| Literature DB >> 32133697 |
Xiaohui Chen1, Ruifeng Yang2, Yongming Liang1, Teng Yuan1, Jiansuo Zhou1, Tiancheng Wang1, Liyan Cui1.
Abstract
BACKGROUND: Laboratory tests play an important role in the diagnosis of syphilis. This study aimed to compare and assess the performance of the Abbott chemiluminescent microparticle immunoassay (CMIA) and the ChIVD light-initiated chemiluminescent assay (LICA) in the detection of Treponema pallidum (TP) antibody.Entities:
Keywords: zzm321990Treponema pallidumzzm321990; Abbott chemiluminescent microparticle immunoassay; ChIVD light-initiated chemiluminescent assay; Treponema pallidum particle agglutination; recombinant immunoblot assay; syphilis
Mesh:
Substances:
Year: 2020 PMID: 32133697 PMCID: PMC7370704 DOI: 10.1002/jcla.23275
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Figure 1Testing algorithms of Treponema pallidum antibody detection. Abbreviations: CMIA, chemiluminescent microparticle immunoassay; LICA, light‐initiated chemiluminescent assay; RIBA, recombinant immunoblot assay; TPPA, treponema pallidum particle agglutination
Evaluation of Abbott CMIA and ChIVD LICA with the diagnose
| Diagnose | Sensitivity | Specificity | PPV | NPV | ||
|---|---|---|---|---|---|---|
| (n) | ||||||
| Positive | Negative | |||||
| Abbott CMIA | ||||||
| Positive | 153 | 11 | 94.44% | 99.89% | 93.29% | 99.91% |
| Negative | 9 | 10 290 | ||||
| ChIVD LICA | ||||||
| Positive | 159 | 20 | 98.15% | 99.81% | 88.83% | 99.97% |
| Negative | 3 | 10 281 | ||||
Abbreviations: CMIA, chemiluminescent microparticle immunoassay; LICA, light‐initiated chemiluminescent assay; NPV, negative predictive value; PPV, positive predictive value.
The comparison of Abbott CMIA and ChIVD LICA
| Abbott CMIA | ||||
|---|---|---|---|---|
| Positive (n) | Negative (n) | Total (n) | ||
| ChIVD LICA | Positive (n) | 150 | 30 | 180 |
| Negative (n) | 48 | 10 270 | 10 318 | |
| Total (n) | 198 | 10 300 | 10 498 | |
Abbreviations: CMIA, chemiluminescent microparticle immunoassay; LICA, light‐initiated chemiluminescent assay.
The characteristics of Abbott CMIA false positive
| Characteristic | False positive | Male | Female |
|---|---|---|---|
| (n) | (n) | (n) | |
| Male infertility | 2 | 0 | 2 |
| Hypertensive disease | 2 | 2 | 0 |
| Female infertility | 1 | 1 | 0 |
| Spinal cord injury | 1 | 1 | 0 |
| Retinal detachment | 1 | 1 | 0 |
| Lung cancer | 1 | 1 | 0 |
| HBV | 1 | 0 | 1 |
| Cirrhosis | 1 | 1 | 0 |
| Chest pain | 1 | 1 | 0 |
| Total | 11 | 8 | 3 |
Abbreviations: CMIA, chemiluminescent microparticle immunoassay; HBV, hepatitis B virus.
The characteristics of ChIVD LICA false positive
| Characteristic | False positive | Male | Female |
|---|---|---|---|
| (n) | (n) | (n) | |
| Male infertility | 5 | 5 | 0 |
| Female infertility | 4 | 0 | 4 |
| Lumbar spinal stenosis | 2 | 0 | 2 |
| HBV | 1 | 1 | 0 |
| HCV | 1 | 0 | 1 |
| Prostate cancer | 1 | 1 | 0 |
| Skin tumor | 1 | 0 | 1 |
| Non‐Hodgkin lymphoma | 1 | 1 | 0 |
| Coronary heart disease | 1 | 0 | 1 |
| Hypertensive disease | 1 | 1 | 0 |
| pregnancy | 1 | 0 | 1 |
| Cervical spondylosis | 1 | 1 | 0 |
| Total | 20 | 10 | 10 |
Abbreviations: HBV, hepatitis B virus; HCV, hepatitis C virus; LICA, light‐initiated chemiluminescent assay.
The characteristics of Abbott CMIA false negative
| Characteristic | False negative | Male | Female |
|---|---|---|---|
| (n) | (n) | (n) | |
| Hypertensive disease | 2 | 1 | 1 |
| Male infertility | 1 | 1 | 0 |
| Female infertility | 1 | 0 | 1 |
| Ligament rupture | 1 | 1 | 0 |
| Hernia | 1 | 1 | 0 |
| Lumbar spinal stenosis | 1 | 1 | 0 |
| Cataract | 1 | 1 | 0 |
| Diabetes | 1 | 1 | 0 |
| Total | 9 | 7 | 2 |
Abbreviation: CMIA, chemiluminescent microparticle immunoassay.
The characteristics of ChIVD LICA false negative
| Characteristic | False negative | Male | Female |
|---|---|---|---|
| (n) | (n) | (n) | |
| Female infertility | 2 | 0 | 2 |
| Gastric cancer | 1 | 0 | 1 |
| Total | 3 | 0 | 3 |
Abbreviation: LICA, light‐initiated chemiluminescent assay.