Abstract
Strip meniscometry tube (SMTube) is useful for assessing tear volume in dry eye disease. We performed an in vitro and a clinical study to examine whether the SMTube can also be used to evaluate tear quantity in lacrimal passage obstruction (LPO). In vitro experiments showed that the SMTube value and the amount of water absorption were significantly correlated (R2 = 0.816, p = 0.0008), but the measured value peaked when 4 μL was absorbed. In the clinical study, 12 eyes of 9 LPO patients and 17 eyes of 17 healthy control subjects were included. The patients' SMTube values were significantly higher than those of the control subjects (p < 0.05). The SMTube value significantly decreased after silicon tube intubation (p < 0.05). Tear meniscus height (TMH) and area (TMA), measured by anterior segment optical coherence tomography, exhibited a significant positive correlation with the SMTube value (TMH, p < 0.001, R2 = 0.396; TMA, p < 0.001, R2 = 0.366). Moreover, the SMTube value significantly correlated positively with the subjective symptom of epiphora (p = 0.012, R2 = 0.255). Thus, SMTube was useful for evaluating the tear volume and therapeutic effects in patients with LPO.Entities:
Keywords: anterior segment optical coherence tomography; epiphora; lacrimal drainage system dysfunction; silicon tube intubation; strip meniscometry tube; tear meniscus
Year: 2020 PMID: 32218304 PMCID: PMC7235822 DOI: 10.3390/diagnostics10040179
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Figure 1Results of in vitro experiments. (a) The strip meniscometry tube (SMTube) value correlated significantly with the saline volume. (b) Similarly, the absorption time correlated significantly with the saline volume.
Figure 2Comparison of tear meniscus parameters measured by optical coherence tomography and SMTube value between the lacrimal passage obstruction patients and control subjects. Tear meniscus height (a) and area (b) in patients with lacrimal passage obstruction was significantly greater than that in the control subjects. SMTube value in the patients with lacrimal passage obstruction was also significantly greater than that in the control subjects (c). * p < 0.05. ** p < 0.001.
Figure 3Time-course changes of tear meniscus height, tear meniscus area, and strip meniscometry value before surgery, during the silicone tube intubation, after removing the silicone tube. (a) Tear meniscus height decreased significantly after removing the silicone tube, compared to before the surgery. (b) The tear meniscus area decreased significantly after removing the silicone tube, compared to before the surgery. (c) The SMTube value decreased significantly after removing the silicone tube compared to before the surgery and during the silicone tube intubation. * p < 0.05.