S Srivastava1, N K Painuly2, S P Mishra3, K Srivastava4, N Singh5, S Singh6, M L B Bhatt7.
Abstract
BACKGROUND: Brachytherapy treatment planning in cervix carcinoma patients using two dimensional (2D) orthogonal images provides only point dose estimates while CT-based planning provides volumetric dose assessment helping in understanding the correlation between morbidity and the dose to organs at risk (OARs) and treatment volume.Entities:
Keywords: Organs at Risk ; Radiography; Brachytherapy
Year: 2019 PMID: 32039090 PMCID: PMC6943852 DOI: 10.31661/jbpe.v0i0.867
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure1Orthogonal radiograph showing ICRU bladder and rectum point along with two points 1.5cm above and below ICRU bladder point and four points 5mm cranially and cadually from ICRU rectal point.
Mean doses to ICRU points , DMax point , VICRU and D2cc volume of bladder and rectum.
| Bladder | Rectum | |
|---|---|---|
|
| 4.92±1.89 (range 9.83) | 4.58±1.15 (range 6.90) |
|
| 6.91±1.58 (range 9.27) | 4.82±0.88 (range 4.14) |
|
| 1.60±0.67 | 1.13±0.40 |
|
| 5.43±1.35 (range 7.11) | 5.59±1.90 (range 8.65) |
|
| 11.05±8.2 | 4.98±3.83 |
DMax- Maximum dose to point situated 1.5cm superior to ICRU bladder point and to point situated 5mm cranially to ICRU rectal point.
VICRU- Volume of bladder and rectum receiving at least the ICRU point dose.
Mean volumes of different isodose curves obtained from 2D plan and CT-based plan.
| Dose Gy (% of prescribed dose) | Vol. From 2D plan (cm3) | Vol. from CT-based plan (cm3) |
|---|---|---|
| 3.5 (50) | 237.79±61.28 | 224.79±58.43 |
| 7 (100) | 80.53±20.93 | 76.77±19.95 |
| 10.5 (150) | 42.95±11.16 | 40.95±10.64 |
| 14 (200) | 27.5±7.14 | 26.22±6.81 |
Mean volumes of bladder, rectum and bowel receiving 50%, 100%, 150% and 200% of the prescribed dose and mean contoured volume of bladder and rectum.
| Dose in % of prescribed dose | Bladder | Rectum | Bowel |
|---|---|---|---|
| 50% | 30.93±17.33 | 9.83±3.64 | 13.13±6.52 |
| 80% | 7.85±7.04 | 3.63±1.95 | 4.77±2.5 |
| 100% | 2.34±2.24 | 0.58±0.51 | 0.99±0.88 |
| 150% | 0.56±0.62 | 0.12±0.28 | 0.28±0.43 |
| 200% | 0.09±0.27 | 0±0 | 0±0 |
| Mean contoured volumes (cm3) | 101.48±70.88 | 53.68±16.58 | 272.75±115.60 |
Figure2Ratio of D2cc, the dose received by the 2 cm3 of the volume of bladder to ICRU bladder reference point for all patients.
Figure3Ratio of D2cc, the dose received by the 2 cm3 of the volume of rectum to ICRU rectal reference point for all patients.
Figure4Scatter diagram depicting correlation between DMax and D2cc bladder.
Figure5Scatter diagram depicting correlation between DMax and D2cc rectum.
Volume and doses of OAR from CT-based plan.
| Volume (cm3) | Bladder | Rectum | Bowel |
|---|---|---|---|
| 2cc | 6.91±1.59 | 4.82±0.88 | 3.58±1.76 |
| 1cc | 7.73±1.56 | 5.32±0.97 | 4.02±1.94 |
| 0.1cc | 8.70±1.58 | 6.31±1.12 | 4.82±2.22 |