Long-term device survival and quality of life outcomes following artificial urinary sphincter placement.

Timothy C Boswell1, Daniel S Elliott1, Laureano J Rangel2, Brian J Linder1.   

Abstract

BACKGROUND: Artificial urinary sphincter (AUS) placement is the standard for treatment of severe male stress urinary incontinence (SUI). While there is evidence to suggest satisfactory device survival, there is a paucity of data addressing long-term quality of life outcomes.
METHODS: We identified patients who underwent primary AUS placement from 1983 to 2016. We assessed rates of secondary surgery (overall, device infection/erosion, urethral atrophy, malfunction) and factors associated with these endpoints. Quality of life was evaluated by pad usage and Patient Global Impression of Improvement (PGI-I) at various time points from primary surgery. Follow-up was obtained in clinic or by phoned/mailed correspondence.
RESULTS: During the study time frame, 1,154 patients were eligible and included in the analysis. Patients had a median age of 70 years (IQR, 65-75 years) and median follow up of 5.4 years (IQR, 1.6-10.5 years). Overall device survival was 72% at 5 years, 56% at 10 years, 41% at 15 years, and 33% at 20 years. On univariate analysis, variables associated with need for secondary surgery were prior cryotherapy (HR 2.7; 95% CI, 1.6-4.6; P<0.01) or radiation therapy (HR 1.4; 95% CI, 1.1-1.7; P=0.01). On multivariable analysis, only cryotherapy remained significantly associated with this endpoint (HR 2.4; 95% CI, 1.3-4.2; P<0.01). While 36% and 23% of patients 5-10 years out from surgery and >10 years out from surgery, respectively, reported using a security pad or less per day, 78% and 81% of those patients, respectively, reported their PGI-I as at least "much better".
CONCLUSIONS: AUS placement has excellent long-term outcomes, and is associated with sustained improvement in patient quality of life. 2020 Translational Andrology and Urology. All rights reserved.

Entities:  

Keywords:  Male; artificial; stress; urinary incontinence; urinary sphincter

Year:  2020        PMID: 32055467      PMCID: PMC6995928          DOI: 10.21037/tau.2019.08.02

Source DB:  PubMed          Journal:  Transl Androl Urol        ISSN: 2223-4683


  17 in total

1.  Impact of the new cuff design on reliability of the AS800 artificial urinary sphincter.

Authors:  J K Light; J C Reynolds
Journal:  J Urol       Date:  1992-03       Impact factor: 7.450

2.  Long-term functional outcomes after artificial urinary sphincter implantation in men with stress urinary incontinence.

Authors:  Priscilla Léon; Emmanuel Chartier-Kastler; Morgan Rouprêt; Vanina Ambrogi; Pierre Mozer; Véronique Phé
Journal:  BJU Int       Date:  2014-10-23       Impact factor: 5.588

3.  Long-term continence and patient satisfaction after artificial sphincter implantation for urinary incontinence after prostatectomy.

Authors:  D K Montague; K W Angermeier; D R Paolone
Journal:  J Urol       Date:  2001-08       Impact factor: 7.450

4.  Artificial urinary sphincter for post-radical prostatectomy urinary incontinence: long-term subjective results.

Authors:  A E Gousse; S Madjar; M M Lambert; I J Fishman
Journal:  J Urol       Date:  2001-11       Impact factor: 7.450

5.  Long-term durability and functional outcomes among patients with artificial urinary sphincters: a 10-year retrospective review from the University of Michigan.

Authors:  Simon P Kim; Zubair Sarmast; Stephanie Daignault; Gary J Faerber; Edward J McGuire; Jerilyn M Latini
Journal:  J Urol       Date:  2008-03-18       Impact factor: 7.450

Review 6.  The artificial urinary sphincter after a quarter of a century: a critical systematic review of its use in male non-neurogenic incontinence.

Authors:  Frank Van der Aa; Marcus J Drake; George R Kasyan; Andreas Petrolekas; Jean-Nicolas Cornu
Journal:  Eur Urol       Date:  2012-11-23       Impact factor: 20.096

7.  EAU guidelines on assessment and nonsurgical management of urinary incontinence.

Authors:  Malcolm G Lucas; Ruud J L Bosch; Fiona C Burkhard; Francisco Cruz; Thomas B Madden; Arjun K Nambiar; Andreas Neisius; Dirk J M K de Ridder; Andrea Tubaro; William H Turner; Robert S Pickard
Journal:  Eur Urol       Date:  2012-08-31       Impact factor: 20.096

8.  Male Sling and Artificial Urethral Sphincter for Male Stress Urinary Incontinence Among Certifying American Urologists.

Authors:  Joceline S Liu; Matthias D Hofer; Jaclyn Milose; Daniel T Oberlin; Sarah C Flury; Allen F Morey; Chris M Gonzalez
Journal:  Urology       Date:  2015-09-04       Impact factor: 2.649

9.  The Michigan Incontinence Symptom Index (M-ISI): a clinical measure for type, severity, and bother related to urinary incontinence.

Authors:  Anne M Suskind; Rodney L Dunn; Daniel M Morgan; John O L DeLancey; Edward J McGuire; John T Wei
Journal:  Neurourol Urodyn       Date:  2013-08-14       Impact factor: 2.696

10.  The impact of incontinence etiology on artificial urinary sphincter outcomes.

Authors:  Adam R Miller; Brian J Linder; Laureano J Rangel; David Y Yang; Daniel S Elliott
Journal:  Investig Clin Urol       Date:  2017-06-27
View more
  3 in total

Review 1.  Artificial Urinary Sphincter Complications: Risk Factors, Workup, and Clinical Approach.

Authors:  Roger K Khouri; Nicolas M Ortiz; Benjamin M Dropkin; Gregory A Joice; Adam S Baumgarten; Allen F Morey; Steven J Hudak
Journal:  Curr Urol Rep       Date:  2021-03-29       Impact factor: 3.092

2.  Efficacy and safety of intradetrusor botulinum toxin injections for idiopathic overactive bladder syndrome in patients with an artificial urinary sphincter.

Authors:  Geoffroy de Sallmard; Marie-Aimée Perrouin-Verbe; Benoit Peyronnet; Xavier Biardeau; Delphine Maucort-Boulch; Emmanuel Chartier-Kastler; Véronique Phé; Alexia Even; Charles Joussain; Inès Dominique; Gérard Amarenco; Xavier Gamé; Christian Saussine; Alain Ruffion
Journal:  World J Urol       Date:  2021-10-09       Impact factor: 4.226

3.  World record: Single AMS 800 artificial urinary sphincter functioning uninterrupted for 29 years in male paraplegic patient.

Authors:  Peter Rehder; Jannik Stuehmeier; Lukas Andrius Jelisejevas; Alexandra Gulacsi; Wolgang Horninger; Marco Pedrini
Journal:  Urol Case Rep       Date:  2020-05-06