| Literature DB >> 32102497 |
Shinji Yoshii1, Marina Kubo1, Mio Matsumoto2, Takefumi Kikuchi3, Yasunari Takakuwa4.
Abstract
BACKGROUND/AIMS: A multifunctional snare SOUTEN has a sharp tip at the top of the snare loop that enables incision of the mucosa, dissection of the submucosal layer, and snaring of lesion. This study assessed the efficacy and safety of complete endoscopic resection of colorectal neoplasia using SOUTEN.Entities:
Keywords: Hybrid endoscopic submucosal dissection; Multifunctional snare; Precutting endoscopic mucosal resection; SOUTEN
Year: 2020 PMID: 32102497 PMCID: PMC7137565 DOI: 10.5946/ce.2019.117
Source DB: PubMed Journal: Clin Endosc ISSN: 2234-2400
Fig. 1.Image of SOUTEN (Kaneka Medix, Tokyo, Japan) snare. (A) The multifunctional snare was designed to achieve precutting endoscopic mucosal resection and hybrid endoscopic submucosal dissection. (B) A 1.5-mm needle-knife with a knob-shaped tip was attached to the top of the loop. (C) The length of the loop was 40 mm.
Fig. 2.(A) A 30-mm granular type laterally spreading tumor was located in the lower rectum. (B) Local injection of sodium hyaluronate solution. (C) Mucosal incision was performed using a needle-knife of the top of the snare. (D) Adequate trimming was performed. (E) The snare was placed on the dissection plane. (F) The tumor was tightly snared. (G) Post-precutting endoscopic mucosal resection. The lesion was completely resected. (H) The resected specimen. En bloc resection was achieved.
Characteristics of the Patients and Lesions
| Precutting EMR | Hybrid ESD | Conventional ESD | |
|---|---|---|---|
| Number of lesions | 50 | 27 | 31 |
| Sex (M/F) | 15/35 | 9/18 | 14/17 |
| Age (yr) (mean±SD) | 61.3±12.1 | 65.4±11.5 | 63.7±14.0 |
| Mean lesion size (mm) (mean±SD) | 14.5±5.4 | 16.4±6.3 | 21.1±7.7[ |
| Range of the lesion size (mm) | 8–28 | 5–28 | 8–39 |
| Number of lesions | |||
| Less than 20 mm (%) | 39 (36.1) | 19 (17.6) | 10 (9.3) |
| More than 20 mm (%) | 11 (10.2) | 8 (7.4) | 21 (19.4) |
EMR, endoscopic mucosal resection; ESD, endoscopic submucosal dissection; SD, standard deviation.
P<0.01 comparing with precutting EMR.
Summary of Specimen Size, Location, Morphology, and Histopathology
| Precutting EMR | Hybrid ESD | Conventional ESD | |
|---|---|---|---|
| Number of lesions | 50 | 27 | 31 |
| Mean specimen size (mm±SD) | 18.8±5.4 | 21.4±6.3 | 25.7±7.7[ |
| Range of the specimen size (mm) | 10–35 | 10–32 | 12–44 |
| Location of lesions | |||
| Right/left/rectum | 36-12-2 | 18-7-2 | 23-3-5 |
| Colonic folds | |||
| Between folds (%) | 37 (74.0) | 17 (62.9) | 22 (71.0) |
| Across a fold (%) | 13 (26.0) | 10 (37.0) | 9 (29.0) |
| Colonic flexure | |||
| In non-flexure (%) | 44 (88.0) | 23 (85.2) | 25 (80.6) |
| In flexure (%) | 6 (12.0) | 4 (14.8) | 6 (19.4) |
| Visibility | |||
| Vertically (%) | 32 (64.0) | 16 (59.3) | 13 (41.9) |
| Tangentially (%) | 18 (36.0) | 11 (40.7) | 18 (58.1) |
| Morphology | |||
| 0-I/0-II | 9/41 | 4/23 | 12/19 |
| Histopathology | |||
| TA or TSA | 29 | 13 | 17 |
| SSA/P | 15 | 7 | 8 |
| Tis | 1 | 4 | 3 |
| T1 | 4 | 2 | 2 |
| Hyperplastic polyp | 1 | 1 | 1 |
The specimen size is the size of the whole resected specimen, whereas the lesion size is the size of the tumor in the resected specimen.
EMR, endoscopic mucosal resection; ESD, endoscopic submucosal dissection; SD, standard deviation; SSA/P, sessile serrated adenoma/polyp; TA, tubular adenoma; TSA, traditional serrated adenoma.
P<0.01 comparing with precutting EMR, p<0.05 comparing with hybrid ESD.
Summary of the Endoscopic Procedure Time and Outcome
| Precutting EMR | Hybrid ESD | Conventional ESD | |
|---|---|---|---|
| Number of samples | 50 | 27 | 31 |
| Procedure time (sec±SD) | 221.7±139.6 | 461.0±466.2 | 1139.1±755.2[ |
| Snaring time (sec) | 52.1 | 58.4 | Not applicable |
| Gross | 100 | 96.4 | 100 |
| Histological complete resection (%) | 94 | 96.4 | 100 |
| Adverse events | |||
| Bleeding | 0 | 0 | 0 |
| Perforation | 0 | 0 | 0 |
EMR, endoscopic mucosal resection; ESD, endoscopic submucosal dissection; SD, standard deviation.