| Literature DB >> 29344577 |
Grazia Aleppo1, Lori M Laffel2, Andrew J Ahmann3, Irl B Hirsch4, Davida F Kruger5, Anne Peters6, Ruth S Weinstock7, Dennis R Harris8.
Abstract
After reviewing previously published methods, we developed a practical approach to adjusting insulin doses based on insulin sensitivity for adult patients with diabetes using rtCGM trend arrow data.Entities:
Keywords: continuous glucose monitoring; diabetes; fine-tuning; insulin dose adjustment; insulin dosing; trend arrows
Year: 2017 PMID: 29344577 PMCID: PMC5760210 DOI: 10.1210/js.2017-00388
Source DB: PubMed Journal: J Endocr Soc ISSN: 2472-1972
Considerations for Recommending rtCGM
| Patients Meeting One or More of the Following Criteria May Be Considered for rtCGM: |
|---|
| ► Patient is 2 years of age or older |
| ► Currently treated by intensive insulin therapy |
| ► Experiencing frequent hypoglycemia |
| ► Hypoglycemia unawareness |
| ► Excessive glucose variability |
| ► Varying and/or intensive activity |
| ► Desire to improve glycemic control |
| ► Understands behaviors that influence glycemic control |
| ► Willing and able to use rtCGM on a nearly daily basis |
| ► Willing and able to learn how to use device and receive ongoing |
| ► Pregnant or wants to get pregnant |
The following is not an exhaustive list of considerations. It is based on prior publications by the Endocrine Society and American Association of Clinical Endocrinologists, as well as the clinical experience of the authors. Additional factors such as fragile and/or elderly patients, or patients at high risk for hypoglycemia-related cardiac events may also be considered. A foundation criterion is that a patient or caretaker must be willing and able to understand, use, and learn more about rtCGM.
Currently, no rtCGM system is indicated for use in pregnancy.
Education Checklist Prior to Using rtCGM
| The Following are Fundamental Principles and Skills That a Patient and/or Caregiver Should Learn When Using rtCGM. At the End of Training, Patients and/or Caregivers Should be Able to: |
|---|
| ❑ Describe the difference between interstitial fluid and capillary glucose and understand the meaning of lag time. |
| ❑ Recognize the importance of handwashing prior to fingerstick monitoring. |
| ❑ Summarize the calibration procedure and explain when calibration is needed. |
| ❑ Summarize the limitations in rtCGM data accuracy within the first 24 hours following insertion and beyond the manufacturer’s recommended wear time. |
| ❑ Demonstrate the procedures for setting alarms/alerts. |
| ❑ Explain the significance of alarms/alerts, glucose trend data, and trend arrows in making treatment decisions. |
| ❑ Explain how to use trend arrows in individualized treatment decisions. |
| ❑ Explain the dangers associated with frequent insulin dosages following meals ( |
| ❑ Explain how to use rtCGM during sick days or illness. |
| ❑ Explain individualized monitoring and treatment strategies when exercising ( |
| ❑ Demonstrate sensor insertion procedure and list appropriate insertion sites. |
| ❑ When share functions are available: Demonstrate the procedure for uploading the rtCGM data ( |
The Dexcom G5 device should be calibrated twice daily according to the manufacturer’s instructions. Persons who check fingersticks frequently should be informed to not enter every fingerstick value. It is important that patients use the proper fingerstick monitoring technique (e.g., thoroughly washing hands with soap and water before checking). Dexcom G5 calibration is reliant on a properly functioning and accurate blood glucose meter. Therefore, we recommend patients use blood glucose meters with proven accuracy and performance.
Dexcom recommends that sensors be placed in subcutaneous tissue on the abdomen and upper buttock (including lipohypertrophic areas); however, a recent study found comparable accuracy with placement on the back of the arm [15]. Importantly, patients should be instructed not to rely solely on their rtCGM data during first 24 hours after inserting the sensor.
When reviewing alarms/alerts, it is important to discuss how to deal with “alert fatigue,” which may prompt patients to switch them off or underutilize their rtCGM system.
rtCGM can be used during periods of illness but will require additional confirmatory fingerstick check. Importantly, patients should be cautioned about use of medications that contain acetaminophen, which can cause the rtCGM system to display false high readings for up to and beyond 6 hours following ingestion [16, 17].
If the patient chooses to use data sharing, it is important that caregivers receive training in rtCGM use, specifically, use of trend arrows, interpretation, and appropriate response.
Commonly Used Over-the-Counter and Prescription Medications Containing Acetaminophen
| Common Over-the-Counter Medicines Containing Acetaminophen | |||
|---|---|---|---|
| ► Actifed® | ► Dayquil® | ► Midol® | ► Sudafed® |
| ► Alka-Seltzer Plus LiquidGels® | ► Dimetapp® | ► Nyquil® | ► Theraflu® |
| ► Anacin® | ► Dristan® | ► Panadol® | ► Triaminic® |
| ► Benadryl® | ► Excedrin® | ► Robitussin® | ► TYLENOL® Brand Products |
| ► Cepacol® | ► Feverall® | ► Saint Joseph® | ► Vanquish® |
| ► Contac® | ► Equation 44® | ► Aspirin-Free Singlet® | ► Vicks® |
| ► Coricidin® | ► Goody’s® Powders | ► Sinutab® | ► Zicam® |
| ► Liquiprin | |||
Acetaminophen is known to interfere with certain rtCGM sensors causing falsely elevated glucose readings. Patients using rtCGM are cautioned to check with the manufacturer’s information and review labels of over-the-counter medicines for acetaminophen and to ask their provider and/or pharmacist whether their prescribed medication(s) contain acetaminophen.
Includes store and other generic brands.
Figure 1.Dexcom G5 trend arrows. Dexcom G5 presents trend arrow data as icons on the Dexcom G5 Receiver and on the Dexcom G5 Mobile and Dexcom Follow apps (App) on compatible smart devices. According to the manufacturer, trend arrows indicate rates of glucose change (mg/dL per minute) and can be described as the anticipated glucose change in 30 minutes. Notably, the FLAT arrow (➡) indicates steady glucose values but does not indicate zero change. Note that trend arrows are determined by recent rtCGM measurements (generally the most recent 10 minutes of glucose values). In general, anticipated glucose may be less accurate when trying to predict changes over extended periods of time (e.g., beyond 20 to 30 minutes) due to the many factors that may influence glucose levels. Conversion: mg/dL × 0.0555 = mmol/L.
Figure 2.Previous methods to adjust insulin doses using trend arrows. Three previously published methods for adjusting insulin dose using rtCGM trend arrow data are compared [DirecNet (Abbott system) [6], Scheiner (Medtronic and Dexcom systems) [7], and Pettus/Edelman (Dexcom system)] [8]. The DirecNet method takes total insulin dosage including carbohydrate consumption (if any) into consideration. Scheiner and Pettus/Edelman methods are based on anticipated change in blood glucose with the Scheiner method being more conservative in insulin adjustment. Notably, the author of the Scheiner method has presented slightly modified values in recent presentations (personal communication) relative to past publication [7]. We use the more recently presented values in this comparison. All three require calculations beyond correction and carbohydrate consumption. All three assume the patient has insulin requiring diabetes and is using rapid-acting insulin for meals and correction. Note that the recently published Klonoff/Kerr formula recommends adjusting insulin doses by 1, 1.5, or 2 U supplements/decrements for rates of change of 1 to 2, 2 to 3, and >3 mg/dL/min, respectively [9]. Conversion: mg/dL × 0.0555 = mmol/L.
Figure 3.New approach to adjust insulin doses using trend arrows in adults with diabetes. This figure outlines our approach to adjusting insulin dose using trend arrow data from the Dexcom G5. The approach is based on anticipated glucose change and typical insulin sensitivity ranges in adults. This simplified, practical approach provides adjustments in terms of insulin units over the range of insulin sensitivities to minimize additional calculations. It is generally recommended to start adjusting conservatively to understand how the recommendations impact individuals. The authors also recommend individuals use the REPLACE-BG study approach to minimize hypo- and hyperglycemia during the 4 hours following a meal (Fig. 4) rather than these insulin dose adjustments. It is essential to understand that adjusting insulin dose using trend arrows does not replace but adds to standard calculations using ICR and CF. The approach assumes the patient has insulin requiring diabetes, is using rapid-acting insulin for meals and correction, and is using ICR and CF factors that have been accurately determined. Conversion: mg/dL × 0.0555 = mmol/L. CF, correction factor in mg/dL indicates glucose lowering per unit of rapid-acting insulin; ICR, insulin to carbohydrate ratio; U, units of rapid-acting insulin.
Figure 4.Approach to postmeal monitoring and treatment using trend arrows. These suggestions are based on the REPLACE-BG trial [19], which demonstrated that the use of nonadjunctive rtCGM for insulin dose decisions was a safe and effective alternative to conventional adjunctive CGM use. In that setting, instructions were given to participants to monitor trend arrows and minimize glucose extremes following meals. It is especially important to take a standard approach to prevent insulin “stacking” and provide correction at appropriate times following meals. Importantly, these suggestions only use the patient’s CF and do not use the adjustments for trend arrows presented in Fig. 3. It is recommended that no corrective action be taken within the first 2 hours of eating to prevent glucose extremes. Recommendations serve as a guide for postprandial monitoring and correction. Beyond 4 hours, it is assumed that most, if not all, carbohydrate has entered the system and that there is no active insulin on board. In this case, the authors recommend using the trend arrows for dose adjustment (Fig. 3). Conversion: mg/dL × 0.0555 = mmol/L. CGM, continuous glucose monitor.
Figure 5.Sensitivity comparison of methods to adjust insulin doses using trend arrows in adult patients. The figure is a visual comparison of insulin dose adjustments according to previous methods based on anticipated glucose (Scheiner and Pettus/Edelman) and our suggested approach based on insulin sensitivity ranges (Endocrine Society approach). The illustration shows that our approach aligns well with existing methods that indirectly use insulin sensitivity to adjust insulin doses while overcoming some of the limitations (e.g., a need for additional calculations and minimum increments possible for MDI-treated patients). When applied to lower CF ranges (e.g., <25), our approach is more conservative, whereas in the midrange (e.g., 50 to <75), it is more aggressive. However, one must consider that the conversions used in our approach are based on 30 minutes. When considering the anticipated glucose at 1 hour, our suggested dose adjustments become more conservative. For example, a single UP trend arrow indicates that glucose is rising 2 to 3 mg/dL/min. At 30 minutes, the anticipated glucose would be 60 to 90 mg/dL higher. However, the anticipated glucose could be as much as 120 to 180 mg/dL higher at 60 minutes if exposed to other perturbations. If an individual’s CF was 60, our approach would recommend adding 1.5 U of rapid-acting insulin to the premeal bolus. The additional 1.5 U of insulin would be expected to provide additional glucose lowering of 60 mg/dL over the 60 minutes. Given that the 60-minute anticipated glucose could potentially be much higher at 1-hour, our suggestion could be considered conservative. The expected glucose would be closer to target, postprandially, without overcorrecting and without increasing risk for hypoglycemia. As noted, these recommendations are starting points and should be readjusted as experience increases and responsiveness is observed and understood. Conversion: mg/dL × 0.0555 = mmol/L. CF, correction factor in mg/dL indicates glucose lowering per unit of rapid-acting insulin; U, units of rapid-acting insulin.
Case Examples to Put Our Approach Into Practice for Adults
| Examples Assume ICR and CF Values Have Been Accurately Determined by the Patient’s Health Care Team and That the Patient is Administering Rapid-Acting Insulin for Boluses and Corrections. Examples Assume the Patients Have Insulin-Requiring Diabetes and Are Using Dexcom G5 rtCGM. | ||||||||
|---|---|---|---|---|---|---|---|---|
| A 35-year-old, MDI-treated man sits down to eat a meal with 35 g carbohydrate. A SINGLE UP arrow is present. A calculated insulin dose is determined using meal and correction parameters. Due to the single UP arrow and no plan to exercise, an adjustment of +2.5 U is suggested. This increases the total insulin dose to 8.0 U. | ||||||||
| 180 mg/dL | ↑ | 120 mg/dL | 35 g | CF–30 | 2.0 U + 3.5 U = 5.5 U | +2.5 U | 8.0 U | |
| ICR–1:10 | ||||||||
| A 52-year-old woman on insulin pump therapy is eating 30 g of carbohydrate for lunch and sees a SINGLE DOWN arrow. The calculated insulin dose is determined. A negative adjustment of −1.5 U is suggested to account for the falling blood glucose and to prevent hypoglycemia. This decreases the total insulin dose to 2.5 U. (Note: An MDI-treated individual may round the total insulin dose to 2.0 U.) | ||||||||
| 220 mg/dL | ↓ | 120 mg/dL | 30 g | CF–50 | 2.0 U + 2.0 U = 4.0 U | −1.5 U | 2.5 U | |
| ICR–1:15 | ||||||||
| A 75-year-old man on MDI therapy is about to eat a meal with 50 g of carbohydrate. His sensor glucose is elevated and a SINGLE DOWN arrow is present. The calculated insulin dose is determined. A negative adjustment of −1.5 U is suggested to account for the falling blood glucose and prevent hypoglycemia. However, due to the frailty of the patient, the adjustment is further reduced by at least 50% ( | ||||||||
| 240 mg/dL | ↓ | 120 mg/dL | 50 g | CF–60 | 2.0 U + 2.5 U = 4.5 U | −2.5 U | 2.0 U | |
| ICR–1:20 | ||||||||
| A 28-year-old man on insulin pump therapy is noticing a mildly elevated sensor glucose and an ANGLE DOWN arrow 90 minutes after eating a meal with 45 g of carbohydrate. He is considering taking a little extra insulin. However, he follows the suggestions for postmeal monitoring and hypoglycemia prevention and determines that neither a correction nor an adjustment is needed at this time. He will recheck his sensor glucose in 30 minutes. | ||||||||
| 170 mg/dL | ↘ | 120 mg/dL | 0 g | CF–50 | NA | NA | NA | |
| ICR–1:10 | ||||||||
| A 23-year-old woman on MDI therapy is experiencing hyperglycemia 2.5 hours after eating a meal with 45 g of carbohydrate. At the time of eating, she took the appropriate amount of insulin and did not need a correction. Now she is seeing DOUBLE UP arrows. Because she is within 4 hours of eating, she will follow the suggestions for postmeal monitoring and hyperglycemia prevention and will not use the adjustment table. Also, she is an MDI-treated patient who is not using a bolus calculator that accounts for insulin-on-board. Therefore, she reduces the calculated correction insulin dose of 2.0 U by 50% to prevent overcorrection. This reduces the total insulin dose to 1.0 U. She will recheck in 1 hour. | ||||||||
| 240 mg/dL | ↑↑ | 120 mg/dL | 0 g | CF–60 | 2.0 U × 50% = 1.0 U | NA | 1.0 U | |
| ICR–1:10 | ||||||||
| A 49-year-old man on insulin pump therapy is working late into a stressful afternoon. Lunch was nearly 5 hours ago. He notices his sensor glucose is elevated with an ANGLE UP arrow. He calculates his insulin dose for correction using predetermined values and then uses the adjustment table to account for the rising blood glucose. He uses the adjustment table because he is beyond 4 hours from his last meal and there shouldn’t be any active insulin-on-board from his mealtime bolus. He adds the adjustment of 1.5 U, which increases his total insulin dose to 3.8 U. | ||||||||
| 180 mg/dL | ↗ | 100 mg/dL | 0 g | CF–35 | 2.3 U | +1.5 U | 3.8 U | |
| ICR–1:8 | ||||||||
| A 35-year-old woman has been struggling with a headache all morning. She takes 1 g of acetaminophen. After 1 hour, she notices a high sensor glucose value and DOUBLE UP arrows. However, she recalls that the Dexcom G5 can yield falsely high readings when acetaminophen is used. She checks her blood glucose by fingerstick and does not use the trend arrow data or adjustment table. The fingerstick reading shows she is actually at 115 mg/dL. No correction is needed. She will continue to use fingerstick monitoring until 6 hours has passed since ingesting the acetaminophen. | ||||||||
| 243 mg/dL | ↑↑ | 120 mg/dL | 0 g | CF–70 | 0.0 U | NA | 0.0 U | |
| ICR–1:20 | ||||||||
| A 38-year-old woman on MDI therapy has had to walk much more than usual during her midday work. She managed to stay near her target glucose; however, as she prepares for her lunch, she sees DOUBLE DOWN arrows. She calculates the insulin dose a slight correction and her meal with 45 g carbohydrate. Then she uses the adjustment table, which suggests a negative adjustment of −3.5 U. After rounding, this is a net total insulin dose of 0.0 U. She will eat without taking an insulin dose and carefully monitor the rest of the afternoon. (Note: If she was consuming less carbohydrate or was below target glucose, a negative total insulin dose may result, in which case, 15 g of fast-acting carbohydrate and close monitoring until trend arrows stabilize may be more appropriate than consuming her typical lunch.) | ||||||||
| 160 mg/dL | ↓↓ | 130 mg/dL | 45 g | CF–45 | 0.7 U + 3.0 U = 3.7 U | −3.5 U | 0.0 U | |
| ICR–1:15 | ||||||||
Calculated Insulin Dose includes insulin needed to cover carbohydrate intake and correction to reach target glucose. The calculations use the predetermined ICR and CF values and assume these values have been accurately determined by the patient’s health care team and that the patient is using rapid-acting insulin for carbohydrate intake and correction.
Abbreviations: CF, correction factor in mg/dL indicates glucose lowering per unit of rapid-acting insulin; U, units of rapid-acting insulin. Conversion: mg/dL × 0.0555 = mmol/L.