Literature DB >> 24497090

Magnet dislocation: an increasing and serious complication following MRI in patients with cochlear implants.

F Hassepass1, V Stabenau1, S Arndt1, R Beck1, S Bulla2, T Grauvogel1, A Aschendorff1.   

Abstract

PURPOSE: Cochlear implantation (CI) represents the gold standard in the treatment of children born deaf and postlingually deafened adults. Initial magnetic resonance imaging (MRI) was contraindicated in CI users. Meanwhile, there are specific recommendations concerning MRI compatibility depending on the type of CI system and the device manufacturer. Some CI systems are even approved for MRI with the internal magnet left in place. The aim of this study was to analyze all magnet revision surgeries in CI patients at one CI center and the relationship to MRI scans over time.
MATERIALS AND METHODS: Between 2000 and 2013, a total of 2027 CIs were implanted. The number of magnet dislocation (MD) surgeries and their causes was assessed retrospectively.
RESULTS: In total 12 cases of MD resulting from an MRI scan (0.59 %) were observed, accounting for 52.2 % of all magnetic revision surgeries. As per the labeling, it was considered safe to leave the internal magnet in place during MRI while following specific manufacturer recommendations: MRI intensity of 1.5 Tesla (T) and compression head bandage during examination.
CONCLUSION: A compression head bandage in a 1.5 T MRI unit does not safely prevent MD and the related serious complications in CI recipients. We recommend a Stenvers view radiograph after MRI with the internal magnet in place for early identification of MD, at least in the case of pain during or after MRI examination. MRI in CI patients should be indicated with restraint and patients should be explicitly informed about the possible risks. Recommendations regarding MRI compatibility and the handling of CI patients issued with MRI for the most common CI systems are summarized. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2014        PMID: 24497090     DOI: 10.1055/s-0033-1356238

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  11 in total

1.  [Prevalence and complications of MRI scans of cochlear implant patients: German version].

Authors:  G Grupe; J Wagner; S Hofmann; A Stratmann; P Mittmann; A Ernst; I Todt
Journal:  HNO       Date:  2016-03       Impact factor: 1.284

2.  Prevalence and complications of MRI scans of cochlear implant patients : English version.

Authors:  G Grupe; J Wagner; S Hofmann; A Stratmann; P Mittmann; A Ernst; I Todt
Journal:  HNO       Date:  2017-01       Impact factor: 1.284

Review 3.  Implementation of a comprehensive MR safety course for medical students.

Authors:  Steffen Sammet; Christina L Sammet
Journal:  J Magn Reson Imaging       Date:  2015-07-14       Impact factor: 4.813

Review 4.  [MRI in patients with auditory implants equipped with implanted magnets-an update : Overview and procedural management].

Authors:  S Nospes; M A Brockmann; A Läßig
Journal:  Radiologe       Date:  2019-01       Impact factor: 0.635

Review 5.  Magnetic resonance safety.

Authors:  Steffen Sammet
Journal:  Abdom Radiol (NY)       Date:  2016-03

6.  Magnetic resonance imaging after cochlear implants.

Authors:  Naif Bawazeer; Hella Vuong; Sophie Riehm; Francis Veillon; Anne Charpiot
Journal:  J Otol       Date:  2018-11-09

7.  Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI.

Authors:  Guy Fierens; Nicolas Verhaert; Farida Benoudiba; Marie-France Bellin; Dennis Ducreux; Jean-François Papon; Jérôme Nevoux
Journal:  PLoS One       Date:  2020-04-09       Impact factor: 3.240

8.  Two-phase survey on the frequency of use and safety of MRI for hearing implant recipients.

Authors:  Paul van de Heyning; Griet Mertens; Vedat Topsakal; Ruben de Brito; Wilhelm Wimmer; Marco D Caversaccio; Stefan Dazert; Stefan Volkenstein; Mario Zernotti; Lorne S Parnes; Hinrich Staecker; Iain A Bruce; Gunesh Rajan; Marcus Atlas; Peter Friedland; Piotr H Skarzynski; Serafima Sugarova; Vladislav Kuzovkov; Abdulrahman Hagr; Robert Mlynski; Joachim Schmutzhard; Shin-Ichi Usami; Luis Lassaletta; Javier Gavilán; Benoit Godey; Christopher H Raine; Rudolf Hagen; Georg M Sprinzl; Kevin Brown; Wolf-Dieter Baumgartner; Eva Karltorp
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-03-31       Impact factor: 2.503

9.  Magnet and receiver-stimulator displacement after cochlear implantation: Clinical characters and management approaches.

Authors:  Asma Alahmadi; Saad Alenzi; Mohammed Alsheikh; Saeed Alghamdi; Mostafa E Morra; Khalid M Badr
Journal:  Saudi Med J       Date:  2021-08       Impact factor: 1.422

Review 10.  Safety of active auditory implants in magnetic resonance imaging.

Authors:  Guy Fierens; Nina Standaert; Ronald Peeters; Christ Glorieux; Nicolas Verhaert
Journal:  J Otol       Date:  2021-01-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.