| Literature DB >> 32089598 |
Anjum Sultana1, Sumit Makkar1, Deepesh Saxena1, Amit Wadhawan2, Chandan Kumar Kusum1.
Abstract
AIM: The present study was done to evaluate and compare the stability of the implant and the loss of crestal bone in the implants placed using OD drilling and traditional drilling technique. SETTING ANDEntities:
Keywords: Crestal bone levels; implant stability; osseodensification; resonance frequency analysis
Year: 2020 PMID: 32089598 PMCID: PMC7008626 DOI: 10.4103/jips.jips_133_19
Source DB: PubMed Journal: J Indian Prosthodont Soc ISSN: 0972-4052
Decision tree for osseodensification protocol
| Implant diameter | Drill | Bur 1 | Bur 2 | Bur 3 | Bur 4 |
|---|---|---|---|---|---|
| 3.5, 3.7, 3.8 | Pilot drill | VT1525 | VT2535 | ||
| 4.0, 4.2, 4.3 | Pilot drill | VT1828 | VT2838 | ||
| 4.5, 4.7, 4.8 | Pilot drill | VT1525 | VT2535 | VT3545 | |
| 5.0, 5.2, 5.3 | Pilot drill | VT1828 | VT2838 | VT3848 |
Figure 1(a) DENSAH KIT. (b) Preoperative frontal view
Figure 2Osteotomy preparation using osseodensifying drilling technique
Figure 3Three-dimensional placement of implant
Figure 4Resonance frequency analysis measurement using Ostell
Figure 5Abutment in place
Figure 6Final prosthesis in place
Graph 1Comparative evaluation of the stability of implants placed using osseodensification and traditional drilling technique
Table with values measured
| Serial number | Osseodensification method | Traditional method | ||
|---|---|---|---|---|
| Immediate postoperative (RFA1) | Before loading (RFA2) | Immediate postoperative (RFA1) | Before loading (RFA2) | |
| 1 | 58 | 57 | 58 | 57 |
| 2 | 72 | 62 | 35 | 61 |
| 3 | 35 | 61 | 75 | 68 |
| 4 | 75 | 68 | 39 | 62 |
| 5 | 69 | 70 | 76 | 75 |
| 6 | 72 | 62 | 35 | 55 |
| 7 | 69 | 68 | 62 | 65 |
| 8 | 73 | 68 | 58 | 65 |
| 9 | 75 | 65 | 76 | 75 |
| 10 | 59 | 75 | 76 | 75 |
| Mean | 65.7 | 65.6 | 59 | 65.8 |
| SD | 12.36 | 5.23 | 17.28 | 7.39 |
| Maximum | 75 | 75 | 76 | 75 |
| Minimum | 35 | 57 | 35 | 55 |
RFA1: Resonance frequency analysis at baseline, RFA2: Resonance frequency analysis at 6 months, SD: Standard deviation
Crestal bone levels from first implant thread after 8 months
| Serial number | Post loading (Osseodensification method) | Post loading (Traditional drilling method) | ||||||
|---|---|---|---|---|---|---|---|---|
| Labial buccal | Palatal | Mesial | Distal | Labial buccal | Palatal | Mesial | Distal | |
| 1 | 0.34 | 0 | −0.21 | 0 | 0.34 | 0 | −0.21 | 0 |
| 2 | −0.81 | 0 | 0.58 | 0 | 0 | 0 | −1.14 | −1.24 |
| 3 | 0 | 0 | −1.14 | −1.24 | 0.38 | 0 | −0.32 | 0 |
| 4 | 0.38 | 0 | −0.32 | 0 | 0 | 0 | −1.18 | −1.23 |
| 5 | 2.31 | 1.02 | −1.02 | 0.88 | 4.57 | 1.51 | 2.08 | 0 |
| 6 | 1.76 | −0.98 | −0.57 | −2.58 | 2.53 | 2.21 | −0.52 | −0.23 |
| 7 | −1.27 | −1.98 | −1.34 | 1.09 | 0 | 0 | 0 | 2.01 |
| 8 | 1.34 | 1.21 | 0.21 | −1.01 | 0 | 0 | 1.02 | 1.62 |
| 9 | 0.53 | 1.01 | −0.21 | 1.21 | 1.35 | 2.01 | −1.01 | 1.02 |
| 10 | 1.34 | 2.01 | 0.3 | 1.01 | 4.57 | 1.51 | 2.08 | 0 |
| Mean | 0.592 | 0.229 | −0.372 | −0.064 | 1.374 | 0.724 | 0.080 | 0.195 |
| SD | 1.122 | 1.151 | 0.644 | 1.230 | 1.867 | 0.957 | 1.236 | 1.076 |
| Maximum | 2.31 | 2.01 | 0.58 | 1.21 | 4.57 | 2.21 | 2.08 | 2.01 |
| Minimum | −1.27 | −1.98 | −1.34 | −2.58 | 0 | 0 | −1.18 | −1.24 |
Positive values denote above and negative values denote bone levels below the first thread. SD: Standard deviation
Graph 2Evaluation of the crestal bone levels on the labial/buccal site using osseodensification and traditional drilling technique
Comparison of implant stability at baseline and 6 months for osseodensification and traditional drilling technique
| Stability | Probable values of unpaired | Significance |
|---|---|---|
| RFA1 | 0.1820**, | NS |
| RFA2 | 0.6428**, | NS |
**The values came out to be not significant i.e., >0.05. NS: Not significant, RFA: Resonance frequency analysis
Comparison of crestal bone levels for labial/buccal, palatal, mesial, and distal sites of implant at baseline, after 6 months, and 8 months between osseodensification and traditional drilling technique
| Sites | Probable values of unpaired | ||
|---|---|---|---|
| At baseline | After 6 months | After 8 months | |
| Labial buccal | 0.7933**, | 0.3998**, | 0.8738**, |
| Palatal | 0.7842**, | 0.0673**, | 0.7935**, |
| Mesial | 0.0503**, | 0.9249**, | 0.5698**, |
| Distal | 0.0158*, | 0.5219**, | 0.4071**, |
**No significant difference between groups, i.e., >0.05, *A significant difference between groups, i.e., <0.05. NS: Not significant, S: Significant
Comparison (by one-way ANOVA-F-test) among the different sites for different time points (at baseline, 6 months and 8 months) for osseodensification and traditional drilling technique
| Time points | Source of variation | df | ||
|---|---|---|---|---|
| At base line | Between groups (traditional drill method) | 0.079**, | 2.448 | 3 |
| Between groups (osseodensifying drill technique) | 0.172**, | 1.764 | 3 | |
| After 6 months | Between groups (traditional drilling technique) | 0.407**, | 0.992 | 3 |
| Between groups (osseodensifying drilling technique) | 0.017*, | 3.872 | 3 | |
| After 8 months | Between groups (traditional drilling technique) | 0.017*, | 3.869 | 3 |
| Between groups (osseodensifying drilling technique) | 0.183**, | 1.707 | 3 |
**No significant difference between groups, i.e., >0.05, *A significant difference between groups, i.e., <0.05. NS: Not significant, S: Significant, P: Probability, F-cal: F calculated statistics are based on the ratio of mean squares, df: Degree of freedom
Crestal bone levels from first implant thread after immediate post op
| Serial number | At baseline (Osseodensification method) | At baseline (Traditional drilling method) | ||||||
|---|---|---|---|---|---|---|---|---|
| Labial buccal | Palatal | Mesial | Distal | Labial buccal | Palatal | Mesial | Distal | |
| 1 | 2.01 | 1.81 | 0.38 | 0.72 | 2.01 | 1.81 | 0.38 | 0.72 |
| 2 | 0 | 0.35 | 0.23 | 1.28 | −4.53 | 0.64 | 0 | −0.97 |
| 3 | −4.53 | 0.69 | 0 | −0.97 | 0.37 | 1.02 | 0 | −0.72 |
| 4 | 0.37 | 1.2 | 0 | −0.72 | −1.63 | 1.02 | −0.38 | −3.09 |
| 5 | 2.62 | 2.43 | −2.34 | −0.88 | 1.08 | 2.01 | 0.35 | 0.61 |
| 6 | 0 | 0.35 | 0.28 | 1.28 | −1.28 | −0.98 | −1.28 | −4.28 |
| 7 | −0.57 | 0 | 0.37 | 0.31 | −3.25 | 1 | −1.82 | 0.81 |
| 8 | −0.81 | 0 | 0.58 | 0 | 1.09 | 0.37 | 2.81 | −1.37 |
| 9 | 2.62 | 2.43 | −2.34 | −0.88 | 1.21 | 1.28 | 2.1 | 2.11 |
| 10 | −2.72 | 1.77 | −1.04 | 1.39 | 1.08 | 2.01 | 0.35 | 0.61 |
| Mean | −0.101 | 1.103 | −0.388 | 0.153 | −0.385 | 1.018 | 0.251 | −0.557 |
| SD | 2.278 | 0.955 | 1.119 | 0.977 | 2.187 | 0.896 | 1.381 | 1.957 |
| Maximum | 2.62 | 2.43 | 0.58 | 1.39 | 2.01 | 2.01 | 2.81 | 2.11 |
| Minimum | −4.53 | 0 | −2.34 | −0.97 | −4.53 | −0.98 | −1.82 | −4.28 |
Positive values denote above and negative values denote bone levels below first thread. SD: Standard deviation
Crestal bone levels from first implant thread after 6 months
| Serial number | After 6 months (Osseodensification method) | After 6 months (Traditional method) | ||||||
|---|---|---|---|---|---|---|---|---|
| Labial buccal | Palatal | Mesial | Distal | Labial buccal | Palatal | Mesial | Distal | |
| 1 | 1.02 | 0.21 | −0.41 | 1.21 | 1.02 | 0.21 | −0.41 | 1.21 |
| 2 | −0.57 | 0 | 0.37 | 0.31 | 3.34 | 0 | 0 | −0.97 |
| 3 | −3.34 | 0 | 0 | −0.97 | −2.16 | 0 | −2.47 | −1.71 |
| 4 | −2.16 | 0 | −2.47 | −1.71 | 0.58 | 1.65 | −1.1 | 1.13 |
| 5 | 1.01 | 1.28 | −1.02 | 1.11 | 0.52 | 1.01 | 0.37 | 0.52 |
| 6 | 1.01 | 1.28 | −1.02 | 1.11 | 3.53 | 2.28 | −4.8 | −1.14 |
| 7 | 2.31 | 1.02 | −1.02 | 0.88 | 0.33 | 2.01 | 0 | −0.73 |
| 8 | 1.11 | 0 | −3.54 | 1.74 | −1.83 | 3.58 | 0 | −1.69 |
| 9 | 1.96 | −0.98 | −0.52 | 1.61 | 1.5 | 1.89 | 0.91 | 2.43 |
| 10 | 2.9 | 2.06 | −2.21 | 2.05 | 0.52 | 1.01 | 0.37 | 0.52 |
| Mean | 0.525 | 0.487 | −1.184 | 0.734 | 0.735 | 1.364 | −0.713 | −0.043 |
| SD | 1.980 | 0.894 | 1.211 | 1.206 | 1.841 | 1.146 | 1.717 | 1.403 |
| Maximum | 2.9 | 2.06 | 0.37 | 2.05 | 3.53 | 3.58 | 0.91 | 2.43 |
| Minimum | −3.34 | −0.98 | −3.54 | −1.71 | −2.16 | 0 | −4.8 | −1.71 |
Positive values denote above and negative values denote bone levels below first thread. SD: Standard deviation