Literature DB >> 32271819

Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI.

Guy Fierens1, Nicolas Verhaert1,2, Farida Benoudiba3, Marie-France Bellin4,5, Dennis Ducreux3,4, Jean-François Papon4,6, Jérôme Nevoux4,6.   

Abstract

Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully implantable AMEI was determined in temporal bones. Eleven temporal bones were implanted with the actuator according to the manufacturer's surgical guidelines. The actuator was coupled on the incus short process as recommended for sensorineural hearing loss. Temporal bones were exposed 10 times to the MRI magnetic field by entering the MRI suite in a clinically relevant way. Computed Tomography (CT) images were acquired before and after the experiment to investigate the risk of actuator dislocation. Based on the electrical impedance of the actuator, the loading of the actuator to the incus was confirmed. Relative actuator displacement was determined on the CT images by comparing the initial with the final actuator position in 3D space. Impedance curves were analyzed after each exposure to check the loading of the actuator to the ossicles. Analysis of CT images with a 0.30.6 mm in-plane resolution indicate no actuator displacement. The maximum detected change in impedance for all actuators was 8.43 Ω at the actuator's resonance frequency. Impedance curves measured when the actuator was retracted from the short process after the experiment still indicate the presence of a clear resonance peak. No actuator displacement or dislocation could be detected in the analysis of CT images and the measured impedance curves. Impedance curves obtained when the actuator was retracted from the incus short process still show a clear resonance peak, indicating the device is still functional after the MRI exposures.

Entities:  

Year:  2020        PMID: 32271819      PMCID: PMC7144975          DOI: 10.1371/journal.pone.0231213

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Over the last decades, a number of different active middle ear implants (AMEI) have shown a succesful outcome in patients with different types of pathologies and hearing losses. The Carina® implant (Cochlear Ltd., Sydney, Australia) is the only fully-implantable AMEI for patients with moderate to severe sensorineural, conductive or mixed hearing loss currently on the market [1]. The Carina AMEI features a sensitive implantable subcutaneous microphone to pick up sound which is amplified in the implant body and converted into mechanical vibrations by the implantable actuator to stimulate a patient. Depending on a patient’s specific needs, the actuator can be coupled to the ossicles, the oval window or the round window to compensate the hearing loss [2]. Coupling the actuator to the body of the incus has been used most in practice as it does not require mounting an additional prosthesis to the actuator tip [1, 3–5]. In addition to an increasing amount of people being implanted, Magnetic Resonance Imaging (MRI) is established as an important imaging technique in many clinical situations [6]. For patients with implantable devices however, getting an MRI scan is not straightforward due to safety concerns [6, 7]. One of the concerns are magnetically-induced forces and torques on magnetic components in the implant [8], which could lead to the displacement of the component resulting in patient harm. Most current hearing implants feature an implantable cylindrical Neodymium magnet which provides alignment between the implant and an externally worn sound processor to allow a transcutaneous, inductive coupling between both components. Magnetically-induced forces and/or torques acting on these magnets have proven to cause discomfort to patients, and in rare cases dislodge the magnet from its silicone housing, creating trauma and pain and an urgent need for revision surgery [9, 10]. Additionally, magnet dislodgement may cause adverse biological events when it goes undetected and untreated such as infections in the skin region surrounding the implant site, transdermal magnet extrusions or meningitis as a result of an infection spreading along the implant [10]. Actuators of AMEI like the Carina implant feature an electromagnetic actuator. Similar magnetically-induced forces could lead to a displacement of the actuator which could result in dislocation of the actuator from the ossicles or even damage the ossicular chain itself. Up until now, the Carina implant is labeled MRI-unsafe, indicating that a patient with this device can never undergo an MRI scan. The only MRI-conditional AMEI on the market at the moment is the Vibrant Soundbridge® 503 implant (MED-EL, Innsbruck, Austria) [11], which can be used for field strengths of 1.5 Tesla (T). In a prior version of this device, several issues have been detected related to the device’s electromagnetic actuator [12]. Studies by Todt et al. [12, 13] indicated that the implanted actuator could move due to magnetically-induced forces, leading to a number of actuator dislocations. Patients have also reported tremendous pain in the middle ear when being scanned with this device, leading to a premature ending of the procedure [12]. Additional experiments in temporal bones by Jesacher et al. [14] however indicated that the induced forces would not lead to irreversible damage to a patient’s ossicular chain. For the Carina implant, knowledge is lacking on the stability of the actuator when being exposed to a 1.5 T magnetic field. The present study will therefore investigate the risk of dislocating the actuator from its standard fixation on the incus body when exposed to a 1.5T magnetic field. The loading of the actuator will be tested before and after exposure to a 1.5 T field to assess the coupling efficiency. An actuator will be rigorously coupled to the body of the incus in order to represent the most commonly used configuration [1, 3–5]. The coupling stability is observed in the case when a patient traverses the fringe field of a 1.5 T MRI scanner in a clinically relevant manner. Magnetically-induced forces are directly proportional to magnetic field gradients [8, 15], which are considerably large when a patient crosses the scanner’s fringe field before or after an examination. After crossing the fringe field, the patient is put in the isocenter of the scanner. In this location the homogeneity of the field is maximal, and the induced torque will be maximal as well [8].

Materials and methods

Data was collected from eleven ears in fresh-frozen, anonymized human specimen obtained from L’école de Chirurgie du Fer à Moulin (Paris, France). Temporal bones (TBs) were harvested and used after the surgical procedure in accordance with the Helsinki declaration and approved by the ethics committee of Le Bicétre hospital (Paris, France) in agreement with article R2213-13 of the “Code Général des collectivités territoriales”.

Surgical procedure

Eleven ears have been implanted with a Carina T2 middle ear actuator (Cochlear Ltd., Sydney, Australia) at L’école de Chirurgie du Fer a Moulin (Paris, France). Implantations have been made for moderate to severe sensorineural hearing loss indications by coupling the actuator to the body of the incus bone. An incision was made superior to the ear canal before performing an atticotomy to expose the incus body. Using an actuator template, the fixation system is placed against the cortex in a position which would allow placing the actuator tip in contact with the incus body following the manufacturer’s instructions for use [2]. Once the correct position is obtained, the actuator template is removed and the fixation system is put in place using four surgical screws. Afterwards, the TB was harvested en bloc and stored refrigerated for approximately 14 hours at -4 °C. The day after implantation, the TBs were transported to Bicétre University Hospital (Le Kremlin-Bicétre, France) where the actuators are delicately mounted into the fixation system and loaded to the incus body (Fig 1) using the Cochlear™Carina® interface, the intraoperative test system provided by the manufacturer. The loading or contact of the actuator is measured by performing impedance measurements which are visualized in the accompanying fitting software. The impedance curve before and after loading are saved for further processing.
Fig 1

Medial (left) and superior or middle fossa (right) view of the actuator loading.

The actuator is loaded by putting it in contact with the incus body. The point of contact is detected using the Cochlear™Carina® interface.

Medial (left) and superior or middle fossa (right) view of the actuator loading.

The actuator is loaded by putting it in contact with the incus body. The point of contact is detected using the Cochlear™Carina® interface.

Experimental procedure

Prior to exposing the TBs to the static magnetic field of the MRI scanner, each TB is scanned using a Siemens Definition AS CT scanner (Siemens Healthineers, Erlangen, Germany) using the following parameters: source voltage 140 kV, current 220 mA, slice thickness 0.6 mm acquired every 0.3 mm, pitch 0.8 mm, in-plane pixel spacing 0.156 x 0.156 mm, exposure time 1 s, convolution kernel U75u, and 512 x 512 pixel matrix. Each TB is then exposed to the 1.5 T static magnetic field of a Magnetom Area Scanner (Siemens Healthineers, Erlangen, Germany) similar to when a patient would enter the MRI suite as shown in Fig 2. The TB is then rotated and put onto the patient bed, simulating when a patient would be lying down. The patient bed is then moved into the scanner at a velocity of approximately 30 cm/s, centering the TB in the scanner’s isocentre where it is kept for 1 minute. The procedure is then reversed to remove the TB from the MRI suite and static magnetic field.
Fig 2

Trajectory of entering the temporal bone (TB) in the magnetic field of the MRI suite on an upper view (left) and on a lateral view (right).

The temporal bone (TB) is exposed to the 1.5T static magnetic field of the MRI scanner similar to when a patient would enter the MRI suite. This figure shows the magnetic field distribution surrounding the scanner using isomagnetic field lines in black. The left image shows the upper view of the magnetic field and the right image shows the lateral view. The TB trajectory is illustrated in red.

Trajectory of entering the temporal bone (TB) in the magnetic field of the MRI suite on an upper view (left) and on a lateral view (right).

The temporal bone (TB) is exposed to the 1.5T static magnetic field of the MRI scanner similar to when a patient would enter the MRI suite. This figure shows the magnetic field distribution surrounding the scanner using isomagnetic field lines in black. The left image shows the upper view of the magnetic field and the right image shows the lateral view. The TB trajectory is illustrated in red. Inside the MRI control room and outside the static magnetic field, the actuator is connected to the Carina® interface in order to check the actuator loading via an electrical impedance measurement. Impedance curves were measured and saved after each exposure. When no resonance peak was observed it was retracted to check the functionality before reloading it. This procedure has been repeated ten times per TB before re-scanning the TB in the CT-scanner.

Data processing

Analysis of CT images

CT images obtained before and after exposing the TBs to the magnetic field were used to check the position of the tip of the actuator with respect to the ossicular chain. This position was analyzed using the distance between the edge of the ossicular chain (head of the malleus or body of the incus) along the longitudinal axis of the actuator and the connection point with the tip of the actuator. The position of the actuator tip and the position of the measurement lines are verified in the three orthogonal planes: axial, coronal and sagittal of the CT scan images (Fig 3). For the statistical analysis, we compared the measurement (length in mm) before and after MRI using a “Wilcoxon matched-pairs signed rank test” for each TB with the software Prism (GraphPad Software, Inc.).
Fig 3

Analysis of CT images.

In each available CT dataset, an axis was defined along the longitudinal axis of the actuator housing, shown in purple in the top left image and in blue on the bottom left figure. The intersection point between this axis and the malleus head was measured (right) and compared between images acquired before/after the experiment.

Analysis of CT images.

In each available CT dataset, an axis was defined along the longitudinal axis of the actuator housing, shown in purple in the top left image and in blue on the bottom left figure. The intersection point between this axis and the malleus head was measured (right) and compared between images acquired before/after the experiment.

Analysis of impedance curves

The loading of the actuator is checked during the experiment using the impedance measurement data acquired with the manufacturer’s software. According to the surgical instructions [2] a decrease of ≥50Ω at the resonance frequency indicates initial contact with the ossicles, after which the MicroDrive™ shall be turned an additional quarter turn to further load the actuator. This additional quarter turn will advance the actuator approximately 62.5 µm and ensure a good coupling efficiency [16, 17]. Impedance curves acquired after each exposure to the static magnetic field are compared to the impedance curve acquired after initial loading of the actuator. Impedance values are compared between both curves near the actuator’s resonance frequency. When a difference of ≥50Ω is observed, the actuator is considered to be unloaded. Impedance curves for one representative actuator after 1 and 10 exposures, both loaded and unloaded are shown in Fig 4.
Fig 4

Actuator impedance curves.

Impedance curves as measured using the Carina® interface are shown both for loaded (dotted) and unloaded (solid) actuators in a dotted and solid line respectively. Results after 1 exposure (red) and after 10 exposures (blue) to the magnetic field are shown.

Actuator impedance curves.

Impedance curves as measured using the Carina® interface are shown both for loaded (dotted) and unloaded (solid) actuators in a dotted and solid line respectively. Results after 1 exposure (red) and after 10 exposures (blue) to the magnetic field are shown.

Results

Actuator coupling as measured using CT images

The measured lengths (mm) between the actuator housing and the anterior edge of the malleus head are stated in Table 1 for each temporal bone. For three TBs the CT data was corrupted and no measurements could be made. No differences were observed in the measured length before and after exposure to the magnetic field (p = 0.265). Using the acquired images, it can be observed that the tip of the actuator is still on the same position related to the ossicles before and after the experiment. An illustration of the performed measurements is shown in Fig 5 for specimen 4093-L.
Table 1

Summary of results.

SpecimenDistance on CT before MRI (mm)Distance on CT after MRI (mm)Mean change in impedance (Ω, Min - Max)Amount of MRI-induced unloadings
3918-L9.4029.4001.88 (0.97 - 2.27)0
3918-R9.6849.6917.49 (5.51 - 8.75)0
4000-L--3.22 (0.92 - 4.38)0
4000-R10.00210.0042.24 (1.30 - 2.92)0
4093-L9.9499.9515.25 (4.54 - 6.16)0
4093-R9.9509.9517.62 (6.81 - 8.43)0
4100-L--2.75 (1.97 - 3.59)0
4100-R--6.73 (6.12 - 7.45)0
4107-L10.03010.0405.77 (5.06 - 6.48)0
4107-R10.00310.0115.92 (4.21 - 7.13)0
X-R10.00310.0136.78 (6.48 - 7.06)0

Summary of the obtained results per TB showing the distance between the edge of the ossicular chain and the entry point of the tip into the actuator body before (1st column) and after 10 exposures (2nd column) of the TB. Changes in impedance before and after scanning and MRI-induced unloadings are shown in the third and fourth column, respectively. Specimen are annotated by their anonymized reference, including a reference to the implanted ear (L: left, R: right). CT data was corrupted for 3 TBs, making it impossible to measure the distance.

Fig 5

Measurements on CT-scan before (9.949 mm, top) and after (9.951 mm, bottom) MRI on specimen 4093-L.

Summary of the obtained results per TB showing the distance between the edge of the ossicular chain and the entry point of the tip into the actuator body before (1st column) and after 10 exposures (2nd column) of the TB. Changes in impedance before and after scanning and MRI-induced unloadings are shown in the third and fourth column, respectively. Specimen are annotated by their anonymized reference, including a reference to the implanted ear (L: left, R: right). CT data was corrupted for 3 TBs, making it impossible to measure the distance.

Actuator coupling as measured using the Carina interfacea TLA

A total of eleven Carina actuators have been implanted in TBs before being exposed ten times to the static 1.5 T magnetic field of an MRI scanner. The third column of Table 1 illustrates a summary of the results obtained using the impedance measurements. Differences in impedances at the resonance frequency between the initial loaded curve and each post-exposure measurement are shown for each specimen as the mean value over all exposures with the minimum-maximum range between brackets. A visual representation of the obtained data per specimen is shown in Fig 6 below. For each specimen a boxplot is shown that visualizes the distribution of measured impedance differences after each exposure. The unloading threshold of ≥50Ω that is specified in the manufacturer guidelines is highlighted with a dotted red line. As shown, no dislocations have been observed. No statistical analyses have been performed on the data as not a single dislocation has been observed.
Fig 6

Summary of impedance differences at the actuator resonance frequency before and after exposure.

For each specimen, a box plot is shown that indicates the spread in impedance difference measured after each exposure. The impedance difference of ≥50Ω that is considered to be the unloading threshold based on manufacturer guidelines is indicated with a dotted red line. Note that the y-axis is organized logarithmically to allow a clear visualization.

Summary of impedance differences at the actuator resonance frequency before and after exposure.

For each specimen, a box plot is shown that indicates the spread in impedance difference measured after each exposure. The impedance difference of ≥50Ω that is considered to be the unloading threshold based on manufacturer guidelines is indicated with a dotted red line. Note that the y-axis is organized logarithmically to allow a clear visualization.

Discussion

Magnetically-induced forces are a significant safety hazard for patients with hearing implants, as most of these devices contain permanent magnets that can migrate during an MRI examination [9, 10]. For patients with cochlear or auditory brainstem implants, this risk is limited to a dislocation of the retention magnet implanted below the patient’s skin. For patients with middle ear implants containing electromagnetic actuators, there is an additional increased risk of migration of the implanted actuator that may possibly damage the recipient’s ossicular chain or functioning inner ear. Prior work by Todt et al. [12, 13] and Jesacher et al. [14] has indicated that this was a severe risk for patient’s implanted with the previous generation of the MED EL Vibrant Soundbridge® implant. The present study has provided more insights on how severe the risk of actuator migration could be for patient’s implanted with a Cochlear™Carina® fully-implantable AMEI. Even though this implant is MRI contraindicated [2], the obtained results provide more insights in the matter which could be very useful when a Carina® patient has to undergo an emergency MRI examination, and for future generations of this device. Experiments were performed by implanting eleven temporal bones with the electromagnetic T2 actuator coupled to the short process of the incus. This surgical position was preferred as it resembles the majority of reported surgical approaches in Carina® patients [1, 3–5]. Each temporal bone was entered and exited into the MRI suite in a clinically relevant way to simulate how a patient would enter the MRI suite during an examination. By crossing the scanner’s fringe field, the actuators are exposed to a varying magnetic field. This will induce Lorentz forces on magnetized components according to [8]: with the magnetization vector and the magnetic field gradient. This force will be maximal at the maximum , which will be at the edge of the scanner bore at the entrance of the scanner. The presented experiments were however limited to the clinically relevant case, as it is highly unlikely that a patient will cross the fringe field this close to the edge of the scanner bore. Once the actuators were put onto the patient bed, the bed was moved into the scanner bore up to the point where the actuators were in the scanner isocenter. In the scanner’s isocenter, the homogeneity of the magnetic field will be maximal, and so will be the induced torque on magnetized components, according to [8]: By combining traversing the fringe field in a clinically relevant manner and putting the actuators in the scanner isocenter in a single experiment, the actuators were exposed to realistic forces and torques that may occur during scanning. Forces and torques due to the dynamic gradient field are not taken into account due to the significantly lower magnitude of this field compared to the scanner’s static magnetic field. In order to determine if the actuator has migrated due to the induced forces and torques, two measurements were used. By comparing CT images acquired before and after the experiment, positional changes larger than the scanner resolution of 0.30.6 mm can be observed. To increase the resolution of this method, electrical impedance curves are measured which can be used to detect if the actuator remained loaded onto the short process of the incus. Using the CT data, no actuator position changes could be observed in any of the temporal bones. By comparing the impedance of the actuator before and after loading, any dislocations could be detected. The threshold for actuator dislocation was a difference of 50 Ω at the actuator resonance, which has not been observed in any of the temporal bones. It should be noted however that the presented methods are only capable of detecting permanent position changes of the actuator. The presented study therefore only focuses on the coupling stability of the T2 actuator when exposed to a 1.5 T MRI magnetic field, and it does not study any transient vibrational or positional changes that could cause patient discomfort or unintended acoustic stimulation of the patient. As sound can be picked up at very low intensities, MRI induced actuator vibrations of all acoustic hearing implant should be investigated in further study. During this study, eleven implanted temporal bones have been exposed over a total of 110 times to the static magnetic field of a 1.5T MRI scanner. Although not a single position change has been observed in this large amount of exposures, care should always be taken to avoid exposing patients with electromagnetic AMEI to high gradient magnetic fields. The presented research did not identify any adverse events for the patients, but that does not imply patient safety [18, 19]. The presented work indicates that the risk of actuator dislocation for the Cochlear Carina middle ear implant system due to magnetically induced forces and torques is low. The Cochlear™Carina® system is MRI contraindicated, and more experiments would be necessary to demonstrate the safety of the full system from a regulatory perspective. Despite its MRI-unsafe label, it could be required to expose a patient with a Carina implant to a 1.5T MRI examination in life threatening conditions. The presented research indicates that the Carina® interface can be used as a post-scan check to verify actuator coupling.

Conclusion

The presented work has provided more insights regarding the risk of dislocating the Carina middle ear actuator due to MRI-induced displacement forces and torques. Temporal bones have been implanted with Cochlear Carina middle ear actuators coupled to the incus short process. Two measurement modalities have been used to investigate any permanent actuator displacement due to these forces or torques. Dislocations of the actuator have not been observed during the analysis of the obtained dataActuator dislocation or positional changes have not been observed in any of the presented experiments, indicating that the risk on dislocating the Carina middle ear actuator when coupled to the incus short process is limited. (XLSX) Click here for additional data file. 20 Jan 2020 PONE-D-19-30029 Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI PLOS ONE Dear Mr. Fierens, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Additional Editor Comments: Dear authors, The decision for your manuscript is "minor revision". Please address the comments made by the reviewers. Also, I have an additional question: The authors concluded that "the risk on dislocating the Carina middle ear actuator when coupled to the incus short process is limited" - Could the authors please provide more information on how this conclusion was reached? Best regards, We would appreciate receiving your revised manuscript by Mar 05 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'. Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Rafael da Costa Monsanto, M.D. Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Please disclose the funding from Cochlear Ltd. in your competing interests statement. 3. In your data availability statement you write, that all relevant data are within the paper. Please ensure you have provided the individual data points used to create the figures and determine means, medians and variance measures presented in the results, tables and figures (http://journals.plos.org/plosone/s/data-availability#loc-faqs-for-data-policy). If these data cannot be publicly deposited or included in the supporting information, e.g. due to patient privacy or ownership by a third party, explain why and explain how researchers may access them. Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Partly Reviewer #2: Yes Reviewer #3: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: I read the manuscript entitled “Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI” with great interest. In this study, the authors aimed to measure the coupling stability of an actuator of a fully implantable AMEI after MRI scans. The subject of the study is clinically relevant, and the results are presented in an intelligible fashion. Although I commend the authors for tackling such an important subject, I do have a few concerns regarding the methodology that the authors could perhaps clarify. 1. For stapedotomy, the angle between the prosthesis used and the incus long process is known to be ideal at 90 degrees (https://www.ncbi.nlm.nih.gov/pubmed/12671422). I wonder that there must be perfect alignment between the actuator and the ossicular chain for optimal device performance. Why did the authors not perform these angle measurements? 2. The data from 3 TB CT scans was corrupted. Therefore, only 8 CT scans could be analyzed. I think this is a small sample to get further conclusions. 3. What does LTA mean? 4. I believe that statistical analysis for mean impedance values (before and after MRI) are necessary. Reviewer #2: Dear Author, your manuscript is well done and interesting. It is the first that evaluate the compatibility of the Carina Implant with the MRI. Your job is ready for the publication but one defect is that only the trasductor is studied. Can you evaluate all the device? You can stabilized implant and microphone with screws and then do the MRI. Please, let me know your opinion about this possibility. Reviewer #3: The authors of the present study conducted an extremely pertinent analysis of the subject in question. They clearly introduced the importance of magnetic resonance imaging in the growing world of implantable hearing aids. The study was conducted with fidelity of technical application, respecting the necessary steps to conduct the imaging exams in order to minimize the bias of selection or variation of radiological technique. The procedure of implantation of the anatomical specimens respected the indication criteria according to the latest manufacturer's guidelines. Actuator T2 dislocations were analyzed not only according to the computed tomography exam, which could lead us to a technical bias, but also according to the average of the impedance curves before and after 10 exams. MRI scan using the manufacturer's TLA. Finally, the authors concluded that and according to the results presented, little effect of the factor (magnetic resonance) on the implant analyzed, when using the classic anvil short coupling method. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: Yes: Henrique Furlan Pauna, MD Reviewer #2: Yes: Luca Bruschini Reviewer #3: Yes: IULO SERGIO BARAUNA FILHO [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. 10 Mar 2020 Dear reviewers, We thank you for your time and effort to review our submission. We have addressed your comments in detail in the attached rebuttal letter. Kind regards, Guy Fierens Submitted filename: Response to Reviewers.pdf Click here for additional data file. 19 Mar 2020 Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI PONE-D-19-30029R1 Dear Dr. Fierens, We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements. Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication. Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. With kind regards, Rafael da Costa Monsanto, M.D. Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: I believe the authors have properly addressed all topics regarding their submission. I believe it is ready for publication. Reviewer #2: This manuscript describ the stability of the Carina transducer during a MRI. It could be the first step to demonstrate that a patient with the Carina implant can perform the MRI. This manuscrip is well done and ready for the publication ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: Yes: Henrique Furlan Pauna Reviewer #2: Yes: Luca Bruschini 23 Mar 2020 PONE-D-19-30029R1 Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI Dear Dr. Fierens: I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. For any other questions or concerns, please email plosone@plos.org. Thank you for submitting your work to PLOS ONE. With kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Rafael da Costa Monsanto Academic Editor PLOS ONE
  17 in total

1.  Magnetic resonance imaging safety of the floating mass transducer.

Authors:  Ingo Todt; Grit Rademacher; Friederike Wagner; Eva Schedlbauer; Jan Wagner; Dietmar Basta; Arne Ernst
Journal:  Otol Neurotol       Date:  2010-12       Impact factor: 2.311

Review 2.  MRI information for commonly used otologic implants: review and update.

Authors:  Roya Azadarmaki; Rhonda Tubbs; Douglas A Chen; Frank G Shellock
Journal:  Otolaryngol Head Neck Surg       Date:  2014-01-07       Impact factor: 3.497

3.  MRI scanning in patients implanted with a Vibrant Soundbridge.

Authors:  Ingo Todt; Jan Wagner; Romy Goetze; Sandra Scholz; Rainer Seidl; Arne Ernst
Journal:  Laryngoscope       Date:  2011-06-06       Impact factor: 3.325

4.  Magnetic resonance imaging in cochlear implant recipients: pros and cons.

Authors:  Emanuel Kanal
Journal:  JAMA Otolaryngol Head Neck Surg       Date:  2015-01       Impact factor: 6.223

Review 5.  The physics of MRI safety.

Authors:  Lawrence P Panych; Bruno Madore
Journal:  J Magn Reson Imaging       Date:  2017-05-19       Impact factor: 4.813

Review 6.  Should patients with brain implants undergo MRI?

Authors:  Johannes B Erhardt; Erwin Fuhrer; Oliver G Gruschke; Jochen Leupold; Matthias C Wapler; Jürgen Hennig; Thomas Stieglitz; Jan G Korvink
Journal:  J Neural Eng       Date:  2018-03-07       Impact factor: 5.379

7.  A Retrospective Multicentre Cohort Review of Patient Characteristics and Surgical Aspects versus the Long-Term Outcomes for Recipients of a Fully Implantable Active Middle Ear Implant.

Authors:  Philippe P Lefebvre; Javier Gisbert; Domenico Cuda; Stéphane Tringali; Arnaud Deveze
Journal:  Audiol Neurootol       Date:  2017-01-05       Impact factor: 1.854

8.  Comparison of Carina active middle-ear implant with conventional hearing aids for mixed hearing loss.

Authors:  V A Savaş; B Gündüz; R Karamert; R Cevizci; M Düzlü; H Tutar; Y A Bayazit
Journal:  J Laryngol Otol       Date:  2016-04       Impact factor: 1.469

9.  Magnet dislocation: an increasing and serious complication following MRI in patients with cochlear implants.

Authors:  F Hassepass; V Stabenau; S Arndt; R Beck; S Bulla; T Grauvogel; A Aschendorff
Journal:  Rofo       Date:  2014-02-04

Review 10.  MRI-conditional pacemakers: current perspectives.

Authors:  António M Ferreira; Francisco Costa; António Tralhão; Hugo Marques; Nuno Cardim; Pedro Adragão
Journal:  Med Devices (Auckl)       Date:  2014-05-07
View more
  1 in total

Review 1.  Safety of active auditory implants in magnetic resonance imaging.

Authors:  Guy Fierens; Nina Standaert; Ronald Peeters; Christ Glorieux; Nicolas Verhaert
Journal:  J Otol       Date:  2021-01-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.