Evaluation of the validity of a rapid method for measuring high and low haemoglobin levels in whole blood donors.

Hayedeh J Shahshahani1, Nahid Meraat, Fatemeh Mansouri.   

Abstract

BACKGROUND: Haemoglobin screening methods need to be highly sensitive to detect both low and high haemoglobin levels and avoid unnecessary rejection of potential blood donors. The aim of this study was to evaluate the accuracy of measurements by HemoCue in blood donors.
MATERIALS AND METHODS: Three hundred and fourteen randomly selected, prospective blood donors were studied. Single fingerstick blood samples were obtained to determine the donors' haemoglobin levels by HemoCue, while venous blood samples were drawn for measurement of the haemoglobin level by both HemoCue and an automated haematology analyser as the reference method. The sensitivity, specificity, predictive values and correlation between the reference method and HemoCue were assessed. Cases with a haemoglobin concentration in the range of 12.5-17.9 g/dL were accepted for blood donation.
RESULTS: Analysis of paired results showed that haemoglobin levels measured by HemoCue were higher than those measured by the reference method. There was a significant correlation between the reference method and HemoCue for haemoglobin levels less than 12.5 g/dL. The correlation was less strong for increasing haemoglobin levels. Linear correlation was poor for haemoglobin levels over 18 g/dL. Thirteen percent of donors, who had haemoglobin levels close to the upper limit, were unnecessarily rejected. DISCUSSION: HemoCue is suitable for screening for anaemia in blood donors. Most donors at Yazd are males and a significant percentage of them have haemoglobin values close to the upper limit for acceptance as a blood donor; since these subjects could be unnecessarily rejected on the basis of HemoCue results and testing with this method is expensive, it is recommended that qualitative methods are used for primary screening and accurate quantitative methods used in clinically suspicious cases or when qualitative methods fail.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23114520      PMCID: PMC3729129          DOI: 10.2450/2012.0031-12

Source DB:  PubMed          Journal:  Blood Transfus        ISSN: 1723-2007            Impact factor:   3.443


  17 in total

1.  Hb screening of blood donors: how close is close enough?

Authors:  Ritchard G Cable
Journal:  Transfusion       Date:  2003-03       Impact factor: 3.157

2.  Estimation of hemoglobin in blood donors: A comparative study using hemocue and cell counter.

Authors:  Shalini Bahadur; Sonal Jain; Manjula Jain
Journal:  Transfus Apher Sci       Date:  2010-07-27       Impact factor: 1.764

3.  A retrospective cohort study of blood hemoglobin levels in blood donors and competitive rowers.

Authors:  P I Johansson; H Ullum; K Jensen; N H Secher
Journal:  Scand J Med Sci Sports       Date:  2008-02-21       Impact factor: 4.221

4.  The difference between fingerstick and venous hemoglobin and hematocrit varies by sex and iron stores.

Authors:  Ritchard G Cable; Whitney R Steele; Russell S Melmed; Bryce Johnson; Alan E Mast; Patricia M Carey; Joseph E Kiss; Steven H Kleinman; David J Wright
Journal:  Transfusion       Date:  2011-10-20       Impact factor: 3.157

5.  HemoCue photometer: a better alternative of hemoglobin estimation in blood donors?

Authors:  Kafil Akhtar; Rana K Sherwani; Khaliqur Rahman; Jaseem Hasan; Mohd Shahid
Journal:  Indian J Hematol Blood Transfus       Date:  2009-01-11       Impact factor: 0.900

6.  Statistical analysis of inappropriate results from current Hb screening methods for blood donors.

Authors:  Virge James; Keith F Jones; Elizabeth M Turner; Robert J Sokol
Journal:  Transfusion       Date:  2003-03       Impact factor: 3.157

7.  Evaluation of hemoglobin of blood donors deferred by the copper sulphate method for hemoglobin estimation.

Authors:  R B Sawant; Z S Bharucha; S B Rajadhyaksha
Journal:  Transfus Apher Sci       Date:  2007-03-26       Impact factor: 1.764

8.  A study of the prevalence of iron deficiency and its related factors in blood donors of Yazd, Iran, 2003.

Authors:  H Javadzadeh Shahshahani; M Attar; M Taher Yavari
Journal:  Transfus Med       Date:  2005-08       Impact factor: 2.019

9.  Quality evaluation of four hemoglobin screening methods in a blood donor setting along with their comparative cost analysis in an Indian scenario.

Authors:  Rashmi Tondon; Anupam Verma; Prashant Pandey; Rajendra Chaudhary
Journal:  Asian J Transfus Sci       Date:  2009-07

10.  Improved strategy for screening prospective blood donors for anaemia.

Authors:  F E Boulton; M J Nightingale; W Reynolds
Journal:  Transfus Med       Date:  1994-09       Impact factor: 2.019

View more
  6 in total

Review 1.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

2.  Diabetes care in the dispersed population of Greenland. A new model based on continued monitoring, analysis and adjustment of initiatives taken.

Authors:  Michael Lynge Pedersen
Journal:  Int J Circumpolar Health       Date:  2019       Impact factor: 1.228

3.  Effects of preanalytical factors on hemoglobin measurement: A comparison of two HemoCue® point-of-care analyzers.

Authors:  Ralph D Whitehead; Ming Zhang; Maya R Sternberg; Rosemary L Schleicher; Bakary Drammeh; Carine Mapango; Christine M Pfeiffer
Journal:  Clin Biochem       Date:  2017-04-12       Impact factor: 3.281

4.  Hemoglobin concentration, total hemoglobin mass and plasma volume in patients: implications for anemia.

Authors:  James M Otto; James O M Plumb; Eleri Clissold; Shriya B Kumar; Denis J Wakeham; Walter Schmidt; Michael P W Grocott; Toby Richards; Hugh E Montgomery
Journal:  Haematologica       Date:  2017-06-08       Impact factor: 9.941

5.  A Comparative Study of Serum Ferritin Levels Among Unfit and Fit Blood Donors.

Authors:  Olusola Festus Olowoselu; Ebele Uche; Olufemi Oyedeji; Oluwakemi E Otokiti; Oluwamuyiwa Anthony Ayanshina; Akinsegun Akinbami; Vincent Osunkalu
Journal:  Niger Med J       Date:  2020-02-24

6.  Reproducibility and validity of portable haemoglobinometer for the diagnosis of anaemia in children under the age of 5 years.

Authors:  Alessandra da Silva Pereira; Inês Rugani Ribeiro de Castro; Flávia Fioruci Bezerra; José Firmino Nogueira Neto; Ana Carolina Feldenheimer da Silva
Journal:  J Nutr Sci       Date:  2020-01-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.