Rik Hj Hendrix1, Alexander Ks Yeung1, Yuri M Ganushchak1, Patrick W Weerwind1.
Abstract
INTRODUCTION: This study analyzed the effect of different flows and pressures on the intraoxygenator flow path in three contemporary oxygenators and its consequences for oxygen transfer efficiency.Entities:
Keywords: flow path; oxygenator; oxygenator blood volume; oxygenator efficiency
Mesh:
Year: 2020 PMID: 32056504 PMCID: PMC7536508 DOI: 10.1177/0267659119899883
Source DB: PubMed Journal: Perfusion ISSN: 0267-6591 Impact factor: 1.972
Flow path parameters per oxygenator at three different post-oxygenator pressures during pulsatile flow.
| Parameter | Oxygenator | Pressure (150 mm Hg) | Pressure (200 mm Hg) | Pressure (250 mm Hg) | p value |
|---|---|---|---|---|---|
| OXBV (mL) | Capiox | 249 ± 7 | 248 ± 6 | 245 ± 8 | 0.397 |
| Inspire | 280 ± 9 | 279 ± 10 | 279 ± 9 | 0.987 | |
| Quadrox | 288 ± 9 | 290 ± 10 | 292 ± 8 | 0.516 | |
| Ratio BA | Capiox | 1.38 ± 0.16 | 1.46 ± 0.16 | 1.48 ± 0.10 | 0.247 |
| Inpsire | 1.86 ± 0.27 | 1.78 ± 0.24 | 1.84 ± 0.24 | 0.736 | |
| Quadrox | 2.45 ± 0.59 | 2.56 ± 0.58 | 2.66 ± 0.70 | 0.715 | |
| Chord 30 | Capiox | 1.23 ± 0.18 | 1.31 ± 0.17 | 1.31 ± 0.11 | 0.461 |
| Inspire | 1.60 ± 0.26 | 1.55 ± 0.20 | 1.62 ± 0.22 | 0.739 | |
| Quadrox | 2.41 ± 0.70 | 2.50 ± 0.66 | 2.56 ± 0.77 | 0.882 | |
| Chord 50 | Capiox | 1.16 ± 0.14 | 1.23 ± 0.14 | 1.22 ± 0.09 | 0.328 |
| Inspire | 1.45 ± 0.26 | 1.44 ± 0.22 | 1.44 ± 0.23 | 0.987 | |
| Quadrox | 2.22 ± 0.61 | 2.31 ± 0.63 | 2.38 ± 0.71 | 0.832 |
OXBV: oxygenator blood volume.
Ratio BA: ratio of the area under the dilution curve before and after maximum dilution in total; Chord 30: ratio of the area under the dilution curve before and after maximum dilution starting from 30% of maximum curve height; Chord 50: ratio of the area under the dilution curve before and after maximum dilution starting from 50% of maximum curve height.
Flow path parameters per oxygenator at three different post-oxygenator pressures during continuous flow.
| Parameter | Oxygenator | Pressure (150 mm Hg) | Pressure (200 mm Hg) | Pressure (250 mm Hg) | p value |
|---|---|---|---|---|---|
| OXBV (mL) | Capiox | 237 ± 7 | 239 ± 4 | 239 ± 5 | 0.513 |
| Inspire | 275 ± 6 | 275 ± 5 | 276 ± 5 | 0.776 | |
| Quadrox | 284 ± 9 | 284 ± 8 | 284 ± 7 | 0.968 | |
| Ratio BA | Capiox | 1.49 ± 0.14 | 1.51 ± 0.13 | 1.46 ± 0.14 | 0.687 |
| Inpsire | 1.68 ± 0.10 | 1.65 ± 0.11 | 1.70 ± 0.13 | 0.552 | |
| Quadrox | 2.45 ± 0.58 | 2.44 ± 0.58 | 2.53 ± 0.47 | 0.901 | |
| Chord 30 | Capiox | 1.31 ± 0.14 | 1.33 ± 0.13 | 1.21 ± 0.11 | 0.559 |
| Inspire | 1.43 ± 0.11 | 1.43 ± 0.10 | 1.45 ± 0.10 | 0.778 | |
| Quadrox | 2.37 ± 0.67 | 2.34 ± 0.66 | 2.42 ± 0.55 | 0.947 | |
| Chord 50 | Capiox | 1.22 ± 0.11 | 1.26 ± 0.12 | 1.21 ± 0.11 | 0.465 |
| Inspire | 1.27 ± 0.09 | 1.28 ± 0.08 | 1.31 ± 0.11 | 0.616 | |
| Quadrox | 2.14 ± 0.61 | 2.16 ± 0.60 | 2.21 ± 0.48 | 0.949 |
OXBV: oxygenator blood volume.
Ratio BA: ratio of the area under the dilution curve before and after maximum dilution in total; Chord 30: ratio of the area under the dilution curve before and after maximum dilution starting from 30% of maximum curve height; Chord 50: ratio of the area under the dilution curve before and after maximum dilution starting from 50% of maximum curve height.
Figure 1.Oxygenator blood volume (OXBV) per oxygenator at a pulsatile flow of 2, 4, and 6 L/min and maximum permitted flow. An asterisk indicates a significantly higher OXBV compared to 4 and 6 L/min, and maximum flow; a hash tag indicates a significantly higher OXBV compared to the maximum flow.
Figure 2.Oxygen transfer (mL/min/m2) per oxygenator at blood flows of 4 and 6 L/min.
Figure 3.Oxygen transfer (ml/dL) per oxygenator at blood flows of 4 and 6 L/min.
Figure 4.ΔpO2 (mm Hg) per oxygenator at blood flows of 4 and 6 L/min.