kuck; Josep Brugada; Alexander Fürnkranz; Andreas Metzner; Feifan Ouyang; K R Julian Chun; Arif Elvan; Thomas Arentz; Kurt Bestehorn; Stuart J Pocock; Jean-Paul Albenque; Claudio Tondo
Journal:  N Engl J Med       Date:  2016-04-04       Impact factor: 91.245

10.  First confirmation of histologic changes in the human heart after cryoballoon ablation.

Authors:  Tatsuhiko Hirao; Junichi Nitta; Akiko Adachi; Yoshihide Takahashi; Masahiko Goya; Kenzo Hirao
Journal:  HeartRhythm Case Rep       Date:  2018-11-04
View more
  1 in total

Review 1.  Procedural Feasibility and Long-Term Efficacy of Catheter Ablation of Atypical Atrial Flutters in a Wide Spectrum of Heart Diseases: An Updated Clinical Overview.

Authors:  Roberto De Ponti; Raffaella Marazzi; Manola Vilotta; Fabio Angeli; Jacopo Marazzato
Journal:  J Clin Med       Date:  2022-06-09       Impact factor: 4.964